Actifuse™ is a bone void filler intended only for orthopaedic applications as a filler for gaps and voids that are not intrinsic to the stability of the bony structure. Actifuse is indicated to be packed gently into bony voids or gaps of the skeletal system, i.e., extremities, pelvis, and spine, including use in posterolateral fusion procedures with appropriate stabilizing hardware and intervertebral disc space. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced by bone during the healing process. When used in the intervertebral disc space, Actifuse must be used with an intervertebral body fusion device cleared by FDA for use with a bone void filler. ALTAPORE is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities, posterolateral spine, intervertebral disc space, and pelvis). ALTAPORE can be used by itself, with autograft as a bone graft extender or with autogenous bone marrow aspirate. These osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. ALTAPORE resorbs and is replaced with bone during the healing process. When used in the intervertebral disc space, ALTAPORE must be used with an intervertebral body fusion device cleared by FDA for use with a bone void filler. ALTAPORE SHAPE is an implant intended to fill bony voids or gaps of the skeletal system (i.e. extremities, posterolateral spine, intervertebral disc space, and pelvis). ALTAPORE SHAPE can be used by itself or with autograft as a bone graft extender in posterolateral spinal fusion procedures. These osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. ALTAPE resorbs and is replaced with bone during the healing process. When used in the intervertebral disc space. ALTAPORE SHAPE must be used with an intervertebral body fusion device cleared by FDA for use with a bone void filler. ALTAPORE MIS is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities, posterolateral spine, intervertebral disc space, and pelvis). ALTAPORE MIS can be used by itself, with autograft as a bone graft extender or with autogenous bone marrow aspirate. These osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. ALTAPORE MIS resorbs and is replaced with bone during the healing process. When used in the intervertebral disc space, ALTAPORE MIS must be used with an intervertebral body fusion device cleared by FDA for use with a bone void filler.
Device Story
Silicate-substituted calcium phosphate bone void filler; osteoconductive/bioactive scaffold; porous structure mimics cancellous bone. Available as granules in aqueous gel (ABX, ALTAPORE) or wax-like putty in polymer carrier (SHAPE). Implanted by surgeons in OR to fill skeletal voids; resorbs and replaced by natural bone. Used as standalone or with autograft/bone marrow aspirate. When used in intervertebral disc space, requires concurrent use of FDA-cleared intervertebral body fusion device. Provides scaffold for new bone growth; facilitates healing of osseous defects.
Clinical Evidence
No clinical data. Substantial equivalence supported by identical design and material composition to the primary predicate device (K232315).
Technological Characteristics
Phase-pure, porous, silicate-substituted calcium phosphate (0.8wt% Si). Composed of Ca, P, O, H, Si. Available as granules in aqueous gel or wax-like putty in absorbable Alkylene Oxide co-polymer carrier. Sterile, single-use. Osteoconductive and bioactive (forms surface apatite layer).
Indications for Use
Indicated for filling surgically created or traumatic osseous defects in the skeletal system (extremities, pelvis, spine, posterolateral fusion, intervertebral disc space) that are not intrinsic to bony stability. Used as standalone filler, autograft extender, or with bone marrow aspirate. Contraindicated for use in the intervertebral disc space without a cleared intervertebral body fusion device.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}------------------------------------------------
August 2, 2024
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Baxter Healthcare Corporation Phillip Romei Manager, Global Regulatory Affairs 25212 W. Illinois Route 120 Round Lake, Illinois 60073
# Re: K241564
Trade/Device Name: Actifuse ABX; Actifuse MIS; Actifuse Shape; ALTAPORE; ALTAPORE SHAPE: ALTAPORE MIS Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable Calcium Salt Bone Void Filler Device Regulatory Class: Class II Product Code: MQV Dated: May 22, 2024 Received: May 31, 2024
# Dear Phillip Romei:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Robert M. Stefani -S 2024.08.02 09:17:50 -04'00'
For: Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
#### K241564
Device Name Actifuse ABX: Actifuse MIS; Actifuse Shape; ALTAPORE; ALTAPORE SHAPE: ALTAPORE MIS
Indications for Use (Describe)
## ACTIFUSE ABX, ACTIFUSE SHAPE and ACTIFUSE MIS:
Actifuse™ is a bone void filler intended only for orthopaedic applications as a filler for gaps and voids that are not intrinsic to the stability of the bony structure. Actifuse is indicated to be packed gently into bony voids or gaps of the skeletal system, i.e., extremities, pelvis, and spine, including use in posterolateral fusion procedures with appropriate stabilizing hardware and intervertebral disc space. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
When used in the intervertebral disc space, Actifuse must be used with an intervertebral body fusion device cleared by FDA for use with a bone void filler.
### ALTAPORE:
ALTAPORE is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities, posterolateral spine, intervertebral disc space, and pelvis). ALTAPORE can be used by itself, with autograft as a bone graft extender or with autogenous bone marrow aspirate. These osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. ALTAPORE resorbs and is replaced with bone during the healing process.
When used in the intervertebral disc space, ALTAPORE must be used with an intervertebral body fusion device cleared by FDA for use with a bone void filler.
### ALTAPORE SHAPE:
ALTAPORE SHAPE is an implant intended to fill bony voids or gaps of the skeletal system (i.e. extremities, posterolateral spine, intervertebral disc space, and pelvis).
ALTAPORE SHAPE can be used by itself or with autograft as a bone graft extender in posterolateral spinal fusion procedures. These osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. ALTAPE resorbs and is replaced with bone during the healing process.
When used in the intervertebral disc space. ALTAPORE SHAPE must be used with an intervertebral body fusion device cleared by FDA for use with a bone void filler.
ALTAPORE MIS: ALTAPORE MIS is an implant intended to fill bony voids or gaps of the skeletal system (i.e.,
{4}------------------------------------------------
extremities, posterolateral spine, intervertebral disc space, and pelvis).
ALTAPORE MIS can be used by itself, with autograft as a bone graft extender or with autogenous bone marrow aspirate. These osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. ALTAPORE MIS resorbs and is replaced with bone during the healing process.
When used in the intervertebral disc space, ALTAPORE MIS must be used with an intervertebral body fusion device cleared by FDA for use with a bone void filler.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) Summary
| Device Trade Name: | ACTIFUSE ABX, ACTIFUSE SHAPE, ACTIFUSE MIS<br>ALTAPORE<br>ALTAPORE SHAPE<br>ALTAPORE MIS |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | Baxter Healthcare Corporation<br>One Baxter Parkway<br>Deerfield, Illinois 60015 |
| Contact: | Phillip Romei<br>Specialist, Regulatory Affairs<br>25212 W. Illinois Route 120<br>Round Lake, IL 60073 |
| Prepared by: | Justin Eggleton, Vice President, Head of Musculoskeletal<br>Regulatory Affairs<br>MCRA, LLC<br>803 7th Street, NW, 3rd Floor<br>Washington, DC 20001<br>Office: 202.552.5800 |
| Date Prepared: | May 22, 2024 |
| Classifications: | 21 CFR §888.3045, Resorbable Calcium Salt Bone Void Filler<br>Device |
| Class: | II |
| Product Codes: | MQV |
| Primary Predicate: | Catalyst Bone Void Filler, OssDsign AB, K232315 |
| Reference Devices: | Actifuse ABX, Actifuse Shape, & Actifuse MIS, Baxter, K082575<br>Altapore, Baxter, K192363<br>Altapore Shape, Baxter, K200640<br>Altapore MIS, Baxter, K201464 |
{6}------------------------------------------------
# Indications For Use:
# ACTIFUSE ABX, ACTIFUSE SHAPE, ACTIFUSE MIS
Actifuse™ is a bone void filler intended only for orthopaedic applications as a filler for gaps and voids that are not intrinsic to the stability of the bony structure. Actifuse is indicated to be packed gently into bony voids or gaps of the skeletal system, i.e., extremities, pelvis, and spine, including use in posterolateral fusion procedures with appropriate stabilizing hardware and intervertebral disc space. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
When used in the intervertebral disc space, Actifuse must be used with an intervertebral body fusion device cleared by FDA for use with a bone void filler.
# ALTAPORE
ALTAPORE is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities, posterolateral spine, intervertebral disc space, and pelvis). ALTAPORE can be used by itself, with autograft as a bone graft extender or with autogenous bone marrow aspirate. These osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. ALTAPORE resorbs and is replaced with bone during the healing process.
When used in the intervertebral disc space, ALTAPORE must be used with an intervertebral body fusion device cleared by FDA for use with a bone void filler.
# ALTAPORE SHAPE
ALTAPORE SHAPE is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities, posterolateral spine, intervertebral disc space, and pelvis).
ALTAPORE SHAPE can be used by itself or with autograft as a bone graft extender in posterolateral spinal fusion procedures. These osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. ALTAPORE SHAPE resorbs and is replaced with bone during the healing process.
When used in the intervertebral disc space, ALTAPORE SHAPE must be used with an intervertebral body fusion device cleared by FDA for use with a bone void filler.
# ALTAPORE MIS
ALTAPORE MIS is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities, posterolateral spine, intervertebral disc space, and pelvis).
ALTAPORE MIS can be used by itself, with autograft as a bone graft extender or with autogenous bone marrow aspirate. These osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. ALTAPORE MIS resorbs and is replaced with bone during the healing process.
{7}------------------------------------------------
When used in the intervertebral disc space, ALTAPORE MIS must be used with an intervertebral body fusion device cleared by FDA for use with a bone void filler.
## Device Description:
# ACTIFUSE ABX, ACTIFUSE SHAPE, ACTIFUSE MIS
Actifuse is a phase-pure, porous, silicate-substituted calcium phosphate osteoconductive bone void filler, comprising a single-phase calcium phosphate scaffold, either granules or granules delivered in a matrix of resorbable polymer. 0.8wt% silicon (Si) is incorporated into the crystalline structure. The interconnected and open porous structure of Actifuse is similar to human cancellous bone. After it is implanted, ACTIFUSE undergoes physiologically-mediated resorption and is replaced by natural bone
ACTIFUSE is osteostimulatory based upon in vitro studies that show cellular responses, such as metabolic activity and proliferation, are accelerated when compared to an identical that did not contain 0.8% silicon by weight..
ACTIFUSE is bioactive based on in vitro studies that show it forms a surface apatite-layer when submerged in simulated body fluid that contains the same ion concentrations as human blood plasma. This apatite layer provides scaffolding onto which the patient's new bone will grow, allowing complete repair of the defect.
ACTIFUSE is osteoconductive based on in vivo animal studies that show it achieves bone healing in a critical defect model as confirmed with radiographic, histolopathological, histomorphometric, and mechanical analyses.
The Actifuse bone void filler is presented in multiple product variants:
### ACTIFUSE ABX
ACTIFUSE ABX presents the ACTIFUSE bioactive and osteoconductive silicate-substituted calcium phosphate bone void filler suspended in an absorbable aqueous gel carrier packaged in a wide-bore applicator. ACTIFUSE ABX is a sterile, single-use device available in volumes of 1.5 ml, 2.5 ml, 5 ml, 10 ml, and 20 ml.
ACTIFUSE ABX is designed for use as a standalone bone graft substitute or as an autograft extender. While not necessary, the product can be mixed with autologous blood, autologous bone marrow aspirate, or autologous bone at the discretion of the surgeon.
### ACTIFUSE MIS
ACTIFUSE MIS System is an ACTIFUSE ABX packaging configuration in which the ACTIFUSE ABX implant material is supplied in a sterile 7.5ml cartridge packed with ACTIFUSE ABX implant material is presented with a delivery applicator and as a separate 7.5ml refill cartridge.
{8}------------------------------------------------
# ACTIFUSE SHAPE
ACTIFUSE SHAPE supplies the ACTIFUSE bioactive and osteoconductive silicate-substituted calcium phosphate bone void filler suspended in an absorbable Alkylene Oxide co-polymer carrier as a shaped wax-like putty in a sterile pouch. ACTIFUSE SHAPE is available as a 1.6 ml small cylinder, 2.6 ml medium cylinder, 8 ml large cylinder, and 15.8 ml large strip.
ACTIFUSE SHAPE is designed for use as a standalone bone graft substitute or as an autograft extender. While not necessary, the product can be mixed with autologous blood or autologous bone at the discretion of the surgeon.
# ALTAPORE
ALTAPORE is a bioactive and osteoconductive silicate-substituted calcium phosphate bone void filler. The interconnected and open porous structure of the silicate-substituted calcium phosphate phase of ALTAPORE is similar to human cancellous bone and is intended to support bone growth with macro and micro- porosity. ALTAPORE is composed solely of elements that exist naturally in normal bone (Ca. P. O. H. Si).
ALTAPORE is supplied in a sterile applicator and contains ALTAPORE microgranules, sized 1-2 mm, 80-85% total porosity, suspended in an absorbable aqueous gel carrier. ALTAPORE does not set in-situ following implantation. ALTAPORE is available in 1.5 ml, 2.5 ml, 5 ml, 10 ml, and 20 ml configurations.
ALTAPORE is designed for use as a standalone bone graft substitute or as an autograft extender. While not necessary, the product can be mixed with Bone Marrow Aspirate (BMA) or autologous bone at the discretion of the surgeon.
ALTAPORE is bioactive based on in vitro studies that show it forms a surface apatite layer when submerged in simulated body fluid that contains the same ion concentrations as human blood plasma. This apatite layer provides scaffolding onto which the patient's new bone will grow, allowing complete repair of the defect.
ALTAPORE is osteoconductive based on in vivo animal studies that show it achieves bone healing in a critical defect model as confirmed with radiographic, histolopathological, histomorphometric, and mechanical analyses. ALTAPORE undergoes cell-mediated remodeling and is replaced by natural bone.
### ALTAPORE MIS
ALTAPORE MIS System is an ALTAPORE packaging configuration in which the ALTAPORE implant material is supplied in a sterile 7.5ml cartridge is presented with a delivery applicator and as a separate 7.5ml refill cartridge.
ALATAPORE MIS System (hereafter referred to as "Altapore MIS) is designed for use as a standalone bone graft substitute or as an autograft extender. While not necessary, the product can be mixed with Bone Marrow Aspirate (BMA) or autologous bone at the discretion of the surgeon.
{9}------------------------------------------------
ALTAPORE MIS is bioactive based on in vitro studies that show it forms a surface apatite layer when submerged in simulated body fluid that contains the same ion concentrations as human blood plasma. This apatite layer provides scaffolding onto which the patient's new bone will grow, allowing complete repair of the defect.
# ALTAPORE SHAPE
ALTAPORE SHAPE supplies the ALTAPORE bioactive and osteoconductive silicate-substituted calcium phosphate bone void filler suspended in an absorbable Alkylene Oxide co-polymer carrier as a shaped wax-like putty in a sterile pouch. ALTAPORE SHAPE is available as a 1.6 ml small cylinder, 2.6 ml medium cylinder, 8 ml large cylinder, and 15.8 ml large strip.
ALTAPORE SHAPE is designed for use as a standalone bone graft substitute or as an autograft extender. While not necessary, the product can be mixed with autologous blood or autologous bone at the discretion of the surgeon.
ALTAPORE SHAPE is bioactive based on in vitro studies that show it forms a surface apatitelayer when submerged in simulated body fluid that contains the same ion concentrations as human blood plasma. This apatite layer provides scaffolding onto which the patient's new bone will grow, allowing complete repair of the defect.
ALTAPORE SHAPE is osteoconductive based on in vivo animal studies that show it achieves bone healing in a critical defect model as confirmed with radiographic, histolopathological, histomorphometric, and mechanical analyses. ALTAPORE SHAPE undergoes cell-mediated remodeling and is replaced by natural bone.
### Predicate Device:
Baxter submits the following information in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical devices, Baxter Bone Void Fillers are substantially equivalent in intended use, indications, basic design, function, and performance to the following predicate device, which has been determined by FDA to be substantially equivalent to preamendment device:
Primary Predicate: Catalyst Bone Void Filler, OssDsign AB, K232315
### Reference Device(s):
Baxter submits the following information in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical devices, Baxter ACTIFUSE ABX, ACTIFUSE SHAPE, ACTIFUSE MIS, ALTAPORE, ALTAPORE SHAPE, and ALTAPORE MIS Bone Void Fillers are substantially equivalent in indications, design principles, and performance to the following predicate devices, which have been determined by FDA to be substantially equivalent to pre-amendment devices:
| Reference Device: | Actifuse ABX, Actifuse Shape, & Actifuse MIS, Baxter, K082575 |
|-------------------|---------------------------------------------------------------|
| | Altapore, Baxter, K192363 |
| | Altapore Shape, Baxter, K200640 |
{10}------------------------------------------------
# Altapore MIS, Baxter, K201464
## Performance Testing Summary:
The purpose of this submission is to expand the indications for use of the Baxter Bone Void Filler devices to include use in the intervertebral disc space in conjunction with an intervertebral body fusion device cleared by FDA for use with a bone void filler.
Because the subject device is identical to the previously cleared primary predicate device (K232315), it was determined that no additional non-clinical data was needed to support the expanded indications proposed in the current submission.
### Conclusion:
The subject device and the predicate devices have intended use, have similar technological characteristics, and are made of similar materials. The subject and predicate devices are packaged in similar materials and are sterilized using similar methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above. Baxter ACTIFUSE ABX, ACTIFUSE SHAPE, ACTIFUSE MIS, ALTAPORE, ALTAPORE SHAPE, and ALTAPORE MIS Bone Void Fillers are as safe, as effective, and performs as well as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.