AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K240680 · Dec 6, 2024
CDM Insights
Oxford Brain Diagnostics, Ltd.
13 different source cohorts of clinical MRI scans
Retrospective clinical cohorts were used to validate the accuracy of normative percentile data and the performance of automated segmentation algorithms against neuroradiologist ground-truth.
Retrospective cohorts; Normative data validation; Segmentation accuracy
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Normative percentile accuracy and segmentation validation; Retrospective analysis of existing clinical cohorts
Cognitively healthy individuals and individuals with Alzheimer's disease, mild cognitive impairment, frontotemporal dementia, and multiple sclerosis (aged 45-90 years); Sample Size: Approximately 2000 test scans for percentile accuracy; 60 cases for segmentation accuracy; Number of Sites: 13 different source cohorts (7 based in the USA)
Not applicable for this study
Accuracy of percentiles; Dice overlap scores for brain structure segmentation
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Brain structure segmentation
—
Dice > 0.58
Dice 0.58 to 0.72
—
—
60 cases comprising cognitively healthy individuals and individuals with Alzheimer's disease, mild cognitive impairment, frontotemporal dementia, and multiple sclerosis
>1 (US board-certified neuroradiologists)
White matter hyperintensity segmentation
—
Dice > 0.58
Dice 0.66 (0.15)
—
—
60 cases comprising cognitively healthy individuals and individuals with Alzheimer's disease, mild cognitive impairment, frontotemporal dementia, and multiple sclerosis
>1 (US board-certified neuroradiologists)
Indications for Use
CDM Insights is a post-processing image analysis software that assists trained healthcare practitioners in viewing, analyzing, and evaluating MR brain images of adults > 45 years of age. CDM Insights provides the following functionalities: - Automated segmentation and quantitative analysis of individual brain structures and white matter hyperintensities - Quantitative comparison of brain structures and derived values with normative data from a healthy population - Presentation of results for reporting that includes numerical values as well as visualization of these results
Device Story
CDM Insights is post-processing software for MR brain images; accepts DICOM inputs. Performs automated segmentation/quantification of brain structures and white matter hyperintensities; conducts post-acquisition analysis of diffusion-weighted imaging (DWI) data; compares patient metrics against normative data from healthy populations. Deployed as a cloud solution; produces DICOM-encapsulated PDF reports with numerical values, graphs, and anatomical color overlays. Used by radiologists and neurologists to assist in clinical evaluation; output supports clinical decision-making by providing quantitative brain morphometry and lesion data. Benefits include standardized, objective assessment of brain structures and longitudinal tracking of metrics.
Clinical Evidence
Bench testing only. Accuracy of segmentation and cortical surfaces validated against gold standard (US board-certified neuroradiologists) using 60 cases (healthy, Alzheimer's, MCI, frontotemporal dementia, MS). Mean Dice overlap for white matter hyperintensities was 0.66 (SD 0.15), exceeding 0.58 criterion. Regional cortical Dice scores ranged 0.58-0.72, all passing 0.58 threshold. Repeatability tested on 121 individuals; reproducibility tested on >1500 subjects (aged 45-90). Percentile accuracy tested on ~2000 independent scans.
Technological Characteristics
Medical image management and processing software. Cloud-based deployment. Inputs: DICOM MR images (T1-weighted, FLAIR, DWI). Outputs: DICOM-encapsulated PDF reports, numerical data, color overlays. Complies with ISO 13485, ISO 14971, IEC 62304. Software-based segmentation and quantification algorithms.
Indications for Use
Indicated for trained healthcare practitioners (radiologists, neurologists) to view, analyze, and evaluate MR brain images of adults > 45 years of age.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 6, 2024
Oxford Brain Diagnostics Ltd % Yulia Nikova Regulatory Affairs Manager Ken Blocking Consulting LLC 3400 N Central Expy Suite #110-225 Richardson, Texas 75080
Re: K240680
Trade/Device Name: CDM Insights Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH Dated: November 4, 2024 Received: November 5, 2024
# Dear Yulia Nikova:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{3}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
D. R. X.
Daniel M. Krainak, PhD Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240680
Device Name CDM Insights
Indications for Use (Describe)
CDM Insights is a post-processing image analysis software that assists trained healthcare practitioners in viewing. analyzing, and evaluating MR brain images of adults > 45 years of age.
CDM Insights provides the following functionalities:
- · Automated segmentation and quantitative analysis of individual brain structures and white matter hyperintensities
- · Quantitative comparison of brain structures and derived values with normative data from a healthy population
- · Presentation of results for reporting that includes numerical values as well as visualization of these results
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for Oxford Brain Diagnostics. The logo features a stylized brain graphic on the left, composed of interconnected circles in white and gray, with one circle in orange. To the right of the brain graphic, the text "oxford brain diagnostics" is written in white against a dark blue background. The text is in a sans-serif font, with "oxford brain" on the first line and "diagnostics" on the second line.
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Devices Act of 1990 and 21 CFR §807.92.
| Applicant/<br>Sponsor: | Oxford Brain Diagnostics Ltd<br>The Oxford Centre for Innovation<br>New Road, Oxford. OX1 1BY United Kingdom | | |
|------------------------|--------------------------------------------------------------------------------------------------------------|------------------------------------------------|--|
| Contact Person: | Terry Pollard<br>Chief Operating Officer<br>TEL: +44 (0)1865 261400 | | |
| Date Prepared: | December 3rd, 2024 | | |
| Subject Device: | Manufacturer: | Oxford Brain Diagnostics Ltd | |
| | Trade Name: | CDM Insights | |
| | Product Name: | CDM Insights | |
| | Classification Name: | Medical Image Management and Processing System | |
| | Classification Panel: | Radiology | |
| | Product Code: | QIH | |
| | Regulation: | 21 CFR §892.2050 | |
| | Class: | II | |
| | Prescription or OTC Use: | Rx | |
| Predicate Device: | Clearance: | K213706 | |
| | Clearance Date: | April 15, 2022 | |
| | Manufacturer: | Siemens Healthcare GmbH | |
| | Trade Name: | Al-Rad Companion Brain MR | |
| | Product Name: | Al-Rad Companion Brain MR | |
| | Classification Name: | Medical Image Management and Processing System | |
| | Classification Panel: | Radiology | |
| | Product Code: | QIH | |
| | Regulation: | 21 CFR §892.2050 | |
| | Class: | II | |
| | Prescription or OTC Use | Rx | |
Device CDM Insights is automated post-processing medical device software Description: that is used by radiologists, neurologists, and other trained healthcare practitioners familiar with the post-processing of magnetic resonance images. It accepts DICOM images using supported protocols and performs: automatic segmentation and quantification of brain structures and lesions, automatic post-acquisition analysis of diffusionweighted magnetic resonance imaging (DWI) data, and comparison of derived image metrics from multiple time-points.
> The values for a given patient are compared against age-matched percentile data from a population of healthy reference subjects. White matter hyperintensities can be visualized and quantified by volume. Output of the software provides numerical values and derived data as graphs and anatomical images with graphical color overlays.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Oxford Brain Diagnostics. The logo consists of a stylized brain graphic on the left, with the words "oxford brain diagnostics" to the right of the graphic. The brain graphic is made up of interconnected circles, some of which are white, gray, and one orange. The text is in a sans-serif font and is white.
CDM Insights output is provided in standard DICOM format as a DICOMencapsulated PDF report.
Indications for CDM Insights is a post-processing image analysis software that assists Use: trained healthcare practitioners in viewing, analyzing, and evaluating MR brain images of adults > 45 years of age.
CDM Insights provides the following functionalities:
- Automated segmentation and quantitative analysis of individual brain structures and white matter hyperintensities
- . Quantitative comparison of brain structures and derived values with normative data from a healthy population
- . Presentation of results for reporting that includes numerical values as well as visualization of these results
Summary of Technological Characteristics: The following aspects of the subject CDM Insights device and the predicate devices are identical:
- Intended Use
- Technological Principle
- Device Classification Name ●
- . Classification Panel
- Regulation Number
- Product Code
- Classification
- Prescription Use
The Indications for Use and Intended Users for the CDM Insights device and the Al-Rad Companion Brain MR device use only slightly different wording:
- . The CDM Insights devices refers to "healthcare practitioners" whereas the Al-Rad Companion Brain MR refers to "clinicians" in the Indications for Use and "healthcare professionals" in the Intended Users. We consider these to be equivalent.
In our Instructions for Use we have, in the same way as our primary predicate, clarified our definition of healthcare practitioners to include "radiologists, neurologists, and other trained healthcare practitioners familiar with the post-processing of magnetic resonance images".
The following features are substantially equivalent between the CDM Insights device and the Al-Rad Companion Brain MRI device (the CDM Insights Device presents percentiles, whereas the Al-Rad Companion Brain MRI device presents z-scores):
- Brain Morphometry Segmentation
- Brain Morphometry Quantification
- Brain Morphometry: Deviation Map ●
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for Oxford Brain Diagnostics. The logo features a stylized brain graphic on the left, composed of interconnected circles in white, gray, and orange. To the right of the brain graphic, the text "oxford brain diagnostics" is written in a sans-serif font. The background of the image is a dark blue color.
| | New Device | Predicate Device | Additional Predicate Device |
|-------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | CDM Insights | Al-Rad Companion Brain MR | OnQ Neuro |
| 510(k) Submitter<br>[Number] | Oxford Brain Diagnostics<br>Ltd | Siemens Healthcare GmBh | CorTechs Labs, Inc |
| 510(k) Number | K240680 | K213706 | K210831 |
| Indications for<br>Use | CDM Insights is a post-<br>processing image analysis<br>software that assists<br>trained healthcare<br>practitioners in viewing,<br>analyzing, and evaluating<br>MR brain images of adults<br>> 45 years of age.<br><br>CDM Insights provides the<br>following functionalities:<br>- Automated<br>segmentation and<br>quantitative analysis of<br>individual brain structures<br>and white matter<br>hyperintensities<br>- Quantitative comparison<br>of brain structures and<br>derived values with<br>normative data from a<br>healthy population<br>- Presentation of results<br>for reporting that includes<br>numerical values as well as<br>visualization of these<br>results | Al-Rad Companion Brain MR<br>is a post- processing image<br>analysis software that<br>assists clinicians in viewing,<br>analyzing, and evaluating<br>MR brain images.<br><br>Al-Rad Companion Brain MR<br>provides the following<br>functionalities:<br>- Automatic segmentation<br>and quantitative analysis of<br>individual brain structures<br>and white matter<br>hyperintensities<br>- Quantitative comparison of<br>each brain structure with<br>normative data from a<br>healthy population<br>- Presentation of results for<br>reporting that includes all<br>numerical values as well as<br>visualization of these results | OnQ Neuro is a fully automated<br>post-processing medical device<br>software intended for analyzing<br>and evaluating neurological MR<br>image data.<br><br>OnQ Neuro is intended to provide<br>automatic segmentation,<br>quantification, and reporting of<br>derived image metrics. OnQ<br>Neuro is additionally intended to<br>provide automatic fusion of<br>derived parametric maps with<br>anatomical MRI data. OnQ Neuro<br>is intended for use on brain<br>tumors, which are<br>known/confirmed to be<br>pathologically diagnosed cancer.<br>OnQ Neuro is intended for<br>comparison of derived image<br>metrics from multiple time-<br>points.<br>The physician retains the ultimate<br>responsibility for making the final<br>diagnosis and treatment decision. |
| Intended Users | The device is intended for<br>healthcare practitioners<br>familiar with the post-<br>processing of magnetic<br>resonance images | The device is intended for<br>healthcare professionals<br>familiar with the post<br>processing of magnetic<br>resonance images | Radiologists, Oncologists |
| Technological<br>Principle | Software | Software | Software |
| Device<br>Description | CDM Insights is automated<br>post-processing medical<br>device software that is<br>used by radiologists,<br>neurologists, and other<br>trained healthcare<br>practitioners familiar with<br>the post-processing of<br>magnetic resonance<br>images. It accepts DICOM<br>images using supported<br>protocols and performs: | Al-Rad Companion Brain MR<br>VA40 is an enhancement to<br>the predicate, Al-Rad<br>Companion Brain MR VA20<br>(K193290). Just as in the<br>predicate, Al-Rad Companion<br>Brain MR addresses the<br>automatic quantification and<br>visual assessment of the<br>volumetric properties of<br>various brain structures based<br>on T1 MPRAGE datasets. In Al- | OnQ Neuro is a fully automated<br>post-processing medical device<br>software that is used by<br>radiologists, oncologists, and<br>other clinicians to assist with<br>analysis and interpretation of<br>neurological MR images. It<br>accepts DICOM images using<br>supported protocols and<br>performs 1) automatic<br>segmentation and volumetric<br>quantification of brain tumors |
| Trade Name | New Device | Predicate Device | Additional Predicate Device |
| | CDM Insights | Al-Rad Companion Brain MR | OnQ Neuro |
| | automatic segmentation<br>and quantification of brain<br>structures and lesions,<br>automatic post-acquisition<br>analysis of diffusion-<br>weighted magnetic<br>resonance imaging (DWI)<br>data, and comparison of<br>derived image metrics<br>from multiple time-points.<br>The values for a given<br>patient are compared<br>against age-matched<br>percentile data from a<br>population of healthy<br>reference subjects. White<br>matter hyperintensities<br>can be visualized and<br>quantified by volume.<br>Output of the software<br>provides numerical values<br>and derived data as graphs<br>and anatomical images<br>with graphical color<br>overlays.<br>CDM Insights output is<br>provided in standard<br>DICOM format as a<br>DICOM-encapsulated PDF<br>report. | Rad Companion Brain MR<br>VA40, the quantification and<br>visual assessment extends to<br>white matter hyperintensities<br>on the basis of T1 MPRAGE<br>and T2 weighted FLAIR<br>datasets. These datasets are<br>acquired as part of a typical<br>head MR acquisition. The<br>results are directly archived in<br>PACS as this is the standard<br>location for reading by<br>radiologist. From a predefined<br>list of 30 structures (e.g.<br>Hippocampus, Left Frontal<br>Grey Matter, etc.), volumetric<br>properties are calculated as<br>absolute and normalized<br>volumes with respect to the<br>total intercranial volume. The<br>normalized values for a given<br>patient are compared against<br>age-matched mean and<br>standard deviations obtained<br>from a population of healthy<br>reference subjects.<br>The white matter<br>hyperintensities can be<br>visualized as a 3D overlay map<br>and the quantification in<br>count and volume as per 4<br>brain regions in the report.<br>As an update to the previously<br>cleared device, the following<br>modifications have been<br>made:<br>1. Modified Intended<br>Use Statement<br>2. Addition of white<br>matter hyperintensities<br>overlay map, count and<br>volume as per 4 brain regions<br>3. Enhanced DICOM<br>Structured Report (DICOM SR)<br>4. Updated deployment<br>structure | which are known/confirmed to<br>be pathologically diagnosed<br>cancer, 2) automatic post-<br>acquisition analysis of diffusion-<br>weighted magnetic resonance<br>imaging (DWI) data and<br>optional automated fusion of<br>derived image data with<br>anatomical MR images, and 3)<br>comparison of derived image<br>metrics from multiple time-<br>points.<br>Output of the software<br>provides values as numerical<br>volumes, and images of derived<br>data as grayscale intensity<br>maps and as graphical color<br>overlays on top of the<br>anatomical image. OnQ Neuro<br>output is provided in standard<br>DICOM format as image series<br>and reports that can be<br>displayed on most third-party<br>commercial DICOM<br>workstations |
| Brain<br>Morphometry<br>Segmentation | Pre-processing<br>functionality for automatic<br>segmentation and<br>volumetry of T1-weighted | Pre-processing functionality<br>for automatic segmentation<br>and volumetry of MPRAGE<br>data | Software performs automatic<br>segmentation |
| Trade Name | New Device | Predicate Device | Additional Predicate Device |
| CDM Insights | MRI data, e.g. MPRAGE or<br>similar acquisition data. | Al-Rad Companion Brain MR | OnQ Neuro |
| | | | |
| Brain<br>Morphometry<br>Quantification | Calculation of label maps<br>(display of brain<br>segmentation) and<br>partially combined label<br>maps (fused with the<br>processed T1-weighted<br>MRI data, e.g. MPRAGE or<br>similar acquisition. | Calculation of label maps<br>(display of brain<br>segmentation) and partially<br>combined label maps (fused<br>with the processed MPRAGE<br>data). | Software performs automatic<br>quantification |
| Brain<br>Morphometry:<br>Deviation Map | Calculation of deviation<br>map (representation of<br>brain status in relation to<br>reference data) and<br>partially combined<br>deviation maps (fused<br>with the processed T1-<br>weighted MRI data, e.g.<br>MPRAGE or similar<br>acquisition) | Calculation of deviation map<br>(representation of brain<br>status in relation to<br>reference data) and partially<br>combined deviation maps<br>(fused with the processed<br>MPRAGE data) User<br>customizable color labels for<br>the overlay map | Unknown |
| Diffusion Analysis | Yes, using a single-<br>compartment diffusion<br>model. | No | Yes, using single and multi-<br>compartment diffusion models. |
| Support<br>Longitudinal<br>Analysis | Yes, performs comparison<br>of derived image metrics<br>from multiple time points | No | Yes, performs comparison of<br>derived image metrics from<br>multiple time points. |
| Brain White<br>Matter<br>Hyperintensities<br>Segmentation | Pre-processing<br>functionality for automatic<br>segmentation and<br>volumetry of T1-weighted<br>data, that includes<br>MPRAGE or similar<br>acquisition, and FLAIR<br>data. (volumetry of white<br>matter hypointensities if<br>FLAIR data is not available) | Pre-processing functionality<br>for automatic segmentation<br>and volumetry of MPRAGE<br>and FLAIR data. | Information on White Matter<br>Hyperintensities is not provided. |
| Brain White<br>Matter<br>Hyperintensities<br>Quantification | Calculation of white<br>matter hyperintensities<br>volume (white matter<br>hypointensities volume if<br>FLAIR data is not<br>available). | Calculation of white matter<br>hyperintensities count and<br>volume as per 4 brain<br>regions.…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.