K212915 · MVision AI · QKB · May 3, 2022 · Radiology
Device Facts
Record ID
K212915
Device Name
MVision AI Segmentation
Applicant
MVision AI
Product Code
QKB · Radiology
Decision Date
May 3, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Automatic contouring of anatomical structures
Deep artificial neural networks
—
—
—
—
—
>1 (radiotherapy experts)
Indications for Use
MVision AI Segmentation is a software system for image analysis algorithms to be used in radiation therapy treatment planning workflows. The system includes processing tools for automatic contouring of CT images using machine learning based algorithms. The produced segmentation templates for regions of interest must be transferred to appropriate image visualization systems as an initial template for a medical professional to visualize, review, modify and approve prior to further use in clinical workflows. MVision AI Segmentation is not intended to detect lesions or tumors. The device is not intended for use with real-time adaptive planning workflows.
Device Story
Software-only device for radiotherapy treatment planning; automates ROI delineation on CT images. Inputs: DICOM CT images from scanners or treatment planning systems (TPS). Processing: Pre-trained, locked deep artificial neural networks generate segmentation templates. Output: Structure sets (contours) exported via DICOM to visualization systems. Usage: Clinical environment (on-premises or cloud); operated by medical professionals. Workflow: Device provides initial contours; clinician reviews, modifies, and approves results before clinical use. Benefit: Reduces manual contouring time/effort, accelerating treatment planning.
Clinical Evidence
Bench testing only. No clinical studies or animal data. Performance evaluation used golden datasets from multiple sources to assess generalizability and robustness. Validation by radiotherapy experts confirmed device reduces upfront contouring effort and time, providing usable templates for clinical workflows.
Technological Characteristics
Software-only device. Deep artificial neural network architecture (locked, pre-trained models). Connectivity: DICOM TCP/IP (on-premises) or secure HTTPS/TLS (cloud). Standards: Adherence to DICOM. Deployment: Standalone software. No physical materials.
Indications for Use
Indicated for automatic contouring of CT images for radiation therapy treatment planning in adult patients. Anatomical sites include Head and Neck, Brain, Breast, Lung and Abdomen, Male Pelvis, and Female Pelvis. Contraindicated for lesion or tumor detection and real-time adaptive planning workflows.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
May 3, 2022
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
MVision AI Kalpana Jha Quality and Compliance Manager C/O Terkko Health Hub Haartmaninkatu 4 Helsinki, 00290 FINLAND
Re: K212915
Trade/Device Name: MVision AI Segmentation Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QKB Dated: March 29, 2022 Received: April 1, 2022
Dear Kalpana Jha:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael O'Hara, Ph.D. Deputy Director Division of Radiological Imaging and Radiation Therapy OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212915
Device Name MVision AI Segmentation
### Indications for Use (Describe)
MVision AI Segmentation is a software system for image analysis algorithms to be used in radiation therapy treatment planning workflows. The system includes processing tools for automatic contouring of CT images using machine learning based algorithms. The produced segmentation templates for regions of interest must be transferred to appropriate image visualization systems as an initial template for a medical professional to visualize, review, modify and approve prior to further use in clinical workflows.
The system creates initial contours of pre-defined structures of common anatomical sites, i.e. Head and Neck, Brain, Breast, Lung and Abdomen, Male Pelvis, and Female Pelvis in adult patients.
MVision AI Segmentation is not intended to detect lesions or tumors. The device is not intended for use with real-time adaptive planning workflows.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Below mentioned information is provided as required by 21 CFR 807.92
Submitter's Information
| Name and Address | MVision AI<br>C/O Terkko Health Hub<br>Haartmaninkatu 4<br>00290 Helsinki<br>Finland<br>Tel: +358 (0) 40 5489 229<br>info@mvision.ai |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Primary Contact Person | Kalpana Jha<br>Quality and Compliance Manager<br>kalpana.jha@mvision.ai |
# Subject Device
| Subject Device | MVision AI Segmentation, New Software Medical Device |
|---------------------------------|-----------------------------------------------------------------------------------------------------|
| Trade Name | MVision Segmentation Service |
| Common Name | Medical Image Segmentation Software |
| Product Code and Classification | Medical image management and processing system.<br>OKB 21 CFR §892.2050 Class II Medical Device |
Tel: +358 44 9214 354
## Predicate Device
Predicate Device - Mirada Workflow Box, Software Medical Device (K181572)
This predicate has not been subject to a design-related recall.
# Device Description
MVision AI Segmentation is a software only medical device which can be used to accelerate region of interest (ROI) delineation in radiotherapy treatment planning by creating automatic segmentation templates on CT images for these ROIs.
The segmentations are produced by pre-trained, locked, and static models that are based on deep artificial neural networks. The produced structure is intended to be used as a template for medical professionals to visualize, modify and approve prior to further use in clinical workflows.
The system is integrated with the customer IT network to receive DICOM images. CT images from, for example, a scanner or a treatment planning system (TPS) are exported to the device. A structure set is created in the device, and the created segmentation results are connected to the original images. These data are sent to the destination DICOM import folder to import the data to, for example, a treatment planning system. The produced structures can then be used as a template for manual ROI editing, review
{5}------------------------------------------------
# 510(k) Summary
and approval workflow. The segmentations are produced by pre-trained and locked models that are based on deep artificial neural networks. To take the device into use, the user does not have to provide any contouring atlases. The models have been trained with the order of hundreds of scans, depending on the ROI in question. The MVision AI Segmentation device creates initial contours of pre-defined structures of common anatomical sites, i.e. Head and Neck, Brain, Breast, Lung and Abdomen, Male Pelvis, and Female Pelvis. Supported anatomical sites and ROI listing of the organs for which the device creates initial contours for, are listed below:
| Lung and Abdomen | Brain | Breast |
|------------------|-------------------|------------------|
| A_Aorta | Body | A_Aorta |
| A_LAD | Brain | A_Carotid_R/L |
| Body | Brainstem | A_LAD |
| Bowel_Large | Cochlea_R/L | Body |
| Bowel_Small | Eye_Ant_R/L | BrachialPlex_R/L |
| BrachialPlex_R/L | Eye_Post_R/L | Breast_R/L |
| Bronchus_Prox | Eye_R/L | Glnd_Thyroid |
| Chestwall_R/L | Glnd_Lacrimal_R/L | Heart |
| Esophagus | Lens_R/L | Humerus_R/L |
| Heart | OpticChiasm | LN_Axillary_R/L |
| Heart+A_Pulm | OpticChiasm_cnv | LN_Breast_L1_R/L |
| Kidney_R/L | OpticNrv_R/L | LN_Breast_L2_R/L |
| Liver | OpticNrv_cnv_R/L | LN_Breast_L3_R/L |
| Lung_R/L | Pituitary | LN_Breast_L4_R/L |
| Pancreas | | LN_IMN_IC4_R/L |
| SpinalCanal | | LN_IMN_R/L |
| SpinalCord | | LN_Intpect_R/L |
| Spleen | | Lung_R/L |
| Stomach | | SpinalCanal |
| Trachea | | SpinalCord |
| Trachea_Prox | | Trachea |
| V_Venacava_I | | V_Venacava_I |
| V_Venacava_S | | |
{6}------------------------------------------------
# 510(k) Summary
| Female Pelvis | Head and Neck | Male Pelvis |
|---------------|-------------------|--------------------|
| Bag_Bowel | A_Carotid_R/L | A_Aorta |
| Bladder | Arytenoid_R/L | Bag_Bowel |
| Body | Body | Bladder |
| Bone_Pelvic | Bone_Mandible | Body |
| Bowel_Large | BrachialPlex_R/L | Bone_Pelvic |
| Bowel_Small | Brain | Bowel_Large |
| Femur_R/L | Brainstem | Bowel_Small |
| Kidney_R/L | Buccal_Mucosa_R/L | Femur_R/L |
| L4_VB | Cavity_Oral | Kidney_R/L |
| L5_VB | Cochlea_R/L | L4_VB |
| Rectum | Cricophar_inlet | L5_VB |
| Sacrum | Esophagus_S | LN_Pivotal |
| SpinalCanal | Eye_Ant_R/L | LN_RTOG |
| UteroCervix | Eye_Post_R/L | Markers |
| | Eye_R/L | Musc_Coccygeus_R/L |
| | Glnd_Lacrimal_R/L | Musc_Iliacus_R/L |
| | Glnd_Submand_R/L | Musc_Obt_Int_R/L |
| | Glnd_Thyroid | Musc_Pirifor_R/L |
| | Glottis | Musc_Psoas_Maj_R/L |
| | LN_Neck_IA | PenileBulb |
| | LN_Neck_IB_R/L | Prostate |
| | LN_Neck_III_R/L | Rectum |
| | LN_Neck_II_R/L | Sacrum |
| | LN_Neck_IVA_R/L | SeminalVes |
| | LN_Neck_IVB_R/L | V_Venacava_I |
| | LN_Neck_IX_R/L | Vessels_Long_R/L |
| | LN_Neck_VC_R/L | Vessels_R/L |
| | LN_Neck_VIA | |
| | LN_Neck_VIB | |
| | LN_Neck_VIIA_R/L | |
| | LN_Neck_VIIB_R/L | |
| | LN_Neck_V_R/L | |
| | LN_Neck_XA_R/L | |
| | LN_Neck_XB_R/L | |
| | Larynx_SG | |
| | Lens_R/L | |
| | Lips | |
| | Lung_R/L | |
| | Musc_Constrict | |
| | OpticChiasm | |
| | OpticChiasm_cnv | |
| | OpticNrv_R/L | |
| | OpticNrv_cnv_R/L | |
| | Parotid_R/L | |
| | Pituitary | |
| | SpinalCanal | |
| | SpinalCord | |
| | Trachea | |
{7}------------------------------------------------
The MVision AI Segmentation software has two deployment modes and is written in a way that allows running the same code in two different environments:
On-premises (local): For the local Healthcare environment, DICOM image and structure set data is transferred via the DICOM TCP/IP protocol for subject device and predicate device.
Cloud: For the cloud environment, de-identified DICOM image and structure set data is transferred via secure HTTPS protocol and HTTPS data is protected by TLS encryption for the subject device.
# Indication for use / Condition of use
MVision AI Segmentation is a software system for image analysis algorithms to be used in radiation therapy treatment planning workflows. The system includes processing tools for automatic contouring of CT images using machine learning based algorithms. The produced segmentation templates for regions of interest must be transferred to appropriate image visualization systems as an initial template for a medical professional to visualize, review, modify and approve prior to further use in clinical workflows.
The system creates initial contours of pre-defined structures of common anatomical sites, i.e. Head and Neck, Brain, Breast, Lung and Abdomen, Male Pelvis, and Female Pelvis in adult patients.
MVision AI Segmentation is not intended to detect lesions or tumors. The device is not intended for use with real-time adaptive planning workflows.
## Intended Use
MVision AI Segmentation is a software system for image analysis algorithms to be used in radiation therapy treatment planning workflows. The system includes processing tools for automatic contouring of CT images using machine learning based algorithms. The produced segmentation templates for regions of interest must be transferred to appropriate image visualization systems as an initial template for a medical professional to visualize, review, modify and approve prior to further use in clinical workflows.
MVision AI Segmentation is not intended to detect lesions or tumors. The device is not intended for use with real-time adaptive planning workflows.
Comparison Summary of Intended use with Predicate device
MVision AI Segmentation and the predicate device are identical with regard to intended use; specifically, for automatic contouring of imaging data to be used in radiotherapy treatment planning workflows. Both the subject and predicate devices are software only medical devices and output contours of both medical devices are intended to be visualized, edited as appropriate and approved by trained medical professionals for further use in clinical workflows. The review and editing of contouring results from both the subject and predicate devices can be performed using different image visualization systems. Both the subject and predicate devices are not intended to be used for automatic diagnosis of disease. Therefore, the intended use of the subject device MVision AI Segmentation is substantially equivalent to the predicate device Workflow Box.
{8}------------------------------------------------
# 510(k) Summa
# Comparison of Technological Characteristics with Predicate Device
MVision AI Segmentation has similar technological characteristics as the predicate device. Automated segmentation of the patient anatomy is the technological principle for both the subject and predicate devices.
At a high level, the subject and predicate devices are based on the following same technological elements:
- Both devices are software only medical devices
- Both devices include automatic segmentation algorithms to process radiological images in ● order to produce contouring of human anatomical regions
- Both devices are intended to be used by trained medical professionals
- Both the subject and the predicate device is not intended to be used for automatic diagnosis of disease
- . Output contours of both devices are intended to be visualized, modified and approved by a medical professional for further use in clinical workflows
- Both the predicate and the subject device do not inherently contain material as they are ● software-only devices
The following technological differences exist between the subject and predicate devices.
- . MVision AI Segmentation is intended for automatic contouring of imaging data using machine learning based algorithms whereas predicate device offers automatic contouring of imaging data using both atlas based deformable image registration as well as machine learning algorithm.
- Subject device MVision AI Segmentation is operated either by the manufacturer in the . manufacturer's cloud or by a healthcare provider in the healthcare provider's IT network. Predicate device Workflow Box can be operated only in a healthcare provider's IT network.
# Performance data
Software verification and validation testing were conducted as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a major level of concern. Verification and Validation for MVision AI Segmentation has been carried out in compliance with the requirements of CFR 21 Part 820 and in adherence to the DICOM standard.
Software verification and validation, performance evaluation for machine learning based algorithms, and implemented Cybersecurity control measures and tests establish that the subject medical device is safe, secure and effective for its user needs and defined intended use.
No animal studies or clinical tests have been included in this pre-market submission.
# Performance Evaluation Summary
Training and test sets (golden dataset) were collected and created to include high granularity in performance evaluation tests. They originate from multiple different sources to make sure the evaluated
{9}------------------------------------------------
# 510(k) Summary
model performance will reflect the real clinical performance in any radiotherapy clinic following the segmentation consensus guidelines the models are trained to comply with.
Performance verification results for various subsets of the golden dataset show the generalizability and robustness of the device for the US patient population and US medical practice. Furthermore, Performance Validation results for machine learning based algorithms carried out by radiotherapy experts show that MVision AI Segmentation assists in reducing the upfront effort and time on typical contouring which can be spent on refining and reviewing the results. Performance validation data further suggests that the subject device produces usable contours (ROIs) as a starting point that will save clinicians' time and it will lead to sooner proceeding to essential parts of radiotherapy treatment planning stages.
## Conclusion
The intended use of the subject device MVision AI Segmentation is substantially equivalent to the predicate device Mirada Workflow Box. MVision AI Segmentation has similar technological characteristics as the predicate device and differences in technological characteristics do not raise different questions of safety and effectiveness.
Software verification and validation and Performance evaluation tests for machine learning based algorithms establish that the subject medical device is non-inferior, performs safely and effectively as the listed predicate device.
Hence, MVision AI Segmentation is substantially equivalent to the predicate device Mirada Workflow Box.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.