K240721 · Globus Medical, Inc. · OLO · Jun 14, 2024 · Neurology
Device Facts
Record ID
K240721
Device Name
ExcelsiusFlex™
Applicant
Globus Medical, Inc.
Product Code
OLO · Neurology
Decision Date
Jun 14, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
ExcelsiusFlex™ when used in conjunction with ExcelsiusHub™ is intended for use as an aid for precisely locating anatomical structures and for spatial positioning and orientation of a tool holder to be used by surgeons for navigating and/or guiding compatible surgical instruments provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or directly acquired anatomical structures. The system is indicated to assist the surgeon in planning the position of the implant components and preparing the bony anatomy during orthopedic procedures. The Total Knee Arthroplasty (TKA) implant systems compatible with ExcelsiusFlex™ are GENflex2® and ACTIFY™ Total Knee System. The ExcelsiusHub™ is intended for use as an aid for precisely locating anatomical structures to be used by surgeons for navigating compatible surgical instruments in open or percutaneous procedures provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans, fluoroscopy or directly acquired anatomical structures. The system is indicated for the planning of orthopedic devices and placement of spinal and orthopedic bone screws and interbody fusion devices.
Device Story
ExcelsiusFlex™ is a robotic positioning system used with ExcelsiusHub™ for orthopedic surgery. Inputs include pre-operative CT scans or directly acquired anatomical data. System uses patient tracking arrays with reflective markers and a positioning camera to map virtual patient data to physical anatomy. Software enables surgical planning and real-time navigation. Robotic arm, controlled by surgeon, provides stereotaxic positioning and orientation for a tool holder and sagittal sawblade. Used in OR by surgeons to guide bone resection and implant placement. Benefits include precise anatomical localization and spatial orientation during TKA and spinal procedures. System constrains robotic arm to planned resection planes, assisting in accurate bone preparation.
Clinical Evidence
No clinical data. Evidence consists of non-clinical system, software, and instrument verification and validation; surgical simulations on bone models; and cadaveric quantitative validation under clinically relevant scenarios. Electrical safety and electromagnetic compatibility testing performed per IEC 60601-1:2020 and IEC 60601-1-2:2014.
Technological Characteristics
Robotic positioning system comprising base station, robotic arm, camera, display, foot pedal, cutting system, and planning laptop. Employs optical tracking of reflective markers on patient/instrument arrays. Software-controlled robotic guidance for stereotaxic surgery. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 62304, and IEC 62366 standards.
Indications for Use
Indicated for surgeons performing orthopedic procedures, specifically Total Knee Arthroplasty (TKA) using GENflex2® or ACTIFY™ systems, and spinal/orthopedic bone screw and interbody fusion device placement. Assists in planning implant position and preparing bony anatomy via navigation and robotic guidance.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
June 14, 2024
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Globus Medical, Inc Jennifer Antonacci Senior Group Manager, Regulatory Affairs Valley Forge Business Center 2560 General Armistead Ave Audubon, Pennsylvania 19403
### Re: K240721
Trade/Device Name: ExcelsiusFlex™ Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: March 15, 2024 Received: March 18, 2024
### Dear Jennifer Antonacci:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
Jesse Muir -S Muir-S
Digitally signed by Jesse
Date: 2024.06.14 13:24:52
-04'00'
For: Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240721
Device Name ExcelsiusFlex™
#### Indications for Use (Describe)
ExcelsiusFlex™ when used in conjunction with ExcelsiusHub™ is intended for use as an aid for precisely locating anatomical structures and for spatial positioning and orientation of a tool holder to be used by surgeons for navigating and/or guiding compatible surgical instruments provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or directly acquired anatomical structures. The system is indicated to assist the surgeon in planning the position of the implant components and preparing the bony anatomy during orthopedic procedures.
The Total Knee Arthroplasty (TKA) implant systems compatible with ExcelsiusFlex™ are GENflex2® and ACTIFY™ Total Knee System.
The ExcelsiusHub™ is intended for use as an aid for precisely locating anatomical structures to be used by surgeons for navigating compatible surgical instruments in open or percutaneous procedures provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans, fluoroscopy or directly acquired anatomical structures. The system is indicated for the planning of orthopedic devices and placement of spinal and orthopedic bone screws and interbody fusion devices.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div style="display:flex; align-items:center;"> <span> <div style="display:inline-block;">☑</div> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"> <span> <div style="display:inline-block;">☐</div> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary: ExcelsiusFlex™
| Company: | Globus Medical Inc.<br>2560 General Armistead Ave.<br>Audubon, PA 19403<br>610-930-1800 |
|--------------------|----------------------------------------------------------------------------------------------------------------|
| Primary Contact: | Jennifer Antonacci, Ph.D.<br>Senior Group Manager, Regulatory Affairs |
| Secondary Contact: | Suganya Gopalakrishnan<br>Regulatory Specialist |
| Date Prepared: | June 12, 2024 |
| Device Name: | ExcelsiusFlex™ |
| Common Name: | Computer-assisted surgical device |
| Classification: | Per 21 CFR as follows:<br>§882.4560 Stereotaxic instrument<br>Product Code(s): OLO<br>Regulatory Class: II |
| Primary Predicate: | ExcelsiusGPS® (K171651) |
| Other Predicates: | Mako Total Knee Application (K193515)<br>VELYS™ Robotic-Assisted Solution (K202769)<br>ExcelsiusHub™ (K211616) |
AQrate™(K182000)
### Purpose:
The purpose of this submission is to request clearance for the ExcelsiusFlex™ and navigated instruments for bone preparation for primary total knee arthroplasty, and expanded indications for use for ExcelsiusHub.
### Device Description:
The ExcelsiusFlex™ is a robotic positioning system with a computer controlled robotic arm, hardware and software that works in conjunction with ExcelsiusHub™ , to enable real time surgical navigation and robotic guidance using patient tracking arrays and a positioning camera. The system assists the surgeon in implant placement planning and intraoperative tracking of patient anatomy by locating anatomical structures and stereotaxic positioning of surgical instruments relative to patient CT images or directly acquired anatomical structures. The navigation and quidance system determines the registration or mapping between the virtual patient (points on the patient images) and the physical patient (corresponding
{4}------------------------------------------------
points on the patient's anatomy), or directly acquired anatomical structures. Once this reqistration is created, the software positions the robotic arm on a planned resection plane.
The ExcelsiusFlex™ is intended for stereotaxic surqery in surgical knee procedures to assist the surgeon in spatial positioning and orientation of a sagittal sawblade, planning the position of the femoral and tibial implants, and for bone preparation during total knee arthroplasty (TKA) procedures. The system constrains the position of the robotic arm to planned resection planes based on the pre-operative or intra-operative plans developed using ExcelsiusFlex™-TKA software on either the ExcelsiusHub™ or a planning laptop.
ExcelsiusFlex™ instruments consist of patient tracking instruments, patient attachment instruments, navigation instruments, end effectors and end effector instruments. Patient attachment instruments provides a point of rigid fixation for the patient tracking instruments. Patient tracking instruments and navigated instruments incorporate unique array patterns with reflective markers, and are used to track patient anatomy and surgical instruments. End effectors attach to the distal end of the robotic arm and provide a rigid connection to the saw blade.
### Indications for Use:
ExcelsiusFlex™ when used in conjunction with ExcelsiusHub™ is intended for use as an aid for precisely locating anatomical structures and for spatial positioning and orientation of a tool holder to be used by surgeons for navigating and/or quiding compatible surqical instruments provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or directly acquired anatomical structures. The system is indicated to assist the surgeon in planning the position of the implant components and preparing the bony anatomy during orthopedic procedures.
The Total Knee Arthroplasty (TKA) implant systems compatible with ExcelsiusFlex™ are GENflex2® and ACTIFY™ Total Knee System.
The ExcelsiusHub™ is intended for use as an aid for precisely locating anatomical structures to be used by surqeons for navigating compatible surgical instruments in open or percutaneous procedures provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans, fluoroscopy or directly acquired anatomical structures. The system is indicated for the planning of orthopedic devices and placement of spinal and orthopedic bone screws and interbody fusion devices.
{5}------------------------------------------------
## Technological Characteristics:
The ExcelsiusFlex™ has similar technological characteristics to the predicate devices including the main system components, workflow, user interface, software features, and design. ExcelsiusFlex™ is comparable to the predicates in terms of intended use, fundamental scientific technology, technological characteristics and principle of operation.
| Device | Subject<br>ExcelsiusFlex™ | Predicate<br>Mako Total Knee<br>Application (K193515) | Predicate<br>VELYS Robotic-<br>Assisted Solution<br>(K202769) |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Principle of<br>operation | - Preoperative images<br>- Imageless registration<br>or pre-op CT-based<br>patient registration<br>- Surgical planning<br>- Real-time tracking of<br>navigated instruments<br>- Active robot tracking | - Preoperative images<br>- Pre-op CT-based<br>patient registration<br>- Surgical planning<br>- Real-time tracking of<br>navigated instruments<br>- Active robot tracking | - Imageless patient<br>registration<br>- Surgical planning<br>- Real-time tracking of<br>navigated instruments<br>- Active robot tracking |
| System<br>Components | Base station, robotic<br>arm, camera, display<br>monitor, foot pedal,<br>cutting system,<br>preoperative planning<br>laptop | Guidance Module,<br>robotic arm, camera<br>stand, cutting system,<br>preoperative planning<br>laptop | Base station: camera,<br>power supply, Robot<br>Control Unit, display<br>monitor, foot pedal |
| Input<br>images | 3D pre-operative exam<br>None (Imageless<br>registration) | 3D preoperative exam | None (Imageless<br>registration) |
| Integrated<br>planning<br>software | Yes, Excelsius™-TKA<br>software | Yes | Yes |
Comparison of Principles of Operation and Technological Characteristics
## Performance Testing:
Verification and validation testing were conducted on ExcelsiusFlex™ alone and combined with ExcelsiusHub™ to confirm that the devices meet performance requirements under the indications for use and to ensure safety and efficacy of the svstem:
- Non-clinical system, software, and instrument verification and validation -. demonstrated compliance with user needs and corresponding design inputs
- Surgical simulations conducted on bone models .
- Cadaveric quantitative validation under clinically relevant scenarios .
- Qualitative validation to confirm intended use ●
- Electrical Safety and Electromagnetic Compatibility .
- o Testing was performed to assure compliance with recognized safety standards for electrical safety and electromagnetic compatibility
- IEC 60601-1:2020 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
{6}------------------------------------------------
- . IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance -Collateral Standard: Electromagnetic disturbances
- . IEC 60601-1-6:2020 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance -Collateral standard: Usability
- . IEC 62304:2015 Medical device software - Software lifecycle processes
- . IEC 62366:2020 Medical devices - Part 1: Application of usability engineering to medical devices
## Software Verification and Validation Testing:
Software validation and verification testing was performed in accordance with the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (June 14, 2023) and IEC 62304:2006-05 Medical Device Software - Software life cycle processes. The Documentation Level for the software in this device is determined to be "Enhanced Documentation Level".
### Basis of Substantial Equivalence:
ExcelsiusFlex™ has been found to be substantially equivalent to the predicate devices with respect to technical characteristics, performance, and intended use. The information provided within this premarket notification supports substantial equivalence to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.