K241802 · Geistlich Pharma AG · MQV · Nov 8, 2024 · Orthopedic
Device Facts
Record ID
K241802
Device Name
Device 300397 Putty
Applicant
Geistlich Pharma AG
Product Code
MQV · Orthopedic
Decision Date
Nov 8, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Device 300397 Putty is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. Device 300397 Putty is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., posterolateral spine, ilium and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. Device 300397 Putty must be used with blood or autogenous bone marrow aspirate and autograft in posterolateral spine.
Device Story
Device 300397 Putty is an osteoconductive, porous, resorbable bone graft substitute; composed of calcium phosphate granules (α-TCP and hydroxyapatite) and porcine collagen (Type I and III). Supplied as a sterile, dry block; hydrated with blood, autogenous bone marrow aspirate, or saline to form a moldable putty. Used by surgeons in clinical settings to fill osseous defects in the skeletal system (posterolateral spine, ilium, pelvis). Acts as a scaffold for bone regeneration; resorbs over time as it is replaced by new bone. Provides a physical filler for voids not intrinsic to structural stability. Benefits patient by facilitating bone healing in surgical or traumatic defects.
Clinical Evidence
Bench testing included chemical/physical characterization, sterilization validation (ISO 11137), packaging validation (ISO 11607, ASTM standards), and biocompatibility (ISO 10993). Animal study: NZ White Rabbit Single Level Spinal Fusion Model evaluated performance as a bone graft extender. Histology confirmed progression/maturation of fusions, new bone formation, and marrow space development with reduced local inflammation when hydrated with BMA or blood.
Technological Characteristics
Osteoconductive bone graft substitute. Materials: α-TCP/hydroxyapatite granules (0.5-2 mm) and porcine collagen (Type I/III). Form factor: sterile dry block, hydrated to moldable putty. Sterilization: irradiation. Standards: ISO 11137 (sterilization), ISO 11607/ASTM (packaging), ISO 10993 (biocompatibility), ICH Q3D (elemental impurities).
Indications for Use
Indicated for patients with bony voids or gaps in the skeletal system (posterolateral spine, ilium, pelvis) not intrinsic to structural stability. Defects may be surgically created or traumatic. Must be used with blood or autogenous bone marrow aspirate and autograft in the posterolateral spine.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services logo. The logo on the right is the FDA logo, which is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 8th, 2024
Geistlich Pharma AG % Roshana Ahmed Sr. Regulatory Specialist TELOS Partners LLC 2850 Frontier Drive Warsaw, Indiana 46582
Re: K241802
Trade/Device Name: Device 300397 Putty Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable Calcium Salt Bone Void Filler Device Regulatory Class: Class II Product Code: MOV Dated: October 4, 2024 Received: October 4, 2024
Dear Roshana Ahmed:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Robert M. Stefani -S 2024.11.08 12:42:36 -05'00'
For: Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K241802
Device Name
Device 300397 Putty
### Indications for Use (Describe)
Device 300397 Putty is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. Device 300397 Putty is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., posterolateral spine, ilium and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process.
Device 300397 Putty must be used with blood or autogenous bone marrow aspirate and autograff in posterolateral spine.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
#### I. Submitter
Geistlich Pharma AG Bahnhofstrasse 40 CH-6110 Wolhusen Switzerland Phone: +41 41 492 55 55
Contact Person: Marco Steiner, Deputy Director Regulatory Affairs Date Prepared: November 1, 2024
#### II. Device
| Device Proprietary Name: | Device 300397 Putty |
|--------------------------|-------------------------------------|
| Common or Usual Name: | Bone Void Filler |
| Classification Name: | Filler, Bone Void, Calcium Compound |
| Regulation Number: | CFR 888.3045 |
| Product Code: | MQV |
| Device Classification | II |
#### III. Predicate Device
Substantial equivalence is claimed to the following device:
- MASTERGRAFT® Putty, K071813, Medtronic Sofamor Danek ●
The following reference devices are cited within the submission:
- Actifuse ABX E-Z-fil Bone Graft Substitute, K081979, ApaTech Limited
- OsteoFlo® NanoPutty® Quadphasic Synthetic Bone Graft, K200064, SurGen Tec LLC ●
- Geistlich Bio-Gide, K192042, Geistlich Pharma AG
{5}------------------------------------------------
#### IV. Device Description
Device 300397 Putty is an osteoconductive, porous, resorbable bone graft substitute which consists of calcium phosphate granules (α-ΤCP and hydroxyapatite) and Type I and III porcine collagen. The single-patient, single-use product is supplied in three (3) sizes as a sterile, dry block. When mixed with autogenous bone marrow aspirate, blood or sterile non-pyrogenic saline, the product forms a moldable non-setting putty (net volume: 2.5 cc, 5 cc or 10 cc).
#### Indications for Use V.
Device 300397 Putty is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. Device 300397 Putty is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., posterolateral spine, ilium and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process.
Device 300397 Putty must be used with blood or autogenous bone marrow aspirate and autograft in posterolateral spine.
#### VI. Comparison of Technological Characteristics
Device 300397 Putty is substantially equivalent to the predicate device in terms of intended use and technological characteristics. There are slight differences between the subject and predicate device with respect to the indications for use statement; however, these differences do not alter the overall intended use of the subject device. Both devices are resorbable bone void fillers intended to fill bony voids or gaps of the skeletal system that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Device 300397 Putty and the predicate device are osteoconductive materials which utilize the same ratio of inorganic to organic material, are sterilized via irradiation, and are biologically safe. The subject device consists of calcium phosphate granules and porcine collagen. The calcium phosphate granules (0.5 - 2 mm) are comprised of a synthetic, porous, biphasic a-TCP/hydroxyapatite material. The use of a different TCP polymorph in the subject device does not raise different questions of safety and effectiveness as the polymorphs share the same structural composition (i.e., the crystal structures are similar and have almost identical unit cell size and volume) and chemical stoichiometry.
The porcine collagen utilized in Device 300397 Putty is used in Geistlich Pharma AG's dental products and the identified raw material sources were previously cleared by the FDA.
{6}------------------------------------------------
The granule size for Device 300397 Putty is slightly larger than that of the predicate device; however, as demonstrated through performance testing, there are no new or different concerns raised by the larger size.
As seen above. Device 300397 Putty is the same or similar to the predicate device in terms of intended use, sterilization and biological safety profile. The technological differences between the subject and predicate device do not raise new questions of safety and effectiveness and the data provided within this submission support that Device 300397 Putty is substantially equivalent to the identified predicate device.
#### Safety & Performance Data VII.
The following studies were successfully executed.
- Characterization of chemical properties ●
- 0 Amino Acid Composition
- Product composition (Residue on ignition) o
- Phase Analysis by X Ray Diffraction O
- Characterization by FT-IR O
- SEM O
- Molecular Weight Distribution of Soluble Protein o
- Onset Temperature and Enthalpy of Protein Denaturation by Differential O Scanning Calorimetry
- Enzymatic Collagen Degradation o
- Characterization of physical properties ●
- Pore Size and Pore Size Distribution O
- Particle Size Distribution O
- Weight, Dimension, and Density O
- o Determination of Hydrated Volume
- Material Handling Studies ●
- . Sterilization Validation per ISO 11137-1:2006, ISO 11137-2:2013, and ISO 11137-3:2017
- Packaging Validation per ISO 11607-1:2019, ASTM F1980:2007, ASTM . F1886/F1886M:2016, ASTM F88:2015, ASTM F1929:2015, and ASTM F2096:2011
- Product stability testing per ICH Q1A(R2) ●
- Transport Simulation per ISTA 3A
- Biocompatibility Studies per ISO 10993-1:2018 ●
- Cytotoxicity per ISO 10993-5:2009 O
- O Sensitization per ISO 10993-10:2010
{7}------------------------------------------------
- Irritation/Intracutaneous Reactivity per ISO 10993-10:2010 O
- Acute Systemic Toxicity per ISO 10993-11:2017 O
- Material Mediated Pyrogenicity per USP <151> / Ph. Eur. Section 2.6.8 o
- Subacute Systemic Toxicity per ISO 10993-6:2016 and ISO 10993-11:2017 O
- Subchronic Systemic Toxicity per ISO 10993-6:2016 and ISO 10993-11:2017 O
- Genotoxicity per ISO 10993-3:2014 O
- Implantation (local tissue reaction) per ISO 10993-6:2016 o
- Hemocompatibility per ASTM F756:2017 and ISO 10993-4:2017 O
- Chronic Systemic Toxicity per ISO 10993-11:2017 O
- Elemental Impurities Analysis ICH Q3D (R2):2022 O
In addition, the safety and performance of Device 300397 Putty was evaluated as a bone graft extender in the NZ White Rabbit Single Level Spinal Fusion Model. The performance of the device was evaluated in accordance with the requirements of the FDA guidance on calcium salt bone void filler guidance while the local tissue reaction was evaluated in accordance with the requirements of ISO 10993-6:2016.
Device 300397 Putty was hydrated with bone marrow aspirate (BMA) or with blood and mixed 1:1 with autograft; the predicate device was hydrated with BMA and mixed 1:1 with autograft. The graft material was placed adjacent to the vertebral body as well as between decorticated transverse processes. Progression in healing versus time was observed when implanted on decorticated transverse processes in the posterolateral space. Further healing was observed at the host decorticated transverse process interfaces with new bone formation with active osteoblasts present directly on the surface of the graft material and autograft material. New bone formation directly on the surfaces of graft material and autograft was also observed. The progression and maturation of the fusions treated with Device 300397 Putty hydrated with BMA or blood were confirmed via histology. Bone remodeling and formation of new marrow spaces was seen at the decorticated host transverse process interface as new bone formation on and through the graft material which was associated with a reduction in local inflammatory cells.
## VIII. Conclusion
The information provided above supports that Device 300397 Putty is as safe and effective as the predicate device. Although minor differences in design and technology exist between the subject and predicate device, performance testing demonstrates that these differences do not raise any new questions of safety and effectiveness. Therefore, it is concluded that Device 300397 Putty is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.