K241447 · Icare Finland OY · HKY · Sep 27, 2024 · Ophthalmic
Device Facts
Record ID
K241447
Device Name
iCare ST500 (TA04)
Applicant
Icare Finland OY
Product Code
HKY · Ophthalmic
Decision Date
Sep 27, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1930
Device Class
Class 2
Indications for Use
The iCare ST500 tonometer is a device intended to be used for the measurement of the intraocular pressure of the human eye. The iCare ST500 tonometer is intended to be used by healthcare professionals.
Device Story
The iCare ST500 is a slit-lamp-mounted rebound tonometer for clinical eye exams. It uses a magnetized, single-use probe launched toward the cornea; the probe bounces off the corneal surface. The device measures probe motion parameters—including deceleration and rebound time—via an induction coil. A proprietary algorithm processes these signals to calculate IOP. The system includes the tonometer, a remote control, and a smart cradle for charging. It is operated by healthcare professionals in a clinic setting. The device provides automated IOP readings, which are displayed on an LCD screen and stored in memory. The output assists clinicians in assessing ocular health and managing conditions related to intraocular pressure. Benefits include rapid, anesthetic-free measurement.
Clinical Evidence
Clinical study (n=156 eyes) compared iCare ST500 to GAT and iCare IC200. Primary endpoint: IOP measurement accuracy per ANSI Z80.10-2014. Results showed the ST500 underestimated IOP by an average of 0.21 mmHg compared to GAT. Regression analysis showed a slope of 1.03 and a squared Pearson correlation coefficient of 0.89. Outlier rate was 2.6%. Bench testing demonstrated high repeatability (SD ≤ 0.78 mmHg) and reproducibility across operators (R-squared 99.7-99.8%).
Technological Characteristics
Rebound tonometry principle using magnetized single-use probes. Slit-lamp mounted form factor. Connectivity via Bluetooth (ANNA-B112 module). Power: 3.6V Li-ion rechargeable battery. Display: 240x320 RGB LCD. Software Class B. Sterilization: Gamma-sterilized probes. Complies with ISO 15004-1, ANSI Z80.36, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 62366-1, IEC 62304, ISO 10993-1, and ANSI Z80.10.
Indications for Use
Indicated for measurement of intraocular pressure (IOP) in human eyes by healthcare professionals. No specific age or gender restrictions; contraindicated for use outside of slit-lamp mounting (not for handheld use).
Regulatory Classification
Identification
A tonometer and accessories is a manual device intended to measure intraocular pressure by applying a known force on the globe of the eye and measuring the amount of indentation produced (Schiotz type) or to measure intraocular tension by applanation (applying a small flat disk to the cornea). Accessories for the device may include a tonometer calibrator or a tonograph recording system. The device is intended for use in the diagnosis of glaucoma.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 27, 2024
Icare Finland Oy Hannes Hyvönen Regulatory Affairs Manager Ayritie 22 Vantaa. 01510 Finland
Re: K241447
Trade/Device Name: iCare ST500 (TA04) Regulation Number: 21 CFR 886.1930 Regulation Name: Tonometer And Accessories Regulatory Class: Class II Product Code: HKY, HKX Dated: August 20, 2024 Received: August 20, 2024
Dear Hannes Hyvönen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Alexander 2024.09.27
Beylin -S 12:03:30 -04'00'
for Elvin Ng
Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K241447
Device Name iCare ST500 (TA04)
Indications for Use (Describe)
The iCare ST500 tonometer is a device intended to be used for the intraocular pressure of the human eye.
The iCare ST500 tonometer is intended to be used by healthcare professionals.
Type of Use (Select one or both, as applicable)
| <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------|
| <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
## 1.1 Contact Details
Applicant Name: Icare Finland Oy Applicant Address: Äyritie 22 Vantaa 01510 Finland Applicant Contact Telephone: 358 9 8875 1150 Applicant Contact: Mr. Hannes Hyvönen Applicant Contact Email: regulatory@icare-world.com
Prepared on: September 25th, 2024
## 1.2 Device Name
| Device Trade Name: | iCare ST500 (TA04) |
|----------------------|---------------------------|
| Common Name: | Tonometer and accessories |
| Classification Name: | Tonometer, Manual |
| Regulation Number: | 886.1930 |
| Product Code(s): | HKY, HKX |
Premarket Notification 510(k) Number: K241447 This premarket notification type is Traditional 510(k).
## 1.3 Legally Marketed Predicate Devices
Predicate #: K220852 Predicate Trade Name (Primary Predicate is listed first): iCare IC200 (TA031) Product Code: HKY
## 1.4 Device Description Summary
The iCare ST500 tonometer is based on a patented, induction-based rebound method, which allows intraocular pressure (IOP) to be measured by a healthcare professional accurately, rapidly and without an anesthetic. The iCare ST500 tonometer measures a
{5}------------------------------------------------
patient's IOP while he or she is in an upright (sitting) position. It is attached to a slit lamp with the ST500 Adapter, making it easy to use during a standard eye exam. This tonometer is designed only for use with a slit lamp not as a handheld device. The tonometer is compatible with most of the common slit lamp models on the market.
The tonometer system consists of the iCare ST500 tonometer (together with ST500 Remote and ST500 Adapter, ST500 SmartCradle and a medical power supply used with ST500 SmartCradle), and a compatible printer and slit lamp.
With the iCare rebound measurement method, a miniature, lightweight, single-use probe is launched in a direction perpendicular to the surface of the center of the cornea of the eye. The probe is composed of a medical grade plastic tip and a metal shaft. The metal shaft is magnetized prior to the measurement. During the measurement, the probe can be considered to act like a moving magnet that induces an electric signal in the surrounding coil allowing for a highly accurate measurement of the probe's motion. After launching. the probe briefly contacts the cornea and bounces back. The tonometer records multiple parameters covering the motion of the probe, including deceleration, and rebound time. Using a proprietary algorithm, the device can calculate the eye's IOP.
The displayed IOP reading is derived from the results from a sequence of six individual probe rebounds. The displayed IOP measurement is also stored in the tonometer's memory for later retrieval.
The iCare ST500 tonometer has a rechargeable battery and can be charged in the ST500 SmartCradle powered by a medical power supply. The tonometer can be operated by using the buttons on the tonometer or by using the wireless battery-operated ST500 Remote control.
{6}------------------------------------------------
## 1.5 | Intended Use / Indications for Use
The iCare ST500 tonometer is a device intended to be used for the measurement of the intraocular pressure of the human eye.
The iCare ST500 tonometer is intended to be used by healthcare professionals.
## 1.6 Indications for Use Comparison
Indications of use statement is equivalent to the predicate device (see Table 1 below).
## 1.7 Technological Comparison
The new iCare ST500 tonometer features the same or equivalent technology as in the predicate device iCare IC200 (K220852) such as:
• Automated measurement sequence, where full measurement sequence can be completed by keeping the measurement button pressed down.
- Wireless connectivity with Bluetooth connection to printer and remote control.
• The Quick Measure feature measures the patient's IOP with fewer rebound measurements and a faster measurement cycle.
Differences with respect to the predicate device iCare IC200 (K220852) are:
• Mounted on a slit lamp with the ST500 adapter, allowing the measurement to be performed only in horizontal (sitting) position.
- Automatic eye side recognition system, an automatic IR-based eye side detection and control functionality. Equivalent to reference device iCare HOME2 (K211355).
- · Probe applicator, a funnel shaped applicator used to ease the loading of the probe. Equivalent to reference device iCare HOME2 (K211355).
- Remote control, the tonometer can be operated by using the buttons on the tonometer or by using the wireless battery-operated ST500 Remote control.
- Rechargeable battery, the iCare ST500 tonometer has a rechargeable battery and can be charged in the ST500 SmartCradle powered by a medical power supply.
{7}------------------------------------------------
| Characteristic | Subject device (K241447) | Predicate Device (K220852) |
|-----------------|----------------------------------------------------|----------------------------------------------------|
| Product | | |
| Appearance | Image: iCare ST500 tonometer with a mannequin head | Image: iCare IC200 Tonometer being used on a child |
| Product/Devic | iCare ST500 tonometer | iCare IC200 Tonometer |
| e | (Type: TA04) | (Type: TA031) |
| Identification | | |
| Intended Use / | Equivalent | IOP Measurement |
| Indications for | "The iCare ST500 tonometer is a device | "The iCare IC200 tonometer is intended to be |
| Use Statement | intended to be used for the measurement of the | used for the measurement of intraocular |
| | intraocular pressure of the human eye. | pressure of the human eye." |
| | The iCare ST500 tonometer is intended to be | |
| | used by healthcare professionals." | |
| Intended users | Same | Healthcare professionals |
| Measurement | Same | Rebound tonometry |
| method | | |
| Measurement | Same | 7-50 mmHg |
| range | | |
| Default | Equivalent, | Default measurement mode: 6 |
| measurement | Default measurement mode: 6 measurements | measurements, calculation of median IOP |
| sequence | are averaged | and standard deviation based on 4 |
| | | measurements (highest and lowest |
| | | excluded) |
{8}------------------------------------------------
# icare
| Measurement<br>positions | Equivalent<br>Tonometer must be oriented horizontally (0°,<br>patient in sitting position), no supine position<br>measurement possibility, only for slit lamp use<br>not as a handheld device | Tonometer can be used in any angle between 0°<br>(sitting, standing) and 90° (patient in supine<br>position) |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| Device Display | Equivalent,<br>240 x 320 RGB LCD display | 128 x 128 full color OLED display |
| Automatic eye<br>recognition<br>system | Difference,<br>but equivalent to reference device iCare<br>HOME2 (K211355) | No |
| Design | Equivalent,<br>Slit-lamp mounted microprocessor based | Handheld microprocessor based |
| Calibration | Same | No maintenance calibration required |
| Contact tip<br>(probe) | Same | Lightweight, disposable, single use, plastic<br>probe (TP01s) |
| Contact tip<br>sterilization | Same | Gamma-sterilized |
| Anaesthesia<br>required | Same | No |
| Power supply | Difference,<br>Li-ion rechargeable battery, 3,6 V / 2200-2400<br>mAh<br>(not replaceable by the user) | 4 x 1,5V AA Alkaline LR6 batteries |
| Device<br>dimensions<br>and weight | Equivalent,<br>Dimensions:<br>46 x 70 x 181 mm (L x W x H)<br>Weight: 226g (including internal power<br>supply) | Dimensions:<br>43 mm (W) * 104 mm (H) * 214 mm (L)<br>Weight: 165g (without batteries) |
| Connectivity<br>interface | Equivalent,<br>Bluetooth (ANNA-B112 Module) | Bluetooth (Microchip RN4678 Module) |
| User interface | Difference,<br>Near Field Communication (NFC) | Audio indications: |
{9}------------------------------------------------
| Equivalent | Beeps for device too near or too far situation,<br>and for other errors. |
|-----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Graphical User Interface:<br>Equivalent,<br>240 x 320 RGB LCD display | Graphical User Interface:<br>128 x 128 full color OLED display |
| Probe base light:<br>Green: same<br>Red: same<br>Blinking red: same | Probe base light:<br>Green: angle correct, device readiness<br>Red: angle incorrect, measurement not possible<br>Blinking red: measurement error messages |
The iCare ST500 has the same basic operation principle, underlying technology (rebound tonometry) and device materials as the predicate device. The device performance is substantially equivalent.
### Non-Clinical Performance Testing 1.8
The device has been tested based on FDA recognized consensus standards. The following information shows its performance and safety, supporting its substantial equivalence.:
-ISO 15004-1 Second edition 2020-5 (10-123)
-ANSI Z80.36-2021 (10-130)
-IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION (19-49)
-ANSI AAMI IEC 60601-1-2:2014 [Including AMD 1:2021] (19-36)
-IEC 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION (5-132)
-ANSI AAMI IEC 62366-1:2015+AMD1:2020 (Consolidated Text) (5-129)
-ANSI AAMI IEC 62304:2006/A1:2016 (13-79)
-ISO 10993-1:2018 (2-258)
-ANSI Z80.10-2014 (10-96)
There were no changes related to sterilization (of probes). Device and software risk analysis has been performed in accordance with ISO 14971:2019 risk management
{10}------------------------------------------------
standard. Device firmware and software level of concern was Class B (Basic Documentation).
Summary of bench testing performed is presented below:
The accuracy and repeatability of the Icare ST500 tonometer was assessed in a bench test. The test was done by measuring manometrically controlled test corneas. To assess accuracy and repeatability, measurements were performed with the test tonometer using Default Measure and Quick Measure.
Icare ST500 met the predetermined acceptance criteria for accuracy and repeatability over the measurement range ensuring substantial equivalence to the previously cleared Icare IC200 (K190316) tonometer both for Default Measure and Quick Measure.
For repeatability with Default Measure, the highest standard deviation was 0.44 mmHg (CV 1.1 %). With Quick Measure, the highest standard deviation was 0.78 (CV 1.9%).
Reproducibility was assessed in a test in which two operators performed three measurements with three different ST500 units. Across all the three iCare ST500 units, the mean differences between the operators were -0.12 mmHg and -0.02 mmHg with a standard deviations of 0.75 mmHg and 0.91 mmHg for Default Measure and Quick Measure, respectively. The R-squared value in regression analysis was 99.8% for Default Measure and 99.7% for Quick Measure, which indicates high reproducibility across the operators for both Default Measure and Quick Measure of iCare ST500 tonometer.
The iCare ST500 has two measurement modes: Single and Series. In the Single Mode, the probe is shot every time the measurement button is pressed. In the Series Mode, the measurement button is pressed and held, and the probe is shot consecutively until the measurement is finished or the measurement button is released. The performance of ST500 is not affected the mode used.
Please be advised that bench testing conditions do not cover all the error sources within a clinical setting and thus higher variability is expected in clinical use.
# 1.9 Clinical Performance Testing
The clinical study was made for iCare ST500 tonometer according to ANSI Z80.10:2014 (in accordance with FDA's extent of recognition). See Table 2 below.
{11}------------------------------------------------
| Mean GAT – 1st ST500 | | | | |
|---------------------------|-----------|-----------------|-----------|-------|
| | ≤ 16 mmHg | 16 – 23<br>mmHg | ≥ 23 mmHg | Total |
| Number of eyes | 59 | 57 | 40 | 156 |
| Number of outlier | 0 | 2 | 2 | 4 |
| Percentage of<br>outliers | 0.0% | 3.5% | 5.0% | 2.6% |
Table 2 – iCare ST500 1st measurement vs. GAT mean
There were 165 subjects enrolled in the study of which 156 had at least one eligible eye. The measurements were made with iCare ST500 and reference tonometers iCare IC200 and GAT. In case both subject's eyes were eligible, the eye with higher GAT reference pressure was used in the analysis (so the number of eyes included in the analysis was 156). The population of the study consisted of subjects having an appointment to the investigational site and from subjects recruited using a study announcement.
On average iCare ST500 underestimated IOP by 0.21 mmHg compared to GAT. The IOP groups (per standard) were low (≤ 16 mmHg), medium (> 16 to < 23 mmHg) and high (≥ 23 mmHg). There was underestimation in low (0.20 mmHg) and medium (0.49 mmHg) IOP groups and overestimation in medium (0.18 mmHg) IOP group. Regression analyses confirmed that iCare ST500 results correlate well compared to GAT results: the slope and the squared Pearson coefficient of correlation were close to 1 (1.03 and 0.89, respectively). In conclusion, iCare ST500 met the requirements of ANSI Z80.10-2014.
## 1.10 Conclusions
Based on the nonclinical and clinical tests iCare ST500 tonometer is safe and performed as intended when used according to its indications for use and in accordance with its labeling. It performs as well as the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.