Centurion™ Vision System (Active Sentry™) (8065753057)
K233902 · Alcon Laboratories, Inc. · HQC · Jan 10, 2024 · Ophthalmic
Device Facts
Record ID
K233902
Device Name
Centurion™ Vision System (Active Sentry™) (8065753057)
Applicant
Alcon Laboratories, Inc.
Product Code
HQC · Ophthalmic
Decision Date
Jan 10, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 886.4670
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Centurion™ Vision System (Active Sentry™) is indicated for emulsification, separation, irrigation, and aspiration of cataracts, residual cortical material and lens epithelial cells, vitreous aspiration and cutting associated with anterior vitrectomy, bipolar coagulation, and intraocular lens injection. The AutoSert™ IOL Injector Handpiece is intended to deliver qualified AcrySof™ intraocular lenses into the eye following cataract removal. The AutoSert™ IOL Injector Handpiece achieves the functionality of injection of intraocular lenses. The AutoSert™ IOL Injector Handpiece is indicated for use with AcrySof ™ lenses SN60WF. SN6AD1, SN6AT3 through SN6AT9, as well as approved AcrySof ™ lenses that are specifically indicated for use with this inserter, as indicated in the approved labeling of those lenses.
Device Story
Ophthalmic surgical system for small incision cataract extraction and IOL injection; utilizes CENTURION™ OZil™, Active Sentry™, and INFINITI™ OZil™ handpieces. System emulsifies and aspirates lens material while replacing with balanced salt solution (BSS™) to maintain stable eye chamber volume. Surgeon controls phaco power, aspiration vacuum, flow rate, vitrectomy cut rate, IOL injection rate, and coagulation power via footswitch. Used in clinical surgical settings by ophthalmologists. Provides real-time fluidics management to assist in cataract removal and lens implantation; benefits patient through stable intraocular pressure and controlled surgical environment.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by the fact that the only change is a labeling modification (new warning statement for accessories), with no changes to the device design, materials, or technology.
Technological Characteristics
Phacofragmentation system; includes ultrasonic handpieces (OZil™), fluidics management, and IOL injection capability. Operates via controlled power, aspiration, and irrigation. System components, materials, and sterilization processes remain identical to the predicate device (K161794).
Indications for Use
Indicated for patients undergoing cataract surgery requiring emulsification, separation, irrigation, and aspiration of cataracts, residual cortical material, and lens epithelial cells; anterior vitrectomy; bipolar coagulation; and intraocular lens (IOL) injection using qualified AcrySof™ lenses.
Regulatory Classification
Identification
A phacofragmentation system is an AC-powered device with a fragmenting needle intended for use in cataract surgery to disrupt a cataract with ultrasound and extract the cataract.
Predicate Devices
CENTURION™ Vision System (Active Sentry™) (K161794)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 10, 2024
Alcon Laboratories, Inc. Rita De Rama Manager, Regulatory Affairs 6201 South Freeway Fort Worth, Texas 76134
Re: K233902
Trade/Device Name: Centurion™ Vision System (Active Sentry™) (8065753057) Regulation Number: 21 CFR 886.4670 Regulation Name: Phacofragmentation System Regulatory Class: Class II Product Code: HQC Dated: December 6, 2023 Received: December 11, 2023
Dear Rita De Rama:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product: and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Claudine H. Krawczyk -S
Claudine Krawczyk Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure
{3}------------------------------------------------
### Indications for Use
Submission Number (if known)
K233902
Device Name
Centurion TM Vision System (Active Sentry TM) (8065753057)
Indications for Use (Describe)
The Centurion™ Vision System (Active Sentry™) is indicated for emulsification, separation, irrigation, and aspiration of cataracts, residual cortical material and lens epithelial cells, vitreous aspiration and cutting associated with anterior vitrectomy, bipolar coagulation, and intraocular lens injection.
The AutoSert™ IOL Injector Handpiece is intended to deliver qualified AcrySof™ intraocular lenses into the eye following cataract removal.
The AutoSert™ IOL Injector Handpiece achieves the functionality of injection of intraocular lenses. The AutoSert™ IOL Iniector Handbiece is indicated for use with AcrySof ™ lenses SN60WF. SN6AD1, SN6AT3 through SN6AT9, as well as approved AcrySof ™ lenses that are specifically indicated for use with this inserter, as indicated in the approved labeling of those lenses.
Type of Use (Select one or both, as applicable)
escription Use (Part 21 CFR 801 Subpart D)
he-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summarv
Prepared on: 2024-01-08
| Contact Details | 21 CFR 807.92(a)(1) | |
|------------------------------------|----------------------------------------------------------|--------------|
| Applicant Name | Alcon Laboratories, Inc. | |
| Applicant Address | 6201 South Freeway Fort Worth TX 76134 United States | |
| Applicant Contact Telephone | 949 886 5787 | |
| Applicant Contact | Mrs. Rita De Rama | |
| Applicant Contact Email | rita.derama@alcon.com | |
| Device Name | 21 CFR 807.92(a)(2) | |
| Device Trade Name | Centurion™ Vision System (Active Sentry™) (8065753057) | |
| Common Name | Phacofragmentation system | |
| Classification Name | Unit, Phacofragmentation | |
| Regulation Number | 886.4670 | |
| Product Code | HQC | |
| Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) | |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K161794 | CENTURION™ Vision System (Active Sentry™) | HQC |
| Device Description Summary | 21 CFR 807.92(a)(4) | |
Alcon's CENTURION™ Vision System is an ophthalmic surgical instrument designed to provide cataract extraction using the CENTURION™ OZil™handpiece, the CENTURION™ Active Sentry™ handpiece, and the INFIN门!" OZZI" handpiece.
The CENTURION™ Vision System is intended for use in small incision cataraction and IOL injection surgical procedures. This system allows the surgeon to emulsify and aspirate the eye, while replacing aspirated fluid and lens material with balanced salt solution (BSS™). This process maintains a stable (inflated) eye chamber volume. Using system controls, the surgeon regulates the amount of power applied to the handpiece tip, the rate of aspiration vacuum, and the flow of BSS™ irrigating solution. The system includes a footswitch to enable the surgeon to control flow of fluidics, aspiration rate, phaco power, vitrectory cut rate, IOL injection rate, and coagulation power.
## Intended Use/Indications for Use
The Centurion™ Vision System (Active Sentry") is indication, separation, irrigation, and aspiration of cataracts, residual cortical material and lens epithelial cells, vitreous and cutting associated with anterior vitrectomy, bipolar coagulation, and intraocular lens injection.
The AutoSert™ IOL Injector Handpiece is intended to deliver qualified AcrySof™
intraocular lenses into the eye following cataract removal.
The AutoSert™ IOL Injector Handpiece achieves the function of intraocular lenses. The AutoSert™ IOL Injector Handpiece is indicated for use with AcrySof ™ lenses SN60WF, SN6AT3 through SN6AT9, as well as approved AcrySof™ lenses that are specifically indicated for use with this inserter, as indicated in the approved labeling of those lenses.
### Indications for Use Comparison
21 CFR 807.92(a)(5)
21 CFR 807.92(a)(5)
{5}------------------------------------------------
The indications for use are the same between the subject device and the predicate device.
## Technological Comparison
21 CFR 807.92(a)(6)
A modification to add a new Warning statement in the labeling for the TurboSonics™ ultrasonic tips, which are accessories for the CENTURION® Vision System, is the only change being made to the predicate device.
The modified CENTURION™ Vision System (Active Sentry") is substantially equivally submitted CENTURION™ Vision System (Active Sentry™) in that both devices:
· Have the same indication for use.
· Use the same operating principle.
· Use the same technology, including components used.
· Are of the same design, incorporate the same materials, have the same shelf life, and are packaged and sterilized using the same materials and processes.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Non-clinical testing was previously performed in the predicate device and can be referenced in K161794. A modification to add a new Warning statement in the labeling for the TurboSonics™ ultrasonic tips, which are accessories for the CENTURION® Vision System, is the only change being made to the predicate device. Thus, the subject device described in this submission is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.