K241803 · Insulcloud S.L. · QOG · Aug 22, 2024 · General Hospital
Device Facts
Record ID
K241803
Device Name
INSULCLOCK® v2.0 PRO
Applicant
Insulcloud S.L.
Product Code
QOG · General Hospital
Decision Date
Aug 22, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Indications for Use
INSULCLOCK® v2.0 pro is intended to detect, store and transfer insulin doserelated data, the time of injection and the temperature to which the disposable insulin pen has been exposed to. All this information is sent to a mobile app via integrated SDK.
Device Story
Reusable data transmitter attaches to disposable insulin pens (Kwikpen or Flextouch); detects/stores insulin dose, injection time, and ambient temperature. Data transfers via Bluetooth to compatible mobile app using SDK. Used in home environment by patients or caregivers. Healthcare providers access data via app to review injection history and temperature exposure; aids in diabetes management and therapy adherence. Rechargeable battery-powered; provides digital record of insulin administration.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing and human factors validation study.
Technological Characteristics
Reusable data transmitter; foiled flex circuit; rechargeable battery (UNE-EN 62133-2, UN 38.3, UL 1642). Bluetooth connectivity. Compatible with Kwikpen and Flextouch pens. Meets ISO 11608-1 for dose accuracy, IEC 60601-1 for electrical safety, and IEC 62304 (Class B) for software. Biocompatibility per ISO 10993-1. Cybersecurity per UL 2900-1/2-1.
Indications for Use
Indicated for patients 18-65 years old with type I or type II diabetes mellitus using disposable insulin pens. Users outside this age range require supervision. Designed for home use with a compatible app.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 22, 2024
INSULCLOUD S.L.
% Dulciana Chan Principal Consultant ROM+ 2790 Mosside Blvd. Suite 800 Monroeville, Pennsylvania 15146
Re: K241803
Trade/Device Name: INSULCLOCK® v2.0 PRO Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: QOG Dated: June 18, 2024 Received: June 21, 2024
Dear Dulciana Chan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{2}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Shruti N. Mistry -S
Shruti Mistry Assistant Director Division of Drug Delivery and General Hospital Devices, and Human Factors Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K241803
Device Name
INSULCLOCK® v2.0 PRO
#### Indications for Use (Describe)
- Insulclock® is for use by patients using disposable insulin diagnosed with type I or type II diabetes mellitus.
- Insulclock® is for use by people from 18 to 65 years old.
Out of this age range, the Insulclock® must be used under supervision.
- Insulclock® is designed for use in any home environment.
- Using a compatible App.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
# DATE PREPARED
June 18, 2024
#### MANUFACTURER AND 510(k) OWNER
INSULCLOUD S.L. Avenida Sabino Arana, #8, 48013 Telephone: Official Contact: Pilar Aranda, Quality Technician
#### REPRESENTATIVE/CONSULTANT
Allison C. Komiyama, Ph.D., R.A.C. Dulciana Chan, MSE RQM+ Telephone: +1 (877) 652-0830 Email: akomiyama@rqmplus.com dchan@rqmplus.com
#### DEVICE INFORMATION
| Proprietary Name/Trade Name: | INSULCLOCK® v2.0 PRO |
|------------------------------|-----------------------------------------------------------|
| Common Name: | Piston Syringe |
| Regulation Number: | 880.5860 |
| Class: | II |
| Product Code: | QOG |
| Premarket Review: | OHT3: Office of Gastrorenal, OBGyn, General Hospital, and |
| | Urology Devices, Office of Product Evaluation and Quality |
| Review Panel: | General Hospital and Personal Use Devices |
#### PREDICATE DEVICE IDENTIFICATION
The INSULCLOCK®v2.0 PRO is substantially equivalent to the following predicate:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary Predicate |
|---------------|--------------------------------------|-------------------|
| K212217 | Tempo Smart Button | ✓ |
The predicate device has not been subject to a design related recall.
#### DEVICE DESCRIPTION
The INSULCLOCK® v2.0 pro is a reusable data transmitter with Software Development Kit (SDK) that detects and stores insulin dose- related data when attached to a disposable insulin pen and then transfers information to a compatible mobile application via Bluetooth® wireless technology which can display the brand of insulin, dose amount, injection time, and ambient temperature.
Models:
- INSULCLOCK® v2.0 PRO for Kwikpen® (Lilly®) ।
- INSULCLOCK® v2.0 PRO for Flex-touch® (Novo Nordisk®) -
{5}------------------------------------------------
# INDICATIONS FOR USE
- Insulclock® is for use by patients using disposable insulin diagnosed with type I or type II diabetes mellitus.
- Insulclock® is for use by people from 18 to 65 years old.
Out of this age range, the Insulclock® must be used under supervision.
- Insulclock® is designed for use in any home environment.
- Using a compatible App.
INTENDED USE: INSULCLOCK® v2.0 pro is intended to detect, store and transfer insulin doserelated data, the time of injection and the temperature to which the disposable insulin pen has been exposed to. All this information is sent to a mobile app via integrated SDK.
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
INSULCLOCK® v2.0 pro is substantially equivalent to the Tempo Smart Button (K212217) predicate device based on the similarities summarized in the following Table of Substantial Equivalence.
{6}------------------------------------------------
| Attribute | Subject Device | Predicate Device | Comparison |
|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|
| | INSULCLOUD S.L.<br>INSULCLOCK® v2.0 pro | Eli Lilly and Company<br>Tempo Smart Button<br>K212217 | |
| Intended Use of<br>Electronic<br>Functionality | Detect, store, and transfer insulin dose- related data from the pen to a compatible application | Detect, store, and transfer insulin dose- related data from the pen to a compatible application | Same |
| Indications for Use | Insulclock® is for use by patients using disposable insulin diagnosed with type I or type II diabetes mellitus.<br>- Insulclock® is for use by people from 18 to 65 years old.<br>Out of this age range, the Insulclock® must be used under supervision.<br>- Insulclock® is designed for use in any home environment.- Using a compatible App. | The Tempo Smart Button is intended to detect, store, and transfer insulin dose-related data from a Tempo Pen to a compatible application such as the myTempo App (App). The tempo Smart button is indicated for single-patient use by patients 18 years or older who are:<br>Diagnosed with type 1 or 2 diabetes mellitusUsing pre-filled insulin Tempo pens (Basaglar® Tempo Pen (insulin glargine) injection 100 units/mL, Lyumjev® Tempo Pen (insulin lispro-aabc) injection 100 units/mL, Humalog®Tempo Pen (insulin lispro injection) 100 units/mL]Using compatible application (APP) | Similar |
| Intended Users | Patients between the ages of 18 and 65 who are diagnosed with type 1 or type 2 diabetes mellitus. If the user is outside this age range, the device must be used under supervision. | People with diabetes age 18 and older | Similar |
| Single Patient Use | Yes | Yes | Same |
| Availability | Prescription only | Prescription only | Same |
| Classification | Class II - 21 CFR 880.5860 | Class II - 21 CFR 880.5860 | Same |
| Drug Contact | None | None | Same |
| Biocompatibility<br>defined by ISO 10993-1 | Meets requirements for intact skin contact | Meets requirements for intact skin contact | Same |
| Dose Accuracy | Meets ISO 11608-1 requirements when evaluated with compatible pens | Meets ISO 11608-1 requirements when evaluated with compatible pens | Same |
| Attribute | Subject Device | Predicate Device | Comparis<br>on |
| | INSULCLOUD S.L.<br>INSULCLOCK® v2.0 pro | Eli Lilly and Company<br>Tempo Smart Button<br>K212217 | |
| Reusable Device | Yes | Yes | Same |
| Digital Dose Display | No | No | Same |
| Electronically<br>Controlled Dosing | No | No | Same |
| Battery | Rechargeable battery meeting<br>UNE-EN 62133-2:2017,UN 38.3<br>and UL 1642. | Non-rechargeable | Different |
| Electronic Structure | Foiled flex circuit | Foiled flex circuit | Same |
| Electrical Safety | Meets IEC 60601-1 Requirements | Meets IEC 60601-1 Requirements | Same |
| Embedded Firmware | Meets IEC 62304 requirements:<br>Class B<br>Enhanced documentation in<br>accordance with Content of<br>Premarket Submissions for<br>Device Software Functions | Meets IEC 62304 requirements:<br>Major level of concern | Same |
| Electronic Data for<br>Insulin Identification | Insulin Identification manually<br>selected by the user during setup. | Insulin Identification detected by<br>system. | Different |
| | Insulin identification displayed on<br>mobile application. | Insulin identification displayed on<br>mobile application. | Same |
| Electronic Data for<br>Dose Information<br>Capture | Dose-related data recorded; dose<br>information displayed on mobile<br>application. | Dose-related data recorded; dose<br>information displayed on mobile<br>application. | Same |
| Dose History | Full; Displayed on mobile<br>application. | Full; Displayed on mobile<br>application. | Same |
| Data Transfer | Bluetooth. | Bluetooth. | Same |
| Insulin Disposable Pen<br>compatibilities | • Kwikpen® model (Lilly)<br>• Flextouch® model (Novo<br>Nordisk) | Tempo Pens (Basaglar® U-100<br>Tempo Pen, Lyumjev® U-100<br>Tempo Pen, Humalog® U-100<br>Tempo Pen) | Different |
| Compatible Mobile<br>Applications | Mobile applications for Android<br>8.0 or higher or iOS 15.0 or<br>higher, using the<br>INSULCLOCK® v2.0 pro<br>Software Development Kit (SDK) | My Tempo App | Different |
| Temperature<br>Measurement | Ambient temperature<br>measurement | No | Different |
{7}------------------------------------------------
{8}------------------------------------------------
The following technological differences exist between the subject and predicate devices:
- Tempo Smart Button is indicated for use in people with diabetes over 18 years of age and -INSULCLOCK® v2.0 pro intended users are patients between the ages of 18 and 65 who are diagnosed with type 1 or type 2 diabetes mellitus. If the user is outside this age range, the device must be used under supervision.
- The INSULCLOCK® v2.0 pro battery is rechargeable, whereas the predicate battery is not rechargeable.
- Both INSULCLOCK® v2.0 pro and Tempo Smart Button display Insulin identification on the mobile application. However, with INSULCLOCK® v2.0 pro, insulin is identified by the user in the app during setup, while it is detected by the system when using Tempo Smart Button.
- -The subject and predicate devices are compatible with different Insulin Disposable Pens. Tempo Smart Button is compatible with Tempo pens (i.e., Basaglar® U-100 Tempo Pen, Lyumjev® U-100 Tempo Pen, Humalog® U-100 Tempo Pen). INSULCLOCK® v2.0 pro features two device models, one compatible with the Kwikpen® model (Lilly) and one with the Flextouch® model (Novo Nordisk).
- -Compatible Mobile Applications: Tempo Smart Button is only compatible with My Tempo App. whereas INSULCLOCK® v2.0 pro is compatible with Mobile applications for Android 8.0 or higher or iOS 15.0 or higher using the INSULCLOCK® v2.0 pro Software Development Kit (SDK).
- -INSULCLOCK® v2.0 pro can measure ambient temperature whereas Tempo Smart Button cannot.
In summary, INSULCLOCK® v2.0 pro has the same intended use and similar indications, technological features, and operating principles to the predicate device. The minor differences in technological characteristics do not raise new or different questions of safety or effectiveness.
# SUMMARY OF NON-CLINICAL TESTING
The INSULCLOCK® v2.0 pro was tested to evaluate its performance based on the following standards:
| Standard | Version | Title |
|----------------------------|---------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility | | |
| ISO 10993-1 | 2018 | Biological evaluation of medical devices — Part 1: Evaluation and testing<br>within a risk management process |
| Software and Cybersecurity | | |
| IEC 62304 | 2006 | Medical device software. Software Life Cycle Processes |
| UL 2900-1 | 2023 | Software Cybersecurity for Networks-Connectable Products, Part 1: General<br>Requirements |
| UL 2900-2-1 | 2023 | Software Cybersecurity for Networks-Connectable Products, Part 2-1:<br>Particular Requirements for Network Connectable Components of Healthcare<br>and Wellness Systems |
| Electrical Safety and EMC | | |
| IEC 60601-1 | 2005 + | Medical electrical equipment - Part 1: General requirements for basic safety |
{9}------------------------------------------------
| Standard | Version | Title |
|-----------------------------------------|-----------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1-2 | Amd1:2012 | and essential performance. |
| IEC 60601-1-2 | 2014 | Medical electrical equipment - Part 1-2: General requirements for safety -<br>Collateral standard: Electromagnetic compatibility - Requirements and tests. |
| IEC 60601-1-<br>2:2020 (4.1<br>Edition) | 2020 | Medical Electrical Equipment - Part 1-2: General Requirements For Basic<br>Safety And Essential Performance - Collateral Standard: Electromagnetic<br>Disturbances - Requirements And Tests |
| ANSI C63.27 | 2017 | American National Standard for Evaluation of Wireless Coexistence |
| AIM 7351731 | N/A | Medical Electrical Equipment & System Electromagnetic Immunity Test for<br>Exposure to RFID Readers |
| AAMI TIR69:2017<br>(R2020) | 2017 | American National Standard for Evaluation of Wireless Coexistence |
| Performance | | |
| UNE-EN 11608-1 | ISO 2022 | Needle-based injection systems for medical use - Requirements and test<br>methods - Part 1: Needle-based injection systems |
## HUMAN FACTORS PERFORMANCE DATA
A human factors validation study was conducted on participants who were representative of intended users. The results of the study showed that intended users were able to operate the device as intended.
#### SUMMARY OF CLINICAL TESTING
Clinical testing was not required to demonstrate the substantial equivalence of the INSULCLOCK® v2.0 pro to its predicate device.
#### CONCLUSION
Based on the information presented in this 510(k) premarket notification, the INSULCLOCK® v2.0 pro is considered substantially equivalent to the predicate device, Tempo Smart Button (K212217). The differences noted between the INSULCLOCK® v2.0 pro and the predicate device do not impact substantial equivalence based on the successfully conducted testing of the subject device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.