K241567 · Oxos Medical · OWB · Dec 6, 2024 · Radiology
Device Facts
Record ID
K241567
Device Name
MC2 Portable X-ray System
Applicant
Oxos Medical
Product Code
OWB · Radiology
Decision Date
Dec 6, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Indications for Use
The MC2 Portable X-ray System is designed to aid clinicians with point-of-care visualization and guidance during X-rays of distal extremities. It is intended for use in clinical environments and is not intended for surgical applications.
Device Story
MC2 Portable X-ray System provides point-of-care diagnostic X-ray imaging for distal extremities. System comprises battery-powered emitter and cassette; interfaces wirelessly with tablet or monitor via OXOS Device App. Operator positions emitter using Infrared Tracking System; LIDAR ensures safe source-to-skin distance (SSD). Supports single radiography, serial radiography, and fluoroscopy. Safety interlocks (SID, SSD, active area, stand-mounted) prevent unauthorized acquisition. Used in clinics/imaging centers by qualified professionals. Output displayed on tablet/monitor for clinical visualization; aids orthopedic decision-making. Stand-mounted configuration required for serial/fluoroscopic modes; handheld allowed for single radiography. Benefits include portable, rapid extremity visualization.
Clinical Evidence
No clinical trials on patients were performed. A task-based image quality study was conducted where five radiologists evaluated radiographic, serial radiographic, and fluoroscopic images acquired by technologists across all intended anatomical indications to confirm clinical adequacy.
Technological Characteristics
Portable X-ray system with iRay Mercu0909X detector (9x9 inch, CsI scintillator). Features LIDAR for SSD, Class 1 laser for alignment. kVp range 40-80. Connectivity via WiFi/USB-C. DICOM 3.0 compliant. Software-controlled via OXOS Device App. Safety standards: IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-28, 60601-2-43, 60601-2-54. IP ratings: Emitter/Cassette IP00, Foot Pedal IPX8, Case IP67.
Indications for Use
Indicated for use by qualified/trained medical professionals on adult patients for orthopedic radiographic, orthopedic serial radiographic, orthopedic fluoroscopic, and orthopedic interventional procedures of extremities distal to the shoulders and distal to the knees. Not for surgical use. Not intended to replace stationary radiographic systems. Used in healthcare facilities (outpatient clinics, urgent cares, imaging centers, sports medicine, occupational medicine). Contraindicated for bariatric patients, mammography, dental applications, non-intact skin, cardiac applications, and use near pacemakers/ICDs.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 6, 2024
Oxos Medical, Inc. % Dhruv Vishwakarma Director of Regulatory Affairs Powers Regulatory Consulting 1100 Peachtree St NE Suite 700 ATLANTA, GA 30309
Re: K241567
Trade/Device Name: MC2 Portable X-ray System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: OWB, IZL, OXO Dated: November 5, 2024 Received: November 7, 2024
Dear Dhruv Vishwakarma:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K241567
Device Name MC2 Portable X-ray System
#### Indications for Use (Describe)
The MC2 Portable X-ray System is indicated for use by qualified/trained medical professionals on adult patients for orthopedic radiographic, orthopedic, orthopedic fluoroscopic, and orthopedic interventional procedures of extremities distal to the shoulders and distal to the knees. The device is not intended for use during surgery. The device is not intended to replace a stationary radiographic system.
The device is to be used in healthcare facilities where qualified operators are present (e.g., outpatient clinics, urgent cares, imaging centers, sports medicine facilities, occupational medicine clinics).
The device is not intended to be used in environments with the following characteristics:
-Aseptic or sterile fields, such as in surgery
- -Home or residential settings or other settings where qualified operators are not present
-Vehicular and moving environments
- -Environments under direct sunlight
-Oxygen-rich environments, such as near an operating oxygenation concentrator
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the letters OXOS in a stylized font. The O's are circles with small gaps at the top, bottom, left, and right. The X is formed by two diagonal lines that intersect in the middle, and the S is a curved letter.
### Contact Details
| | Applicant Information | Correspondent Information |
|---------|--------------------------------------------------------|--------------------------------------------------------|
| Name | OXOS Medical, Inc. | OXOS Medical, Inc. |
| Address | 1100 Peachtree St NE<br>Suite 700<br>Atlanta, GA 30309 | 1100 Peachtree St NE<br>Suite 700<br>Atlanta, GA 30309 |
| Contact | Mr. Dhruv Vishwakarma | Mr. Dhruv Vishwakarma |
| Phone | 770-375-7840 | 770-375-7840 |
| Email | dhruv.vishwakarma@oxos.com | dhruv.vishwakarma@oxos.com |
### Device Information
| Device Information | |
|---------------------|------------------------------------------------------------------------------------|
| Trade Name | MC2 Portable X-ray System |
| Common Name | Interventional fluoroscopic x-ray system |
| Classification Name | Image-Intensified Fluoroscopic X-Ray System, Mobile<br>Regulation Number: 892.1650 |
| Product Codes | OWB, OXO, IZL |
### Legally Marketed Predicate Devices
| Type | K-Number | Trade Name | Product Code |
|-----------|----------|--------------------------------------|--------------|
| Predicate | K211191 | Virtual C DRF Digital Imaging System | OWB |
| Reference | K182207 | MinXray, Inc. TR90BH | IZL |
| Reference | K193225 | KDR U-ARM | KPR |
| Reference | K221714 | Flat Panel Detector | MQB |
| Reference | K150150 | Rayence 1717SCN | MQB |
{5}------------------------------------------------
## Device Description Summary
The MC2 Portable X-ray System ("MC2 System" or "MC2") is a portable and handheld X-ray system designed to aid clinicians with point-of-care visualization through diagnostic X-rays of extremities distal to the shoulders and distal to the knees. The device allows clinicians to select desired technique factors best suited for their patient's anatomy. The MC2 consists of two major system components: the emitter and the cassette. The MC2 emitter and cassette are battery-powered and are charged via a wired charger. The system is intended to interface wirelessly to an external tablet when used with the OXOS Device App or to a monitor with an off-the-shelf ELO Backpack and the OXOS Device App. The MC2 utilizes an Infrared Tracking System to allow the emitter to be positioned above the patient's anatomy and aligned to the cassette by the operator. The MC2 also utilizes a LIDAR system to ensure patient safety by maintaining a safe source-to-skin distance.
The MC2 is capable of three X-ray imaging modes: single radiography, serial radiography, and fluoroscopy. In single and serial radiography modes, the user can utilize the entire range of kV values (40-80kV), while fluoroscopy mode is limited to 40-64kV. In single radiography mode, the user can utilize the entire range of mAs values, while serial radiography and fluoroscopy are limited to 0.04-0.08 mAs. Single radiography acquisitions may be performed handheld, while serial radiography and fluoroscopy require the emitter to be in a stand-mounted configuration.
The MC2 contains various safety features to ensure patient and operator safety. The primary interlocks that ensure system geometry is maintained include a source-to-image distance interlock, an active area interlock, a source-to-skin distance interlock, and a stand-mounted interlock.
The source-to-image distance interlock uses the Tracking System to disallow X-ray acquisition when the device is outside the bounds of source-to-image distance (SID). This acts concurrently with the source-to-skin distance (SSD) interlock which uses the LIDAR system to disallow X-ray acquisition below 30cm source-to-skin distance. Both conditions must be met for X-ray acquisition to be allowed. The active area interlock uses the Tracking System to prevent the X-ray field from extending beyond the bounds of the active area. The stand-mounted interlock prevents handheld X-ray acquisition in serial radiography and fluoroscopy modes.
In addition to the components listed above, the MC2 includes accessories, such as a clinical cart and a wireless foot pedal should be used for stand-mounted imaging when initiating single, serial or fluoroscopic acquisitions remotely. The clinical cart supports the MC2 for stand-mounted operation and allows the user to position anatomy easily. An accessory stand such as the clinical cart is required to facilitate stand-mounted imaging modes. Radiography and Photo modes may be used without a stand.
{6}------------------------------------------------
### Intended Use/Indications for Use
### Intended Use:
The MC2 Portable X-ray System is designed to aid clinicians with point-of-care visualization and guidance during X-rays of distal extremities. It is intended for use in clinical environments and is not intended for surgical applications.
### Indications for Use:
The MC2 Portable X-ray System is indicated for use by qualified/trained medical professionals on adult patients for orthopedic radiographic, orthopedic serial radiographic, orthopedic fluoroscopic, and orthopedic interventional procedures of extremities distal to the shoulders and distal to the knees. The device is not intended for use during surgery. The device is not intended to replace a stationary radiographic system.
The device is to be used in healthcare facilities where qualified operators are present (e.g., outpatient clinics, urgent cares, imaging centers, sports medicine facilities, occupational medicine clinics).
The device is not intended to be used in environments with the following characteristics:
- Aseptic or sterile fields, such as in surgery
- Home or residential settings or other settings where qualified operators are not present ●
- Vehicular and moving environments ●
- Environments under direct sunlight ●
- Oxygen-rich environments, such as near an operating oxygenation concentrator ●
### Contraindications
The MC2 System is contraindicated as follows:
- The MC2 System is NOT intended for bariatric patients. ●
- The MC2 System is NOT intended for mammography. ●
- The MC2 System is NOT intended for dental applications. ●
- The MC2 System is NOT intended to come in contact with non-intact skin.
- The MC2 System is NOT intended for cardiac applications. ●
- The MC2 System is not intended for use in proximity to pacemakers or implantable . cardioverter-defibrillators (ICDs).
{7}------------------------------------------------
## Indications for Use Comparison
The MC2 Portable X-ray System and the predicate device are both intended for use by qualified/trained medical professionals on adult patients for diagnostic radiographic and interventional fluoroscopic procedures. The MC2 Portable X-ray System has a narrower anatomical, procedural, and environmental scope than the Virtual C DRF Digital Imaging System. The MC2 Portable X-ray System is indicated for use by qualified/trained medical professionals on adult patients for orthopedic radiographic, orthopedic serial radiographic, orthopedic fluoroscopic, and orthopedic interventional procedures of extremities distal to the shoulders and distal to the knees. The device is not intended for use during surgery. The device is not intended to replace a stationary radiographic system.
### Technological Comparison
The MC2 Portable X-ray System shares most technological characteristics with the primary predicate device, the Virtual C DRF Digital Imaging System, including principles of operation and design features; however, there are some differences in detailed technological characteristics. Both devices offer single radiographic and fluoroscopic imaging. The MC2 device has longer exposure time capabilities. The MC2 features a higher resolution (pixel pitch) detector and a slightly larger focal spot. MTF and DQE values differ slightly between the two devices. The subject device and the primary predicate both utilize LIDAR to provide the means to maintain a safe source-to-skin distance.
The MinXray, Inc. TR90BH is used as a reference device due to its handheld nature. It is used for performing static radiographic imaging for similar anatomical indications.
The Konica Minolta KDR U-Arm features serial radiographic (DDR) capabilities similar to those of the subject MC2 device. The DDR functionality in the MC2 includes various mitigations to disallow use for guidance. These include a 2 second delay from image acquisition to display of each preview frame, a low-resolution preview display, a lower frame rate during image preview, and a salient disclaimer informing the user that the mode is not to be used for fluoroscopy.
The iRay Venu1717X is an FDA-cleared detector that is very similar in design and technological features to the detector found in the subject MC2 device. The X-ray detection technology is identical in both detectors, with similar DQE and MTF. The reference device has a large 17 x 17" active area.
The Rayence 1717SCN reference device is an FDA-cleared X-ray detector similar to the iRay Mercu0909X found in the subject device.
{8}------------------------------------------------
The following table provides a comparison of the technological characteristics of the MC2 with the predicate and reference devices:
| | Subject Device:<br>OXOS MC2 (K241567) | Predicate Device:<br>Virtual C DRF<br>(K211191) | Relevant Reference<br>Device | Comparison |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Display | The MC2 system interfaces via<br>the OXOS Device App with a<br>tablet or monitor. An<br>off-the-shelf tablet is optionally<br>supplied with the MC2 or can<br>be supplied by the user. | Display on<br>workstation monitor. | | The subject device's display is similar<br>to that of other devices, including the<br>reference devices, in that it can be<br>displayed on monitors or tablets. |
| Detector | iRay Mercu0909X<br>Active Area: 9 x 9 inches<br>Frame rate: 5 fps @ 1×1 | Vivix-D1212G<br>Active Area: 12" x<br>12"<br>Frame rate: 20 FPS<br>@ 1x1 | Konica Minolta KDR<br>U-Arm:<br>348.8 x 425.6 mm<br>(13.7 x 16.8 in)<br>15 FPS<br>Rayence 1717SCN:<br>43.18cm x 43.18 cm<br>(17 in x 17 in) | The subject and predicate devices<br>have detectors with equivalent<br>technological characteristics. |
| MTF | MTF @ 1 Lp/mm, RQA5: 55%<br>MTF @ 2 Lp/mm, RQA5: 23% | MTF @ 2.0 Lp/mm:<br>30% | Rayence 1717SCN:<br>MTF @ 1 Lp/mm:<br>59%<br>MTF @ 2 Lp/mm:<br>26% | The subject device has a similar MTF<br>compared to the predicate device. |
| DQE | DQE @ 0 Lp/mm: 72%<br>DQE @ 1 Lp/mm: 51%<br>DQE @ 2 Lp/mm: 31%<br>RQA5 | DQE @ 1 Lp/mm:<br>56% | Konica Minolta KDR<br>U-Arm:<br>DQE @ 0 Lp/mm:<br>72%<br>RQA5<br>Rayence 1717SCN:<br>DQE @ 0 Lp/mm:<br>61%<br>DQE @ 1 Lp/mm:<br>53%<br>DQE @ 2 Lp/mm:<br>28%<br>RQA5 | The subject device has a similar DQE<br>compared to the predicate device. |
| Pixel<br>Pitch | 139µm | 145µm | Rayence 1717SCN:<br>127µm | Similar. |
| Scintillator | Cesium lodide (Csl) | Cesium lodide (Csl) | Cesium lodide (Csl) | Identical. |
| Collimator | Removable fixed collimators<br>(pucks) and an automatic<br>collimator with manual stepped<br>adjustment capability for fine<br>adjustments. | Machine-Vision<br>Collimator<br>(Accession<br>2010848-000) | | The subject device has a collimation<br>design similar to the predicate device.<br>Both devices use device tracking<br>information to confine the X-ray field to<br>the active area. |
| Source to<br>Skin<br>Distance<br>(SSD) | LIDAR | LIDAR | | Identical. |
| Field<br>Visualizati<br>on | Viewfinder on emitter screen<br>UI. | A visual display<br>provides real time<br>video images of the<br>patient, and a | | The subject device has a Viewfinder<br>on the emitter to aid patient<br>positioning. The predicate device uses<br>a similar technique to display the x-ray |
| | Subject Device:<br>OXOS MC2 (K241567) | Predicate Device:<br>Virtual C DRF<br>(K211191) | Relevant Reference<br>Device | Comparison |
| | | the video images<br>represents the<br>location and size of<br>the radiation beam<br>with respect to the<br>patient. | | field with respect to the patient in<br>real-time. |
| Laser | Class 1 laser device compliant<br>with IEC 60825-1:2014 -3rd<br>edition used for Light Detection<br>and Ranging (LIDAR) to<br>indicate the distance from the<br>x-ray source to the target to be<br>diagnosed.<br>Class 1 laser devices compliant<br>with IEC 60825-1:2014 - 3rd<br>edition display a red crosshair<br>used to align patient anatomy. | Unknown. | MinXray TR90BH:<br>Two lasers:<br>Class II; laser diode;<br>max output ≤1.0 mW,<br>wavelength 635 nm | The subject and reference device b<br>similarly utilize lasers for patient<br>positioning. |
| Focal<br>Spot<br>(IEC) | 0.8 | SB-80-2K-LD: 0.6 | MinXray TR90BH: 0.8 | The subject device's focal spot is<br>similar to the predicate and reference<br>devices. |
| Exposure<br>Time /<br>Pulse<br>Duration | 40-200 ms | 10 ms | MinXray TR90BH:<br>0.01-0.3 sec (10-300<br>ms).<br>0.01 sec (10 ms) Step | The subject device exposure times are<br>similar to the predicate and reference<br>device's exposure time. |
| mA | 1.0-2.0 mA | 0.1-2.0 mA | MinXray TR90BH:<br>15 mA @ 82-90<br>kVDC. (2 kVP steps) | The subject device's mA is identical to<br>the predicate device. |
| kVp | 40–80 kVp | 30–80 kVp | MinXray TR90BH:<br>40-90kVp | The kVps of all devices are similar. |
| Software | Custom software is embedded<br>within the emitter, cassette, and<br>foot pedal to operate the<br>device. The OXOS Device App<br>(non-embedded) is for image<br>viewing, study management,<br>collimation control, image<br>processing for DICOM 3.14<br>greyscale calibration, and<br>display of messages and alerts. | Virtual C DRF<br>software | N/A | All predicate and reference devices<br>have software that supports the<br>device. |
| Ingress<br>Protection<br>Rating | IP00 (E1 Emitter, C1 Cassette,<br>H1 Wired Charger) IPX8 (F1<br>Foot Pedal)<br>IP67 (P1 Case) | Unknown | N/A | N/A. |
| Image<br>Processin<br>g | User Interface can drag, zoom,<br>rotate, adjust brightness,<br>contrast, and sharpness. | Includes software<br>for image capture<br>and analysis. | All include software<br>for image capture and<br>analysis. | Similar. |
| Connectiv<br>ity<br>Options | WiFi, USB-C ports | Ethernet or Wi-Fi | MinXray TR90BH:<br>RJ45 port for DR<br>interface | The predicate and reference devices<br>all have various connectivity options. |
| DICOM | DICOM 3.0 compliant | DICOM | All are DICOM 3.0<br>compliant | Identical. |
| Electrical<br>Safety | IEC60601-1:2005 + A2 (2020)<br>IEC60601-1-2:2020 | IEC60601-1:2005 +<br>A1 (2012) | - | Identical. |
| Subject Device:<br>OXOS MC2 (K241567) | Predicate Device:<br>Virtual C DRF<br>(K211191) | Relevant Reference<br>Device | Comparison | |
| IEC60601-1-3:2008 + A2 | IEC60601-1-3:2008 | | | |
| (2021) | IEC60601-2-28:201 | | | |
| IEC60601-2-28:2017 | 0 | | | |
| IEC60601-2-43:2022 | IEC60601-2-43:201 | | | |
| IEC60601-2-54:2022 | 0 | | | |
| NEMA PS 3.1-3.20 | IEC60601-2-54:200 | | | |
| | 9 | | | |
| | NEMA PS 3.1-3.20 | | | |
{9}------------------------------------------------
{10}------------------------------------------------
## Cybersecurity
Following the FDA guidance titled Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions (2023), the MC2 was risk assessed for security threats separately from device risk assessment using the methodology prescribed in the quidance.
Threat modeling was performed in accordance with the STRIDE methodology. Security controls were implemented in the following categories:
- Authentication controls ●
- Authorization controls ●
- . Cryptography controls
- Code, data, and execution integrity controls ●
- . Confidentiality controls
- Event detection and logging controls ●
- Resiliency and recovery controls
- Firmware and software update controls ●
Security controls were verified and demonstrated adequacy within the MC2's usage environments. Additionally, fuzz testing of vulnerable interfaces and robust third-party penetration testing of the entire system were performed, demonstrating the MC2's security.
## Performance Testing
All components of the MC2, including software, were verified and validated to demonstrate compliance with the appropriate regulations and in accordance with the risk profile analysis. A third-party testing laboratory performed performance testing of the MC2 in accordance with the applicable standards. The third-party testing laboratory and formal internal testing also tested all safety interlocks. All forms of testing showed the MC2 to be compliant with the relevant standards and safe and effective in the procedures and scenarios outlined in the Indications for Use.
{11}------------------------------------------------
A comprehensive, task-based image quality study was conducted to assess the clinical adequacy of the device's imaging performance. This study collected radiographic images, radiographic series, and fluoroscopic series for all relevant anatomical indications stated in the Indications for Use. Images were acquired for static radiography, serial radiography, and fluoroscopic modes. Radiologic technologists acquired images in all acquisition modes with appropriately selected techniques, and five radiologists clinically evaluated the image quality.
# Non-Clinical and Clinical Tests Summary and Conclusions
OXOS performed nonclinical bench testing on the MC2 Portable X-ray System to demonstrate the system performance met required design inputs and user needs and is safe for its intended use. The testing included the verification and validation of: the design and software per OXOS specifications, the adequacy of image quality for the MC2 intended use, the conformance to voluntary standards evaluating device safety and performance, and the usability of the device, demonstrating that MC2 is safe and effective for its intended users.
The MC2 Portable X-ray System met bench testing acceptance criteria as defined in the test protocols. This testing supports substantial equivalence of the MC2 technological characteristics and performance compared to the predicate device.
### Standards
| Regulation or Standard | K241567 - 510(k) Summary |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 21 CFR 801 | Labeling |
| 21 CFR 1020.30(a), (b), (d), (e), (g), (j), and (q) Sections (c), (h), (k), (l), (m), (n), and (o) met via conformity to IEC 60601-1-3 and 60601-2-54 | Diagnostic X-ray systems and their major components |
| 21 CFR 1020.31 met via conformity to IEC 60601-1-3 and IEC 60601-2-54 | Radiographic equipment |
| 21 CFR 1020.32(d)(3)(v), and (g) Sections (a), (b), (c), (d)(1), (d)(2), (d)(3)(i) – (iv), (d)(4), (f), (h), (i), (j), and (k) met via conformity to IEC 60601-1-3, IEC 60601-2-54, and IEC 60601-2-43 | Fluoroscopic equipment |
| 47 CFR 18 | Telecommunication |
| ISO 14971 Third Edition 2019-12 | Medical devices - Application of risk<br>management to medical devices |
| IEC 60601-1 Edition 3.2 2020-08<br>CONSOLIDATED VERSION | Medical electrical equipment - Part 1: General<br>requirements for basic safety and essential<br>performance |
| 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED<br>VERSION | Medical electrical equipment - Part 1-2: General<br>requirements for basic safety and essential<br>performance - Collateral Standard:<br>Electromagnetic disturbances - Requirements<br>and tests |
| 60601-1-3 Edition 2.2 2021-01 CONSOLIDATED<br>VERSION | Medical electrical equipment - Part 1-3: General<br>requirements for basic safety and essential<br>performance - Collateral Standard: Radiation<br>protection in diagnostic X-ray equipment |
| 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED<br>VERSION | Medical electrical equipment - Part 1-6: General<br>requirements for basic safety and essential<br>performance - Collateral standard: Usability |
| IEC 60601-2-28 Edition 3.0 2017-06 | Medical electrical equipment - Part 2-28:<br>Particular requirements for the basic safety and<br>essential performance of X-ray tube assemblies<br>for medical diagnosis |
| IEC 60601-2-43 Edition 2.2 2019-10<br>CONSOLIDATED VERSION | Medical electrical equipment -<br>Part 2-43: Particular requirements for the basic<br>safety and essential performance of X-ray<br>equipment for interventional procedures |
| IEC 60601-2-54 Edition 2.0 2022-09 | Medical electrical equipment - Part 2-54:<br>Particular requirements for the basic safety and<br>essential performance of X-ray equipment for<br>radiography and radioscopy |
| IEC 62304 Edition 1.1 2015-06 CONSOLIDATED<br>VERSION | Medical device software - Software life cycle<br>processes |
| ISO 10993-1 Fifth edition 2018-08 | Biological evaluation of medical devices - Part 1:<br>Evaluation and testing within a risk management<br>process |
| NEMA PS 3.1 - 3.20 2022d<br>Digital Imaging and Communications in Medicine<br>(DICOM) Set | Digital Imaging and Communications in Medicine<br>(DICOM) Set |
| IEC 62563-1 Edition 1.2 2021-07<br>CONSOLIDATED VERSION | Medical electrical equipment - Medical image<br>display systems - Part 1: Evaluation methods |
| IEC 61910-1 Edition 1.0 2014-09 | Medical electrical equipment - Radiation dose<br>documentation - Part 1: Radiation dose<br>structured reports for radiography and<br>radioscopy |
| IEEE ANSI USEMCSC C63.27-2021 | American National Standard for Evaluation of<br>Wireless Coexistence |
| AAMI TIR69:2017/(R2020) | Technical Information Report Risk management<br>of radio-frequency wireless coexistence for<br>medical devices and systems. |
| IEC 62366-1 Edition 1.1 2020-06<br>CONSOLIDATED VERSION | Medical devices - Part 1: Application of usability<br>engineering to medical devices |
| IEC 62133-2 Edition 1.0 2017-02 | Secondary cells and batteries containing alkaline<br>or other non-acid electrolytes - Safety<br>requirements for portable sealed secondary<br>cells, and for batteries made from them, for use<br>in portable applications - Part 2: Lithium systems |
| Section 38.3 of the UN Manual of Tests and<br>Criteria | Lithium metal and lithium ion batteries |
The MC2 complies with the following standards, as applicable:
{12}------------------------------------------------
{13}------------------------------------------------
The MC2 complies with 21 CFR Subchapter J, including 21 CFR 1020.30 to 1020.32, partially by conforming with IEC 60601-1-3 Edition 2.2 (2021), IEC 60601-2-54 Edition 2.0 (2022), and IEC 60601-2-43 Edition 2.2 (2019), in accordance with section 14(c)(1)(A) of the FD&C Act as outlined in "Medical X-Ray Imaging Devices: Conformance with IEC Standards," dated February 21, 2023.
O
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.