This device is used to dilate strictures in the pancreatobiliary systems and to dilate openings via the transgastric or transduodenal wall. This device is indicated for adult use only.
Device Story
Tornus ES is a rotary-operated, screw-type dilator for pancreatobiliary strictures and transgastric/transduodenal openings. Device consists of stainless-steel dual-coil shaft and dilation segments, coated with resin and hydrophilic material, with a tapered metal tip. Physician advances device by rotating grip, utilizing spiral-wound coil design to screw through target site; tapered shape expands stricture. Used in clinical settings by physicians. Benefits include effective dilation of anatomical strictures. Device is provided sterile via ethylene oxide.
Clinical Evidence
Bench testing only. Testing included appearance, tensile strength, guidewire pass-through, dilation ability, slide durability, radio-detectability, corrosion resistance, polyurethane strength, trackability, kink resistance, and simulated use/torsional strength. All tests met acceptance criteria. Biocompatibility testing performed per ISO 10993 (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity) confirmed non-cytotoxic, non-sensitizing, non-irritant, and non-pyrogenic status.
Technological Characteristics
Materials: Stainless steel, polyurethane elastomer, acrylic resin, PVP, nitrocellulose, polypropylene. Design: Dual-coil (rope/wire) rotary dilator with tapered metal tip. Dimensions: 7 Fr (2.62 mm) O.D., 195 cm length, 0.018" or 0.025" guidewire compatibility. Energy: Manual mechanical rotation. Sterilization: Ethylene Oxide (ISO 11135).
Indications for Use
Indicated for adult patients requiring dilation of pancreatobiliary strictures or dilation of openings via the transgastric or transduodenal wall.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 17, 2024
ASAHI Intecc Co., Ltd. % Candace Cederman Principal Consultant CardioMed Device Consultants 1783 Forest Drive Suite 254 Annapolis, Maryland 21401
Re: K241801
Trade/Device Name: Tornus ES Regulation Number: 21 CFR 876.5010 Regulation Name: Biliary Catheter And Accessories Regulatory Class: Class II Product Code: FGE Dated: November 15, 2024 Received: November 18, 2024
Dear Candace Cederman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{3}------------------------------------------------
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Anthony Lee -S
Anthony Lee, Ph.D., M.B.A. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K241801
Device Name
Tornus ES
Indications for Use (Describe)
This device is used to dilate strictures in the pancreatobiliary systems and to dilate openings via the transgastric or transduodenal wall. This device is indicated for adult use only.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
#### 510(k) Summary [as required by 21 CFR 807.92(c)]
# Tornus ES 510(k) K241801
| DATE PREPARED: | December 10, 2024 |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT | ASAHI INTECC CO., LTD.<br>3-100 Akatsuki-cho, Seto<br>Aichi 489-0071<br>Japan |
| CONTACT | Yoshi Terai<br>President/CEO<br>ASAHI INTECC USA, INC.<br>3002 Dow Avenue, Suite 212<br>Tustin, CA 92780 US<br>Tel: (949) 756-8252, FAX: (949) 756-8165<br>e-mail: ASAHI.ra-fda@ASAHI-intecc.com |
| TRADE NAME: | Tornus ES |
| DEVICE CLASSIFICATION: | Class II per 21 CFR §876.5010 |
| CLASSIFICATION NAME: | Biliary catheter and accessories |
| PRODUCT CODE | FGE |
| PREDICATE DEVICE: | Soehendra Biliary Dilatation Catheter (K171548) |
| REFERENCE DEVICE: | Cook Dilator Sets (K183036)<br>Zimmon Pancreatic Stent Sets (K172057)<br>Soehendra Stent Retriever (K161203)<br>Cysto-Gastro Sets (K211909) |
## INDICATIONS FOR USE:
This device is used to dilate strictures in the pancreatobiliary systems and to dilate openings via the transgastric or transduodenal wall. This device is indicated for adult use only.
{6}------------------------------------------------
# DETAILED DESCRIPTION
The Tornus ES is a rotary-operated dilator designed for use in bile duct/pancreatic duct strictures and via transgastric/transduodenal dilation of openings.
The Tornus ES is comprised of stainless-steel coils and polymeric materials. The Dilation and Shaft segments are composed of coils materials that are welded together.
The Tornus ES is available for prescription use only.
## DILATION SEGMENT
The coil of the Dilation portion of the device is a dual coil design consisting of an upper laver Wire coil and a Rope coil under layer. The distal side of the dilation segment is tapered, and the surface is coated with Resin and a Hydrophilic coating.
A tapered metal tip is welded to the tip of the coil. This metal tip is available in two applicable guidewire diameters: 0.018 inch and 0.025 inch.
## SHAFT
The Shaft coil is also a dual coil design consisting of a rope coil in both the upper and under lavers.
The proximal end of the coil is coated with polyurethane, and a grip is bonded to it using an adhesive.
## Principal of Operation
The Tornus ES is advanced as the physician rotates the grip, in a fashion similar to that of a screw, due to the spiral-wound wire coil design. Once the target site is reached, the site is expanded via the outer diameter of the device's tapered shape.
## COMPARISON WITH PREDICATE DEVICES:
The subject device has similar intended use and indication for use as compared with the predicate device. The subject device has similar technological characteristics as the predicate devices. While differences exist between the subject and predicate device, with respect to the access point, those differences do not raise different/new questions of safety and/or effectiveness
{7}------------------------------------------------
# Comparison with Predicate Device
| | Subject | | Predicate | | | Reference | | Subject | Predicate | Reference | | | |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|-----------------------------------------------------------------------------------------------------------|------------------------------------------------|------------------------------------------------------------------|-----------------------------------------------------------|---------------------------------------------------------------------|---------------------------------------------------------------------------------|
| Device Name | Tornus ES | Soehendra Biliary<br>Dilation Catheter | Cook Dilator Sets | Zimmon<br>Pancreatic Stent<br>Sets | Soehendra Stent<br>Retriever | Cysto-Gastro Sets | Device Name | Tornus ES | Soehendra Biliary<br>Dilation Catheter | Cook Dilator Sets | Zimmon<br>Pancreatic Stent<br>Sets | Soehendra Stent<br>Retriever | Cysto-Gastro Sets |
| Manufacturer | ASAHI INTECC | Wilson-Cook Medical | Cook Incorporated | Cook Ireland Ltd. | Wilson-Cook<br>Medical | G-Flex Europe SPRL | Design<br>Feature/Main<br>Components | Rope coil<br>Resin coating<br>Hydrophilic coating<br>Metal tip<br>Grip | Catheter shaft<br>Marker ring<br>Luer lock hub | Dilator shaft,<br>coating<br>(hydrophilic),<br>guidewire<br>port | Stent Introducer | Handle<br>Rope coil<br>Tapered threaded<br>screw type distal<br>tip | Metal Tip<br>Needle<br>(Depends on lineup)<br>Catheter Handle<br>with electrode |
| 510(k) | K241801 | K171548 | K183036 | K172057 | K161203 | K211909 | Material | Stainless Steel,<br>Polyurethane<br>elastomer, Acrylic<br>resin, PVP,<br>Nitrocellulose,<br>Polypropylene | Unknown | Resin<br>Hydrophilic coating | Polyethylene<br>EVA | Stainless Steel<br>Resin | Stainless Steel,<br>Teflon,<br>POM + MABS |
| Classification<br>Panel | Gastroenterology/<br>Urology | Gastroenterology/<br>Urology | Cardiovascular/<br>Gastroenterology/<br>Urology | Gastroenterology/<br>Urology | Gastroenterology/<br>Urology | Gastroenterology/<br>Urology | O.D.<br>(Dilation)^ | 7 Fr<br>(2.62 mm) | 6 — 10 Fr<br>(2.0 - 3.3 mm) | 3.0 - 26.0 Fr<br>(1.0 - 8.7 mm) | 3 - 11.5 Fr<br>(1.66 - 2.66 mm) | 8 Fr<br>(2.70 mm) | 6, 8.5 and 10 Fr<br>(2.0 - 3.3 mm) |
| Regulation | 21 CFR 876.5010 | 21 CFR 876.5010 | 21 CFR 870.1310<br>21 CFR 876.5010 | 21 CFR 876.5010 | 21 CFR 876.5010 | 21 CFR 876.4300 | Effective<br>Length | 195 cm | 200 cm | 6 – 65 cm | Stent length:<br>2 – 15 cm<br>Introducer length:<br>170cm | 168 cm | 180 and 210 cm |
| Common<br>Name | Dilatation Catheter | Dilatation Catheter | Dilator, vessel, for<br>percutaneous<br>catheterization /<br>stents, drains and<br>dilators for the<br>biliary ducts | Drainage Catheter | Stent Retriever | Endoscopic<br>electrosurgery device | Guidewire<br>Diameter (in) | 0.018" | 0.035" | Unknown | 0.035" | 0.018" | 0.035" |
| Product Code | FGE | FGE | DRE, FGE | FGE | FGE | KNS | | 0.025" | | | | 0.035" | |
| Class | Class II | Class II | Class II | Class II | Class II | Class II | Sterilization | EO | Unknown | EO | EO | EO | EO |
| Indication for<br>Use | This device is used<br>to dilate strictures in<br>the pancreatobiliary<br>systems and to<br>dilate openings via<br>the transgastric or<br>transduodenal wall.<br>This device is<br>indicated for adult<br>use only. | Soehendra Biliary<br>Dilatation Catheters<br>are used to dilate<br>biliary strictures.<br>These devices are<br>indicated for adult<br>use only | Intended to be<br>used for dilating<br>puncture sites or<br>catheter tracts for<br>percutaneous<br>placement of<br>devices for vascular<br>and non-vascular<br>applications such<br>as in the venous,<br>arterial, biliary and<br>renal systems. | Zimmon Pancreatic<br>Stent sets are used<br>to drain obstructed<br>pancreatic ducts | Used to remove<br>stents from the<br>biliary and<br>pancreatic duct(s)<br>while maintaining<br>wire guide<br>placement | Cysto-Gastro Sets<br>are intended to be<br>used to<br>electrosurgically<br>cannulate pancreatic<br>pseudo-cysts<br>endoscopically (via<br>the transgastric or<br>transduodenal wall)<br>as an alternative to<br>surgical or<br>percutaneous<br>treatment | Shelf Life | 3 years | Unknown | 3 years | 3 years | 3 years | 3years |
{8}------------------------------------------------
^ The shaft O.D of the Tornus ES is 2.64mm, with a Tip ID of either 0.018" or 0.025"
{9}------------------------------------------------
#### Non Clinical testing / Performance Data:
| Test item | Test Results |
|----------------------------------------|--------------|
| Appearance | Pass |
| Tensile strength | Pass |
| Guidewire pass-through ability | Pass |
| Dilation ability | Pass |
| Slide durability | Pass |
| Radio-detectability | Pass |
| Corrosion resistance | Pass |
| Polyurethane strength | Pass |
| Guide wire trackability | Pass |
| Kink resistance | Pass |
| Dimension measurement | Pass |
| Simulated use and (torsional strength) | Pass |
Non clinical laboratory testing was performed on the Tornus ES to determine substantial equivalence. The following testing/assessments were performed:
The in vitro bench tests demonstrated that the Tornus ES met all acceptance criteria. Performance data demonstrate that the device functions as intended and is substantially equivalent to the predicate and reference devices.
## BIOCOMPATIBILITY:
The Tornus ES was tested in accordance with ISO 10993 and found to be biocompatible. The following tests were performed:
| Test Method | Standard | Acceptance Criteria | Results |
|-----------------------------------------------------------------|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|
| Cytotoxicity<br>MEM Elution Test | ISO 10993-5<br>No deviations | The test system is considered suitable if no signs of<br>cellular reactivity (Grade 0) are noted for both the<br>negative control article and the medium control. | Non-cytotoxic |
| Sensitization<br>KLIGMAN<br>Maximization Test | ISO 10993-10<br>No deviations | The extracts should show no evidence of causing<br>delayed dermal contact sensitization in the guinea pig. | Non-Sensitizing |
| Irritation<br>Intracutaneous<br>Injection Test | ISO 10993-10<br>No deviations | The test extract and the negative control must exhibit<br>similar edema and erythema scores. | Non-Irritant |
| Systemic Toxicity<br>Acute System<br>Toxicity Test | ISO 10993-11<br>No deviations | The test article must not show significantly greater<br>biological activity than the control. | Non-toxic |
| Systemic Toxicity<br>Rabbit Pyrogen Test<br>(material mediated) | ISO 10993-11<br>No deviations | The test article should not increase the rectal<br>temperature of any of the animals by more than 0.5<br>degrees Celsius. | Non-pyrogenic |
{10}------------------------------------------------
#### Sterilization and Shelf Life:
The Tornus ES sterilization process using Ethylene Oxide (EO) has been validated in accordance with ISO 11135: 2014 to achieve a sterility assurance level (SAL) of 10-9. EO and Ethylene Chlorohydrin (ECH) residuals were below the limits specified in ISO 10993-7:2008.
Both baseline and accelerated shelf-life testing were conducted demonstrating the device will perform as intended to support the proposed 3-year shelf-life.
#### CONCLUSION:
The Tornus ES have the similar intended use and indications, and the same or similar technological characteristics such as design, materials, sterilization method, performance, and operating principles as the predicate and reference device. Performance data demonstrates that the device functions as intended.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.