Local and global reconstruction algorithms for modeling and tractography
—
—
—
—
—
—
Indications for Use
ANDI is intended for display of medical images and other healthcare data. It includes functions for image review, basic measurements, planning, and visualization (MPR reconstructions), Modules are available for image processing and atlas-assisted visualization and segmentation, where an output can be generated for use by a system capable of reading DICOM image sets. ANDI is indicated for use in the processing of diffusion-weighted MRI sequences into 3D maps that represent white matter tracts. The information presented by ANDI is for measurement of brain white matter tracts only without making a prediction, diagnosis, or interpretation of brain health, It is the responsibility of the physician to review all clinical information associated with a patient in order to make a diagnosis and to determine next steps in the clinical care of the patient. Typical users of ANDI are medical professionals, including but not limited to neurologists and radiologists. ANDI should be used only as adjunctive information. The decision made by trained medical professionals will be considered final. It is not a diagnostic aid.
Device Story
ANDI is cloud-based quantitative imaging software; extracts features from medical images to provide adjunctive information. Inputs: diffusion-weighted MRI (dMRI) and T1-weighted images (optional reversed-phase encoded B=0 images). Processing: local (per-voxel) and global reconstruction algorithms; modeling; tractography; white matter bundling; maps microstructural properties of white matter along specific bundles. Output: DICOM-encapsulated PDF report containing quantitative insights into white matter microstructure. Used in clinical settings by neurologists/radiologists; integrates via API with RIS/PACS. Output serves as reference information for comparison against control populations or longitudinal patient data; assists physicians in clinical decision-making. Benefits: provides quantitative visualization of white matter tracts to support standard-of-care evaluations.
Clinical Evidence
No clinical studies were performed. Evidence consists of non-clinical performance testing, including software verification and validation, assessment of measured endpoints, AI-based brain extraction, and robustness of the processing pipeline. Summative evaluation confirmed report structure and content.
Indicated for processing diffusion-weighted MRI sequences into 3D maps representing white matter tracts for measurement purposes. Intended for use by medical professionals (e.g., neurologists, radiologists) as adjunctive information. Not a diagnostic aid; does not provide predictions, diagnoses, or interpretations of brain health.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the logo for the Department of Health & Human Services - USA, featuring a stylized human figure. To the right is the logo for the U.S. Food & Drug Administration (FDA), with the FDA acronym in a blue square and the full name "U.S. FOOD & DRUG ADMINISTRATION" written in blue text.
July 25, 2023
Imeka Solutions, Inc. % Valerie Lacroix Executive Vice President 195. Belvédère Nord #201 Sherbrooke, QC J1H 4A7 Canada
Re: K230913
Trade/Device Name: ANDI Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH, LLZ Dated: June 26, 2023 Received: June 27, 2023
Dear Valerie Lacroix:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
D.K.
Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
| Submission Number (if known) | |
|------------------------------|--|
|------------------------------|--|
K230913
Device Name
ANDI
Indications for Use (Describe)
ANDI is intended for display of medical images and other healthcare data. It includes functions for image review, basic measurements, planning, and visualization (MPR reconstructions), Modules are available for image processing and atlas-assisted visualization and segmentation, where an output can be generated for use by a system capable of reading DICOM image sets.
ANDI is indicated for use in the processing of diffusion-weighted MRI sequences into 3D maps that represent white matter tracts. The information presented by ANDI is for measurement of brain white matter tracts only without making a prediction, diagnosis, or interpretation of brain health, It is the responsibility of the physician to review all clinical information associated with a patient in order to make a diagnosis and to determine next steps in the clinical care of the patient.
Typical users of ANDI are medical professionals, including but not limited to neurologists and radiologists. ANDI should be used only as adjunctive information. The decision made by trained medical professionals will be considered final. It is not a diagnostic aid.
Type of Use (Select one or both, as applicable)
X | Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments reqarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
#### Sponsor Information:
IMEKA Solutions, Inc. 195. Belvédère Nord Sherbrooke, QC, Canada J1H 4A
| Contact Person | Valerie Lacroix |
|----------------|------------------------------------------------------------------------|
| | Executive Vice President |
| | (888) 311-0599 |
| | <a href="mailto:valerie.lacroix@imeka.ca">valerie.lacroix@imeka.ca</a> |
#### Date of Summary: March 31, 2023
### Device Name and Classification
| Common or Usual Name: | Automated Radiological Image Processing Software |
|------------------------------|-----------------------------------------------------------------------------------------------------------|
| Proprietary Name: | ANDI |
| Classification Name: | System, Image Processing, Radiological Picture Archiving and<br>Communications System (21 CFR § 892.2050) |
| Classification Product Code: | QIH, LLZ |
| Predicate Device: | Quicktome (K203518) |
### Description of Device
ANDI is quantitative imaging software that extracts features from medical images to provide adjunctive information for use with the complete standard of care evaluation of the patient. The device processes diffusion weighted images using local (on a per-voxel basis) and global (for the whole brain) reconstruction algorithms, respectively called modeling, tractography, and white matter bundling, to map microstructural properties of the white matter. ANDI achieves its intended use by extracting white matter bundles that connect specific regions of the brain and then performing a microstructure analysis along those bundles.
ANDI combines two families of features derived from diffusion Magnetic Resonance Imaging (dMRI) along with T1-weighted images as an auxiliary input to augment processing. T1-weighted imaging is a general imaging modality that highlights differences in tissue types (skull, cerebrospinal fluid, white / gray matter), whereas dMRI is sensitive to the direction of white matter structures, based on the movements of water protons. Combining information from these MRI types gives ANDI data needed to quantify and visualize local water diffusion properties and to map these microstructural properties of the white matter along specific white matter bundles.
The ANDI analysis techniques provide quantitative insight into white matter microstructure,
{5}------------------------------------------------
#### 510(k) Summary - ANDI
corresponding to the local environment of each neurological fiber population. The resulting report provides trained medical professionals with reference information as an adjunct to care. The information included in the report is intended to be used by the trained medical professionals as a comparison between a patient and a control population as well as a comparison to the subject himself, with an optional longitudinal comparison.
#### Indications for Use
ANDI is intended for display of medical images and other healthcare data. It includes functions for image review, basic measurements, planning, and visualization (MPR reconstructions). Modules are available for image processing and atlas-assisted visualization and segmentation, where an output can be generated for use by a system capable of reading DICOM image sets.
ANDI is indicated for use in the processing of diffusion-weighted MRI sequences into 3D maps that represent white matter tracts. The information presented by ANDI is for measurement of brain white matter tracts only without making a prediction, diagnosis, or interpretation of brain health. It is the responsibility of the physician to review all clinical information associated with a patient in order to make a diagnosis and to determine next steps in the clinical care of the patient.
Typical users of ANDI are medical professionals, including but not limited to neurologists and radiologists. ANDI should be used only as adjunctive information. The decision made by trained medical professionals will be considered final. It is not a diagnostic aid.
Comparative Data for Determining Substantial Equivalence of New Device to Predicate Device Information provided in this 510(k) submission shows that ANDI is substantially equivalent to the predicate device, Quicktome, cleared under K203518, in terms of intended use, indications for use, physical characteristics, workflow, anatomical location, and technological characteristics.
Both devices are cloud-based image processing software that apply an algorithm for modeling diffusion data and generating tractography (whole brain CSD tractography) reports for clinicians to use in planning and visualization. Both devices accept similar inputs. T1 and multi-directional diffusion weighted images (DWI), to process MRI information and output a report using CSD as the underlying diffusion modeling technique used by a tractography process to reconstruct white matter pathways. The output of both devices is intended as adjunctive information for medical professionals.
While both devices share many characteristics, there are identified differences, none of which have a negative impact on the device safety or effectiveness, nor do they raise any potential safety risks:
- The predicate utilizes the Glasser atlas, which applies surfacic definitions delineating 180 areas per hemisphere bounded by sharp changes in cortical architecture, function, connectivity, and/or topography. In contrast, ANDI uses a proprietary white matter bundles atlas defined by 3D models to segment the major white matter pathways from the subject's tractogram. The differences in the atlas types are minimal, with both approaches used to identify white matter bundles.
- The predicate device accepts T1, T2, FLAIR, and DWI images. The subject device does not accept T2 or FLAIR images because the subject device does not generate endpoints from T2 or
{6}------------------------------------------------
FLAIR images. Likewise, ANDI accepts reversed-phase encoded B=0 image as an optional input. This difference in input format does not present concerns for safety or effectiveness given that the main information source for both devices is the DWI, and the T1, T2, or FLAIR images are auxiliary images used to augment the quality of processing. Given that the subject device uses a robust T1-based refinement of processing, there is no performance impact in not supporting T2 or FLAIR.
- . Output data for the two devices are similar, however the output format between the two devices is different. The predicate device outputs a DICOM Structured Report, while the subject device output is a DICOM encapsulated PDF report. This difference does not raise concerns for safety or effectiveness because both are DICOM-compliant files that are supported by RIS/PACS and both formats can contain the same information.
- . While the predicate features a user interface for planning and visualization, ANDI relies on an API for image input and report output. This difference does not present a safety or efficacy concern because the output is sent to the RIS/PACS in a DICOM-compliant format.
- . The predicate device uses the DICOM format for its input. The subject device uses the NIFTI file format for its input. This difference does not present a safety or effectiveness concern because the image data in both formats is identical.
Comparisons to the Predicate device in conjunction with design verification and validation activities described in the 510(k) submission support substantial equivalence of ANDI.
### Non-Clinical Test Summary
Non-clinical performance testing has been performed in compliance with the following International and FDA recognized consensus standards and FDA guidance document:
- ISO 14971:2019, Medical Devices Application of Risk Management to Medical Devices .
- ANSI/AAMI/IEC 62304:2006/A1:2016, Medical Device Software Software Life Cycle ● Processes
- Digital Imaging and Communications in Medicine (DICOM) Set (NEMA PS 3.1 3.20) ●
- . Guidance for Industry and FDA Staff - Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, November 2021
Performance tests were conducted to assess the measured end points, AI-based brain extraction, and robustness of the processing pipeline. A summative evaluation was performed to assess the content and structure of the ANDI report.
Software testing and validation were done according to written test protocols established before testing was conducted. Test results were reviewed by designated technical professionals before software proceeded to formalize after ensuring that the software fully satisfies all expected and previously defined system requirements and features.
Through the performance test. it was confirmed that ANDI meets all performance test criteria and that all functions work as intended. Test results support the conclusion that device performance satisfies the design intent and is equivalent to its predicate device.
### Clinical Test Summary
No clinical studies were considered necessary and performed.
{7}------------------------------------------------
### Conclusion
In conclusion, the tests conducted, as well as all verification activities, demonstrate that the design specifications and technological characteristics of ANDI meet applicable requirements and standards for the safety and effectiveness of the device for its intended use. There are differences in technological characteristics between the predicates and the proposed device, but the nature of those differences do not raise new or different questions of safety or effectiveness as compared to the predicate devices. Therefore, ANDI is substantially equivalent to the currently marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.