K241490 · Mvision AI OY · QKB · Oct 18, 2024 · Radiology
Device Facts
Record ID
K241490
Device Name
Contour+ (MVision AI Segmentation)
Applicant
Mvision AI OY
Product Code
QKB · Radiology
Decision Date
Oct 18, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Automatic contouring of anatomical structures
Deep artificial neural networks
Minimum agreement against ground truth segmentations determined through DSC and S-DSC@2mm
—
Datasets originating from multiple EU and US clinical sites with broad diversity of patients and medical imaging technology.
—
Golden datasets originating from multiple EU and US clinical sites.
>1 (radiotherapy experts)
Indications for Use
Contour+ (MVision Al Segmentation) is a software system for image analysis algorithms to be used in radiation therapy treatment planning workflows. The system includes processing tools for automatic contouring of CT and MR images using machine learning based algorithms. The produced segmentation templates for regions of interest must be transferred to appropriate image visualization systems as an initial template for a medical professional to visualize, review, modify and approve prior to further use in clinical workflows. The system creates initial contours of pre-defined structures of common anatomical sites, i.e., Head and Neck, Brain, Breast, Lung and Abdomen, Male Pelvis, and Female Pelvis. Contour+ (MVision Al Segmentation) is not intended to detect lesions or tumors. The device is not intended for use with real-time adaptive planning workflows.
Device Story
Software-only device for radiotherapy treatment planning; automates ROI delineation on CT and MR images. Inputs: DICOM CT/MR images via local network (TCP/IP) or cloud (HTTPS/TLS). Processing: Pre-trained, locked, static deep artificial neural networks perform automatic contouring. Output: Segmentation structure set transferred to visualization systems (e.g., TPS) as initial template. Usage: Clinicians review, modify, and approve contours before clinical use. Benefits: Reduces upfront contouring effort/time; improves workflow efficiency. Deployment: On-premises or cloud-hosted; supports Ubuntu and Windows platforms.
Clinical Evidence
Bench testing only. Performance evaluated using golden datasets from multiple EU/US clinical sites. Metrics included Dice Score (DSC) and Surface-Dice Score (S-DSC@2mm) compared against ground truth. Results demonstrated robustness and generalizability for US patient populations. Radiotherapy experts validated that the software reduces upfront contouring effort. Software verification and validation followed IEC 62304, IEC 62366-1, and ISO 14971 standards.
Indicated for patients prescribed radiation therapy. Used by clinicians for automatic contouring of CT and MR images to create initial segmentation templates for regions of interest in Head and Neck, Brain, Breast, Lung, Abdomen, and Pelvis. Not for lesion/tumor detection or real-time adaptive planning.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
October 18, 2024
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around a stylized image of a human figure. To the right of the seal, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked above the word "ADMINISTRATION".
MVision AI Ov Kalpana Jha VP of Regulatory and Market Strategy Paciuksenkatu 29, 6th Floor Helsinki, 00270 Finland
Re: K241490
Trade/Device Name: Contour+ (MVision AI Segmentation) Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QKB Dated: June 25, 2024 Received: September 24, 2024
Dear Kalpana Jha:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{3}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Locon Weidner
Lora D. Weidner, Ph.D. Assistant Director Radiation Therapy Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
Submission Number (if known)
K241490
Device Name
Contour+ (MVision Al Segmentation)
#### Indications for Use (Describe)
Contour+ (MVision Al Segmentation) is a software system for image analysis algorithms to be used in radiation therapy treatment planning workflows. The system includes processing tools for automatic contouring of CT and MR images using machine learning based algorithms. The produced seqmentation templates for reqions of interest must be transferred to appropriate image visualization systems as an initial template for a medical professional to visualize, review, modify and approve prior to further use in clinical workflows.
The system creates initial contours of pre-defined structures of common anatomical sites, i.e., Head and Neck, Brain, Breast, Lung and Abdomen, Male Pelvis, and Female Pelvis.
Contour+ (MVision Al Segmentation) is not intended to detect lesions or tumors. The device is not intended for use with real-time adaptive planning workflows.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
### The following information is provided as required by 21 CFR 807.92
| Date Prepared: | June 25, 2024 |
|----------------|---------------|
|----------------|---------------|
**Submitter's Information**
| Company Name and Address: | MVision AI Oy |
|------------------------------------|--------------------------------------------------------------------|
| | Paciuksenkatu 29, 6th floor |
| | 00270 Helsinki, Finland |
| | Tel: +358 (0) 40 721 3783 |
| | <a href="mailto:info@mvision.ai">info@mvision.ai</a> |
| Establishment Registration Number: | 3022745617 |
| Contact Person: | Kalpana Jha |
| | VP of Regulatory and Market Strategy |
| | <a href="mailto:kalpana.jha@mvision.ai">kalpana.jha@mvision.ai</a> |
| | Tel: +358 44 9214 354 |
### Subject Device
| Device Trade Name: | Contour+ (MVision Al Segmentation) |
|-----------------------------|--------------------------------------------------------------|
| Common Name: | Medical Image Segmentation Software |
| Device Classification Name: | Radiological Image Processing Software For Radiation Therapy |
| Product Code: | QKB |
| Device Class: | Class II |
| Review Panel: | Radiology |
| Regulation Description: | Medical Image Management And Processing System |
| Regulation Number: | 21 CFR §892.2050 |
### Predicate Device
| Device Name: | MVision Al Segmentation |
|----------------|-------------------------|
| 510(k) Number: | K212915 |
| Manufacturer: | MVision AI Oy |
This predicate has not been subject to a design-related recall.
### Device Description
Contour+ (MVision Al Segmentation) is a software-only medical device (software system) that can be used to accelerate region of interest (ROI) delineation in radiotherapy treatment planning by automatic contouring of predefined ROIs and the creation of segmentation templates on CT and MR images.
The Contour+ (MVision Al Segmentation) software system is integrated with a customer IT network and configured to receive DICOM CT and MR images, e.g., from a CT or MRI scanner or a treatment planning
{6}------------------------------------------------
system (TPS). Automatic contouring of predefined ROIs is performed by pre-trained, locked, and static models that are based on machine learning using deep artificial neural networks. The models have been trained on several anatomical sites, including the brain, head and neck, bones, breast, lung and abdomen, male pelvis, and female pelvis using hundreds of scans from a diverse patient population. The user does not have to provide any contouring atlases. The resulting segmentation structure set is connected to the original DICOM images and can be transferred to an image visualization system (e.g., a TPS) as an initial template for a medical professional to visualize, modify and approve prior to further use in clinical workflows.
The following is a listing of supported anatomical sites and ROIs of the organs for which the device creates initial contours:
| CT Bones | CT Breast | CT Head and Neck |
|-----------|--------------------|-------------------|
| C1 | A_Aorta | A_Carotid_L/R |
| C2 | A_Aorta_v2 | Arytenoid_L/R |
| C3 | A_Carotid_L/R | Body |
| C4 | A_LAD | Bone_Mandible |
| C5 | A_Pulm | BrachialPlex_L/R |
| C6 | Atrium_L/R | Brain |
| C7 | Bag_Bowel | Brainstem |
| L1 | Body | Buccal_Mucosa_L/R |
| L2 | Br_1234_RTOG_L/R | Cavity_Oral |
| L3 | Br_234_RTOG_L/R | Cochlea_L/R |
| L4 | BrachialPlex_L/R | Cricophar_inlet |
| L5 | Breast_L/R | Esophagus_S |
| Rib01_L/R | Breast_RTOG_L/R | Eye_Ant_L/R |
| Rib02_L/R | BrTW_1234_RTOG_L/R | Eye_Post_L/R |
| Rib03_L/R | BrTW_234_RTOG_L/R | Eye_L/R |
| Rib04_L/R | BrTW_RTOG_L/R | Glnd_Lacrimal_L/R |
| Rib05_L/R | Esophagus | Glnd_Submand_L/R |
| Rib06_L/R | Glnd_Thyroid | Glnd_Thyroid |
| Rib07_L/R | Heart | Glottis |
| Rib08_L/R | Heart+A_Pulm | LN_Neck_IA |
| Rib09_L/R | Humerus_Head_L/R | LN_Neck_IB_L/R |
| Rib10_L/R | Humerus_L/R | LN_Neck_III_L/R |
| Rib11_L/R | Liver | LN_Neck_II_L/R |
| Rib12_L/R | LN_Axilla_RTOG_L/R | LN_Neck_IVA_L/R |
| T1 | LN_Axillary_L/R | LN_Neck_IVB_L/R |
| T10 | LN_B_RTOG_L1_L/R | LN_Neck_IX_L/R |
| T11 | LN_B_RTOG_L2_L/R | LN_Neck_VC_L/R |
| T12 | LN_B_RTOG_L3_L/R | LN_Neck_VIA |
| T2 | LN_B_RTOG_L4_L/R | LN_Neck_VIB |
| T3 | LN_B_RTOG_L5_L/R | LN_Neck_VIIA_L/R |
| T4 | LN_Breast_L1_L/R | LN_Neck_VIIB_L/R |
| T5 | LN_Breast_L2_L/R | LN_Neck_V_L/R |
| T6 | LN_Breast_L3_L/R | LN_Neck_XA_L/R |
{7}------------------------------------------------
| LN_Breast_L4_L/R | LN_Neck_XB_L/R |
|------------------|------------------|
| LN_IMN_IC4_L/R | Larynx_SG |
| LN_IMN_L/R | Lens_L/R |
| LN_Intpect_L/R | Lips |
| LN_RTOG_IMN_L/R | Lung_L/R |
| Lung_L/R | Musc_Constrict |
| SpinalCanal | OpticChiasm |
| SpinalCord | OpticChiasm_cnv |
| Spleen | OpticNrv_L/R |
| Stomach | OpticNrv_cnv_L/R |
| Trachea | Parotid_L/R |
| V_Venacava_I | Pituitary |
| V_Venacava_S | SpinalCanal |
| Ventricle_L/R | SpinalCord |
| Wire | Trachea |
| CT Lung and Abdomen | CT Female Pelvis | CT Male Pelvis |
|---------------------|--------------------|--------------------|
| A_Aorta | A_Aorta | A_Aorta |
| A_Aorta v2 | Bag_Bowel | Bag_Bowel |
| A_LAD | Bladder | Bladder |
| A_Pulm | Body | Body |
| Atrium L/R | Bone_Pelvic | Bone_Pelvic |
| Bag_Bowel | Bowel_Large | Bowel_Large |
| Body | Bowel_Small | Bowel_Small |
| Bones | CTV_Central | CaudaEquina |
| Bowel_Large | CTV_Param | Duodenum |
| Bowel_Small | CTV_Pelvis | Femur_Head_L/R |
| BrachialPlex L/R | Duodenum | Femur_Implant_L/R |
| Bronchus_Prox | Femur_Head_L/R | Femur_L/R |
| Chestwall L/R | Femur_Implant_L/R | Kidney_L/R |
| Duodenum | Femur_L/R | L4_VB |
| Esophagus | Kidney_L/R | L5_VB |
| Heart | L4_VB | Liver |
| Heart+A_Pulm | L5_VB | LN_Inguinal_L/R |
| Humerus_Head_L/R | Liver | LN_NRG |
| Humerus_L/R | LN_Gyn_RTOG | LN_Pivotal |
| Kidney_L/R | LN_Inguinal_L/R | LN_RTOG |
| Liver | LN_PAN | Markers |
| Lung_L/R | LN_PAN_Long | Musc_Coccygeus_L/R |
| Pancreas | LN_Pivotal | Musc_Iliacus_L/R |
| SpinalCanal | LN_RTOG | Musc_Obt_Int_L/R |
| SpinalCord | Musc_Coccygeus_L/R | Musc_Pirifor_L/R |
T7 T8 T9
{8}------------------------------------------------
| Spleen | Musc_Iliacus_L/R | Musc_Psoas_Maj_L/R |
|---------------|--------------------|--------------------|
| Stomach | Musc_Obt_Int_L/R | Pancreas |
| Trachea | Musc_Pirifor_L/R | PenileBulb |
| Trachea_Prox | Musc_Psoas_Maj_L/R | Prostate |
| V_Venacava_I | Pancreas | RectoSigmoid |
| V_Venacava_S | RectoSigmoid | Rectum |
| Ventricle_L/R | Rectum | Sacrum |
| | Sacrum | SeminalVes |
| | SpinalCanal | SpinalCanal |
| | SpinalCord | SpinalCord |
| | Spleen | Spleen |
| | Stomach | Stomach |
| | UteroCervix | V_Venacava_I |
| | V_Venacava_I | Vessels_L/R |
| | Vagina | Vessels_Long_L/R |
| | Vessels_L/R | |
| | Vessels_Long_L/R | |
| CT Brain | MR Brain | MR T2 Male Pelvis |
|-------------------|--------------------|-------------------|
| A_Carotid_L/R | Amygdala_L/R | Bladder |
| Body | Body | BladderTrigone |
| Brain | Brain | PenileBulb |
| Brainstem | Brainstem | Prostate |
| Cochlea_L/R | Cerebellum | Rectum |
| Eye_Ant_L/R | CorpusCallosum | SeminalVes |
| Eye_L/R | Eye_L/R | Spacer |
| Eye_Post_L/R | Glnd_Lacrimal_L/R | |
| Fossa_Pituitary | Hippocampus_L/R | |
| Glnd_Lacrimal_L/R | Hypothalamus | |
| Lens_L/R | MedullaOblongata | |
| OpticChiasm | Midbrain | |
| OpticChiasm_cnv | OpticChiasm | |
| OpticNrv_cnv_L/R | OpticChiasm_cnv | |
| OpticNrv_L/R | OpticNrv_cnv_L/R | |
| Parotid_L/R | OpticNrv_L/R | |
| Pituitary | OpticTract_cnv_L/R | |
| SpinalCanal | OpticTract_L/R | |
| SpinalCord | Pituitary | |
| | Pons | |
| | Thalamus_L/R | |
{9}------------------------------------------------
The Contour+ (MVision Al Segmentation) software system has two deployment modes and is written in a way that allows running the same code in two different environments:
- On-premises (local): For the local Healthcare environment, DICOM image and structure set data ● is transferred via the DICOM TCP/IP protocol.
- Cloud: For the cloud environment, de-identified DICOM image and structure set data are ● transferred via secure HTTPS protocol. HTTPS data is protected by TLS encryption.
### Indication for use / Intended use
Contour+ (MVision Al Segmentation) is a software system for image analysis algorithms to be used in radiation therapy treatment planning workflows. The system includes processing tools for automatic contouring of CT and MR images using machine learning based algorithms. The produced segmentation templates for regions of interest must be transferred to appropriate image visualization systems as an initial template for a medical professional to visualize, review, modify and approve prior to further use in clinical workflows.
The system creates initial contours of pre-defined structures of common anatomical sites, i.e., Head and Neck, Brain, Breast, Lung and Abdomen, Male Pelvis, and Female Pelvis.
Contour+ (MVision Al Segmentation) is not intended to detect lesions or tumors. The device is not intended for use with real-time adaptive planning workflows.
### Comparison with the Predicate Device
Contour+ (MVision AI Segmentation) is an incremental minor version release with the same intended use, indications for use, and principles of operation as the predicate software version.
The following table outlines the similarities and differences between the two software versions:
| | MVision AI Segmentation (K212915) | Contour+ (MVision AI Segmentation) | Comparison |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | MVision AI Segmentation is a software<br>system for image analysis algorithms to be<br>used in radiation therapy treatment<br>planning workflows. The system includes<br>processing tools for automatic contouring<br>of CT images using machine learning<br>based algorithms. The produced<br>segmentation templates for regions of<br>interest must be transferred to appropriate<br>image visualization systems as an initial<br>template for a medical professional to<br>visualize, review, modify and approve<br>prior to further use in clinical workflows.<br>The system creates initial contours of pre-<br>defined structures of common anatomical<br>sites, i.e. Head and Neck, Brain, Breast,<br>Lung and Abdomen, Male Pelvis, and<br>Female Pelvis in adult patients.<br>MVision AI Segmentation is not intended<br>to detect lesions or tumors. The device is<br>not intended for use with real-time<br>adaptive planning workflows. | Contour+ (MVision AI Segmentation) is<br>a software system for image analysis<br>algorithms to be used in radiation therapy<br>treatment planning workflows. The<br>system includes processing tools for<br>automatic contouring of CT and MR<br>images using machine learning based<br>algorithms. The produced segmentation<br>templates for regions of interest must be<br>transferred to appropriate image<br>visualization systems as an initial<br>template for a medical professional to<br>visualize, review, modify and approve<br>prior to further use in clinical workflows.<br>The system creates initial contours of<br>pre-defined structures of common<br>anatomical sites, i.e., Head and Neck,<br>Brain, Breast, Lung and Abdomen, Male<br>Pelvis, and Female Pelvis.<br>Contour+ (MVision AI Segmentation) is<br>not intended to detect lesions or tumors.<br>The device is not intended for use with<br>real-time adaptive planning workflows. | Same intended use.<br>The only difference<br>is the inclusion of<br>processing tools for<br>automatic contouring<br>of MR images, in<br>addition to CT<br>images — this is a<br>change in<br>technological<br>characteristics that<br>does not affect the<br>intended use of the<br>device. |
| | MVision AI Segmentation (K212915) | Contour+ (MVision Al Segmentation) | Comparison |
| Intended<br>Users | Designed to be used by clinicians trained<br>in radiation therapy workflows. | Designed to be used by clinicians trained<br>in radiation therapy workflows. | Same. |
| Target Patient<br>Population | Patients who have been prescribed<br>radiation therapy. | Patients who have been prescribed<br>radiation therapy. | Same |
| Deployment<br>Environment | • Cloud based software Application<br>• Local (on customer premises)<br>installation in healthcare provider's IT<br>network / server | • Cloud based software Application<br>• Local (on customer premises)<br>installation in healthcare provider's IT<br>network / server | Same |
| Operating<br>Platform | • Ubuntu | • Ubuntu<br>• Windows | Contour+ may also<br>run in a Windows<br>operating platform |
| Communications /<br>Networking | DICOM image and structure set data<br>transfers<br>• TCP/IP, SCP, and HTTP (Local)<br>• TCP/IP, SCP, and HTTPS (Cloud) | DICOM image and structure set data<br>transfers<br>• TCP/IP, SCP, and HTTP (Local)<br>• TCP/IP, SCP, and HTTPS (Cloud) | Minor changes:<br>Added API script to<br>send scans directly<br>from the Varian<br>Eclipse Treatment<br>Planning System;<br>Provided read-only<br>Import/Export UI to<br>upload/download<br>DICOM data via a<br>web browser |
| Imaging<br>modalities | • CT | • CT<br>• MR | Added support for<br>automatic contouring<br>of MR images |
| Segmentation<br>algorithms | • Machine learning-based contouring<br>using deep learning-based models<br>• Does not support segmentation based<br>on atlas-based techniques | • Machine learning-based contouring<br>using deep learning-based models<br>• Does not support segmentation based<br>on atlas-based techniques | Same algorithms;<br>Updated certain<br>CT segmentation<br>models; Added 1 CT<br>and 2 MR models |
| Registration based<br>re-contouring | No | No | Same |
{10}------------------------------------------------
{11}------------------------------------------------
The different technological characteristics of the two software versions do not raise different questions of safety and effectiveness.
- Both software versions are intended to accelerate region of interest delineation in radiological images . by automatic contouring and the creation of segmentation templates to be used by qualified healthcare professionals in radiation therapy treatment planning workflows.
- Both software versions are not intended to detect lesions or tumors, and they are not intended for use with real-time adaptive planning workflows.
- Both software versions are intended to create initial contours of predefined structures of common ● anatomical sites, such as head and neck, brain, breast, lung and abdomen, male pelvis, and female pelvis on scans that are appropriate for treatment planning. The addition of a model to segment specifically bone structures on CT images and the addition of models for the segmentation of MR images does not change the device intended use and indications for use.
- Both software versions are intended to produce segmentation templates for regions of interest that must ● be transferred to appropriate image visualization systems as an initial template for a medical professional to visualize, review, modify and approve prior to further use in clinical workflows.
- As radiotherapy treatment planning may be performed effectively using CT and/or MR images, 0 automatic contouring for the creation of segmentation templates of predefined structures of anatomical regions of interest in images from either modality offers the same benefits and presents the same risks.
- In both software versions, automatic contouring of images (segmentation of anatomical regions of ● interest) is performed by pre-trained, locked, and static models that are based on machine learning using the same deep artificial neural networks. The segmentation algorithms remain the same.
## Performance data - Verification and Validation Testing
Verification and validation testing of this software release was also conducted as per FDA's Guidance for the "Content of Premarket Submissions for Device Software Functions (2023)," including compliance with recognized consensus standards (e.g., IEC 62304, IEC 62366-1, ISO 14971, DICOM) and FDA guidance for "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions (2023)," and "Design Considerations and Pre-market Submission Recommendations for Interoperable Medical Devices (2017)."
Device performance with the updated CT models and the additional CT and MR models, i.e., the effectiveness of the automatic segmentation methods (machine learning-based models) in the modified software version, was evaluated using the same data management practices (including data sources, sample size, and model generalizability) and established test protocols as previously reviewed by the FDA.
The training and the test (golden) datasets were chosen to achieve high granularity in performance evaluation tests. The datasets originate from multiple EU and US clinical sites (with over 50% of data coming from US sites) with a broad diversity of patients and medical imaging technology that are deemed representative of the US population and medical practice. This also ensures that the evaluated model performance reflects the real clinical performance in any radiotherapy clinic following the segmentation consensus guidelines the models are trained to comply with.
The performance of both CT and MR automatic segmentation models was evaluated by comparing the produced auto-segmentations to ground truth segmentations and calculating scores DSC (Dice Score) and S-DSC@2mm (Surface-Dice Score) for all regions of interest (ROI). The ROI acceptance criteria are based on a set level of minimum agreement against ground truth segmentations determined through clinically relevant similarity metrics DSC and S-DSC(@2mm.
{12}------------------------------------------------
Performance verification and validation results for various subsets of the golden dataset show the generalizability and robustness of the device for the US patient population and US medical practice.
Performance validation of machine learning-based algorithms for automatic segmentation was also carried out by radiotherapy experts. The results show that Contour+ (MVision AI Segmentation) assists in reducing the upfront effort and time required for contouring CT and MR images, which can instead be devoted by clinicians on refining and reviewing the software-generated contours.
The additional functional interoperability changes (the Varian Eclipse TPS, and the Import/Export UI) and the labeling change to support deployment of the software on a Windows operating platform do not have an impact on device safety or effectiveness. Successful software execution in a Windows operating environment and the functional interoperability changes was performed using the same established design verification test protocols as for the predicate software version per Product Life Cycle Procedure, based on FDA recognized consensus standards and design control regulation 21 CFR 820.30.
### Conclusion
Contour+ (MVision AI Segmentation) is an incremental minor version release with the same intended use, indications for use, and principles of operation as the predicate software version. The different technological characteristics of the two software versions do not raise different questions of safety and effectiveness. Software verification and validation testing, including the performance of all CT and MR models using established protocols and recognized standards demonstrate that Contour+ (MVision AI Segmentation) fulfills the same acceptance criteria, provides the intended benefits, and it is as safe and as effective as the predicate software version (MVision AI Segmentation).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.