The Outlook Surgical Versa One System is intended for use in examination and visualization in otolaryngology and head and neck procedures.
Device Story
Video nasopharyngoscopy system for otolaryngology and head/neck procedures; used in OR, clinics, or medical offices by clinical practitioners. System includes endoscope with semi-flexible shaft, connection cable, wireless display unit, and wireless HDMI receiver. User attaches instruments to endoscope shaft; flexible portion becomes rigid upon attachment. Integrated LED light source illuminates glass fiber bundles; distal CMOS sensor captures images; custom PCB in handle converts signal to live video. Video stream displayed on handheld unit or external monitor via HDMI/wireless. Device powered by rechargeable Li-ion battery or AC. Enables visualization of internal anatomy during procedures; aids clinical decision-making by providing real-time anatomical views. Reusable device requires manual cleaning and APS STERRAD sterilization between uses.
Clinical Evidence
Bench testing only. Includes simulated use/cycling, optical performance (ISO 8600-X, 12233, 15739), electrical safety/EMC (IEC 60601-1/2-18, 60601-1-2), RF safety (IEC 63195-1, ANSI/IEEE C95.1), photobiological safety (IEC 62471), cleaning validation (AAMI TIR12/ST98), biocompatibility (ISO 10993), and sterilization validation (ISO 22441, ISO 11135). User validation performed per IEC 62366-1 using anatomical and cadaver models confirmed usability.
Technological Characteristics
Materials: 304 stainless steel, glass, borosilicate, crosslinked PVDF, epoxy. Sensing: Distal CMOS sensor. Illumination: Integrated LED with glass fiber bundles. Power: Rechargeable Li-ion battery or AC. Connectivity: Wireless HDMI/data transfer. Form factor: 3.0mm/3.8mm diameter, 102-105mm length. Sterilization: Manual cleaning + APS STERRAD. Software: Embedded firmware for video processing/transmission.
Indications for Use
Indicated for examination and visualization in otolaryngology and head and neck procedures for patients requiring such clinical assessment.
Regulatory Classification
Identification
A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 11, 2025
ResnENT, LLC
% Billi-Jo Pfalzgraf
Regulatory Consultant
MedEdge Consulting
7005 Evergreen Pl
Roswell, Georgia 30076
Re: K241731
Trade/Device Name: Outlook Surgical Versa One System (8900139)
Regulation Number: 21 CFR 874.4760
Regulation Name: Nasopharyngoscope (Flexible Or Rigid) And Accessories
Regulatory Class: Class II
Product Code: EOB
Dated: July 29, 2025
Received: July 29, 2025
Dear Billi-Jo Pfalzgraf:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K241731 - Billi-Jo Pfalzgraf
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K241731 - Billi-Jo Pfalzgraf
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Joyce C. Lin -S
for Shu-Chen Peng, Ph.D.
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K241731
Device Name
Outlook Surgical Versa One System (8900139)
Indications for Use (Describe)
The Outlook Surgical Versa One System is intended for use in examination and visualization in otolaryngology and head and neck procedures.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
> "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
Proprietary and Confidential
Page 1 of 7
# Outlook Surgical Versa One System 510(k)
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
Date: August 8, 2025
## 1.0 SUBMITTER:
ResnENT, LLC
3210 Mulligan Drive
Bloomington, IL 61704
## 1.1 PRIMARY CONTACT PERSON:
Billi-Jo Pfalzgraf
Regulatory Consultant
MedEdge Consulting
(704) 840-8453
bjpfalzgraf@mededge.io
## 1.2 SECONDARY CONTACT PERSON:
Tiffini Wittwer
Regulatory Consultant and Owner
MedEdge Consulting
(707) 799-6732
twittwer@mededge.io
## 2.0 DEVICE:
Trade Name: Outlook Surgical Versa One System
Common/Usual Name: Flexible Nasopharyngoscope System
Classification Name: Nasopharyngoscope (flexible or rigid) (21 CFR 874.4760)
Regulatory Class: II
Product Code: EOB
## 3.0 PREDICATE DEVICE:
The subject device is equivalent to the following devices:
- K191102 Stryker Precision S 4K Sinuscope
## 4.0 DEVICE DESCRIPTION:
The Outlook Surgical Versa One System is a video nasopharyngoscopy system that provides illumination and live video stream of a patient's internal anatomy during otolaryngology and head and neck examinations and procedures. The device is composed of an endoscope, a connection cable available in 6ft or 12ft lengths, a wireless enabled display unit, and a wireless HDMI receiver. Users attach accessory instruments to the semi-flexible shaft of the endoscope enabling the user to visualize with the instrument attached.
The device will be used by clinical practitioners in a sterile field in an operating room, healthcare clinic, or medical office. The Outlook Surgical Versa One System is a reusable device initially
{5}
supplied as non-sterile to the user and requiring the user to process (i.e., clean and sterilize) the device for initial use, as well as to reprocess the device after each use.
The Versa One Endoscope has a semi-flexible shaft that enables the user to attach instruments. The semi-flexible shaft is rigid near the adapter end and semi-flexible at the distal end to conform the shape of the attached instrument. When an instrument is attached to the scope, the flexible portion becomes rigid and the scope functions as a rigid endoscope. The Versa One endoscope is compatible with Versa One straight and curved sheaths.
The Outlook Surgical Versa One System live stream video can be viewed on the display unit and an off-the-shelf monitor connected to the display unit via an HDMI cable or wirelessly. The off-the shelf monitor is not provided with the device. The device operates off rechargeable battery power.
The Versa One Endoscope uses an integrated LED light source to illuminate glass fiber bundles in the shaft and a CMOS sensor at the distal tip to send the signal to a custom PCB in the endoscope handle that converts the signal to live video.
# 5.0 INDICATIONS FOR USE:
The Outlook Surgical Versa One System is intended for use in examination and visualization in otolaryngology and head and neck procedures.
# 6.0 SUBSTANTIAL EQUIVALENCE:
The Outlook Surgical Versa One System is similar to the predicate in several ways. Both devices have the same intended use, indication for use, and the same use environment. Both devices illuminate and provide visualization of the patient's anatomy. The subject and predicate devices perform as rigid endoscopes with the same direction of view and lack of articulation control and internal channel. Both devices are reusable and use similar methods of reprocessing. Performance and safety testing has demonstrated that technical differences raise no new questions of safety and effectiveness. Thus, the subject Outlook Surgical Versa One System is substantially equivalent to the predicate Stryker Precision S 4K Sinuscope.
The table below compares the intended use and technological characteristics of the subject and predicate device.
| Feature | Outlook Surgical Versa One System (Subject Device) | Stryker Precision S 4K Sinuscope (Predicate Device) | Similarities and Differences |
| --- | --- | --- | --- |
| 510(k) | K241731 | K191102 | |
| Common Name | Nasopharyngoscope System | Sinuscope | |
| Classification Name | Nasopharyngoscope (flexible or rigid) and Accessories | Nasopharyngoscope (flexible or rigid) and Accessories | Same |
| Device Class | Class II | Class II | Same |
| Regulation | 21 CFR 874.4760 | 21 CFR 874.4760 | Same |
| Product Code | EOB | EOB | Same |
| Indication for Use | The Outlook Surgical Versa One System is intended for use in examination and | The Precision S 4K Sinuscope is intended for use in otolaryngology and | Similar. Narrower indication for use does not |
Proprietary and Confidential
{6}
Proprietary and Confidential
Page 3 of 7
| Feature | Outlook Surgical Versa One System (Subject Device) | Stryker Precision S 4K Sinuscope (Predicate Device) | Similarities and Differences |
| --- | --- | --- | --- |
| | visualization in otolaryngology and head and neck procedures. | Head and Neck procedures, including rhinology, and endoscopic plastic and reconstructive surgery. | raise new questions of safety or effectiveness. |
| Use Environment | Medical facilities | Medical facilities | Same |
| Principle of Operation | Attach proximal end to provided power source, attach scope shaft to instrument to create rigid scope configuration and insert into nose or mouth | Attach proximal end to light source and camera and insert rigid scope into nose or mouth | Performance testing and User Validation demonstrate this difference does not raise new questions of safety and effectiveness. |
| Image transmission | CMOS sensor at distal tip transfers images to integrated handheld monitor with video processor to display live stream images. | Rod lenses with camera/monitor connected to sinuscope eyepiece | Optical performance testing demonstrates this difference does not raise new questions of safety and effectiveness. |
| Wireless video transmission | Yes. Optional wireless data transfer between handheld monitor and wireless receiver. | No | Safety testing in accordance with IEC 60601-1, IEC 60601-2-18, IEC 63195-1, and ANSI/IEEE C95.1 demonstrates this difference does not raise new questions of safety and effectiveness. |
| Internal working channel | No | No | Same |
| Shaft type | Rigid when in use (attached to instrument) | Rigid shaft | Same |
| Required Accessories | Sheath required for use | No | User Validation and Performance testing demonstrate this difference does not raise new questions of safety and effectiveness. |
| Shaft Outer Diameter | 3.0mm, 3.8mm | 3.1mm, 4.0mm | Performance and User Validation demonstrates this difference does not raise new questions of safety or effectiveness. |
| Shaft Working Length | 102mm - 105mm | 125mm - 180mm | Performance and User Validation testing demonstrate this difference does not raise new questions of safety and effectiveness. |
| Articulation Control | No | No | Same |
{7}
Proprietary and Confidential
Page 4 of 7
| Feature | Outlook Surgical Versa One System (Subject Device) | Stryker Precision S 4K Sinuscope (Predicate Device) | Similarities and Differences |
| --- | --- | --- | --- |
| Scope Direction of View | 0° | 0° - 70° | Same |
| Scope Field of View | 120° | 80° - 105° | User Validation and Performance testing demonstrate this difference does not raise new questions of safety and effectiveness. |
| Scope Power Source | Rechargeable Li-ion battery or AC power from control box through cable connected to scope. | Light source powered externally and coupled to scope. | Electrical safety testing in accordance with IEC 60601-1 and 60601-2-18 demonstrate this difference does not raise new questions of safety and effectiveness. |
| Illumination Output | Illumination fibers; commercially available source | Illumination fibers, commercially available source | Same |
| Light Source | Integrated LED | External light source coupled to scope | Electrical safety testing in accordance with IEC 60601-1 and 60601-2-18 and Photobiological testing in accordance with IEC 62471 demonstrates this difference does not raise new questions of safety and effectiveness. |
| Key Patient-Contacting Materials | 304 stainless steel, glass, borosilicate, crosslinked polyvinylidene fluoride, epoxy | Stainless Steel, optical glass, glass fibers | Biocompatibility testing in accordance with ISO 10993 and performance testing demonstrates this difference does not raise new questions of safety and effectiveness. |
| Single Use or Reusable | Reusable – scope
Single Use – accessories | Reusable | Sterilization validation in accordance with ISO 11135 and ISO 10993-7 demonstrates this difference does not raise new questions of safety and effectiveness. |
| Scope Reprocessing | Manual cleaning followed by APS STERRAD Sterilization | Manual and automated cleaning, manual and automated disinfection, autoclave, Steris VPRO, and APS STERRAD sterilization | Similar |
{8}
Proprietary and Confidential
Page 5 of 7
# 7.0 PERFORMANCE DATA:
The table below lists all of the testing that has been performed on the Outlook Surgical Versa One System.
| Test | Test Method Summary | Results |
| --- | --- | --- |
| Performance Testing: Simulated Use/Cycling | The Versa One endoscope was tested to show the device performs as intended through multiple simulated use procedures and repeated instrument attachment/detachment cycles. | Passed. All acceptance criteria were met. The simulated use testing is relevant to determining substantial equivalent because it demonstrates the subject device performs as intended as a rigid scope when attached to an instrument. The cycling testing demonstrates that the subject device performs as expected and can be used safely throughout its minimum expected use life. |
| User Validation | FDA Guidance document “Applying Human Factors and Usability Engineering to Medical Devices”, IEC 62366-1. Multiple users were presented with all device components and followed the IFU to set up and use the system using an anatomical model and to confirm results in a cadaver model. | Passed. All users were able to choose the appropriate attachment for a procedure, attach the instrument to the scope, and perform the simulated procedure with the device. This test is relevant to determining substantial equivalence because it demonstrates users can use the device as intended in the expected use environment. |
| Performance Testing: Optical Performance | Testing was conducted according to ISO 8600-X, ISO 12233, and ISO 15739 for optical performance. Image capture and analysis software was used to evaluate signal to noise ratio and dynamic range, image quality (geometric distortion, MTF, and FOV), image intensity uniformity, and color accuracy. Endoscope image latency was also measured. | Passed. All devices met performance criteria. This testing is relevant in determining substantial equivalence because the subject device and predicate are both intended to illuminate and visualize patient anatomy. |
| Performance Testing: Simulated Use | The Outlook Surgical Versa One System was tested for automatic exposure control, wireless link distance, Display power up, battery charge, battery status indicator, image rotation, and HDMI port activity. Versa One Endoscopes were evaluated for shaft durability and bend radius using an automated process. | Passed. All systems met performance criteria. This testing is relevant in determining substantial equivalence because it demonstrates the subject device reliably performs as intended. The reliability of the wireless link distance supports substantial equivalence of the subject device to the predicate device with wireless transmission of images. |
{9}
Proprietary and Confidential
Page 6 of 7
| Performance Testing: Electrical Requirements (non 60601-1) | The Versa One Display and Wireless HD Receiver were inspected for electrical requirements not included in IEC 60601-1 or IEC 60601-2-18 testing. Device charger and Display MOPP, isolation barriers, and HDMI connectors were inspected. | Passed. All devices met criteria. This test is relevant in determining substantial equivalence because is demonstrates the subject device safely and reliably performs as intended. |
| --- | --- | --- |
| Performance Testing: Dimension, Material, Environmental | Versa One Endoscopes were tested for patient-contacting surface temperature, water splash testing, ambient temperature testing, dimensional and material requirements, cable force measurement and leak testing of the distal end. | Passed. This is relevant to determining substantial equivalence because the subject device and predicate used to illuminate and visualize patient anatomy and have the same use environment. |
| Electrical Safety, electromagnetic compatibility (EMC), and RF Safety | Testing was conducted in compliance with the IEC 60601-1 and IEC 60601-2-18 for electrical safety, IEC 60601-1-2 for EMC, IEC 63195-1, and ANSI/IEEE C95.1-1 for RF Safety | Passed. This test is relevant in determining substantial equivalence because the it demonstrates the electrical safety, electromagnetic compatibility, and RF safety of the subject device. |
| Photobiological Safety | Testing was conducted in accordance with IEC 62471 for photobiological safety. | Passed. This test is relevant in determining substantial equivalence because it demonstrates the photobiological safety of the subject device. |
| Cleaning Validation | FDA Guidance document Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, AAMI TIR12 and AAMI ST98. | Passed. This test is relevant in determining substantial equivalence because the subject device and predicate device are both cleaned manually. |
| Biocompatibility Testing | FDA Guidance, “Use of International Standard ISO 10993-1, ‘Biological evaluation and testing within a risk management Process’ and 10993-1: Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. (ISO 10993-5, ISO 10993-10, ISO 10993-23) | Passed. Non-cytotoxic, non-sensitizing, non-irritating, non-systemically toxic, and non-pyrogenic. This test is relevant in determining substantial equivalence because it demonstrates any material differences between the subject device and the predicate do not raise new questions of safety. |
| Sterilization Validation | For scope: ISO 22441:2022 Sterilization of health care products, low temperature vaporized hydrogen peroxide – Requirements for the development, validation and routine control of sterilization process for medical devices. The Overkill test method was used. For sheaths: ISO 11135:2014 Sterilization of health care products – | Passed. This test is relevant in determining substantial equivalence because the subject device and predicate device are both reprocessed with APS STERRAD sterilization. Sterilization validation for the sheaths is relevant in determining substantial equivalence because it demonstrates the sterile, single-use accessories do not raise |
{10}
| | Ethylene oxide – Requirements for the development, validation and routine control of a sterilization process for medical devices. | new questions of safety or effectiveness. |
| --- | --- | --- |
## 8.0 CONCLUSION:
The results of the performance testing described above demonstrate that the Outlook Surgical Versa One System is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Proprietary and Confidential
Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.