K240600 · East End Medical I, LLC · DYB · Sep 6, 2024 · Cardiovascular
Device Facts
Record ID
K240600
Device Name
SafeCross Vascular Introducer System (4001)
Applicant
East End Medical I, LLC
Product Code
DYB · Cardiovascular
Decision Date
Sep 6, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
The SafeCross™ Vascular System is a Steerable Balloon Introducer System with Access Dilator used for introducing various cardiovascular catheters into the vasculature including the heart. In addition, the device can be used for monitoring intracardiac pressures, sampling blood, and infusing solutions.
Device Story
Steerable balloon introducer system for cardiovascular catheter placement; includes steerable introducer sheath (SIS), long access dilator (LAD), and short access dilator (SAD). SIS features distal positioning balloon for stability and bi-directional deflection (0° to 180°). Physician operates device in clinical setting to navigate vasculature; balloon inflated with contrast/saline mix for visualization and stability. System facilitates catheter introduction, pressure monitoring, blood sampling, and fluid infusion. No RF puncture member included. Benefits include accurate positioning and stable access for diagnostic/therapeutic procedures.
Clinical Evidence
Bench testing only. Biocompatibility testing performed on the new short access dilator (SAD) per ISO 10993-1, including cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemocompatibility, and material-mediated pyrogenicity. All tests met specifications.
Technological Characteristics
Steerable introducer sheath with distal positioning balloon; bi-directional deflection (0-180°). Materials: Pebax, barium, polycarbonate, polypropylene, HDPE. Dimensions: 8.5F compatible, 75 cm length. Connectivity: None. Sterilization: Not specified. Software: None.
Indications for Use
Indicated for introducing cardiovascular catheters into the vasculature, including the heart, and for monitoring intracardiac pressures, blood sampling, and fluid infusion. Intended for general vascular access; not for transseptal puncture.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
Predicate Devices
SafeCross Transseptal Puncture Device and Introducer (TSP/I) System (K203459)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
September 6, 2024
East End Medical I, LLC Carlos Cruz OA/RA Director 10320 USA Today Way Miramar, Florida 33025
Re: K240600
Trade/Device Name: SafeCross Vascular Introducer System (4001) Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: February 22, 2024 Received: March 4, 2024
Dear Carlos Cruz:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Kalkidan Molla -S
Kalkidan A. Molla, MS Acting Assistant Director DHT2B: Division of Circulatory Support, Structural, and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240600
### Device Name
SafeCross Vascular System is a Steerable Balloon Introducer System with Access Dilator
Indications for Use (Describe)
The SafeCross Vascular System is a Steerable Balloon Introducer System with Access Dilator used for introducing various cardiovascular catheters into the vasculature including the heart. In addition, the device can be used for monitoring intracardiac pressures, sampling blood, and infusing solutions.
| Type of Use (Select one or both, as applicable) |
|--------------------------------------------------------------------------------------|
| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for EastEnd Medical. The word "EastEnd" is in dark gray, while the word "Medical" is in light gray. To the left of the text is a peach-colored circle.
# K240600
## 510(k) SUMMARY
#### 1. GENERAL INFORMATION
#### 1.1 Submitter and 510(k) Owner
East End Medical I LLC 10320 USA Today Way Miramar, FL 33025
#### 1.2 Official Correspondent
David Joseph Neugaard 10320 USA Today Way Miramar, FL 33025 +1 (954) 507-7887
#### 1.3 Date of Preparation
September 5, 2024
#### 2. NAME OF THE DEVICE
#### 2.1.1 Trade/Proprietary Name
SafeCross Vascular Introducer System
#### 2.1.2 Common/Usual Name
Introducer Catheter Septostomy Catheter
#### 2.1.3 Classification Information
| Classification Name: | Introducer Catheter |
|----------------------------|---------------------------|
| Classification Regulation: | 21 CFR§870.1340 |
| Class: | 2 |
| Product Code: | DYB, Introducer, Catheter |
| Panel: | Cardiovascular |
#### 3. PREDICATE DEVICES
Predicates: SafeCross Transseptal Puncture Device and Introducer (TSP/I) System, K203459. The predicate has not been the subject to a design-related recall.
References: Oscor Destino™ Reach Steerable Sheath, K151951
#### 4. DESCRIPTION OF THE DEVICE
The SafeCross Vascular Introducer System is used to introduce various cardiovascular catheters into the vasculature including the heart. Unlike the predicate device, it does not include an RF Puncture Member and is not intended to cross the atrial septum. The subject device system includes three (3) components: the Steerable Introducer Sheath (SIS), a Long Access Dilator (LAD), and a Short Access Dilator (SAD). In the subject device, the SIS and
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for EastEndMedical. The word "EastEnd" is in a dark gray font, while the word "Medical" is in a lighter gray font. To the left of the text is a circular graphic that is light orange in color.
LAD are identical to their respective corresponding components of the predicate device (K203459). The Vascular SIS features a compliant, specially shaped, atraumatic overhanging Positioning Balloon on its distal end to facilitate the accurate positioning and stability of the SIS while navigating the patient's vasculature. The physician is able to bi-directionally deflect the distal segment in a range of 0° to 180°. The subject device includes a new component, the SAD (12 Fr, 20 cm in length).
#### INTENDED USE 5.
The SafeCross™ Vascular System is a Steerable Balloon Introducer System with Access Dilator used for introducing various cardiovascular catheters into the vasculature including the heart. In addition, the device can be used for monitoring intracardiac pressures, sampling blood, and infusing solutions.
#### INTENDED USE COMPARED TO THE PREDICATES 6.
The SafeCross Vascular Introducer System has an intended use statement that is a subset of the intended uses from the Predicate. The devices also share the same target patient population, the same users and conditions of use (Table 1).
| | Subject Device<br>SafeCross Vascular<br>Introducer System<br>East End Medical I<br>Inc. | Predicate<br>SafeCross Transseptal<br>Puncture Device and<br>Introducer (TSP/I)<br>System, K203459 | Same or Different | |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use<br>Statement | The SafeCross™ Vascular<br>System is a Steerable<br>Balloon Introducer System<br>with Access Dilator used for<br>introducing various<br>cardiovascular catheters into<br>the vasculature including the<br>heart. In addition, the device<br>can be used for monitoring<br>intracardiac pressures,<br>sampling blood, and infusing<br>solutions. It is intended for<br>use in general vascular<br>access without requiring<br>transseptal puncture. | The SafeCross Transseptal<br>Puncture Device and Introducer<br>(TSP/I) System is used to<br>introduce various cardiovascular<br>catheters to the heart, including<br>the left side of the heart. The<br>system enables left heart access<br>through a puncture of the atrial<br>septum during a transseptal<br>catheterization procedure. In<br>addition, the device can be used<br>for monitoring intracardiac<br>pressures, sampling blood, and<br>infusing solutions | Different<br>(Expanded indication for<br>general vascular access). | |
| Comparison<br>to Reference<br>Device | The Subject Device's<br>intended use aligns with the<br>Oscor Destino™ Reach<br>Steerable Sheath (K151951)<br>Reference Device, which is<br>used for introducing<br>diagnostic and therapeutic<br>devices into the human<br>vasculature, including<br>intracardiac, renal, or other<br>peripheral placements, but<br>not neural placements. | N/A | Similar<br>(The use of the Subject Device<br>in general vascular access<br>aligns with the intended use of<br>the Reference Device.) | |
| | Predicate Device - Steerable<br>Introducer Sheath<br>SafeCross TSP/I System K203459<br>East End Medical I Inc. | Reference Device -<br>Oscor Destino™ Reach<br>Steerable Sheath<br>(K151951) | Subject Device - SafeCross<br>Vascular Introducer System<br>East End Medical I Inc. | Comparison of Subject<br>Device to Predicate or<br>Reference |
| Deflectable<br>Distal Segment<br>Length | S: 5.5 mm<br>M: 6.6 mm<br>L: 9.5 mm | S: 1.7 cm<br>M: 2.2 cm<br>L: 5.0 cm | S: 5.5 mm | Similar, subset |
| Materials | Biocompatible materials | Biocompatible materials | Biocompatible materials | Same |
| Radiopacity | Shaft is radiopaque Positioning<br>balloon is inflated<br>with 1 cc of 20% contrast - 80%<br>saline through the Positioning Balloon<br>Inflation Port | Radiopaque distal tip | Shaft is radiopaque Positioning<br>balloon is inflated with 1 cc of 20%<br>contrast - 80%<br>saline through the Positioning<br>Balloon Inflation Port | Same |
| Proximal end | Hemostasis valve and 2 side ports with<br>3-way stopcock for Positioning Balloon<br>Inflation Port and Flushing Line Port<br>(injection /aspiration of fluids) | Hemostasis valve and a side port<br>with 3-way stopcock for injection<br>or aspiration of fluids | Hemostasis valve and 2 side ports with<br>3-way stopcock for Positioning Balloon<br>Inflation Port and Flushing Line Port<br>(injection /aspiration of fluids) | Same |
| Distal end/tip | Inflatable positioning balloon at the<br>distal tip to facilitate positioning at the<br>fossa ovalis and stabilize the introducer<br>system for precise use of the RF<br>Puncture Member | No balloon | Inflatable positioning balloon at the<br>distal tip to facilitate positioning and<br>stability of the Steerable Introducer<br>Sheath assembly within the<br>vasculature. | Similar, appropriate for<br>indication |
| OD | 0.160" | 0.157", 0.185" or 0.208" | 0.160" | Same |
| ID | 8.5F compatible | 8.5F, 10F or 12F | 8.5F compatible | Same |
| Sheath effective length | 75 cm | 67 to 77 cm | 75 cm | Same |
| Compatible guidewire | 0.035" | 0.035" | 0.035" | Same |
| Distal curve(s) | Min 0° - Max 180° 1 | bi-deflection distal tip | Min 0° - Max 180° 1 | Same |
| | Predicate Device - Steerable<br>Introducer Sheath<br>SafeCross TSP/I System K203459<br>East End Medical I Inc. | Reference Device -<br>Oscor Destino™ Reach<br>Steerable Sheath<br>(K151951) | Subject Device - SafeCross<br>Vascular Introducer System<br>East End Medical I Inc. | Comparison of Subject<br>Device to Predicate or<br>Reference |
| Access Dilator | 103 cm TSP/I AD provided with kit;<br>compatible with 0.035" guidewire | 85 cm to 94cm<br>The dilator has a tapered distal tip | LAD Length: 103cm<br>SAD Length: 20cm<br>Both provided with kit; compatible with<br>0.035" guidewire | Similar, addition of the small<br>access dilator |
| RF Puncture Member | RF Puncture Member: Working length:<br>103 cm, provided with kit; compatible<br>with 0.035" guidewire | Not Included | Not Included | Different: Subject device<br>does not include RF Puncture<br>member in the kit, the<br>Vascular system consists of<br>three components co-<br>packaged in a product kit.<br>1. Vascular Device<br>2. Long Access Dilator<br>(LAD)<br>3. Short Access Dilator<br>(SAD) |
| Endovascular<br>Navigation Path | Venous: From femoral vein to inferior<br>vena cava to right atrium (to left atrium). | Venous: From femoral vein to<br>inferior vena cava to right atrium.<br>Arterial: From femoral artery to<br>abdominal aorta bifurcation to<br>either thoracic aorta or distally to<br>contralateral femoral/popliteal<br>artery. | Venous: From femoral vein to inferior<br>vena cava to right atrium.<br>Arterial: From femoral artery to<br>abdominal aorta bifurcation to either<br>thoracic aorta or distally to contralateral<br>femoral/popliteal artery. | Same<br>Similar2 |
# Table 1. Intended Use / Indications for Use Comparison
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for EastEndMedical. The words "EastEnd" are in a dark gray font, while the word "Medical" is in a lighter gray font. To the left of the text is a circular shape in a light orange color.
# 7. TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE PREDICATES
A comparison of the technological features between the subject device and the Predicate is shown in Table 2 below for the Steerable Introducer Catheter.
## Table 2. SafeCross Vascular Introducer System Comparison to Predicate
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for EastEndMedical. The word "EastEnd" is in a dark gray font, while the word "Medical" is in a lighter gray font. To the left of the text is a circular graphic in shades of orange.
¹ Per bench deflection angle testing.
2 Venous vaculature navigation from the groin (femoral vein) up to the 3 devices. Arterial vascular navigation is the same between Reference Device and similar between the Predicate and Subject device in terms of complexity and placement.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for EastEndMedical. The word "EastEnd" is in a dark gray font, while the word "Medical" is in a lighter gray font. To the left of the text is a circular shape that is colored in shades of orange.
| | (LAD) Access Dilator<br>(8.5 Fr x 103 cm)<br>East End Medical I Inc. | (SAD) Access Dilator<br>(12 Fr x 20 cm)<br>East End Medical I Inc. | Same or Different |
|----------------|----------------------------------------------------------------------|--------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Access Dilator | Shaft, Pebax, barium | Shaft, Polypropylene | Different: The SAD was subjected to a series of<br>biocompatibility tests in accordance with FDA<br>guidance, using International Standard ISO 10993-1.<br>• Cytotoxicity<br>• Sensitization<br>• Intracutaneous Reactivity |
| | Hub, Polycarbonate | Hub, High Density Polyethylene | • Acute Systemic Toxicity<br>• Hemocompatibility (Hemolysis, Complement Activation,<br>Partial Thromboplastin Time, In Vivo Thrombogenicity,<br>Heparinized Blood Platelet and Leukocyte Count Assay)<br>• Material Mediated Pyrogenicity<br>Results: The test Article (SAD) meet all the test specifications. |
### Table 3. Component Materials (AD)
#### 7.1 Similarities and Differences in Technology Comparison
The SafeCross Vascular Introducer System is similar to the SafeCross TSP/I System (K203459) in terms of components and operational use. As outlined in detail in Tables 1-3, technological differences compared to the predicate SafeCross TSP/I System include that the subject SafeCross Vascular Introducer System contains an SAD, does not include an RF puncture member, and incorporates two modifications to the indication for use: (1) removal of transseptal crossing to the left heart and (2) inclusion of peripheral vasculature. While these modifications do not alter the intended use of the device, the subject device's inclusion of general vasculature to the indication necessitated a risk assessment using a reference device (Oscor Destino Reach Steerable Sheath - K151951) to help address safety and effectiveness of the new technological characteristics. A detailed assessment of anatomical navigation paths used for both the reference and subject device led to an SIS deflection angle testing ("distal curve(s)" in Table 2) specification change from 90° to 149.6°. There have been no changes to manufacturing or sterilization of the device. While the packaging materials remain unchanged, the configuration differs between the subject and predicate devices. The subject device, SafeCross Vascular Introducer System, includes both a Long Access Dilator (LAD) and a Short Access Dilator (SAD), whereas the predicate device is co-packaged with a single LAD and an RF puncture member.
The technology of the Subject Steerable Introducer Sheath is identical to the Predicate Introducer Sheaths. Both the SafeCross Steerable Introducer Sheath and the Predicate are copackaged with an access dilator. The technology of the Long Access Dilator (LAD) is identical between the predicate and subject devices.
#### Performance Testing 8.
Design Verification and Validation testing that was conducted on the predicate device was leveraged to support the subject device, due to a risk-based assessment of the similarity or equivalence in geometry and functionality of the device components.
While no new design verification or validation testing was performed on the subject
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for East End Medical. The words "EastEnd" are in a dark gray font, while the word "Medical" is in a lighter gray font. To the left of the words is a circular shape in a light orange color.
device, in using technical information from the reference device (K151951), a new specification for the SIS deflection angle (149.6°) was retroactively applied to predicate device prior SIS deflection testing to ensure the subject device can be curved to navigate through all potential navigation pathways. Testing was not repeated, as the predicate and subject device have an identical SIS component that met the 149.6° specification.
Additionally, biocompatibility testing was conducted on the newly proposed short access dilator (SAD) for the subject device to provide evidence that the new technological characteristic (i.e., new material of the SAD) does not raise different questions of safety and effectiveness compared to the predicate.
The SAD was subjected to a series of biocompatibility tests in accordance with FDA guidance, using International Standard ISO 10993-1.
- · Cytotoxicity
- · Sensitization
- · Intracutaneous Reactivity
- · Acute Systemic Toxicity
- · Hemocompatibility (Hemolysis, Complement Activation, Partial Thromboplastin Time, in vivo thrombogenicity, Heparinized Blood Platelet and Leukocyte Count Assay)
- · Material Mediated Pyrogenicity
#### 9. Conclusions
The SafeCross Vascular Introducer System has the same intended use as the SafeCross TSP/I System predicate device, and similar technological characteristics. The differences in technological characteristics (i.e., new material of the SAD) do not raise new or different questions of safety and effectiveness since all biocompatibility testing conducted meet the specifications. The Oscor Destino™ Reach Device was used as a reference device for some of the performance evaluations. Therefore, the SafeCross Vascular Introducer System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.