Total knee arthroplasty is a total joint replacement surgery designed to provide increased patient mobility and reduced pain by replacing the damaged knee joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. Total knee arthroplasty may include supplemental fixation through stems, sleeves, and/or modular augments where bone loss requires said fixation in the opinion of the surgeon. Total knee arthroplasty may also include more constrained bearing surfaces when necessary to provide stability where musculoligamentous supporting structures are insufficient. The ATTUNE Revision Total Knee is designed to accommodate knee flexion to 130 degrees in those patients able to attain a high degree of knee flexion.
Device Story
ATTUNE Revision Knee System is a modular total knee arthroplasty prosthesis; replaces damaged knee joint articulation to reduce pain and improve mobility. Components include femoral components, fixed-bearing tibial inserts, tibial bases, stems, femoral sleeves, augments, and offset adaptors. Surgeon selects and assembles modular components intraoperatively based on patient anatomy and bone quality; components fixed via cement or press-fit (porous-coated). Device provides structural support and stability for patients with ligamentous insufficiency or bone loss. Used in clinical orthopedic settings by surgeons. Benefits include restoration of joint function and stability in complex revision or primary cases.
Clinical Evidence
No clinical data. Substantial equivalence was determined based on non-clinical bench testing, including fatigue testing, patellofemoral/tibiofemoral contact area and pressure testing, constraint testing, and component interlock strength testing, in compliance with FDA guidance for knee joint prostheses.
Technological Characteristics
Materials: Cast Co-Cr-Mo (ASTM F75), Ti-6Al-4V (ASTM F136/F620), Ti-6Al-4V ELI (ASTM F136), UHMWPE (ASTM F648), wrought Co-Cr-Mo (ASTM F1537/F799). Design: Modular system with femoral components, tibial bases, stems, sleeves, augments, and offset adaptors. Fixation: Cemented or press-fit (porous-coated). Connectivity: None (mechanical implant). Sterilization: Not specified.
Indications for Use
Indicated for patients with severely painful/disabled joints due to osteoarthritis, post-traumatic arthritis, or rheumatoid arthritis; moderate valgus, varus, or flexion deformities; avascular necrosis of the femoral condyle; or previous unsuccessful knee procedures. Also indicated for patients requiring total knee arthroplasty with absence/loss of both cruciate ligaments, moderate varus-valgus or flexion instability requiring increased constraint, or bone loss requiring supplemental fixation.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
ATTUNE Total Knee System (Posterior Stabilized) (K111433)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a human figure embracing a bird, which is a symbol of care and protection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 10, 2016
DePuy Orthopaedics Incorporated Soraya Hori Regulatory Affairs Associate 700 Orthopaedic Drive Warsaw, Indiana 46582
Re: K160700 Trade/Device Name: ATTUNE® Revision Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH, MBH Dated: March 10, 2016 Received: March 14, 2016
Dear Soraya Hori:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### 4. INDICATIONS FOR USE
K160700 510(k) Number (if known):
Device Name: ATTUNE® Revision Knee System
Indications for Use:
Candidates for total knee replacement include patients with:
- A severely painful and/or severely disabled joint resulting from osteoarthritis, post-● traumatic arthritis, or rheumatoid arthritis
- Moderate valgus, varus, or flexion deformities
- Avascular necrosis of the femoral condyle
- A previous unsuccessful knee replacement, osteotomy, or other knee procedure ●
ATTUNE Revision Knee System implants are designed for use in total knee arthroplasty for patients with:
- Absence or loss of both cruciate ligaments ●
- Moderate varus-valgus or flexion instability that requires a bearing surface with increased ● constraint in the clinical judgment of the surgeon
- Bone loss that requires supplemental fixation in the clinical judgment of the surgeon
The porous-coated metaphyseal sleeves are intended for either cemented or cementless applications.
Any non porous-coated components are intended for cemented use only.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
{3}------------------------------------------------
## 5. 510(K) SUMMARY
| Submitter: | DePuy Orthopaedics, Inc.<br>700 Orthopaedic Drive<br>Warsaw, IN 46582 |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment<br>Registration Number: | 1818910 |
| Contact Person: | Soraya L. Hori<br>Sr. Regulatory Affairs Associate<br>Phone: (574) 372-7491<br>Fax: (574) 371-4987 |
| Date Prepared: | March 10, 2016 |
| Proprietary Name: | ATTUNE® Revision Knee System |
| Common Name: | Total Knee Replacement Prosthesis |
| Classification Name: | 21 CFR 888.3560 Knee joint patellofemorotibial<br>polymer/metal/polymer semi-constrained cemented prosthesis<br>21 CFR 888.3565 Knee joint patellofemorotibial metal/polymer<br>porous-coated uncemented prosthesis |
| Product Codes: | JWH, MBH |
| Predicate Devices: | The ATTUNE Revision Knee System components are substantially<br>equivalent to legally marketed devices including: |
| | Primary Predicates:<br>● ATTUNE Total Knee System (Posterior Stabilized) -K111433*<br>● Darwin Knee System (TC3) – K952830**<br>Reference Predicates:<br>● ATTUNE Total Knee System (Cruciate Retaining) –K101433*<br>● SIGMA Co-Cr Tibial Trays – K032151**<br>● Universal Knee Stem Extensions and Universal Metaphyseal<br>Sleeves - K063633**<br>● PFC Modular Plus Tibial Wedge – K923807**<br>● DePuy SIGMA Knee Prosthesis - K060515**<br>● DePuy LPS Metaphyseal Sleeves -K071417<br>● SROM Noiles PS Total Knee System – K941769<br>● Triathlon® TS Knee System – K070095 |
*These 510(k)s are encompassed within the ATTUNE Total Knee Product Family.
** These 510(k)s are encompassed within the SIGMA Product Family.
{4}------------------------------------------------
#### ATTUNE REVISION CRS FEMORAL COMPONENTS:
The ATTUNE Constrained Revision System (CRS) Femoral Components are manufactured from cast Co-Cr-Mo alloy conforming to ASTM F75. The ATTUNE Revision CRS Femoral Components have an asymmetric trochlear groove and are available in sizes 1 to 10 in both right and left options. The components contain a fixed boss to allow attachment of the ATTUNE Revision Femoral Sleeves or the ATTUNE Revision Stems. Augment holes are also present to allow attachment of the ATTUNE Revision Femoral Augments. The ATTUNE Revision CRS Femoral Components have recessed cement pockets and a textured fixation surface. A removable plug made of Ultra High Molecular Weight Polyethylene (UHMWPE) conforming to ASTM F648 is pre-assembled in the stem boss.
#### ATTUNE REVISION CRS FIXED BEARING INSERTS:
The ATTUNE Revision CRS Fixed Bearing Tibial Inserts articulate with the ATTUNE Revision CRS Femoral Components. The UHMWPE Antioxidant (AOX) Tibial Inserts are manufactured from raw materials conforming to ASTM F648 and are available in sizes 1 to 10 and in thicknesses of 6, 8, 10, 12, 14, 16, 18, 20, 22, 24 and 26mm.
#### ATTUNE REVISION FIXED BEARING TIBIAL BASES:
The ATTUNE Revision Fixed Bearing Tibial Bases are manufactured from cast Co-Cr-Mo alloy conforming to ASTM F75 and includes a Ti-6Al-4V removable end cap. The Tibial Bases are available in sizes 1 to 10. Augment holes are also present to allow attachment of the ATTUNE Revision Tibial Augments. The ATTUNE Revision Fixed Bearing Tibial Bases fixation surface is textured. It incorporates a stem and keel for additional stability and recessed cement pockets for cement fixation. The ATTUNE Revision Fixed Bearing Tibial Bases utilize a central, universal locking mechanism that engages anterior and posterior tabs in the Tibial Insert. Additionally, the ATTUNE Revision Fixed Bearing Tibial Bases have a Base Protector assembled with the device to minimize potential marring of the proximal bearing surface of the component during impaction. After impaction is completed, the Base Protector is then removed and discarded prior to inserting the ATTUNE Revision CRS Fixed Bearing Inserts.
{5}------------------------------------------------
#### ATTUNE REVISION STEMS:
The ATTUNE Revision Cemented and Press-fit Stems are manufactured from Ti-6Al-4V ELI alloy conforming to ASTM F136. The ATTUNE Revision Stems are optional components that may be used to provide additional fixation of the knee prosthesis construct and are compatible with the ATTUNE Revision CRS Femoral, ATTUNE Revision Fixed Bearing Tibial Base. ATTUNE Revision Offset Adaptor, and the ATTUNE Revision Femoral Sleeve components. All ATTUNE Revision Stems utilize a threaded junction with mating components. The ATTUNE Revision Press-fit Stems are cross-slotted and are designed with a coneshaped distal tip to facilitate insertion into the medullary canal of the distal femur or proximal tibia. The ATTUNE Cemented Stems are designed with a cylindrical or tapered body geometry depending on length and recessed cement pockets. The fixation surface is blasted and fluted.
#### ATTUNE REVISION FEMORAL SLEEVES:
The ATTUNE Revision Femoral Sleeves are manufactured from Ti-6Al-4V alloy conforming to ASTM F620 or ASTM F136. The ATTUNE Revision Femoral Sleeve is an optional component intended to provide improved fit of the femoral prosthesis where the metaphyseal bone in the distal femur is either absent or of poor quality. The Sleeve has a symmetric taper geometry with steps in the body to provide initial fixation and bone loading. There are six component sizes - 30, 35, 40, 45, 50 and 55mm - and are available in non-porous cemented (30mm only), partially porous coated, or fully porous coated. The ATTUNE Revision Femoral Sleeves utilize a locking taper to attach to the ATTUNE Revision CRS Femoral Component and a threaded junction with the ATTUNE Revision Stem components. A removable plug manufactured from UHMWPE conforming to ASTM F-648 is pre-assembled to the sleeve component.
#### ATTUNE REVISION FEMORAL AUGMENTS:
The ATTUNE Revision Femoral Augments are designed to build up the distal and/or posterior interior surfaces to accommodate bony insufficiency and aid in flexion/extension gap balancing. The augments are manufactured from cast Co-Cr-Mo conforming to ASTM F75 and are available in sizes 1 to 10 with varying thicknesses depending on size. Each augment attaches to either the distal or posterior interior surface of the ATTUNE Revision CRS Femoral Component.
#### ATTUNE REVISION TIBIAL AUGMENTS:
The ATTUNE Revision Tibial Augments are manufactured from wrought Co-Cr-Mo alloy conforming to ASTM F1537 or cast Co-Cr-Mo alloy conforming to ASTM F75 or forged Co-Cr-Mo alloy conforming to ASTM F799 and are available from sizes 1 to 10 and in thicknesses of 5, 10, and 15mm. The Tibial Augments are designed to be compatible with the ATTUNE Revision Fixed Bearing Tibial Bases.
{6}------------------------------------------------
#### ATTUNE REVISION OFFSET ADAPTOR:
The ATTUNE Revision Offset Adaptors are manufactured from wrought Co-Cr-Mo alloy conforming to ASTM 1537. They consist of two components: an offset body and a lock nut. The ATTUNE Revision offset adaptors are available in three offsets: 2 mm, 4 mm, and 6 mm. The Offset Adaptors are compatible with the ATTUNE Revision CRS Femoral Components, the ATTUNE Revision Fixed Bearing Tibial Bases, and the ATTUNE Revision Stems.
- Intended Use: Total knee arthroplasty is a total joint replacement surgery designed to provide increased patient mobility and reduced pain by replacing the damaged knee joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components.
Total knee arthroplasty may include supplemental fixation through stems, sleeves, and/or modular augments where bone loss requires said fixation in the opinion of the surgeon. Total knee arthroplasty may also include more constrained bearing surfaces when necessary to provide stability where musculoligamentous supporting structures are insufficient.
The ATTUNE Revision Total Knee is designed to accommodate knee flexion to 130 degrees in those patients able to attain a high degree of knee flexion.
Indications for Use: Candidates for total knee replacement include patients with:
- A severely painful and/or severely disabled joint resulting from osteoarthritis, post-traumatic arthritis, or rheumatoid arthritis
- Moderate valgus, varus, or flexion deformities
- Avascular necrosis of the femoral condyle
- 0 A previous unsuccessful knee replacement, osteotomy, or other knee procedure
ATTUNE Revision Knee System implants are designed for use in total knee arthroplasty for patients with:
- Absence or loss of both cruciate ligaments ●
- Moderate varus-valgus or flexion instability that requires a bearing surface with increased constraint in the clinical judgment of the surgeon
- Bone loss that requires supplemental fixation in the clinical ● judgment of the surgeon
The porous-coated metaphyseal sleeves are intended for either cemented or cementless applications.
Any non porous-coated components are intended for cemented use only.
{7}------------------------------------------------
#### Summary of Technologies/ Substantial Equivalence:
The ATTUNE Revision Knee System has the same indications, manufacturing methods, and sterilization methods as previously cleared DePuy total knee replacement systems. The design, materials, and packaging are similar to those of the predicate devices.
#### Non-Clinical Testing:
The table below lists the functional testing that was conducted in compliance with FDA Guidance, Class II Special Controls Guidance Knee Joint Patellofemorotibial Caand F Femorotibial Document: Metal/Polymer Porous-Coated Uncemented Prostheses: Guidance for Industry and FDA, to verify that the implant performance is substantially equivalent to predicate devices for anticipated in-vivo loading via various constraint, contact, wear, and fatigue tests. The results demonstrated that the subject devices are substantially equivalent to the legally marketed predicate devices.
The proposed devices also meet the requirement of bacterial endotoxin testing as specified in ANSI AAMI ST-72:2011.
| Fixed Bearing Tibial Base Fatigue Testing | Patellofemoral Contact Area and Contact Pressure Testing |
|--------------------------------------------------------|---------------------------------------------------------------------------------------|
| A-P Shear Spine Fatigue Testing | Component Interlock Strength Testing: Offset Adaptor to Stem and Stem and Tibial Base |
| Rollback Layout for Design Verification | Component Interlock Strength Testing: Femoral Sleeve to Femoral Component |
| Rotation Layout for Design Verification | Component Interlock Strength Testing: Stem to Tibial Base |
| Tibiofemoral Constraint Testing | Component Interlock Strength Testing: Stem to Femoral Sleeve |
| Medialized Dome Lateral Stability Testing | Component Interlock Strength Testing:<br>Varus/Valgus Loading of Femur and Insert |
| Tibiofemoral Contact Area and Contact Pressure Testing | Component Interlock Strength Testing: Insert Insertion and Extraction |
#### Clinical testing was not necessary to determine substantial equivalence Clinical Testing: between the ATTUNE Revision Knee System and the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.