CELERITY HP Indicator Tape

K240760 · Steris · JOJ · Apr 10, 2024 · General Hospital

Device Facts

Record IDK240760
Device NameCELERITY HP Indicator Tape
ApplicantSteris
Product CodeJOJ · General Hospital
Decision DateApr 10, 2024
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.2800
Device ClassClass 2

Intended Use

The Indicator Tape is intended to secure items wrapped in synthetic wrap materials until used and to distinguish between processed and unprocessed units through a color change from the start color (pink/magenta) to orange, yellow or lighter. The Indicator Tape is also intended to be used as an external process indicator when applied to synthetic wrap materials and/or Tyvek pouches. The tape may be used in the following sterilization cycles: · Lumen, Non Lumen, Fast Non Lumen, Fast, Flexible, and Specialty Cycles of the V-PRO: 1, 1 Plus, maX 2, 60, and s2 Low Temperature Sterilization Systems. · Standard and Advanced Cycles of the STERRAD® NX Sterilizer with or without ALLClear · Standard, Flex Scope, Express and DUO Cycles of the STERRAD® 100NX Sterilizer with or without ALLClear.

Device Story

CELERITY HP Indicator Tape is a 3/4" wide crepe paper tape coated with vaporized hydrogen peroxide (VHP) reactive ink. Used in clinical settings to secure sterilization wraps and provide visual confirmation of VHP exposure. Upon exposure to VHP sterilization cycles, the ink undergoes a chemical reaction, changing color from pink/magenta to orange, yellow, or lighter. The tape is applied manually by sterile processing staff. It serves as a Type 1 process indicator per ANSI/AAMI/ISO 11140-1:2014. The color change allows healthcare providers to distinguish between processed and unprocessed items, ensuring that only sterilized items are used in clinical procedures. The device is a passive chemical indicator; it does not involve electronic processing or software.

Clinical Evidence

Bench testing only. Testing included simulated use, ISO 11140-1 Type 1 performance verification, light stability, endpoint stability, and shelf-life testing. Results confirmed complete color change in full cycles, incomplete change in partial cycles, and adherence through worst-case cycles.

Technological Characteristics

Crepe paper tape with VHP-reactive ink and protective varnish coating. Dimensions: 3/4" wide, 60 yards long. Complies with ANSI/AAMI/ISO 11140-1:2014 for Type 1 process indicators. Passive chemical sensing mechanism; no energy source, software, or connectivity.

Indications for Use

Indicated for use as an external process indicator to secure synthetic wrap materials and Tyvek pouches during VHP sterilization cycles (V-PRO and STERRAD systems) and to provide visual confirmation of processing via color change.

Regulatory Classification

Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. April 10, 2024 Steris Corporation Anthony Piotrkowski Director, Regulatory Affairs 5960 Heisley Rd Mentor, Ohio 44060 Re: K240760 Trade/Device Name: CELERITY HP Indicator Tape Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: JOJ Dated: March 20, 2024 Received: March 20, 2024 Dear Anthony Piotrkowski: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). > Sincerely, Stephen A. Digitally signed by Stephen A. Anisko -S Date: 2024.04.10 Anisko -S 11:42:33 -04'00' For: Christopher Dugard Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices {2}------------------------------------------------ OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K240760 Device Name CELERITY HP Indicator Tape Indications for Use (Describe) The Indicator Tape is intended to secure items wrapped in synthetic wrap materials until used and to distinguish between processed and unprocessed units through a color change from the start color (pink/magenta) to orange, yellow or lighter. The Indicator Tape is also intended to be used as an external process indicator when applied to synthetic wrap materials and/or Tyvek pouches. The tape may be used in the following sterilization cycles: · Lumen, Non Lumen, Fast Non Lumen, Fast, Flexible, and Specialty Cycles of the V-PRO: 1, 1 Plus, maX 2, 60, and s2 Low Temperature Sterilization Systems. · Standard and Advanced Cycles of the STERRAD® NX Sterilizer with or without ALLClear · Standard, Flex Scope, Express and DUO Cycles of the STERRAD® 100NX Sterilizer with or without ALLClear. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| Prescription Use (Part 21 CFR 801 Subpart D) |X | Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # K240760 510(k) Summary for CELERITY HP Indicator Tape ### Sponsor Facility STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 639-4459 Contact: Director, Regulatory Affairs Telephone: (440) 392-7437 Fax No: (440) 357-9198 Tony piotrkowski@steris.com Anthony Piotrkowski Submission Date: April 10, 2024 STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600 {5}------------------------------------------------ #### 1. Predicate Device Trade Name: Common/Usual Name: Classification: Classification Name: 510(k) Submitter/Holder: 510(k) Number: VERIFY VH2O2 Indicator Tape Chemical Indicator Class II Physical/chemical Sterilization Process Indicator STERIS Corporation K231488 #### 2. Device Description The CELERITY HP Indicator Tape (Indicator Tape) is a ¾' wide crepe paper tape printed with diagonal stripes of a vaporized hydrogen peroxide (VHP) reactive ink. The ink is sealed with a varnish, the function of which is to inhibit removal of the ink via transference or by the adhesive (as the tape is dispensed). The reactive ink meets the performance specifications for a Type 1 process indicator for vaporized hydrogen peroxide as defined in ANSI/AAMI/ISO 11140-1:2014. #### 3. Indications for Use: The Indicator Tape is intended to secure items wrapped in synthetic wrap materials until used and to distinguish between processed and unprocessed units through a color change from the start color (pink/magenta) to orange, yellow or lighter. The Indicator Tape is also intended to be used as an external process indicator when applied to synthetic wrap materials and/or Tyvek pouches. The tape may be used in the following sterilization cycles: - . Lumen, Non Lumen, Fast Non Lumen, Fast, Flexible, and Specialty Cycles of the V-PRO: 1, 1 Plus, maX, maX 2, 60, and s2 Low Temperature Sterilization Systems. - . Standard and Advanced Cycles of the STERRAD® NX Sterilizer with or without ALLClear - . Standard, Flex Scope, Express and DUO Cycles of the STERRAD® 100NX Sterilizer with or without ALLClear. #### 4. Technological Characteristics The proposed and predicate devices are chemical indicators in accordance with ISO 11140-1:2014 for use in monitoring Vaporized Hydrogen Peroxide sterilization cycles. The mechanism of action and endpoint are the same and when exposed to the defined processing conditions, the proposed and predicate devices exhibit a visible color change. {6}------------------------------------------------ | Feature | Proposed: CELERITY HP Indicator Tape | Predicate: VERIFY VH2O2 Indicator<br>Tape (K231488) | Comparison | |-----------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use<br>/ Indications<br>for Use | The Indicator Tape is intended to secure items<br>wrapped in synthetic wrap materials until used<br>and to distinguish between processed and<br>unprocessed units through a color change<br>from the start color (pink/magenta) to orange,<br>yellow or lighter. The Indicator Tape is also<br>intended to be used as an external process<br>indicator when applied to synthetic wrap<br>materials and/or Tyvek pouches. | The Indicator Tape is intended to secure items<br>wrapped in synthetic wrap materials until used<br>and to distinguish between processed and<br>unprocessed units through a color change<br>from the start color (pink) to peach, yellow or<br>lighter. The Indicator Tape is also intended to<br>be used as an external process indicator when<br>applied to synthetic wrap materials and/or<br>Tyvek pouches. | Similar, a slight<br>change to the<br>start and end<br>color descriptor<br>to match color<br>description of<br>other Celerity<br>chemical<br>indicators and | | | The tape may be used in the following<br>sterilization cycles:<br>• Lumen, Non Lumen, Fast Non Lumen,<br>Fast, Flexible, and Specialty Cycles of the<br>V-PRO: 1, 1 Plus, maX, maX 2, 60, and<br>s2 Low Temperature Sterilization<br>Systems.<br>• Standard and Advanced Cycles of the<br>STERRAD® NX Sterilizer with or<br>without ALLClear<br>• Standard, Flex Scope, Express and DUO<br>Cycles of the STERRAD® 100NX<br>Sterilizer with or without ALLClear | The tape may be used in the following<br>sterilization cycles:<br>• Lumen, Non Lumen, Fast Non Lumen,<br>Fast, Flexible, and Specialty Cycles of the<br>V-PRO: 1, 1 Plus, maX, maX 2, 60, and<br>s2 Low Temperature Sterilization<br>Systems.<br>• STERRAD® 100S Sterilizer (Default<br>Cycle)<br>• Standard and Advanced Cycles of the<br>STERRAD® NX Sterilizer with or<br>without ALLClear<br>• Standard, Flex Scope, Express and DUO<br>Cycles of the STERRAD® 100NX<br>Sterilizer with or without ALLClear | removal of the<br>STERRAD<br>100S Default<br>Cycle | | Chemical<br>Indicator<br>Agent and<br>Performance<br>Specification | The indicator agent is a vaporized hydrogen<br>peroxide reactive ink that meets the<br>requirement specified in ANSI/AAMI/ISO<br>11140-1:2014 for a Type 1 Process Indicator<br>for a vaporized hydrogen peroxide process | The indicator agent is a vaporized hydrogen<br>peroxide reactive ink that meets the<br>requirement specified in ANSI/AAMI/ISO<br>11140-1:2014 for a Type 1 Process Indicator<br>for a vaporized hydrogen peroxide process | Identical | | Device<br>Design | Tape: 3/4" wide by 60 yards long crepe<br>(masking) tape which is wound around a 3"<br>core. A hydrogen peroxide reactive ink is laid<br>down on the non-adhesive surface. The ink is<br>protected from transfer to the adhesive via a<br>coating. | Tape: 3/4" wide by 60 yards long crepe<br>(masking) tape which is wound around a 3"<br>core. A hydrogen peroxide reactive ink is laid<br>down on the non-adhesive surface. The ink is<br>protected from transfer to the adhesive via a<br>coating. | Identical | | Shelf Life | 5 months - real-time testing is ongoing | Tape 24 months | real-time<br>testing is<br>ongoing | | Performance<br>Limitations<br>(taken from<br>instructions<br>for use) | Do not overlap the tape onto itself as this may<br>prevent the underlying layer from being<br>appropriately exposed to hydrogen peroxide<br>resulting in a failed indicator response. | Do not overlap the tape onto itself as this may<br>prevent the underlying layer from being<br>appropriately exposed to hydrogen peroxide<br>resulting in a failed indicator response. | Identical | # Table 1 Comparison of CI Physical Description and Technological Properties {7}------------------------------------------------ ## STERIS Traditional 510(k) PREMARKET NOTIFICATION K240760 CELERITY HP Indicator Tape #### 5. Performance Testing Performance testing was completed using well established methods such as those described in ISO 11140-1 and those used to verify the predicate design. | Testing | Acceptance Criteria | Study Result | |--------------------------|--------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------| | Simulated Use<br>Testing | • Complete color change in full cycle<br>• Incomplete color change in partial cycle<br>• Remain adhered after worst-case cycle | • Complete color change in full cycle<br>• Incomplete color change in partial cycle<br>• Remains adhered after worst-case cycle | | ISO 11140-1 Testing | Meets requirement of Type 1 VHP Indicator | Meets requirement of Type 1 VHP Indicator | | Light Stability | Ink performs as intended when exposed to<br>light for labeled period | Ink performs as intended when exposed to<br>light for labeled period | | Endpoint Stability | Endpoint color remains stable for labeled<br>period | Endpoint color remains stable for labeled<br>period | | Shelf Life | Meets all performance requirements at<br>labeled shelf-life | Meets all performance requirements at labeled<br>shelf-life | | Table 2. Verification Results Summary | | |---------------------------------------|--| |---------------------------------------|--| The results of the performed testing demonstrate that the CELERITY HP Indicator Tape performs as intended. #### 6. Conclusion The conclusion drawn from the non-clinical testing demonstrates that the subject device, CELERITY HP Indicator Tape, submitted under K240760, is as safe, as effective, and performs as well as the legally marketed predicate device cleared under K231488.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...