The CENTURION® Vision system (Active SentryTM) is indicated for emulsification, separation, irrigation, and aspiration of cataracts, residual cortical and lens epithelial cells, vitreous aspiration and cutting associated with anterior vitrectomy, bipolar coagulation, and intraocular lens injection. The AutoSert IOL Injector Handpiece is intended to deliver qualified ACRYSOF intraocular lenses into the eye following cataract removal. The AutoSert IOL Injector Handpiece achieves the function of intraocular lenses. The AutoSert IOL Injector Handpiece is indicated for use with the ACRYSOF lenses SN6OWF. SNWAD1. SN6AT3 through SN6AT9. as well as approved ACRYSOF lenses that are specifically indicated for use with this inserter, as indicated in the approved labeling of those lenses.
Device Story
Ophthalmic surgical system for cataract extraction and IOL injection; utilizes phacoemulsification handpieces (OZil, Active Sentry) to emulsify/aspirate lens material while maintaining stable intraocular chamber volume via balanced salt solution (BSS) irrigation. Surgeon controls phaco power, aspiration rate, vacuum, and fluid flow via footswitch, touch screen, or remote. System provides real-time display and self-diagnostics. Used in clinical settings by surgeons; facilitates small-incision cataract surgery; improves surgical efficiency and chamber stability.
Clinical Evidence
Bench testing only. Evidence includes biocompatibility testing (cytotoxicity, sensitization, acute systemic toxicity, ocular irritation, material characterization) and electrical/electromagnetic safety testing (IEC 60601-1, 60601-1-2, 60601-2-2, 60601-1-8, 80601-2-58, 60601-1-6, 62366). All tests passed.
Technological Characteristics
Modular ophthalmic surgical system; microprocessor-based; Windows-based software. Features: peristaltic pump, longitudinal (44kHz) and torsional (32kHz) ultrasound modes, gravity/pressurized irrigation. User interface: touch screen, footswitch, remote. Connectivity: wired/wireless footswitch. Sterilization: EtO and Gamma for disposables. Complies with IEC 60601-1, 60601-1-2, 60601-2-2, 60601-1-8, 80601-2-58, 60601-1-6, and 62366.
Indications for Use
Indicated for patients undergoing cataract surgery requiring emulsification, separation, irrigation, aspiration, anterior vitrectomy, bipolar coagulation, or IOL injection. Compatible with specific ACRYSOF intraocular lenses.
Regulatory Classification
Identification
A phacofragmentation system is an AC-powered device with a fragmenting needle intended for use in cataract surgery to disrupt a cataract with ultrasound and extract the cataract.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an emblem featuring a stylized human figure with three overlapping profiles, creating a sense of depth and connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 14, 2017
Alcon Research, Ltd. Mr. Paul Swift, RAC Director, Global Regulatory Affairs Project 6201 South Freeway Dr. Fort Worth, TX 76134-2099
Re: K161794
Trade/Device Name: Centurion Vision System (Active Sentry") Regulation Number: 21 CFR 886.4670 Regulation Name: Phacofragmentation System Regulatory Class: Class II Product Code: HQC Dated: March 6, 2017 Received: March 8, 2017
Dear Mr. Swift:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Bradley S. Cunningham -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K161794
Device Name
CENTURION VISION SYSTEM (Active Sentry™)
#### Indications for Use (Describe)
The CENTURION® Vision system (Active SentryTM) is indicated for emulsification, separation, irrigation, and aspiration of cataracts, residual cortical and lens epithelial cells, vitreous aspiration and cutting associated with anterior vitrectomy, bipolar coagulation, and intraocular lens injection. The AutoSert IOL Injector Handpiece is intended to deliver qualified ACRYSOF intraocular lenses into the eye following cataract removal.
The AutoSert IOL Injector Handpiece achieves the function of intraocular lenses. The AutoSert IOL Injector Handpiece is indicated for use with the ACRYSOF lenses SN6OWF. SNWAD1. SN6AT3 through SN6AT9. as well as approved ACRYSOF lenses that are specifically indicated for use with this inserter, as indicated in the approved labeling of those lenses.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 5 510(k) Summary
This 510(k) summary document has been prepared in accordance with section 21 CFR 807.92.
### Submitter of the 510(k)
| Company: | Alcon Laboratories, Inc.<br>6201 South Freeway<br>Fort Worth, TX 76134-2099, USA |
|-------------------------|----------------------------------------------------------------------------------|
| Primary Contact Person: | Paul Swift, RAC<br>Director, Global Regulatory Affairs |
| Phone: | 817.615.2952 |
| Fax: | 817.551.4630 |
| Email: | Paul.Swift@alcon.com |
| Date Prepared: | December 16, 2016 |
### Devices Subject to this 510(k)
| Trade Names: | CENTURION® Vision System (Active Sentry® |
|--------------------|------------------------------------------|
| Regulation Number: | 21 CFR 886.4670 |
| Regulation Name: | Phacofragmentation System |
| Regulatory Class: | Class II |
| Product Code: | HQC |
## A. Predicate Device
CENTURION Vision System K121555 December 21, 2012
## B. Device Description
Alcon's CENTURION ® Vision System is an ophthalmic surgical instrument designed for use in cataract extraction using the CENTURION® OZil® handpiece, the CENTURION® Active Sentry® handpiece, and the INFINITI® OZil® handpiece.
The CENTURION® Vision System is intended for use in small incision cataract lens extraction and IOL injection surgical procedures. This system allows the surgeon to emulsify and aspirate
{4}------------------------------------------------
the lens in the eye, while replacing aspirated fluid and lens material with balanced salt solution. This process maintains a stable (inflated) eye chamber volume. Using system controls, the surgeon regulates the amount of power applied to the handpiece tip, the rate of aspiration, vacuum, and the flow of BSS® irrigating solution. The system includes a footswitch to enable the surgeon to control flow of fluidics, aspiration rate, phaco power, vitrectomy cut rate, IOL injection rate, and coagulation power.
## C. Indications for Use
The CENTURION® Vision system (Active Sentry®) is indicated for emulsification, separation, irrigation, and aspiration of cataracts, residual cortical material and lens epithelial cells, vitreous aspiration and cutting associated with anterior vitrectomy, bipolar coagulation, and intraocular lens injection. The AutoSert IOL Injector Handpiece is intended to deliver qualified ACRYSOF intraocular lenses into the eye following cataract removal.
The AutoSert IOL Injector Handpiece achieves the functionality of injection of intraocular lenses. The AutoSert IOL Injector Handpiece is indicated for use with the ACRYSOF lenses SN6OWF, SNWAD1, SN6AT3 through SN6AT9, as well as approved ACRYSOF lenses that are specifically indicated for use with this inserter, as indicated in the approved labeling of those lenses.
{5}------------------------------------------------
# D. Comparison to Technological Characteristics with the Predicate Device
| Characteristic | CENTURION® Vision System | CENTURION® Vision System<br>(Active Sentry®) |
|----------------------------------------------------|------------------------------------------------|------------------------------------------------|
| 510 (k) # | K121555 | K161794 |
| Intended Use | Anterior Segment Ophthalmic<br>Surgery | Anterior Segment Ophthalmic<br>Surgery |
| Hardware (Modular) Design | Yes | Yes |
| Microprocessor based | Yes | Yes |
| Programmable for multiple<br>surgeons | Yes | Yes |
| User Interface | Footswitch, Touch Screen and<br>Remote control | Footswitch, Touch Screen and<br>Remote control |
| Self-Diagnostics | Yes | Yes |
| Real-Time Display | Yes | Yes |
| Software platform | Windows based | Windows based |
| U/S Pulse Mode | Yes | Yes |
| Heads-Up Display | Yes (optional only) | Yes (optional only) |
| Anterior Vitrectomy | Yes | Yes |
| Posterior Vitrectomy | No | No |
| Occlusion Mode | | |
| Longitudinal | Yes | Yes |
| Torsional | Yes | Yes |
| Operating U/S Frequency | | |
| Longitudinal | 44kHz nominal | 44kHz nominal |
| Torsional | 32kHz nominal | 32kHz nominal |
| Support for ALCON®<br>UltraChopper® Tip | Yes (K091777) | Yes (K091777) |
| Power Watch Mode | Yes | Yes |
| IOL Injection Mode | Yes | Yes |
| Coagulation/Cautery/Diathermy | Yes | Yes |
| Irrigation | | |
| Gravity | Yes | Yes |
| Pressurized | Yes | Yes |
| Maximum Irrigation Pressure<br>Threshold (in mmHg) | Less than 207 | Less than 207 |
| Aspiration Pump Type | Peristaltic | Peristaltic |
| Vacuum Range (in mmHg) | 0-650(+) | 0-650(+) |
| Console Screen | Active Matrix color LCD | Active Matrix color LCD |
| Multi-function Foot Pedal | | |
| Wired | Yes | Yes |
| Wireless | Yes | Yes |
| Characteristic | CENTURION®<br>Vision System | CENTURION® Vision System<br>(Active Sentry®) |
| Remote Control | Yes(if selected) | Yes(if selected) |
| Automatic IV Pole | Yes | Yes |
| Voice Confirmation | Yes | Yes |
| Prime Mode | Yes | Yes |
| Machine Height Width and<br>Depth | 63" x 23" x 30" max | 63" x 23" x 30" max |
| Machine Weight | 235 lbs max | 235 lbs max |
| Electrical Power Specifications | 100-240 V AC, 50/60 Hz | 100-240 V AC, 50/60 Hz |
| Operating Temperature Range | 10 - 35°C | 10 - 35°C |
| Maximum Humidity<br>(RH = Relative Humidity) | 95% RH noncondensing | 95% RH noncondensing |
| Disposable Pack - Provided<br>Sterile | Yes | Yes |
| Disposable Pack - Method of<br>sterilization | EtO and Gamma | EtO and Gamma |
# Table 1 Substantial equivalence comparison
{6}------------------------------------------------
# E. Performance Data
Performance of the CENTURION® Vision System (Active Sentry®) has been demonstrated through various performance validations, biocompatibility testing and electromagnetic compatibility and electrical safety testing.
Table 2: Summary of biocompatibility studies for the Active Sentry® Handpiece
| Biocompatibility Testing for<br>Active Sentry Handpiece | Cell Line/<br>Species | Test Media Extract | Test<br>Result |
|--------------------------------------------------------------|-----------------------|------------------------------------------------------|----------------|
| Cytotoxicity<br>ISO Elution Method 1X MEM | L-929 Mouse | MEM Extract | PASS |
| Cytotoxicity<br>ISO MTT Method | L-929 Mouse | MEM Extract | PASS |
| Sensitization<br>ISO Maximization Sensitization<br>Study | Guinea Pig | Saline & Sesame Oil Extracts | PASS |
| Acute Systemic Toxicity<br>ISO Acute Systemic Toxicity Study | Mouse | Saline & Sesame Oil Extracts | PASS |
| Ocular Irritation<br>ISO (Intraocular Injection) Method | Rabbit | Balanced Salt Solution (BSS® Irrigating<br>Solution) | PASS |
| Material Characterization<br>(Leachables)<br>ISO Method | Not Applicable | Isopropanol and Purified Water | PASS |
{7}------------------------------------------------
# Table 3: CENTURION® Vision System (Active Sentry®) electromagnetic compatibility and electrical safety
| Document<br>Number | Revision | Document Name | Result |
|--------------------|----------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| EN 60601-1 | 2013 | Medical Electrical Equipment, Part 1 – General requirements for basic safety<br>and essential performance. (Including A1:2013)<br>(Equivalent to IEC 60601-1: 2005+ A1: 2012) | PASS |
| EN 60601-1-<br>2 | 2007 | Medical electrical equipment Part 1: General requirements for basic safety<br>and essential performance - 2. Collateral Standard: Electromagnetic<br>compatibility-Requirements and test.<br>(Equivalent to IEC 60601-1-2: 2007) | PASS |
| EN 60601-2-<br>2 | 2009 | Medical electrical equipment – Part 2: General requirements for basic safety<br>and essential performance - Particular requirements for the Safety of high<br>frequency surgical equipment.<br>(Equivalent to IEC 60601-2-2: 2009) | PASS |
| EN 60601-1-<br>8 | 2007 | Medical Electrical Equipment, Part 1: General requirements for basic safety<br>and essential performance - 6. Collateral standard for alarm systems<br>(Equivalent to IEC 60601-1-8: 2006) | PASS |
| EN 80601-2-<br>58 | 2009 | Medical Electrical Equipment, Part 2-58: Particular requirements for the basic<br>safety and essential performance of lens removal devices and vitrectomy<br>devices for ophthalmic surgery (Equivalent to IEC 80601-2-58:2008) | PASS |
| IEC 60601-1-<br>6 | 2010 | Medical Electrical Equipment Part 1-6: General requirements for safety -<br>collateral standard: usability (Including A1:2013) | PASS |
| IEC 62366 | 2007 | Medical Devices - Application of usability engineering to medical devices<br>(Including A1:2014) | PASS |
#### F. Conclusions
The modified CENTURION® Vision System (Active Sentry®) is substantially equivalent to the originally submitted CENTURION® Vision System in that both devices:
- . have the same indicated use.
- . use the same operating principle,
- . incorporate the same basic phaco handpiece, same consumables and accessories cleared in the original submission as a system
- . use the same technology including components used.
- o are of the same design, incorporate the same materials, have the same shelf life, and are packaged and sterilized using the same materials and processes;
In summary, the changes made to the CENTURION® Vision System (Active Sentry®) described in this submission are substantially equivalent to the predicate device: CENTURION® Vision System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.