POLARSTEM Collared (Standard and Lateral) and Valgus Femoral Stem with Ti/HA
K143739 · Smith & Nephew, Inc. · LZO · Apr 20, 2015 · Orthopedic
Device Facts
Record ID
K143739
Device Name
POLARSTEM Collared (Standard and Lateral) and Valgus Femoral Stem with Ti/HA
Applicant
Smith & Nephew, Inc.
Product Code
LZO · Orthopedic
Decision Date
Apr 20, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The POLARSTEM Collared and Valgus Stem with Ti/HA (INTEGRATION-PLUSY) is indicated for: - Advanced degeneration of the hip joint as a result of degenerative, post-traumatic or rheumatoid arthritis. - Fracture or avascular necrosis of the femoral head . - . Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement The POLARSTEM with Ti/HA is intended for single use only and is to be implanted without bone cement.
Device Story
The POLARSTEM Collared and Valgus Femoral Stem is a cementless hip prosthesis component. It is implanted by orthopedic surgeons during total hip replacement or revision surgery to replace the femoral head. The device is manufactured from titanium alloy (Ti-6Al-4V) with a porous titanium plasma/hydroxyapatite (Ti/HA) coating to promote osseointegration. It features a 12/14 taper, a polished neck, and distal grooves for rotational stability. The device is intended for single use. By providing a stable, biocompatible femoral component, it restores hip joint function, reduces pain, and improves mobility for patients suffering from degenerative joint disease or previous surgical failure.
Clinical Evidence
Bench testing only. Evidence includes hip stem fatigue testing and range of motion analysis to demonstrate mechanical integrity and functional equivalence to the predicate.
Technological Characteristics
Materials: Titanium alloy (Ti-6Al-4V) with porous titanium plasma/hydroxyapatite (Ti/HA) coating. Design: Cementless femoral stem with 12/14 taper, polished neck, and distal grooves. Sterilization: Not specified. Connectivity: None (mechanical implant).
Indications for Use
Indicated for patients with advanced hip joint degeneration (degenerative, post-traumatic, or rheumatoid arthritis), femoral head fracture or avascular necrosis, or failure of previous hip surgeries (reconstruction, fixation, arthrodesis, hemiarthroplasty, surface replacement, or total hip replacement).
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
Predicate Devices
POLARSTEM Standard and Lateral Femoral Stems with Ti/HA (K130728)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo features a stylized image of a human face in profile, with three overlapping profiles suggesting a sense of community and support. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 20, 2015
Smith & Nephew, Incorporated Mr. Jeff Sprague Regulatory Affairs Specialist 7135 Goodlett Farms Parkway Cordova, Tennessee 38016
Re: K143739
Trade/Device Name: POLARSTEM Collared (Standard and Lateral) and Valgus Femoral Stem with Ti/HA Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis. Regulatory Class: Class II Product Code: LZO, KWY, LWJ, MEH Dated: March 10, 2015 Received: March 11, 2015
Dear Mr. Sprague:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{2}------------------------------------------------
### Page 2 - Mr. Jeff Sprague
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# INDICATIONS FOR USE
#### Smith & Nephew POLARSTEM Collared and Valgus Femoral Stem with Ti/HA
510(k) Number (if known): K143739
#### Device Name: POLARSTEM Collared (Standard and Lateral) and Valgus Femoral Stem with Ti/HA
#### Indications for Use:
The POLARSTEM Collared and Valgus Stem with Ti/HA (INTEGRATION-PLUSY) is indicated for:
- Advanced degeneration of the hip joint as a result of degenerative, post-traumatic or rheumatoid arthritis.
- Fracture or avascular necrosis of the femoral head .
- . Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement
The POLARSTEM with Ti/HA is intended for single use only and is to be implanted without bone cement.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{4}------------------------------------------------
## 510(k) Summary Smith & Nephew POLARSTEM Collared and Valgus Femoral Stem with Ti/HA
| Submitted by: | Smith & Nephew, Inc.<br>Advanced Surgical Devices Division<br>7135 Goodlett Farms Parkway<br>Cordova, Tennessee 38016 |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date of Summary: | April 1, 2015 |
| Contact Person | Jeff Sprague, Regulatory Affairs<br>T (901) 399-5215 F (901) 721-2736 |
| Name of Device: | POLARSTEM Collared (Standard and Lateral)<br>and Valgus Femoral Stem with Ti/HA |
| Common Name: | Total Hip Joint, Femoral Component,<br>Cementless |
| Device Classification Name and Reference: | 21 CFR 888.3353 – Hip joint<br>metal/polymer/metal semi-constrained<br>cemented or nonporous uncemented<br>prosthesis<br><br>21 CFR 888.3390 – Hip joint femoral (hemi-hip<br>metal/polymer cemented or uncemented<br>prosthesis<br><br>21 CFR 888.3360 – Hip joint femoral (hemi-hip<br>metallic cemented or uncemented prosthesis |
| Device Class: | Class II |
| Panel Code: | Orthopaedics/87 |
| Product Code: | LZO, KWY, LWJ, MEH |
# Device Description
The POLARSTEM Collared (Standard and Lateral) and Valgus femoral stems designed for noncemented use are made of a titanium alloy (Ti-6Al-4V) with a fully porous titanium plasma/hydroxyapatite coating (Ti/HA). The design incorporates a highly polished neck area, distal grooves to increase rotational stability, and a 12/14 taper. The collared and valgus femoral stems are substantially equivalent to the POLARSTEM standard and lateral femoral stem with Ti/HA cleared via K130728.
#### Intended Use
The POLARSTEM Collared (Standard and Lateral) and Valgus Stem with Ti/HA (INTEGRATION-PLUS®) is indicated for:
{5}------------------------------------------------
- Advanced degeneration of the hip joint as a result of degenerative, post-traumatic or rheumatoid arthritis.
- . Fracture or avascular necrosis of the femoral head
- Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement
The POLARSTEM with Ti/HA is intended for single use only and is to be implanted without bone cement.
The above indications are substantially equivalent to the indications cleared for the POLARSTEM standard and lateral femoral stem with Ti/HA (K130728). The only difference is the removal of the following bullet points as they are considered subsets of the more general indications seen above. Removal of the following points does not change the indications.
- All forms of osteoarthritis
- Patients with hips at risk of dislocation
- Femoral neck fracture or proximal fracture to hip joint
# Technological Characteristics
A review of the mechanical data indicates that the POLARSTEM Collared (Standard and Lateral) and Valgus Femoral Stem with Ti/HA is capable of withstanding expected in vivo loading without failure.
# Substantial Equivalence Information
The overall design, materials, and indications for use for the POLARSTEM Collared (Standard and Lateral) and Valgus Femoral Stem with Ti/HA are substantially equivalent to the following commercially available predicate devices. The Ti/HA coating on the POLARSTEM Collared and Valgus femoral stem is identical in composition, properties, and manufacturing location (Smith & Nephew AG) to the predicate below.
| Manufacturer | Description | Submission<br>Number | Clearance<br>Date |
|-----------------------------------|------------------------------------------------------------|----------------------|-------------------|
| Smith & Nephew<br>Orthopaedics AG | POLARSTEM Standard and Lateral Femoral<br>Stems with Ti/HA | K130728 | 10/10/2013 |
The following tests were used as a basis for the determination of substantial equivalence:
- . Hip Stem Fatigue Testing
- . Range of Motion Analysis
All tests which are in relation to the surface characterization (physical, chemical or mechanical) are discussed in detail in the Ti/HA Coating Master File MAF - 1762 Amendment 1 and 2 and are not included in this 510(k).
{6}------------------------------------------------
#### Conclusion
As previously noted, this 510(k) Premarket Notification is being submitted to request clearance for the POLARSTEM Collared (Standard and Lateral) and Valgus Femoral Stem with Ti/HA coated by Smith & Nephew AG. Based on the similarities to the predicate component and a review of the mechanical testing performed, the devices are substantially equivalent to the commercially available predicate device listed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.