K240381 · Smith & Nephew, Inc. · LZO · Jun 27, 2024 · Orthopedic
Device Facts
Record ID
K240381
Device Name
CATALYSTEM Femoral Stems
Applicant
Smith & Nephew, Inc.
Product Code
LZO · Orthopedic
Decision Date
Jun 27, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CATALYSTEM Femoral Stems are intended for total and partial hip arthroplasty in skeletally mature patients for the following indications: Hip components are indicated for individuals undergoing primary surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of degenerative joint disease or any of its composite diagnoses of osteoarthritis, avascular necrosis, and traumatic arthritis. Hip components are also indicated for inflammatory degenerative joint disease including rheumatoid arthritis, congenital dysplasia, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques, and fracture-dislocation of the hip. The CATALYSTEM Femoral Stems are intended for use without bone cement.
Device Story
CATALYSTEM Femoral Stems are non-cemented hip prostheses for primary total or partial hip arthroplasty. Implants are manufactured from forged titanium alloy with CP-Ti and HA coating. Available in collared and collarless variants with standard and high neck offset options. Used by orthopedic surgeons in clinical settings; intended for implantation into the proximal femur. Device functions as a structural replacement for the femoral head/neck; integrates with Smith & Nephew femoral heads and acetabular components. Provides mechanical stability for the hip joint; restores joint function; alleviates pain associated with degenerative or inflammatory joint disease. Provided sterile via gamma irradiation.
Clinical Evidence
No clinical tests were performed. Safety and efficacy are supported by bench testing, including distal and neck fatigue testing (ISO 7206-4/6, ASTM F2996-20), ROM evaluation (ISO 21535), HA coating performance (ASTM F1147-05, F1044-05, F1160-14, ISO 13779), MR compatibility, and biocompatibility (ISO 10993-1).
Technological Characteristics
Materials: Forged titanium alloy, CP-Ti, and HA coating. Design: Non-cemented femoral hip stems, collared/collarless variants, standard/high neck offsets. Standards: ISO 7206-4/6, ISO 21535, ISO 13779, ASTM F2996-20, ASTM F1980-21, ASTM F1147-05, ASTM F1044-05, ASTM F1160-14, ASTM F2182-19e2, ASTM F2052-15, ASTM F2213-2017, ASTM F2119-07, ISO 10993-1. Sterilization: Gamma irradiation.
Indications for Use
Indicated for skeletally mature patients undergoing primary total or partial hip arthroplasty for degenerative joint disease (osteoarthritis, avascular necrosis, traumatic arthritis), inflammatory degenerative joint disease (rheumatoid arthritis), congenital dysplasia, femoral neck fracture, trochanteric fractures of the proximal femur with head involvement, or fracture-dislocation of the hip, where other treatments have failed.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
Predicate Devices
POLARSTEM Collared (Standard and Lateral) and Valgus Femoral (K143739)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 27, 2024
Smith & Nephew Inc. Chelsea Bagley Sr. Regulatory Affairs Specialist 1450 Brooks Rd Memphis, Tennessee 38116
Re: K240381
Trade/Device Name: CATALYSTEM Femoral Stems Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO, MEH, LPH Dated: February 7, 2024 Received: May 31, 2024
Dear Chelsea Bagley:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Christopher Ferreira -S
for
Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K240381
Device Name CATALYSTEM Femoral Stems
### Indications for Use (Describe)
The CATALYSTEM Femoral Stems are intended for total and partial hip arthroplasty in skeletally mature patients for the following indications:
Hip components are indicated for individuals undergoing primary surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of degenerative joint disease or any of its composite diagnoses of osteoarthritis, avascular necrosis, and traumatic arthritis.
Hip components are also indicated for inflammatory degenerative joint disease including rheumatoid arthritis, congenital dysplasia, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques, and fracture-dislocation of the hip.
The CATALYSTEM Femoral Stems are intended for use without bone cement.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|--|
| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | |
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 510(k) #: K240381 | 510(k) Summary | Prepared on: 2024-06-27 |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|-------------------------|
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | Smith & Nephew Inc. | |
| Applicant Address | 1450 Brooks Rd Memphis TN 38116 United States | |
| Applicant Contact Telephone | 4705058820 | |
| Applicant Contact | Mrs. Chelsea Bagley | |
| Applicant Contact Email | chelsea.bagley@smith-nephew.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | CATALYSTEM Femoral Stems | |
| Common Name | Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous<br>uncemented prosthesis | |
| Classification Name | Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or<br>Non-Porous, Uncemented | |
| Regulation Number | 888.3353 | |
| Product Code(s) | LZO, MEH, LPH | |
| Legally Marketed Predicate Devices<br>21 CFR 807.92(a)(3) | | |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K143739 | POLARSTEM Collared (Standard and Lateral) and Valgus Femoral | LZO |
| K052792 | ANTHOLOGY Hip Stem | JDI |
| Device Description Summary<br>21 CFR 807.92(a)(4) | | |
| The purpose of this Traditional 510(k) is to notify the FDA of our intent to market the Smith & Nephew CATALYSTEM Femoral Stems. The<br>CATALYSTEM Femoral Stems consists of femoral stem implants intended for primary hip arthroplasty in skeletally mature patients. The | | |
CATALYSTEM Femoral Stems are provided sterile to the user via gamma irradiation.
The subject CATALYSTEM Femoral Stems are comprised of two variants: CATALYSTEM Collared and CATALYSTEM Collarless. Both variants are available in standard and high neck offset options. The CATALYSTEM Femoral Stems are non-cemented femoral hip stems manufactured from forged titanium alloy and feature a CP-Ti and HA coating.
The CATALYSTEM Femoral Stems are intended to be used with Smith & Nephew femoral heads and acetabular components for total hip arthroplasty and hip hemiarthroplasty.
# Intended Use/Indications for Use
21 CFR 807.92(a)(5)
The CATALYSTEM Femoral Stems are intended for total and partial hip arthroplasty in skeletally mature patients for the following indications:
Hip components are indicated for individuals undergoing primary surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of degenerative joint disease or any of its composite diagnoses of osteoarhitis, avascular necrosis, and traumatic arthritis.
{4}------------------------------------------------
Hip components are also indicated for inflammatory degenerative joint disease including rheumatoid arthritis, congenital dysplasia, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques, and fracture-dislocation of the hip.
The CATALYSTEM Femoral Stems are intended for use without bone cement.
## Indications for Use Comparison
The indications for use of the subject device were updated to clarify the type of hip arthroplasty they are indicated for.
## Technological Comparison
The overall technological characteristics including device design, function, materials and intended use for the CATALYSTEM Femoral Stems are substantially equivalent to the predicate Smith & Nephew systems cleared under the premarket notifications ANTHOLOGY Hip Stems (K052792, 10/07/2005) and POLARSTEM Hip Stems (K143739, 04/20/2015). The indications for use of the subject device were updated to clarify the type of hip arthroplasty they are indicated for.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Mechanical Testing and Analyses
- · Distal Fatigue Testing per ISO 7206-4 and ASTM F2996-20
- · Neck Fatigue Testing per ISO 7206-6 and ASTM F2996-20
- · Range of Motion (ROM) Evaluation per ISO 21535
- · Engineering Analysis of Impingement Performance
- · Fretting and Corrosion Performance Rationale
- · Femoral Head Disassembly Performance Rationale
Hydroxyapatite (HA) Coating Testing
- · Chemical Characterization per ASTM F1980-21, ISO 13779-2 and ISO 13779-3
- · Static Tensile Attachment Strength Testing per ASTM F1 147-05 and ISO 13779-4
- · Static Shear Attachment Testing per ASTM F1044-05R17e01
- · Shear Fatigue Strength Testing per ASTM F1044-05R17e01 and ASTM F1160-14
- · Dissolution Rate and Solubility Product
MR Compatibility per ASTM F2182-19e2, ASTM F2052-15, ASTM F2213-2017 and ASTM F2119-07 Biocompatibility per ISO 10993-1
No clinical tests were performed to support safety and efficacy of the subject device.
The performance mechanical and coating testings and analyses were used as a basis for the determination of substantial equivalence. The results of each of these tests show that when compared to the subject device has met the required acceptance criteria and there are no additional risks. As such, the CATALYSTEM Femoral Stems are substantially equivalent to the predicate devices for the intended use.
## 21 CFR 807.92(a)(6)
## 21 CFR 807.92(a)(5)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.