K160469 · Gowoonsesang Cosmetics Co., Ltd. · GEI · Jan 5, 2017 · General, Plastic Surgery
Device Facts
Record ID
K160469
Device Name
AGNES
Applicant
Gowoonsesang Cosmetics Co., Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Jan 5, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
AGNES is indicated for use in dermatological and general surgical procedures for electrocoagulation and hemostasis.
Device Story
AGNES is an RF electrosurgical device consisting of an LCD screen, RF generator, and SMPS. It uses disposable needle-type RF electrodes to deliver 1MHz RF energy to target tissue. Operated by a physician in a clinical setting, the device is controlled via a footswitch. It features temperature, impedance, continuity, and electrode monitoring. The device applies heat to tissue for coagulation and hemostasis. By destroying sebaceous glands, it aids in treating acne and reducing inflammation. The healthcare provider monitors the procedure via the device interface and visual assessment, using the output to perform precise tissue coagulation. Benefits include effective lesion treatment and potential reduction in acne recurrence.
Clinical Evidence
No human clinical data provided. Evidence consists of bench testing (electrical safety, EMC, RF output, mechanical/electrical integrity of electrodes, biocompatibility per ISO 10993, and sterilization validation) and an animal study. The animal study (mini-pigs and rabbit ear acne model) evaluated penetration depth, coagulation efficacy, and histological destruction of sebaceous glands, demonstrating successful reduction of inflammatory lesions and acne-related pathology.
Technological Characteristics
RF electrosurgical device; 1MHz output frequency; monopolar/bipolar capability. Electrodes: STS 304 stainless steel with p-xylylene dimer C insulating coating. Dimensions: 290mm x 455mm x 271.7mm; Weight: 5.8kg. Power: 100-250VAC. Includes temperature, impedance, continuity, and electrode monitoring sensors. Sterilization: EO gas. Connectivity: Standalone unit with footswitch.
Indications for Use
Indicated for dermatological and general surgical procedures requiring electrocoagulation and hemostasis. No specific age or gender restrictions provided.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
INTRAcel Premium Fractional RF Micro Needle (FRM) System (K153727)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized caduceus or a family of three people.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 5, 2017
GOWOONSESANG Cosmetics Co., Ltd % DongHa Lee Regulatory Affairs Consultant KMC. Inc. Room No. 409, ACE Technotower 1-Cha, 38-9, Digital-ro 31-gil, Guro-gu Seoul, Korea 152-766 KR
Re: K160469
Trade/Device Name: AGNES Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI, KCW Dated: November 4, 2016 Received: November 8, 2016
Dear DongHa Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, Jennifer R. Stevenson -A For Binita S. Ashar, M.D., M.B.A., F.A.C.S.
Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160469
Device Name AGNES
Indications for Use (Describe)
AGNES is indicated for use in dermatological and general surgical procedures for electrocoagulation and hemostasis.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary (K160469)
This summary of 510(K) - safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
## Date: January 3, 2017
## 1. Applicant / Submitter
GOWOONSESANG COSMETICS CO., LTD Address: (Seohyeon-dong, 4, 5F Cocoplaza), 20, Seohyeon-ro 210beon-gil, Bundang-gu, Seongnam-si, Gyeonggi-do, Korea (Postal code: 13591) Tel : +82-31-724-9009 Fax : +82-31-724-9093
### 2. Submission Contact Person
DongHa Lee (Consultant, KMC, Inc.) Address: Room no. 904, 27, Digital-ro 27ga-gil, Guro-gu, Seoul, Korea Tel: +82-70-8965-5554 Fax: +82-2-2672-0579 Email: dhlee@kmcerti.com
### 3. Device Information
- Trade Name: AGNES
- Common Name: RF Electrosurgical Device
- Classification Name: Electrosurgical cutting and coagulation device and accessories
- Classification Product Code and Regulation: GEI, 21CFR 878.4400
- Subsequent Product Code and Regulation: KCW, 21CFR878.5350
- Device Class: 2
## 4. Predicate Device
| | Predicate Device(1) | Predicate Device(2) |
|---------------|-------------------------------------------------------------|---------------------|
| Manufacturer | Jeisys Medical Incorporated | IME Co., Ltd. |
| Device Name | INTRAcel Premium Fractional<br>RF Micro Needle (FRM) System | HR-5000 |
| 510(k) number | K153727 | K881276 |
#### 5. Description
AGNES is a RF electrosurgical device. It consists of LCD screen, radiofrequency generator and SMPS. The accessories are a hand-piece, Single use RF electrodes (needle type). Disposable
{4}------------------------------------------------
neutral electrode pad and a footswitch.
## 6. Indication for use
AGNES is indicated for use in dermatological and general surgical procedures for electrocoagulation and hemostasis.
## 7. Substantial Equivalence
AGNES is substantially equivalent to the predicate device, HR-5000 (K881276) by IME Co., Ltd. The following comparison table is presented to demonstrate substantial equivalence.
| Descriptive Information | Subject Device | Predicate Device (1) | Predicate Device (2) | |
|----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| Manufacturer | Gowoonsesang<br>Cosmetics Co., Ltd. | Jeisys Medical<br>Incorporated | IME Co., Ltd. | |
| Device Name | AGNES | INTRAcel Premium<br>Fractional RF Micro<br>Needle (FRM) System | HR-5000 | |
| 510(k) number | K160469 | K153727 | K881276 | |
| Classification Product Code /<br>Regulatory Number | GEI / 878.4400 | GEI / 878.4400 | KCW / 878.5350 | |
| Subsequent Product Code | KCW / 878.5350 | - | - | |
| Regulatory Class | 2 | 2 | 1 | |
| Indications for Use | AGNES is indicated<br>for use in<br>dermatological and<br>general surgical<br>procedures for<br>electrocoagulation and<br>hemostasis. | INTRAcel Premium<br>Fractional RF Micro<br>Needle (FRM) System<br>device is indicated for<br>use in dermatological<br>and general surgical<br>procedures for<br>electrocoagulation and<br>hemostasis. | Epilation.<br>Dermatologic<br>indications,<br>specifically:<br>telangiectasia, port<br>wine stain, adenoma<br>sebaceum,<br>angiokeratoma,<br>angioma, nevus cell<br>nevi, syringoma, acne,<br>rosacea, senile ectasia,<br>rhinophyma, and<br>lymphangiomas. | |
| Prescription or OTC | Prescription | Prescription | Prescription | |
| Operation | The device uses RF<br>energy delivered<br>through micro needle<br>electrode to apply heat<br>to target tissue for<br>coagulating. | The device uses RF<br>energy delivered<br>through micro needle<br>electrode to apply heat<br>to target tissue for<br>coagulating. | The device uses RF<br>energy delivered<br>through micro needle<br>electrode to apply heat<br>to target tissue for<br>coagulating. | |
| Electrosurgical<br>Unit (ESU) | Monopolar<br>or Bipolar | Monopolar | Bipolar | |
| | Temperature<br>sensors | None | None | |
| | Impedance<br>monitor | None | None | None |
| | Continuity<br>monitor | Checking the<br>connection between the<br>neutral electrode and<br>the electrosurgical unit. | None | Unknown |
| | Electrode<br>monitor | Provide a camera to<br>monitor the electrode<br>coating condition and<br>shape before using by<br>the user. | None | None |
| | Output<br>frequency | 1MHz | 1MHz | 1MHz |
| | Waveform | Oscillating rectangular<br>wave | Unknown | Oscillating rectangular<br>wave |
| | Power<br>output | 2 to 46 W at 200 ohm | 12.5 to 50W at 200<br>ohm | 5 to 50W at 200 ohm |
| | Voltage<br>output | 25 to 104V | Unknown | 10 to 90V |
| | Dimensions | 290mm(W)x455mm(L)<br>x271.7mm(H) | 350mm(W)x405mm(L)<br>x1080mm(H) | 247mm(W)x270mm(L)<br>x127mm(H) |
| | Weight | 5.8Kg | 63Kg | 5.2kg |
| | Power Input | 100-250VAC, 50-60Hz | 120VAC, 50/60 Hz | 100-240VAC, |
| | FDA<br>Approval | None | None | None |
| | Monopolar<br>or Bipolar | Monopolar | Bipolar | Monopolar |
| | Electrode<br>type | Micro Needle type | Micro Needle type | Micro Needle type |
| Active<br>accessory<br>(RF Electrode) | Physical<br>Dimensions | Needle length :<br>0.8/1.25/1.5/ 2.0 mm<br>Thickness : 0.2mm | Needle length :<br>0.5/0.8/1.5/2.0mm<br>Thickness : 0.25mm | Needle length: 1.5mm |
| | Materials | Electrode: STS 304,<br>Insulating Coating: p-<br>xylylene dimer C | ABS + SUS304 | Material: STS 304,<br>Insulating Coating:<br>Teflon PTFE |
| | Single Use<br>or Reusable | Single Use | Single Use | Single Use |
| | Sterilization | EO gas | EO gas | EO gas |
| Neutral<br>electrode pad | Approval | 510(k) cleared by FDA<br>(Κ102372). | None | Unknown |
| | Conductive<br>Or<br>Capacitive | Conductive area:<br>132cm² | None | Unknown |
| | Single Use<br>or Reusable | Single Use | None | Unknown |
| | Physical<br>Specification | Mono-polar | None | Mono-polar |
| | Materials | Conductive material:<br>Metal foil<br>Backing material:<br>Foam backing | None | Metal plate |
| Miscellaneous<br>accessory<br>(Foot switch) | Functions | For emitting RF energy<br>into electrode. | For emitting RF energy<br>into electrode. | For emitting RF energy<br>into electrode. |
| | Performance<br>Specifications | Single pole, single<br>throw | Single pole, single<br>throw | Single pole, single<br>throw |
| | Physical<br>Specification | Single pedal, IP68 | Single pedal, IPX8 | Single pedal, Unknown |
| | Materials | Plastic, SPCC | Unknown | Plastic, SPCC |
{5}------------------------------------------------
{6}------------------------------------------------
### 7.1 The same between the subject device and the predicates device.
- 1) Product Code
: The proposed product code of the subject device is GEI for electrosurgical function. It is the same product code as the predicate device in K153727. The proposed product also has subsequent product code as KCW for epilation function. It is the same product code as the predicate device in K881276.
## 2) Regulatory number and classification
: The proposed regulatory number of the subject device is 878.4400 and the classification is 2. It is the same regulatory number and classification as the predicate device in K153727.
#### 3) Indications for Use
: The subject device is indicated for use in dermatological and general surgical procedures for electrocoagulation and hemostasis. It is the same as the predicate device in K153727.
#### 4) Prescription Use
: The subject device is a prescription use device. It is the same as the predicate devices in K153727 and K881276.
#### 5) Operation
: The principle of operation of the subject device is using RF energy delivered through micro needle electrode to apply heat to target tissue for coagulating. It is the same as the predicate devices in K153727 and K881276.
#### 6) Technical characteristic
: The subject device is monopolar function. It is the same as the predicate device in K881276. The subject device has 1MHz frequency output. It is the same as the predicate devices in K153727 and K881276.
#### 7) Electrode
: The subject device uses disposable sterilized needle type electrode to delivery RF energy to the
{7}------------------------------------------------
target tissue. It is the same as the predicate devices. For monopolar function, the subject device uses a neutral electrode to be attached the patient back. It is the same as the predicate device in K881276.
#### 8) Miscellaneous accessory
: The subject device uses a footswitch (Single pole, single throw) to emit RF energy into electrode. It is the same as the predicate devices.
## 7.2 Differences between Subject and Predicates Devices
1) ESU
: The subject device has different RF output power (2 to 46 W). The range of output power is within the predicate devices and the output was also evaluated by safety and performance tests according to IEC 60601-1, IEC 60601-2-2 and animal testing. The results show that these difference does not raise any problems in the safety and effectiveness.
## 2) Electrode
: The subject device has different electrode physical dimensions (Needle length: 0.8/1.25/1.5/ 2.0 mm, Thickness: 0.2mm). The target tissue is located in dermis (average depth from 0.5mm to 2.0mm). The predicate devices (needle length 0.5/0.8/1.5/2.0mm in K153727, 0.5 to 2.0 mm in K12336 and 1.5mm in K881276) also has target tissue located in dermis. The electrode is coated by different insulating coating material p-xylylene dimer C. Thermal effect was evaluated by animal testing. The results show that these difference does not raise any problems in the safety and effectiveness.
## 8. Electrical Safety and Electromagnetic compatibility
The Electrical Safety and Electromagnetic compatibility tests were performed in accordance with the following standards.
- -IEC 60601-1: 2005, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-2-2: 2009, Medical electrical equipment Part 2-2: Particular requirements for the basic safety and essential performance of Radio frequency surgical equipment and Radio frequency.
- IEC 60601-1-2: 2014, Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests
## 9. Performance Testing - Nonclinical
- 9.1 Electrosurgical Unit
- 1) RF output power testing
{8}------------------------------------------------
The tests were performed in accordance with the FDA recognized standard, IEC 60601-2-2:2009, Part 2-2: Particular requirements for the basic safety and essential performance of Radio frequency surgical equipment and Radio frequency, Clause 201.11 and 201.12
9.2 Active Component - Disposable RF electrode (Needle type)
1) Mechanical testing (mechanical failure and short circuiting)
2) Electrical performance (insulation integrity)
The tests were performed with the RF surgical unit in accordance with the FDA recognized standard,
- -IEC 60601-2-2:2009, Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of Radio frequency surgical equipment and Radio frequency, Clause 201.8, 201.13 and 201.15
- 3) Biocompatibility
The tests were performed in accordance with the FDA recognized standards,
- ISO 10993-1: 2009, Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process.
- ISO 10993-5: 2009, Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10: 2010, Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization.
- 4) Sterility and packing
The sterilization and packing is verified and validated in in accordance with the FDA recognized standards
- ISO 11135: 2014, Sterilization of health care products Ethylene oxide Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices
- ISO 11138-2: 2006, Sterilization of health care products Biological indicators Part 2: -Biological indicators for ethylene oxide sterilization processes
- ISO 10993-7: 2008, Biological evaluation of medical devices Part 7: Ethylene oxide sterilization residuals
- ISO 11607-1: 2006, Packaging for terminally sterilized medical devices Part 1: Requirements for materials, sterile barrier systems and packaging systems
- ASTM F1980-07: 2011, Standard guide for accelerated aging of sterile barrier systems for medical devices.
- 9.3 Disposable Neutral Electrode Pad
{9}------------------------------------------------
The disposable neutral electrode pad is received prior 510(k) clearance by FDA (K102372).
1) Thermal performance, contact impedance, and adhesion testing
The tests were performed with the RF electrosurgical unit in accordance with the FDA recognized standard,
- IEC 60601-2-2:2009, Medical electrical equipment Part 2-2: Particular requirements for the basic safety and essential performance of Radio frequency surgical equipment and Radio frequency, Sub-clause 201.15.101.
9.4 Miscellaneous component - Foot Switch
1) Design specification and performance test.
The tests were performed with the RF electrosurgical unit in accordance with the FDA recognized standard,
- -IEC 60601-2-2:2009, Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of Radio frequency surgical equipment and Radio frequency, Sub-clause 201.8.10.4. 201.11.6.5 and 201.12.2
### 10. Performance Testing - Animal
This animal study was conducted to measure the penetration depth and do a macroscopic check on the improvement of acne and histological pathology analyses on the animal model of acne through clinical assessment (photo assessment) and histological evaluation.
#### 1) Microscopic assessment
- Macroscopic assessment by dermatologist on damage and safety on the skin surface and inside the corium
- Photo taking with a camera (EOS D3000, Cannon, Tokyo, Japan)
- Check the skin conditions using a folliscope magnifying glass (LeedM. Seoul, Korea)
- Assess reduction of acne lesion using 3 dimensional measuring instrument PRIMOS Pico (GFM, Germany)
#### 2) Historical assessment
Extract skin on the rabbit's ear at the end of the test and fix with a 4% paraformaldehyde. Dehydrate with alcohol and xylene by stage and embed with paraffin. Make a fragment of 5 m or smaller using a microtome, and remove paraffin with alcohol and xylene again. Dye with hematoxylin & eosin and check the condition of the epidermis and dispersion of sebaceous glands.
{10}------------------------------------------------
This animal study results as the below.
1) We checked the depth of skin penetration and discovered that the penetration depth was relatively accurate. For mini pigs, we found distinct coagulation of skin on the area applied with RF.
2) In order to make the rabbit's ear model, we conducted intradermal injection of acne bacteria P. acnes and induced oleic acid for 4 weeks. As a result, we found acne lesions such as closed comedos and papules or pustules.
3) We selected groups by each grade and conducted the Agnes treatment, and discovered that the sebaceous glands were destroyed and inflammatory cells, papule and pustule on the lesion area were greatly reduced.
4) The destruction of sebaceous glands by Agnes killed acne bacteria and induced fast relief of inflammation, and helped quick treatment that it is considered as a valid method to prevent reoccurrence.
## 11. Conclusion
In comparing between the subject device and the predicate devices, there are indications for use, the operating principle and technological characteristics (monopolar function, output frequency). Although there are some differences as output power, electrode physical dimensions and electrode insulating coating material, the safety and performance test reports are supported to the safety and effectiveness of the subject device. The result of animal study is also supported to thermal effects on tissue and the intended use of the subject device.
In this regard, we conclude that the subject device is substantially equivalent to the predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.