The Resorbable Mesh Device is intended to fill voids and gaps in the skeletal system that are not intrinsic to the stability of the bony structure. The product is indicated for use with autograft extender in the posterolateral spine and pelvis. The voids or gaps may be surgically created defects or the result of traumatic injury to the bone. The Resorbable Mesh Device is resorbed/remodeled and replaced by host bone during the healing process.
Device Story
Resorbable Mesh Device consists of demineralized human cortical bone particles contained within a knitted poly(lactic-co-glycolic) acid (PLGA) mesh pouch. Device acts as bone graft extender; intended to fill non-structural skeletal voids or gaps in posterolateral spine and pelvis. During healing, mesh and bone particles are resorbed/remodeled and replaced by host bone. Device is provided sterile, ready-to-use, and implanted by surgeons during orthopedic procedures. Benefits include containment of bone graft material at the surgical site, facilitating bone fusion and healing.
Clinical Evidence
Bench testing included biocompatibility (ISO 10993-1), sterilization (ISO 11137, SAL 10^-6), bacterial endotoxin (AAMI ST72, USP<85>), and viral inactivation/clearance. Osteoinductive potential verified via athymic rodent model. In vivo animal study demonstrated comparable resorption, remodeling, and fusion rates versus autograft control.
Technological Characteristics
Materials: Demineralized human cortical bone, PLGA mesh pouch. Sterilization: Electron beam irradiation (SAL 10^-6). Biocompatibility: ISO 10993-1 compliant. Endotoxin: AAMI ST72/USP<85> compliant (≤20 EU/Device). Form factor: Mesh pouch containing particulate bone.
Indications for Use
Indicated for patients requiring bone void filling in the skeletal system (posterolateral spine and pelvis) where the defect is not intrinsic to structural stability. Used with autograft as a bone graft extender for surgically created defects or traumatic bone injuries.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is a stylized representation of human services. To the right of the symbol is a blue square containing the acronym "FDA" in white letters. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
November 16, 2017
SeaSpine Orthopedics Corporation Ms. Caryn Sailor Regulatory Affairs Specialist 5770 Armada Drive Carlsbad, California 92008
Re: K172130
Trade/Device Name: Resorbable Mesh Device Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MOV, MBP Dated: October 18, 2017 Received: October 19, 2017
Dear Ms. Sailor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to Mav 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Katherine D. Kavlock -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K172130
Device Name Resorbable Mesh Device
#### Indications for Use (Describe)
The Resorbable Mesh Device is intended to fill voids and gaps in the skeletal system that are not intrinsic to the stability of the bony structure. The product is indicated for use with autograft extender in the posterolateral spine and pelvis. The voids or gaps may be surgically created defects or the result of traumatic injury to the bone. The Resorbable Mesh Device is resorbed/remodeled and replaced by host bone during the healing process.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------|---------------------------------------------|
|----------------------------------------------|---------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
| Contact Details | |
|-----------------------|---------------------------------------------|
| Owner Operator Name: | SeaSpine Orthopedics Corporation |
| Address: | 5770 Armada Drive, Carlsbad CA 92008 |
| Establishment Name: | IsoTis OrthoBiologics, Inc. |
| Address: | 2 Goodyear, Irvine, CA 92618 |
| Contact person: | Caryn Sailor, Regulatory Affairs Specialist |
| Email address: | caryn.sailor@seaspine.com |
| Phone number: | (949) 855-7174 |
| Fax number: | (949) 595-8711 |
| Date Prepared: | October 18, 2017 |
| Subject Device | |
| Trade Name: | Resorbable Mesh Device |
| Common Name: | Demineralized Bone Matrix Allograft |
| Classification Name: | Resorbable Bone Void Filler |
| Regulatory Class: | II |
| Regulation: | 21 CFR 888.3045 |
| Classification Panel: | Orthopedic |
| Product Code: | MOV, MBP |
## Legally Marketed Predicate Devices
| 510(k) | Product Code | Trade Name | Manufacturer |
|------------------------------|--------------|----------------------------|----------------------------------|
| Number | | | |
| PRIMARY PREDICATE Device | | | |
| K123691 | MQV, MBP | MAGNIFUSE Bone<br>Graft | Medtronic Sofamor<br>Danek, Inc. |
| Additional PREDICATE Devices | | | |
| K082615 | MQV | GRAFTON II eDBM | OsteoTech, Inc. |
| K122513 | MQV, MBP | MAGNIFUSE II Bone<br>Graft | Medtronic Sofamor<br>Danek, Inc. |
## Device Description
The Resorbable Mesh Device is comprised of a resorbable mesh pouch containing demineralized human cortical bone in particulate form processed by FDA registered, AATB-accredited tissue banks. The resorbable mesh pouch is knitted from poly(lactic-co-glycolic) acid (PLGA) fibers which have a high glycolide content.
The Resorbable Mesh Device is offered in various sizes and provided in individually sterile packaged units and ready-to-use. The product is terminally sterilized by
{4}------------------------------------------------
electron beam irradiation and validated to ensure a minimum Sterility Assurance Level (SAL) of 10-6.
#### Indications for Use
The Resorbable Mesh Device is intended to fill voids and gaps in the skeletal system that are not intrinsic to the stability of the bony structure. The product is indicated for use with autograft as a bone graft extender in the posterolateral spine and pelvis. The voids or gaps may be surgically created defects or the result of traumatic injury to the bone. The Resorbable Mesh Device is resorbed/remodeled and replaced by host bone during the healing process.
#### Summary of Non-Clinical Testing to Support Substantial Equivalence
Non-clinical testing performed on the Resorbable Mesh Device includes tests for: biocompatibility, sterilization, bacterial endotoxin, viral inactivation/clearance, osteoinductive potential and in vivo (animal) safety and performance. Biocompatibility was performed in accordance with ISO 10993-1, Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process and test results were acceptable for a permanent, bone/tissuecontacting implantable device. Sterilization complies with ISO 11137, Sterilization of Sterilization of Health Care products - Radiation and has been validated to ensure a sterility assurance level (SAL) of 10°. Bacterial endotoxin testing complies with AAMI ST72 Bacterial Endotoxins – Test Methods, Routine Monitoring, and Alternatives to Batch Testing and USP<85> Bacterial Endotoxin Test and has been validated ensure a BET limit of ≤20 EU/Device. A viral inactivation/clearance study evaluated a panel of model viruses representing various virus types, sizes, shapes and genomes. The DBM processing method was determined to provide significant viral inactivation/clearance potential for a wide range of potential viruses.
Each lot of DBM in the Resorbable Mesh Device is tested for osteoinductive potential using the athymic rodent model. Only DBM demonstrating osteoinductive potential is used to manufacture the Resorbable Mesh Device. It is unknown how osteoinductive potential measured in the athymic rodent model correlates with clinical performance.
An in vivo study of the Resorbable Mesh Device demonstrated comparable resorption, remodeling and rates of fusion when compared to an autograft control. The study employed various analyses and endpoints were assessed at several time points.
## Conclusions
The submitted data demonstrate that the Resorbable Mesh Device is substantially equivalent to the predicate in terms of all relevant characteristics including intended use, indications for use, materials, performance, biocompatibility, osteoinductive potential and viral inactivation/clearance. Specifically, both the Resorbable Mesh Device and the MAGNIFUSE Bone Graft contain demineralized human cortical bone contained in a sealed resorbable polymer mesh pouch, intended as a bone graft
{5}------------------------------------------------
extender in the posterolateral spine or pelvis. Like the MAGNIFUSE Bone Graft, the demineralized bone in the Resorbable Mesh Device has been validated using standard test methods for viral inactivation/clearance and for osteoinductive potential. The mesh pouch of the Resorbable Mesh Device and the MAGNIFUSE Bone Graft are composed of resorbable PLGA and PGA, respectively, which are comparable in terms of material characteristics, resorption profile and function (to contain DBM).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.