ARVIS® Shoulder is indicated for assisting the surgeon in the positioning and alignment of implants relative to reference alignment axes and landmarks in stereotactic orthopedic surgery. The system aids the surgeon in making intraoperative measurements and locating anatomical structures of the shoulder joint based on the patient's preoperative imaging. ARVIS® Shoulder is indicated for total shoulder arthroplasty using the Enovis AltiVate implant system.
Device Story
ARVIS® Shoulder is a computer-controlled surgical navigation system for total shoulder arthroplasty. It uses preoperative CT scans as input; AI-based algorithms perform automatic image segmentation and landmarking to create a surgical plan. During surgery, the system uses electronic hardware (head/waist-mounted) to track surgical instruments relative to the patient's anatomy. The surgeon views intraoperative measurements and guidance on the system display to assist in implant positioning and alignment. The device is used in the OR by surgeons to improve accuracy in shoulder joint replacement procedures, specifically with the Enovis AltiVate implant system. It provides real-time navigation feedback, aiding clinical decision-making regarding implant placement, potentially improving surgical outcomes.
Clinical Evidence
No clinical data. Bench testing included positional accuracy (ASTM F2554), accuracy after mechanical disturbances (ASTM F3107), and cadaveric validation. Software validation of AI segmentation and landmarking used 60 CT scans (36-86 years old). Segmentation metrics: Dice Similarity Coefficient (DSC) > 0.98, Mean Absolute Distance (MAD) ≤ 0.352mm, Hausdorff Distance (HD) ≤ 0.917mm. Landmarking metrics: Mean Euclidean Distance (ED) ≤ 3.29mm, Angular Separation (AS) ≤ 8.70°, and Successful Point/Angular Classification Rates (SPCR/SACR) ≥ 83.3%.
Indicated for patients undergoing total shoulder arthroplasty using the Enovis AltiVate implant system. Assists surgeons in positioning/aligning implants and locating shoulder anatomical structures via preoperative imaging.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{1}------------------------------------------------
April 29, 2024
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Insight Medical Systems, Inc. % Stefanie Auf Der Mauer Regulatory Affairs Consultant Stefanie Michele Auf der Mauer Asmuss (sole trader) 22 Monfield Rochestown, Cork T12C65D Ireland
Re: K240062
Trade/Device Name: ARVIS® Shoulder Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: March 29, 2024 Received: March 29, 2024
Dear Stefanie Auf Der Mauer:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Shumaya Ali-S
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices
{3}------------------------------------------------
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
#### K240062
Device Name
ARVIS® Shoulder
#### Indications for Use (Describe)
ARVIS® Shoulder is indicated for assisting the surgeon in the positioning and alignment of implants relative to reference alignment axes and landmarks in stereotactic orthopedic surgery. The system aids the surgeon in making intraoperative measurements and locating anatomical structures of the shoulder joint based on the patient's preoperative imaging. ARVIS® Shoulder is indicated for total shoulder arthroplasty using the Enovis AltiVate implant system.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) Summary
#### Prepared: 2024-03-29
# 1. Submitter Information
| Submitter: | Insight Medical Systems, Inc.<br>9801 Metric Blvd, Suite 200<br>Austin, TX, 78758<br>United States |
|--------------------|----------------------------------------------------------------------------------------------------|
| Submitter Contact: | Matthew Ryan, VP R&D Technology-Insight<br>Email: matthew.ryan@enovis.com<br>Phone: 949-290-8210 |
| Correspondent: | Stefanie Auf der Mauer<br>Email: stefanie.aufdermauer@enovis.com<br>Phone: +353857591774 |
# 2. Subject Device
| Trade Name: | ARVIS® Shoulder |
|---------------|------------------------------------------|
| Common Name: | Orthopedic Stereotaxic Instrument |
| Regulation: | 21 CFR 882.4560 – Stereotaxic Instrument |
| Product Code: | OLO, LLZ, QIH |
# 3. Predicate Device
| 510(k) Number: | K203115 |
|----------------|-----------------------------------|
| Trade Name: | ARVIS® Surgical Navigation System |
| Manufacturer: | Insight Medical Systems, Inc. |
| Product Code: | OLO |
#### Reference Devices 4.
| 510(k) Number: | K213546 |
|----------------|-----------------------------------------------------------------------|
| Trade Name: | ExactechGPS Total Shoulder Application, Equinoxe Planning<br>Software |
| Manufacturer: | Blue Ortho |
| Product Code: | OLO, LLZ |
| 510(k) Number: | K222767 |
| Trade Name: | PeekMed web |
| Manufacturer: | Peek Health, S.A. |
| Product Code: | QIH, LLZ |
{6}------------------------------------------------
# 5. Device Description Summary
ARVIS® Shoulder is a computer-controlled surgical navigation system intended to provide intra-operative measurements to the surgeon to aid in selection and positioning of orthopedic implant components.
The subject device is the equivalent shoulder system of the predicate ARVIS® Surgical Navigation System used for indicated knee and hip arthroplasties. ARVIS® Shoulder combines software, electronic hardware and surgical instruments to intraoperatively track tools and locate anatomical structures based on the patient's preoperative imaging.
The navigation platform uses the same electronic hardware, mounted on the surgeon's head and waist, as the predicate device. A new equivalent navigation application and a new equivalent surgical instrument set are provided to allow surgeons to navigate instruments in shoulder arthroplasty procedures.
The ARVIS® Shoulder workflow involves CT based reconstruction of the patient's shoulder anatomy and preoperative planning to enable image-based navigation. The surgeon uses the plan data as guidance to navigate and help position shoulder instruments and implants.
The preoperative planning platform uses Al-based automatic image segmentation and landmarking algorithms. The data used to train and test the algorithms was labeled and validated in advance by trained experts. The total data consisted of 300 CT scans (from 300 patients) acquired from candidates for shoulder joint replacement surgery. The cohort was partitioned into two disjoint subsets through random sampling, with 80% producing a training dataset and 20% constituting the test dataset. The training dataset consisted of 240 CT scans (from 240 patients). Patient ages ranged from 36 to 89 years (mean age of 70), with 46% male and 54% female. All CT scans were acquired using FDA cleared CT scanners.
The navigation system is intended to be used with the Enovis AltiVate implant system.
ARVIS® Shoulder displays measurements as described in Performance Claims.
## 6. Indications for Use
ARVIS® Shoulder is indicated for assisting the surgeon in the positioning and alignment of implants relative to reference alignment axes and landmarks in stereotactic orthopedic surgery.
The system aids the surgeon in making intraoperative measurements and locating anatomical structures of the shoulder joint based on the patient's preoperative imaging.
ARVIS® Shoulder is indicated for total shoulder arthroplasty using the Enovis AltiVate implant system.
## 7. Indications for Use Comparison
The subject device is indicated specifically for total shoulder arthroplasty whereas the predicate device is indicated specifically for knee and hip related arthroplasties. The difference in the Indications for Use does not result in a new intended use; the systems
{7}------------------------------------------------
are both intended to provide intra-operative measurements to the surgeon to aid in selection and positioning of orthopedic implant components. The difference in the Indications for Use does not affect the safety and effectiveness of the subject device as demonstrated by safety and performance testing.
#### Technological Comparison 8.
The subject device ARVIS® Shoulder is the same as the predicate ARVIS® Surgical Navigation System in Intended Use, electronic components, materials, and navigation tracking technology and accuracy.
The workflow, patient interface and components used in the subject and predicate device are nearly identical during surgical navigation. The primary technological difference between the subject device and predicate is that the subject device is an image-based system (requiring preoperative imaging and planning for navigation inputs) whereas the predicate device is an imageless system that solely uses intraoperative registrations for development of surgeon quidance. Similar image processing and planning software functions exist in identified reference devices. The additional software functions do not change the Intended Use of the subject device, nor do they raise different questions of safety and effectiveness.
#### 9. Non-Clinical Tests Summary and Conclusions
The following performance tests were completed to demonstrate substantial equivalence of safety and effectiveness with the predicate device:
- Positional Accuracy Testing (ASTM F2554) .
- . Accuracy After Mechanical Disturbances Testing (ASTM F3107)
- Benchtop Accuracy Testing ●
- Cadaveric Accuracy Testing ●
- Cadaver Validation ●
- . Software Verification and Validation, including validation of the deep learning algorithms used for automatic image segmentation and landmarking. Details about this performance test are as follows:
- o Study steps: The reference, ground-truth label for each CT volume in the testing dataset was obtained by a manual process, reviewed, and approved by a trained engineer with more than 2 years' experience with medical image processing and 2 orthopaedic surqeons with subspecialty qualifications in upper limb surgery. The algorithm-generated outputs for every CT scan of the test dataset were compared to the manual ground truth segmentation and landmarking outputs by calculating applicable metrics and assessing against predefined acceptance criteria.
- o Testing dataset: Validation was performed on a set of 60 CT scans (from 60 patients) acquired from patients who were candidates for shoulder joint replacement surgery. To ensure independence of training and test datasets, the latter was always stored separately and only made available for performance validation.
{8}------------------------------------------------
The testing dataset is composed of 55% female and 45% male patient CT scans. Age ranges from 36 to 86 years, the mean age is 67 years, and the standard deviation is 10.5 years. Ethnicity was not recorded.
The cohort of CT scans were acquired using FDA cleared CT scanners with settings defined in the CT Protocol or comparable. CT scanner manufacturers include Toshiba, Siemens, Philips, GE Medical Systems and Canon.
- Testing criteria: The difference between the automatic seqmentation outputs O and manual outputs were quantified using the Dice Similarity Coefficient (DSC), Mean Absolute Distance (MAD) and Hausdorff Distance (HD). The pre-defined acceptance criteria used are specified in the results table below. The difference between the automatic landmarking outputs and the manual outputs were quantified using the Euclidean Distance (ED) and the Angular Separation (AS), and the Successful Point and Angular Classification Rates (SPCR, SACR). The pre-defined acceptance criteria used are specified in the results table below.
- Testing results: The AI algorithms used for automatic image segmentation O and landmarking were able to successfully execute all required steps and met all pre-defined acceptance criteria.
| | DSC | | MAD | | HD | |
|---------|------------------|--------|------------------|---------|-----------------|---------|
| | AC | Result | AC | Result | AC | Result |
| Scapula | Avg DSC<br>> 0.9 | 0.981 | Avg MAD<br>≤ 2mm | 0.229mm | Avg HD<br>≤ 5mm | 0.824mm |
| Humerus | > 0.9 | 0.989 | ≤ 2mm | 0.352mm | ≤ 5mm | 0.917mm |
### Segmentation Validation Results Summary
## Landmarking Validation Results Summary
| | Mean ED / AS | | SPCR / SACR | |
|------------------------|--------------|--------|-------------|--------|
| | AC | Result | AC | Result |
| Glenoid Center | | 1.79mm | | 95.0% |
| Trigonum | | 1.86mm | | 95.0% |
| Inferior Point | ≤ 3.72mm | 2.11mm | ≥75% | 94.9% |
| Medial Epicondyle | | 3.19mm | | 83.3% |
| Lateral Epicondyle | | 3.29mm | | 83.3% |
| Neck Plane Position | | 2.01mm | | 90.0% |
| Neck Plane Orientation | ≤ 10° | 8.70° | | 86.7% |
Clinical testing was not required to demonstrate substantial equivalence.
The performance data provided in this submission demonstrate that ARVIS® Shoulder is as safe, effective, and performs as well as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.