The Rx Knee System is intended for temporary use (maximum 180 days) as a total knee replacement in patients undergoing a two-stage procedure due to a septic process. This is the same intended use as the primary predicate OsteoRemedies Remedy Stemmed Knee Spacer.
Device Story
The Rx Knee System is a temporary knee prosthesis for two-stage revision of infected total knee implants. It consists of femoral, all-polyethylene tibial, and patellar components, plus femoral/tibial augments and stems. Components are implanted by orthopedic surgeons in a clinical setting. The device is used with gentamicin-loaded bone cement (Palacos G) to provide local antibiotic elution while maintaining joint space and mobility. Patients must use mobility aids (crutches/walkers) to reduce physiological loading during the 180-day maximum implantation period. After 180 days, the device must be explanted and replaced with a permanent implant or alternative treatment. The system leverages components from the previously cleared World Knee System, adapted for temporary use. It benefits patients by managing infection while preserving joint function during the interval between revision stages.
Clinical Evidence
Bench testing only. Engineering evaluations verified mechanical performance (range of motion, contact area, stress testing), biocompatibility, and sterilization validation by comparing the subject device to the previously cleared World Knee System. Antibiotic elution was evaluated via a side-by-side comparison of total gentamicin dosage against the primary predicate (K183017) to ensure the subject device does not represent a worst-case scenario.
Technological Characteristics
Materials: Cast CoCr alloy (ASTM F75, ISO 5832-4), Wrought Ti6Al4V ELI (ASTM F136-13), UHMWPE or Vit-E HXLPE, and TiN DLC coating (DMF 1647). Components: Femoral (PS/CR), all-poly tibial, patellar, augments, and universal stems. Fixation: Cemented using gentamicin-loaded bone cement. Connectivity: None. Sterilization: Validated for single-use.
Indications for Use
Indicated for skeletally mature patients undergoing a two-stage total knee replacement procedure due to a septic process, where gentamicin is the appropriate antibiotic. Intended for temporary use (max 180 days) in conjunction with systemic antimicrobial therapy and gentamicin-loaded bone cement (Palacos G). Patients must consistently use traditional mobility assist devices (e.g., crutches, walkers, canes) throughout the implantation period.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
September 11, 2024
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left, featuring a stylized depiction of a human figure. To the right of the HHS logo is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Signature Orthopaedics Pty Ltd Declan Brazil Managing Director 7 Sirius Road Lane Cove, NSW 2066 Australia
Re: K240683
Trade/Device Name: Rx Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH, MBB Dated: August 14, 2024 Received: August 14, 2024
Dear Dr. Brazil:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely, JESSE MUIR - MUIR - MUIR - S MUIR -S
Date: 2024.09.11 10:55:43 S -04'00'
Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K240683
Device Name
Rx Knee System
#### Indications for Use (Describe)
The patient should be skeletally mature to receive a knee replacement. The Rx Knee System is indicated for temporary use (maximum 180 days) as an adjunct to total knee replacement (TKR) in patients undergoing a two-stage procedure due to a septic process and where gentamicin is the most appropriate antibiotic based on the susceptibility pattern of the infecting micro-organism(s). The device is intended for use in conjunction with systemic antimicrobial therapy (standard treatment approach to an infection).
The device is intended for use with the gentamicin-loaded cement. Palacos® G.
The Rx Knee System is not intended for use for more than 180 days, at which time it must be explanted and a permanent device implanted or another appropriate treatment performed (e.g., resection arthroplasty, fusion etc.).
The device is only indicated for patients who will consistently use traditional mobility assist devices (e.g., crutches, walkers, canes) throughout the implantation period, allowing basic joint mobility.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Signature Orthopaedics. The logo features a green circle with a white figure inside, followed by the word "Signature" in a cursive green font. Below "Signature" is the word "ORTHOPAEDICS" in a smaller, sans-serif font, and to the right of that are three horizontal green lines.
# 510(k) SUMMARY
| Manufacturer: | Signature Orthopaedics Pty Ltd<br>7 Sirius Road<br>Lane Cove, NSW 2066<br>Australia |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Signature Orthopaedics Europe Ltd<br>Unit A, IDA Business & Technology Park Garrycastle<br>Athlone Westmeath N37 DY26<br>IRELAND |
| Device Trade Name: | Rx Knee System |
| Common Name: | Temporary Knee Prosthesis |
| Contact: | Dr. Declan Brazil<br>Managing Director of Signature Orthopaedics |
| Prepared By: | Signature Orthopaedics Pty Ltd<br>7 Sirius Road<br>Lane Cove, NSW 2066<br>Australia<br>Phone: +61 (2) 9428 5181<br>Fax: +61 (2) 8456 6065 |
| Date Prepared: | September 10, 2024 |
| Classification: | Primary Product Code<br>Knee joint patellofemorotibial polymer/metal/polymer semi-<br>constrained cemented prosthesis (JWH, 21CFR 888.3560)<br><br>Secondary Product Code<br>Polymethylmethacrylate (PMMA) bone cement (MBB, 21CFR 888.3027) |
| Primary Devices: | Primary Predicate<br>• Remedy Stemmed Knee Spacer (K183017)<br>Predicate/Reference Device<br>• ATTUNE Revision Fixed Bearing (K160700)<br>• World Knee System (K181530, K220737, K223062) |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Signature Orthopaedics. The logo features a green circle with a white figure inside, followed by the word "Signature" in a cursive green font and the word "ORTHOPAEDICS" in a smaller, sans-serif font. Underneath the text is a horizontal line with two green stripes.
# Device Description:
The Rx Knee System is a temporary knee device consisting of femoral, all polyethylene tibial and patellar components, with femoral and tibial augments and a tibial or universal stem to accommodate for variants in patient anatomy. The femoral component is available in posterior stabilised and cruciate retaining variants intended for use with corresponding all polyethylene tibia variants that have been previously cleared as part of Signature Orthopaedics' World Knee System via K181530. The femoral and tibial components have been previously cleared as part of Signature Orthopaedics' World Knee System via K181530, K220737 and K223062. All Rx Knee System components are intended for use with FDA-cleared gentamicin-loaded bone cement.
The device is intended to be used for patients undergoing a two-stage revision procedure for infection of a total knee implant. The Rx Knee System components are sterile, single-use devices intended for temporary use as a total knee replacement.
Materials: Cast Cobalt Chromium (CoCr) alloy (ASTM F75 and ISO 5832-4) for the femoral component, Wrought Titanium-6Aluminium-4Vanadium ELI Alloy (Ti6Al4V ELI, ASTM F136-13) for femoral and tibial augment components, tibial stem, stem extension adapter and locking screw, Ultra-High-Molecular-Weight Polyethylene (UHMWPE) or Vit-E HXLPE for the femoral and tibial components and All-poly tibial components previously cleared as part of predicate World Knee System, and titanium nitride (TiN) diamond-like carbon (DLC) for coating femoral components which is the subject of FDA device master file no. 1647.
### Indications for Use:
The patients should be skeletally mature to receive a knee replacement. The Rx Knee is indicated for temporary use (maximum 180 days) as an adjunct to a total knee replacement (TKR) in patients undergoing a two-stage procedure due to a septic process and where gentamicin is the most appropriate antibiotic based on the susceptibility pattern of the infecting microorganism(s).
The device is intended for use in conjunction with systemic antimicrobial antibiotic therapy (standard treatment approach to an infection).
The device is intended for cemented use with the gentamicin-loaded cement, Palacos® G.
The Rx Knee System is not intended for use for more than 180 days, at which time it should be explanted and a permanent device implanted or another appropriate treatment performed (e.g. resection arthroplasty, fusion etc.).
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Signature Orthopaedics. The logo consists of a green circle with a white figure inside, followed by the word "Signature" in green cursive font and the word "ORTHOPAEDICS" in smaller, sans-serif font. Below the text is a green line, with a lighter green line above it.
The device is only indicated for patients who will consistently use traditional mobility assist devices (e.g., crutches, walkers, canes) throughout the implantation period, allowing basic joint mobility.
### Intended Use:
The Rx Knee System is intended for temporary use (maximum 180 days) as a total knee replacement in patients undergoing a two-stage procedure due to a septic process. This is the same intended use as the primary predicate OsteoRemedies Remedy Stemmed Knee Spacer.
#### Comparison of Technological Characteristics:
The subject device femoral components are identical in geometry and main material to predicate World Knee System symmetrical femoral components which have been cleared as part of the World Knee System via K223062 and originally K181530. The all polyethylene tibial components have been previously cleared as part of the World Knee System via K220737 and originally K181530.
The subject device femoral and tibial components have similar overall dimensions and condylar geometry to the femoral and tibial components of Remedy Stemmed Knee Spacer and ATTUNE Revision Knee. The only difference between the subject device and the primary predicate Remedy Stemmed Knee Spacer is the material composition since the predicate device is composed entirely of PMMA and MMA bone cement. The subject femoral and tibial components are intended for cemented fixation using gentamicin-loaded bone cement, which is similar to the constituting material of the predicate femoral and tibial components.
The subject device femoral and tibial augment components of various sizes are similar to the augments that are part of the ATTUNE Revision Knee System. The predicate World Knee System does not consist of augments since it is intended for for primary knee replacement. The predicate Remedy Stemmed Knee Spacer does not consist of augments, however subject device femoral and tibial augments are designed to achieve the same function as the independent components (femoral component, tibial component and stem extension components) of the predicate device that can be combined with each other depending on the anatomy of the patient.
The subject device universal stem component is similar to the stem component of predicates Remedy Stemmed Knee Spacer and ATTUNE Revision Knee System. The stem has the same straight design and material as the ATTUNE Revision Knee System, with similar diameter options. The predicate World Knee System does not consist of a stem component since it is intended for primary knee replacement. The subject device stem component varies from predicate Remedy Stemmed Knee Spacer with regards to the material as the predicate component is PMMA bone cement with gentamicin with stainless steel reinforcement. However, this difference in technological characteristic does not impact the intended use. The subject device universal stem is intended to be used with gentamicin-loaded bone cement. Hence, the
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Signature Orthopaedics. The logo consists of a green circle with a white figure inside, followed by the word "Signature" in green cursive font and the word "ORTHOPAEDICS" in smaller, block letters. Below the text is a horizontal line with two green stripes.
overall structure of both device stem components will have a metallic core with gentamicinloaded bone cement.
# Performance Testing:
Engineering evaluations were conducted to verify that the performance of the Rx Knee System with respect to mechanical performance, biocompatibility, sterility and antibiotic elution, is as safe and effective as the predicate devices and therefore adequate for anticipated in-vivo use. The following V&V activities were conducted:
- Review of performance testing of the referenced predicate World Knee System ● components previously cleared as part of K223062, K220737 and K181530, to verify that the Rx Knee System device falls within scope of the following performance testing:
- Range of motion analysis о
- o Component contact area and stress testing
- o Tibial post stress testing
The subject device is intended for temporary use (<180 days), and the expected physiological loading on the subject device is less than the reference predicate World Knee System because the Rx Knee System is intended for use with traditional mobility assistance devices for the entire duration of the implant, reducing the amount of direct loading on the patient's knee joint.
- . Comparison of subject device components to the referenced predicate World Knee System components previously cleared as part of K220737, K223062 and K181530 with respect to materials and manufacturing methods (formulation, processing, sterilisation and body/fluid contact) to verify that the device does not raise biocompatibility concerns.
- Evaluation of subject components for sterilisation difficulty to verify that the subject ● device does not present a new worst case for sterilisation validation.
- Volumetric evaluation, including a side by side comparison of the total dosage of . antibiotics released into the body for the subject device and the primary predicate cement spacer (K183017) using the indicated gentamicin bone cement to verify that the device does not represent a worst-case for antibiotic dosage.
The results of this non-clinical testing including engineering evaluations show that the subject Rx Knee System is as safe, as effective and is expected to perform at least as well as the primary predicate cement spacer (K183017).
### Substantial Equivalence Conclusion:
Based on comparability of indications for use and pertinent technological characteristics the Rx Knee System is substantially equivalent to its predicate devices cited herein. Any differences do not raise new questions of safety and effectiveness as established with performance testing; the subject device is at least as safe and effective as the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.