K240765 · Ventris Medical · MQV · Dec 18, 2024 · Orthopedic
Device Facts
Record ID
K240765
Device Name
Porous Biologic Scaffold
Applicant
Ventris Medical
Product Code
MQV · Orthopedic
Decision Date
Dec 18, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Porous Biologic Scaffold is a bone void filler intended for use in bony voids or gaps of the skeletal system that are not intrinsic to the stability of the bony structure. These defects may be surgically created or a result of traumatic injury to the bone. Porous Biologic Scaffold is indicated to be packed gently into bony voids or gaps of the pelvis and posterolateral spine and is used combined with autograft as a bone graft extender. The device is resorbed and replaced with host bone during the healing process.
Device Story
Resorbable bone void filler; two versions: Version 1 (collagen mesh with human demineralized allograft bone matrix) and Version 2 (collagen mesh with biphasic ceramic granules). Supplied prefilled or as separate components for clinician assembly/suturing. Hydrated with blood, saline, or bone marrow aspirate; combined with autograft before implantation. Used in orthopedic procedures (pelvis, posterolateral spine) by surgeons to fill non-structural bony defects. Device resorbs and is replaced by host bone during healing. Benefits include providing a scaffold for bone growth and acting as a graft extender to reduce autograft requirements.
Clinical Evidence
No human clinical data. Bench testing included physical/chemical characterization (XRD, FTIR, porosity, etc.), ISO 10993 biocompatibility, and bacterial endotoxin limits. Viral inactivation testing confirmed clearance potential for DBM component. Osteoinductive potential verified via athymic rodent model (score ≥ 1). Performance evaluated in rabbit posterolateral spine fusion model compared to ICBG and predicate; results showed substantially equivalent new bone formation and fusion rates via radiographic, biomechanical, and histological analysis.
Technological Characteristics
Resorbable collagen mesh scaffold containing either human demineralized allograft bone matrix or biphasic ceramic granules. Available in sizes 18x50mm to 24x100mm. Terminally sterile. Biocompatibility per ISO 10993. Endotoxin limits < 20 EU/device per ANSI/AAMI ST72. Mechanical/chemical properties characterized by XRD, FTIR, porosity, surface area, and density.
Indications for Use
Indicated for patients with bony voids or gaps in the skeletal system (pelvis and posterolateral spine) not intrinsic to structural stability, resulting from surgery or trauma. Used as a bone graft extender in combination with autograft.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 18, 2024
Ventris Medical
John Brunelle
Chief Operating Officer
1201 Dove Street, Suite 470
Newport Beach, California 92660
Re: K240765
Trade/Device Name: Porous Biologic Scaffold
Regulation Number: 21 CFR 888.3045
Regulation Name: Resorbable Calcium Salt Bone Void Filler Device
Regulatory Class: Class II
Product Code: MQV, MBP
Dated: March 23, 2024
Received: November 4, 2024
Dear Mr. Brunelle:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K240765 - John Brunelle
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K240765 - John Brunelle
Page 3
Sincerely,
JESSE MUIR-S
Digitally signed by JESSE MUIR
Date: 2024.12.18 10:37:14
-05'00"
Jesse Muir, Ph.D.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K240765
Device Name
Porous Biologic Scaffold
Indications for Use (Describe)
Porous Biologic Scaffold is a bone void filler intended for use in bony voids or gaps of the skeletal system that are not intrinsic to the stability of the bony structure. These defects may be surgically created or a result of traumatic injury to the bone. Porous Biologic Scaffold is indicated to be packed gently into bony voids or gaps of the pelvis and posterolateral spine and is used combined with autograft as a bone graft extender. The device is resorbed and replaced with host bone during the healing process.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
> "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
Ventris Medical
Porous Biologic Scaffold – Traditional 510(k)
K240765
# 510(k) Summary
## Submitter Information
Name: Ventris Medical
Address: 1201 Dove Street, Suite 470
Newport Beach, CA 92660
Contact Person: John Brunelle, PhD
Chief Operating Officer
Telephone: (949) 706-5551
(617) 448-6570
Date Prepared: March 23, 2024
## Device Information
Trade Name: Porous Biologic Scaffold
Common Name: Bone void filler/ Bone graft substitute
Classification: Class 2
Regulation: 888.3045, Resorbable calcium salt bone void filler device
Product Code: MQV, MBP
## Predicate Device(s)
### Primary Predicate (Intended Use):
1. K140375: Mastergraft Putty (Medtronic)
### Reference Predicates:
2. K172130: Resorbable Mesh Device (Seaspine)
3. K122513: Magnifuse II (Medtronic)
4. K082002: Progenix Plus (Medtronic)
5. K100461: PGA Absorbable Surgical Suture (Riverpoint Medical)
## Indications For Use
Porous Biologic Scaffold is a bone void filler intended for use in osseous defects of the skeletal system that are not intrinsic to the stability of the bony structure. These defects may be surgically created or the result of traumatic injury to the bone. Porous Biologic Scaffold is indicated to be packed gently into bony voids or gaps of the pelvis and posterolateral spine and is used combined with autograft as a bone graft extender. The device is resorbed and replaced with host bone during the healing process.
## Device Description
Porous Biologic Scaffold is a resorbable bone void filler that is available in two versions:
Version 1 is comprised of a porous collagen mesh that is either prefilled or separately packaged with human demineralized allograft bone matrix. In the prefilled format, the demineralized bone matrix is completely
510(k) Summary
{5}
Ventris Medical
Porous Biologic Scaffold – Traditional 510(k)
K240765
enclosed inside the collagen mesh. In the separately packaged format, an open-ended collagen mesh bag is supplied with separately packaged demineralized bone matrix and is filled and sewn closed with absorbable suture by the clinician. Version 1 of the Porous Biologic Scaffold may be hydrated with blood, saline, or bone marrow aspirate, and are combined with autograft bone prior to implantation.
Version 2 is comprised of a porous resorbable collagen mesh bag either prefilled or separately packaged with activated biphasic ceramic granules. In the prefilled format, the biphasic granules are completely enclosed inside the collagen mesh. In the separately packaged format, an open-ended collagen mesh bag is supplied with separately packaged biphasic granules and is filled and sewn closed with absorbable suture by the clinician. Version 2 of the Porous Biologic Scaffold may be hydrated with blood, saline, or bone marrow aspirate, and are combined with autograft bone prior to implantation.
Both formats of the Porous Biologic Scaffold are supplied terminally sterile and are available in sizes of 18 x 50 mm, 18 x 100 mm, 24 x 50 mm, and 24 x 100 mm.
## Summary of Testing
Non-clinical testing was performed in accordance with FDA guidance documents and recognized consensus standards as applicable.
Physical and chemical characterization of the implant materials and finished device was conducted as recommended in the FDA class II bone void filler guidance document. Tests conducted include XRD, FTIR, Trace elements, porosity, surface area, granule size distribution, bulk density, pH, moisture content and handling properties evaluation.
Porous Biologic Scaffold has met all relevant requirements of ISO 10993 biocompatibility testing for permanent bone-contacting implants. Tests for cytotoxicity, genotoxicity, sensitization, irritation/ intracutaneous reactivity, and materials-mediated pyrogenicity were conducted directly on the finished device according to applicable methods within the ISO 10993 panel. Evaluations of sub-acute/sub-chronic toxicity as well as implant reactivity were conducted as part of the rabbit posterolateral spine fusion study. The final device also meets bacterial endotoxin limits of < 20 EU/device according to ANSI/AAMI ST72.
Viral inactivation testing was conducted on the demineralized allograft bone matrix (DBM) component of the Porous Biologic Scaffold, which evaluated a panel of model viruses representing various virus types, sizes, shapes and genomes. The DBM processing methods were determined to provide significant viral clearance potential for a wide range of potential viruses.
The osteoinductive potential of the DBM fibers contained in the Porous Biologic Scaffold is lot tested and verified in an in vivo athymic rodent model, based on demonstrating an osteoinduction score of ≥ 1. Osteoinduction assay results using the athymic rodent assay, or other specified test that has been validated and correlated with the athymic rodent assay, should not be interpreted to predict clinical performance in human subjects.
Implant performance testing was conducted using a posterolateral spine fusion rabbit model to evaluate the safety and performance of the Porous Biologic Scaffold as directly compared to ICBG and the predicate device. Porous Biologic Scaffold demonstrated substantially equivalent new bone formation and spine fusion rates as compared to ICBG and the predicate device via radiographic, biomechanical and histological evaluations across multiple time points.
510(k) Summary
Page 2 of 3
{6}
Ventris Medical
Porous Biologic Scaffold – Traditional 510(k)
K240765
## Substantial Equivalence
Porous Biologic Scaffold has the same intended use, and the same or similar technological characteristics, principles of operation and indications as the predicate devices. Any technological differences presented by the Porous Biologic Scaffold do not raise new questions of safety or effectiveness, as demonstrated by the comparative evaluation in the animal study.
510(k) Summary
Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.