Bioactive Bone Graft Putty is a bone void filler device intended for use in bony voids or gaps that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Bioactive Bone Graft Putty is indicated to be packed gently into bony voids or gaps of the skeletal system (i.e., extremities, pelvis and posterolateral spine fusion procedures). Bioactive Bone Graft Putty can also be used with autograft as a bone graft extender in the posterolateral spine. The device provides a bone void filler that is resorbed and replaced with host bone during the healing process.
Device Story
Synthetic bone void filler; mixture of calcium phosphate granules and bioactive glass granules in resorbable polymer carrier. Supplied as putty in syringe applicator; sterile; single-use. Used by surgeons in clinical settings to fill surgically created or traumatic osseous defects. Acts as osteoconductive matrix; resorbed and replaced by host bone during healing. Bioactivity confirmed via in vitro apatite layer formation in simulated body fluid. Performance evaluated in rabbit femoral cancellous defect and posterolateral spine fusion models; demonstrated equivalent safety, resorption, and bone formation compared to predicate devices.
Clinical Evidence
No human clinical data. Evidence based on non-clinical bench testing and animal performance studies. Bench testing included raw material characterization (ASTM F1538-03, F1185-03, F1088-04a, F1929/F1926M), biocompatibility (ISO 10993), and packaging/sterilization/shelf-life (ISO 11607, ISO 11137, ASTM F2096-11, F88-09, F1929-98, F1980). In vitro studies demonstrated apatite layer formation. Animal studies in rabbit femoral cancellous defect and posterolateral spine fusion models showed equivalent safety and bone formation compared to predicate.
Technological Characteristics
Synthetic bone void filler; calcium phosphate and bioactive glass granules in resorbable polymer carrier. Supplied in syringe applicator. Sterilized via ISO 11137. Materials conform to ASTM F1538-03, F1185-03, F1088-04a. Biocompatible per ISO 10993. Non-electronic, non-software device.
Indications for Use
Indicated for patients requiring bone void filling in non-structural bony defects of the skeletal system, including extremities, pelvis, and posterolateral spine. Can be used as a bone graft extender with autograft in the posterolateral spine.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}------------------------------------------------
K 132071
Page 1 of 2
Submitter:
BioStructures, LLC
Bioactive Bone Graft Putty
# 510(k) SUMMARY
| Submitter Name: | BioStructures, LLC |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Address: | 1201 Dove Street, Suite 470<br>Newport Beach, CA 92660<br>JAN 10 2014 |
| Contact Person: | John Brunelle, Ph.D.<br>Director of R&D/ Manufacturing |
| Phone Number: | 949.553.1717 |
| Date Prepared: | July 2, 2013 |
| Device Trade Name: | Bioactive Bone Graft Putty |
| Device Classification: | Class II |
| Classification Name: | Filler, Bone Void, Calcium Compound |
| Classification Number: | 21 CFR 888.3045 |
| Product Code: | MQV |
| Predicate Device(s): | K071206, Actifuse® ABX, ApaTech Limited<br>K043005, MBCP™™, Biomatlante<br>K112857, Interface Bone Void Filler, BioStructures, LLC<br>K020078, Exactech Resorbable Bone Paste, Exactech, Inc. |
| Indications for Use<br>Statement: | Bioactive Bone Graft Putty is a bone void filler device intended for<br>use in bony voids or gaps that are not intrinsic to the stability of<br>the bony structure. These defects may be surgically created<br>osseous defects or osseous defects created from traumatic injury<br>to the bone. Bioactive Bone Graft Putty is indicated to be packed<br>gently into bony voids or gaps of the skeletal system (i.e.,<br>extremities, pelvis and posterolateral spine fusion procedures).<br>Bioactive Bone Graft Putty can also be used with autograft as a<br>bone graft extender in the posterolateral spine. The device<br>provides a bone void filler that is resorbed and replaced with host<br>bone during the healing process. |
| Device Description: | Bioactive Bone Graft Putty is a synthetic bone void filler comprised<br>of a mixture of calcium phosphate granules and bioactive glass<br>granules suspended in a resorbable polymer carrier that facilitates<br>handling and delivery of the granule components. The device is<br>supplied as putty, pre-loaded in a syringe applicator and packaged<br>in a sterile barrier foil pouch. The device is provided sterile, for<br>single use, in a variety of sizes. |
| Summary of Testing: | Non-clinical testing was performed in accordance with FDA<br>recognized consensus standards and FDA guidance documents<br>as applicable.<br>Raw materials characterization testing was performed according<br>to ASTM standards: F1538-03, F1185-03, F1088-04a,<br>F1929/F1926M and USP NF. |
.
{1}------------------------------------------------
| Submitter: | BioStructures, LLC | |
|------------|--------------------|--|
|------------|--------------------|--|
Bioactive Bone Graft Putty Traditional 510(k)
Biocompatibility testing, according to ISO 10993 for a long-term implant product. demonstrated the device is biocompatible and non-toxic.
Packaging, sterilization and shelf life testing according to the following standards were presented in the 510(k):
- Packaging: ISO11607, ASTM F2096-11, F88-09 and . F1929-98; and ISO 11137.
- Sterilization and Shelf Life; ISO 11137 and ASTM F1980. .
The device is considered bioactive based on in vitro studies that show apatite laver formation on the surface of the implant following immersion in simulated body fluid (SBF). These results have not been correlated to clinical performance.
Animal performance testing was performed in femoral cancellous defect and posterolateral spine fusion (PLF) rabbit models to evaluate the safety and performance of the Bioactive Bone Graft Putty compared to a predicate device. The test results showed equivalent in vivo performance in safety, graft resorption and new bone formation.
Bioactive Bone Graft Putty has the same intended use and the same principles of operation as all the predicates, which serve as osteoconductive matrices for new bone formation.
Bioactive Bone Graft Putty is similar technologically to the four predicates.
The technological differences presented by the composition of materials in Bioactive Bone Graft Putty do not raise new issues of safety or effectiveness, as demonstrated by the side-by-side evaluation in the animal performance studies.
The comparisons and study data presented in the 510(k) lead to the conclusion that Bioactive Bone Graft Putty is substantially equivalent to the predicate devices.
Comparison to the Predicate Devices:
Substantjal Equivalence Conclusion:
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle with three overlapping wings, representing health, services, and people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the eagle symbol in a circular fashion.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 10, 2014
BioStructures, LLC % Patsy J. Trisler, JD, RAC Trisler Consulting 5600 Wisconsin Avenue #509 Chevy Chase, Maryland 20815
Re: K132071
Trade/Device Name: Bioactive Bone Graft Putty Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MOV Dated: December 2, 2013 Received: December 3, 2013
Dear Ms. Trisler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
## Page 2 - Patsy J. Trisler. JD, RAC
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
# Ronalden - Jean - S for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known):
K132071
Device Name:
Bioactive Bone Graft Puttv
### Indications for Use:
Bioactive Bone Graft Putty is a bone void filler device intended for use in bony voids or gaps that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Bioactive Bone Graft Putty is indicated to be packed gently into bony voids or gaps of the skeletal system (i.e., extremities, pelvis and posterolateral spine fusion procedures). Bioactive Bone Graft Putty can also be used with autograft as a bone graft extender in the posterolateral spine. The device provides a bone void filler that is resorbed and replaced with host bone during the healing process.
Prescription Use X
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
### Concurrence of CDRH, Office of Device Evaluation (ODE)
Laurence D. Coyne -S
(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K132071
Section 4.0
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.