OsteoFlo NanoPutty-Quadphasic Synthetic Bone Graft
Applicant
SurGenTec, LLC
Product Code
MQV · Orthopedic
Decision Date
Aug 14, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
OsteoFlo® NanoPutty® - Quadphasic Synthetic Bone Graft is indicated to fill bony voids or gaps of the skeletal system (i.e. the extremities and pelvis) that are not intrinsic to the stability of the bony structure. These osseous defects may be surgically created or the result of traumatic injury to the bone. The device resorbs and is replaced with bone during the healing process.
Device Story
OsteoFlo NanoPutty is an osteoconductive, non-hardening bone void filler; composed of macroporous quadphasic calcium phosphate particulates (HA, α-TCP, β-TCP, bioactive glass) and micro-sized HA particles in a bioresorbable polymer binder. Device is supplied sterile, ready-to-use, and requires no mixing. Applied by clinicians into osseous defects; binder resorbs post-implantation; particles act as scaffold for new bone formation; particles gradually resorb. Benefits include ease of packing and adherence to bone surfaces; facilitates bone regeneration in non-structural voids.
Clinical Evidence
Bench testing only. Included biocompatibility (ISO 10993-1), sterilization validation (ISO 11137), packaging validation (ISO 11607), shelf-life (ASTM 1980), and material characterization. In vivo evaluation in a critical-size rabbit femoral defect model compared performance to NovaBone Putty, showing equivalent bone formation and no adverse events.
Technological Characteristics
Quadphasic calcium phosphate (HA, α-TCP, β-TCP, bioactive glass) and micro-sized HA particles in bioresorbable polymer binder. Sterile (E-beam, SAL 10^-6), single-use. Dimensions: 1.0 ml – 10 ml. Standards: ISO 10993-1, ISO 11137-1/2, ISO 11607-1/2, ASTM 1980-16.
Indications for Use
Indicated for patients with bony voids or gaps in the skeletal system (extremities and pelvis) not intrinsic to structural stability, resulting from surgery or traumatic injury.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}------------------------------------------------
August 14, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
SurGenTec, LLC Travis Greenhalgh CEO 911 Clint Moore Rd Boca Raton, Florida 33487
Re: K200064
Trade/Device Name: OsteoFlo® NanoPutty® - Quadphasic Synthetic Bone Graft Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV Dated: July 23, 2020 Received: July 24, 2020
Dear Mr. Greenhalgh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Laura C. Rose, Ph.D. Acting Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200064
Device Name
OsteoFlo® NanoPutty®- Quadphasic Synthetic Bone Graft
Indications for Use (Describe)
OsteoFlo® NanoPuty - Quadphasic Synthetic Bone Graft is indicated to fill bony voids or gaps of the skeletal system (i.e. the extremities and pelvis) that are not intrinsic to the stability of the bony structure. These osseous defects may be surgically created or the result of traumatic injury to the bone. The device resorbs and is replaced with bonedus ing the healing process.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other spect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
EF
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Surgentec logo. The logo consists of a blue and gray circular design on the left, followed by the word "SURGENTEC" in bold, black letters. A registered trademark symbol is located in the upper right corner of the word.
# Premarket Notification 510(k) Summary OsteoFlo NanoPutty Bone Void Filler - K200064
#### Submitter Information: 1.
| Owner: | SurGenTec, LLC |
|-----------------|--------------------------------------------|
| Address: | 911 Clint Moore Rd<br>Boca Raton, FL 33487 |
| Contact: | Travis Greenhalgh, CEO |
| Telephone: | (916) 759-7999 |
| Email: | travis@SurGenTec.com |
| Date Submitted: | December 9, 2019 |
#### 2. Name of Device:
| Trade Name: | OsteoFlo® NanoPutty® - Quadphasic Synthetic Bone Graft |
|----------------------|--------------------------------------------------------|
| Common Name: | Bone Void Filler |
| Classification Name: | Resorbable Calcium Salt Bone Void Filler |
| Device Regulation: | 21 CFR 888.3045 |
| Regulatory Class: | Class II |
| Product Code: | MQV |
#### 3. Legally Marketed Predicate Device:
Predicate: NovaBone Putty - Bioactive Synthetic Bone Graft NovaBone Products, LLC K110368, K101860, K082672
The following devices also are cited as reference devices within the submission:
- Bioactive Bone Graft Putty: K132071; Biostructures, LLC
- NovaBone MacroPor-Si+: K110925; NovaBone Products, LLC
- NanOss BVF-E: K081558; Pioneer Surgical Technology
#### Device Description 4.
The OsteoFlo® NanoPutty® - Quadphasic Synthetic Bone Graft is an osteoconductive, non-hardening bone void filler. The device is comprised of macroporous calcium phosphate particulates in a bioresorbable polymer binder. The quadphasic particles are composed of HA, a-TCP, B-TCP and bioactive glass. The device also contains micro-sized HA particles. On implantation, the binder is resorbed and bone forms on and between the porous particles as they are gradually resorbed.
{4}------------------------------------------------
OsteoFlo® NanoPutty can be easily packed into osseous defects and adheres to bone surfaces. The device is ready to use, requiring no mixing before application. The single-use device is supplied sterile in multiple package formats.
#### 5. Indications for Use
The OsteoFlo® NanoPutty® - Quadphasic Synthetic Bone Graft is indicated to fill bony voids or gaps of the skeletal system (i.e. the extremities and pelvis) that are not intrinsic to the stability of the bony structure. These osseous defects may be surgically created or the result of traumatic injury to the bone. The device resorbs and is replaced with bone during the healing process.
#### Technological Characteristics and Substantial Equivalence 6.
OsteoFlo® NanoPutty - Ouadphasic Synthetic Bone Graft is substantially equivalent to the predicate NovaBone Putty Bioactive Synthetic Graft. The devices have the same intended use and basic method of operation. The table below compares the indications and key technological attributes of the OsteoFlo NanoPutty device to the predicate.
| Comparator | New Device | Predicate |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | OsteoFlo NanoPutty | NovaBone Putty [K110368] |
| Product Code;<br>Common Name | Resorbable Calcium Salt Bone Void Filler Device<br>(21 CFR 888.3045) | |
| Indications for<br>Use | OsteoFlo® NanoPutty –<br>Quadphasic Synthetic Bone<br>Graft is indicated to fill bony<br>voids or gaps of the skeletal<br>system (i.e. the extremities and<br>pelvis) that are not intrinsic to<br>the stability of the bony<br>structure. These osseous defects<br>may be surgically created or the<br>result of traumatic injury to the<br>bone. The device resorbs and is<br>replaced with bone during the<br>healing process. | NovaBone Putty is indicated for bony<br>voids or gaps that are not intrinsic to<br>the stability of the bony structure.<br>NovaBone Putty is indicated to be<br>gently packed into bony voids or gaps<br>of the skeletal system (i.e. the<br>extremities, posterolateral spine, and<br>pelvis). These defects may be<br>surgically created osseous defects or<br>osseous defects created from traumatic<br>injury to the bone. The product<br>provides a bone void filler that resorbs<br>and is replaced with bone during the<br>healing process. |
| Intended Use | Both devices are intended for use as bone void fillers for repair of defects<br>in the skeletal system. The devices are intended to be resorbed during<br>defect healing and replaced with new bone tissue. | |
| Mechanism of<br>Operation | The device is applied as a pre-<br>mixed putty. The synthetic<br>binder is resorbed to expose<br>quadphasic particles. The<br>particles act as osteoconductive<br>scaffolds for new bone<br>formation as they are slowly<br>resorbed | The device is applied as a pre-mixed<br>putty. The synthetic binder is resorbed<br>to expose bioactive glass particles.<br>The particles act as osteoconductive<br>scaffolds for new bone formation as<br>they are slowly resorbed |
| Material<br>Composition | Quadphasic macroporous<br>calcium phosphate particulates<br>in a bioresorbable polymer | Bioactive glass particles, 90-710 μm,<br>in a bioresorbable polymer binder |
{5}------------------------------------------------
| Comparator | New Device<br>OsteoFlo NanoPutty | Predicate<br>NovaBone Putty [K110368] |
|--------------|----------------------------------------------------|-----------------------------------------------|
| | binder | |
| Sterility | Sterile; SAL 10-6;<br>E-beam radiation; Single-Use | Sterile, SAL 10-6;<br>γ radiation; Single-Use |
| Product Size | 1.0 ml – 10 ml | 2.5 ml – 15 ml |
#### 7. Performance Data
The following non-clinical performance data were provided to demonstrate substantial equivalence of the subject device to the predicate.
- Biocompatibility per ISO 10993-1:2018
- . Sterilization validation per ISO 11137-1:2006 and ISO 11137-2:2013
- Packaging validation per ISO 11607-1:2009 and ISO 11607-2:2006 ●
- Shelf-life testing per ASTM 1980-16 ●
- Material characterization, including x-ray diffraction, particle size, and ● particle porosity and surface area
- . In vivo evaluation in a critical-size rabbit femoral defect model
The in vivo evaluation compared the performance of the device to the NovaBone Putty predicate. Evaluation of local biological effects and bone formation demonstrated performance substantially equivalent to the predicate, with no evidence of adverse events or device related failures.
#### 8. Conclusion
The design characteristics of the OsteoFlo NanoPutty device do not raise different questions of safety and effectiveness. Non-clinical study data supports that the device is safe and effective. These data support that the OsteoFlo NanoPutty device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.