The Precision S 4K Sinuscope is intended for use in otolaryngology and Head and Neck procedures, including rhinology, and endoscopic plastic and reconstructive surgery.
Device Story
Precision S 4K Sinuscope is a reusable, rigid tubular optical instrument; used in otolaryngology and head/neck procedures. Device guides light to illuminate internal anatomy; uses series of lenses (objective, relay rod, ocular) to relay images to external camera system. Operated by clinicians in clinical settings. Output is visual image displayed on camera system monitor; assists in examination, diagnosis, and therapy. Benefits include high-resolution visualization of surgical site.
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993), cleaning (AAMI TIR30), disinfection (AAMI TIR 12), sterilization (ISO 14937, ANSI/AAMI ST79), electrical safety (IEC 60601-1), packaging (ASTM D4149), optical performance (ISO 8600-1), and design validation (AAMI/ANSI HE75). All tests passed.
Technological Characteristics
Rigid rod lens optical system; stainless steel, optical glass, and glass fiber construction. Dimensions: 4.0mm/3.1mm outer diameter; 125mm-180mm working length; 0°-70° direction of view; 80°-105° field of view. Reusable; manual/automated cleaning and disinfection. Sterilization: Autoclave, Steris VPRO, APS Sterrad (SAL 10^-6).
Indications for Use
Indicated for patients undergoing otolaryngology, Head and Neck, rhinology, and endoscopic plastic and reconstructive surgical procedures.
Regulatory Classification
Identification
A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 6, 2019
Stryker Jessie Duong Sr. Staff, Regulatory Affairs Specialist 5900 Optical Court San Jose, California 95138
Re: K191102
Trade/Device Name: Precision S 4K Sinuscope Regulation Number: 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (flexible or rigid) and accessories Regulatory Class: Class II Product Code: EOB Dated: April 23, 2019 Received: April 25, 2019
Dear Jessie Duong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas Nandkumar, Ph.D. Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name Precision S 4K Sinuscope
Indications for Use (Describe)
The Precision S 4K Sinuscope is intended for use in otolaryngology and Head and Neck procedures, including thinology, and endoscopic plastic and reconstructive surgery.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
|X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. The letters are all lowercase and connected, giving the word a modern and sleek appearance. The color of the text is black, contrasting with the white background.
## 510(k) Summary
## Submitter:
| Applicant | Stryker Endoscopy<br>5900 Optical Court<br>San Jose, CA 95138 |
|----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Jessie Duong<br>Senior Staff Regulatory Affairs Specialist<br>Phone: (408) 754-2077<br>Facsimile: (408) 754-2598<br>Email: jessie.duong@stryker.com |
| Date Prepared | April 23, 2019 |
### Subject Device:
| Name of Device | Stryker Precision S 4K Sinuscope |
|----------------------|------------------------------------|
| Common or Usual Name | Sinuscope |
| Classification Name | Nasopharyngoscope, 21 CFR 874.4760 |
| Regulatory Class | Class II |
| Product Code | EOB |
### Predicate Device:
Note: The predicate device has not been subject to a design-related recall.
### Device Description:
Stryker's Precision S 4K Sinuscopes are tubular optical instruments used to provide a view of internal patient anatomy for examination, diagnosis, and therapy during otorhinolaryngology procedures. The devices are available in a variety of outer diameters and working lengths.
The sinuscopes are reusable devices initially supplied as non-sterile to the user and requiring the user to process (i.e. clean and sterilize) the device for initial use, as well as to reprocess the device after each use. The sinuscope connects to the camera system through a separate coupler. The coupler is provided by the camera system and is not subject of this premarket notification.
The sinuscopes achieve their intended use by guiding light to illuminate and image a patient's internal anatomy, then relaying the image out of the surgical site for processing and display by a separate camera system. The sinuscope's optical system consists of a series of lenses, which includes an objective lens to image the intended object, a relay rod lens system to transmit the image along the working length, and an ocular lens to form the final image size.
{4}------------------------------------------------
## Indications for Use:
The Precision S 4K Sinuscope is intended for use in otolaryngology and Head and Neck procedures, including rhinology, and endoscopic plastic and reconstructive surgery.
| Feature | Subject Device | Predicate Device |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Precision S 4K Sinuscope | Schoelly Sinuscope |
| Manufacturer | Stryker Endoscopy<br>5900 Optical Court<br>San Jose, CA 95138<br>USA | Schoelly Fiberoptic GmbH<br>Robert-Bosch-Str. 1-3<br>79211 Denzlingen<br>Germany |
| Submission Reference | Current Submission | K142249 |
| Indications for Use<br>Statement | The Precision S 4K Sinuscope is<br>intended for use in otolaryngology<br>and Head and Neck procedures,<br>including rhinology, and endoscopic<br>plastic and reconstructive surgery | The Schoelly Sinuscope is intended<br>for use in otolaryngology and Head<br>and Neck procedures, including<br>rhinology, and endoscopic plastic<br>and reconstructive surgery |
| Image Transmission | Rigid rod lenses | Rigid rod lenses |
| Outer diameter | 4.0mm, 3.1mm | 4.0mm, 2.7mm |
| Working Length | 125mm - 180mm | 110mm - 175mm |
| DOV | 0°-70° | 0°-70° |
| FOV | 80°-105° | 80°-100° |
| Key Patient-Contacting<br>Materials | Stainless Steel, Optical Glass,<br>Glass Fibers | Stainless Steel, Optical Glass,<br>Glass Fibers, Co-Cr-Ni Alloy |
| Single Use or Reusable | Reusable | Reusable |
| Cleaning | Manual and Automated | Manual and Automated |
| Disinfection | Manual and Automated | Not available |
| Sterilization<br>Methods | Autoclave, Steris VPRO,<br>APS Sterrad | Autoclave, APS Sterrad |
| Sterility Assurance Level | 10-6 | 10-6 |
Comparison of Technological Characteristics with the Predicate Device:
{5}------------------------------------------------
## Performance Testing:
The following performance data were provided in support of the substantial equivalence determination.
| Test Category | Applicable Standards / Guidance | Result |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|--------|
| Biocompatibility | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10<br>ISO 10993-11<br>FDA Guidance: Use of International<br>Standard ISO 10993-1 | Pass |
| Cleaning | AAMI TIR30<br>AAMI ST15883-5 | Pass |
| Disinfection | AAMI TIR 12<br>ISO 15883-2 | Pass |
| Sterilization | ISO 14937<br>ANSI/AAMI ST79<br>ANSI/AAMI ST58<br>ISO 17665-1 | Pass |
| Electrical Safety &<br>Temperature Testing | IEC 60601-1<br>IEC 60601-2-18 | Pass |
| Packaging Qualification | ASTM D4149 | Pass |
| Optical Performance | ISO 8600-1 | Pass |
| Design Validation | AAMI/ANSI HE75<br>AAMI/IEC 62366<br>FDA Guidance: Applying Human<br>Factors and Usability Engineering to<br>Medical Devices | Pass |
| Performance - Bench | In accordance with device<br>performance specifications | Pass |
### Conclusions:
The Precision S 4K Sinuscope is substantially equivalent in design, intended use, principles of operation, technological characteristics and safety features to the predicate device. There are no new issues of safety and/or effectiveness introduced by the Precision S 4K Sinuscope when used as instructed.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.