K233533 · Biocircuit Technologies, Inc. · JXI · Feb 12, 2024 · Neurology
Device Facts
Record ID
K233533
Device Name
NerveTape
Applicant
Biocircuit Technologies, Inc.
Product Code
JXI · Neurology
Decision Date
Feb 12, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5275
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
NerveTape is indicated for the repair of peripheral nerve discontinuities where gap closure can be achieved by flexion of the extremity.
Device Story
NerveTape is a bioabsorbable, extracellular collagen matrix (porcine small intestinal submucosa) featuring integrated nickel-titanium (NiTiNOL) microhooks. It is used by surgeons in clinical settings to repair peripheral nerve injuries. The device is wrapped around severed nerve stumps; the microhooks provide mechanical fixation and apposition, creating a tubular scaffold (entubulation) that protects the nerve and supports axonal growth as the patient's cells infiltrate and remodel the matrix. It is supplied sterile for single-use. The device benefits patients by facilitating nerve healing through stable coaptation without the need for traditional sutures.
Clinical Evidence
No human clinical studies were conducted. Evidence consists of bench testing (performance verification, tensile strength, specification compliance), biocompatibility testing (ISO 10993 series), and an animal study (rabbit tibial nerve model at 4 and 12 weeks). The animal study confirmed the device met all acceptance criteria and was substantially equivalent or superior to the predicate.
Technological Characteristics
Materials: Porcine small intestinal submucosa (SIS) (ISO 22442) and NiTiNOL microhooks. Form factor: Rectangular wrap (1.2–4.5cm width, 1.1–2.2cm length) forming a tube for 1.5–7mm diameter nerves. Thickness: 100-650 μm. Sterilization: Ethylene Oxide (ISO 11135). Connectivity: None (standalone).
Indications for Use
Indicated for repair of peripheral nerve discontinuities in patients where gap closure is achievable via extremity flexion.
Regulatory Classification
Identification
A nerve cuff is a tubular silicone rubber sheath used to encase a nerve for aid in repairing the nerve (e.g., to prevent ingrowth of scar tissue) and for capping the end of the nerve to prevent the formation of neuroma (tumors).
{0}------------------------------------------------
February 12, 2024
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white. To the right of the square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
BioCircuit Technologies, Inc. Jack Griffis Scientific Advisor 1819 Peachtree Road NE. Suite 205 Atlanta, Georgia 30309
## Re: K233533
Trade/Device Name: NerveTape Regulation Number: 21 CFR 882.5275 Regulation Name: Nerve Cuff Regulatory Class: Class II Product Code: JXI Dated: November 2, 2023 Received: November 2, 2023
### Dear Jack Griffis:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Image /page/1/Picture/6 description: The image shows a digital signature. The signature is from Adam D. Pierce. The date of the signature is 2024.02.12, and the time is 17:00:53 -05'00'.
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical. Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices
{2}------------------------------------------------
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233533
Device Name NerveTape
Indications for Use (Describe)
NerveTape is indicated for the repair of peripheral nerve discontinuities where can be achieved by flexion of the extremity.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
510(k) Number: K233533
February 12, 2024 Date:
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
- A. Submitter: BioCircuit Technologies, Inc. 1819 Peachtree Rd NE, Suite 205 Atlanta, Georgia 30309 (800) 905-2204
- B. Company Contact: Jack Griffis Scientific Advisor (404) 583-6889 (direct) jgriffis@biocircuit.com
- C. Device Information: Trade Name: NerveTape Common Name: Nerve Cuff
- D. Classification: Nerve Cuff 21 CFR §882.5275 (JXI) Class II
- E. Predicate Device: PRIMARY: BioCircuit Technologies, Inc. Nerve Tape™ (K210665) REFERENCE: AxoGuard® Nerve Connector (K162741)
- F. Physical Description:
The proposed BioCircuit Technologies, Inc. NerveTape device is composed of a bioabsorbable, extracellular collagen matrix (small intestinal submucosa, SIS) with integrated microhooks made of a nickel-titanium alloy, commonly referred to as NiTiNOL (identical to the predicate), for mechanical fixation and apposition of nerve ends. The NerveTape is implanted around an injured nerve to provide a scaffold which becomes infiltrated and remodeled by the patient's cells. The device protects the damaged or severed nerve while the nerve heals.
The device is packaged and supplied sterile in a clamshell container inside a sealed pouch. The device is identical to the predicate, except for the SIS substrate manufacturer and the addition of a smaller size. The dimensions of the finished device range from 11mm x 12mm x 22mm. The device is intended for the repair of nerves of diameters ranging from 1.5mm to 7mm.
- G. Indications for Use:
NerveTape is indicated for the repair of peripheral nerve discontinuities where gap closure can be achieved by flexion of the extremity.
### Comparison to Predicate Device(s):
The BioCircuit Technologies NerveTape device is substantially equivalent with respect to intended use, structure, and technological characteristics to the BioCircuit Technologies device
{5}------------------------------------------------
cleared under K210665. The devices both have a porcine small intestinal submucosa (SIS) backing (certified to ISO 22442) and operate under the same principle of axonal growth support with an identical form factor (tubular wrap with embedded microhook columns).
The devices differ in the source of the SIS backing substrate and the addition of a smaller sized option (1.5mm nerve compatible), which is equivalent to the original predicate (K162741). The proposed NerveTape device is manufactured with SIS supplied by Smithfield Bioscience, Inc., while the predicate and reference devices are composed of SIS supplied by Cook Biotech, Inc. BioCircuit Technologies, Inc., asserts that any differences from the predicate device do not affect safety or efficacy.
| Parameter | NerveTape<br>(BioCircuit Proposed) | Predicate<br>(BioCircuit) | Reference Predicate<br>(AxoGuard® Nerve Connector) |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | BioCircuit Technologies, Inc. | BioCircuit Technologies, Inc. | Cook Biotech Inc. |
| 510(k) Number | K233533 | K162741 | K162741 |
| Product Code | JXI | JXI | JXI |
| Material (wrap) | Porcine small intestinal submucosa:<br>primarily collagen types I, III, IV, and VI<br>(manufactured by Smithfield Bioscience) | Porcine small intestinal submucosa:<br>primarily collagen types I, III, IV, and VI<br>(manufactured by Cook Biotech) | Porcine small intestinal submucosa: primarily<br>collagen types I, III, IV, and VI (manufactured<br>by Cook Biotech) |
| Material (hook) | NITINOL (NO CHANGE) | | N/A - device is secured to nerves using non-<br>resorbable nylon sutures |
| Shape | Rectangular wrap (rolls into a hollow tube) (NO CHANGE) | | Hollow tube with a slit |
| Supplied Sterile? | Yes (NO CHANGE) | | Yes |
| Sterilization<br>method | Ethylene Oxide (cycle by Parter Sterilization<br>Services via Bioseal Inc. validated cycle) | Ethylene Oxide (validated cycle by Cook<br>Biotech Inc.) | Ethylene Oxide (validated cycle by Cook<br>Biotech Inc.) |
| Intended for<br>single use? | Yes (NO CHANGE) | | Yes |
| Packaging<br>Configuration | Clamshell tray in Tyvek-poly pouch with an<br>outer box (packaged by Bioseal Inc.) | Clamshell tray in Tyvek-poly pouch with an<br>outer box (packaged by Cook Biotech) | Clamshell tray in Tyvek pouch with an outer<br>box (packaged by Cook Biotech) |
| Shelf Life | 18 months (NO CHANGE) | | 18 months |
| Intended use | Intended for peripheral nerve injuries where a gap closure is achieved by flexion of the<br>extremity (NO CHANGE) | | Intended for peripheral nerve injuries where<br>there is no gap or where a gap closure is<br>achieved by flexion of the extremity. |
| Dimensions<br>(Wrapped) | 1.5 – 7mm diameter nerve<br>1.2 – 4.5cm width x 1.1 – 2.2cm length<br>(ADDITION OF SMALLER SIZE) | 2 – 7mm diameter nerve<br>1.4 x 4.5cm width x 1.4 - 2.2cm length | 1.5 – 7mm diameter nerve<br>1.5 - 7mm diameter tube x 1 - 1.5 cm length |
| Thickness<br>(Wrapped) | 100-650 μm¹ | 100-750 μm | 100-1000 μm |
| Microhook<br>Height | 360 – 550 μm<br>(ADDITION OF SMALLER SIZE USING<br>SHORTER MICROHOOKS) | 400 - 550 μm | N/A - device is secured to nerves using non-<br>resorbable nylon suture |
| Microhook<br>Column Qty | 3 - 8 Columns<br>(ADDITION OF SMALLER SIZE USING 3<br>COLUMNS) | 4-8 Columns | N/A - device is secured to nerves using non-<br>resorbable nylon suture |
### Table 1. Table of Substantial Equivalence
<sup>4</sup> NOTE: change in thickness is due to the L and L+ sizes of the proposed NerveTape device with only two (2) SIS layers (which is identical to all other sizes). The original (predicate) Nerve Tape device deared under K210665 utilized three (3) SIS layers for the L and L+ sizes and only two (2) SIS layers for all other sizes.
{6}------------------------------------------------
| Parameter | NerveTape<br>(BioCircuit Proposed) | Predicate<br>(BioCircuit) | Reference Predicate<br>(AxoGuard® Nerve Connector) |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Mechanism of<br>Action | Maintains coaptation via integrated microhooks within the SIS substrate layers, and<br>wrapped around the target site of nerve repair to complete entubulation of the positioned,<br>severed nerve stumps (NO CHANGE) | | Maintains coaptation via non-resorbable<br>nylon sutures used to reconnect the ends of<br>the nerves while the SIS is positioned around<br>the repair site to complete entubulation of the<br>sutured nerve stumps. |
#### H. Summary of Non-Clinical Tests:
Product characterization using known standards and / or clinically relevant acceptance criteria was performed on the proposed device. A summary of this testing is provided in Table 2.
| Test | Test Method Summary | Results |
|---------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance<br>Verification: Ease of Use<br>(Simulated Use in<br>Cadaver) | Surgeon handling of representative final product; acceptance criteria:<br>product possesses acceptable characteristics for handling, trimming and<br>implantation | The handling characteristics of the subject device<br>are substantially equivalent or superior to the<br>predicate device based on acceptability to the end<br>user. All samples met acceptance criteria. |
| Performance<br>Verification: Monotonic<br>Tensile Strength | Repair strength as assessed via device retention strength on repaired<br>cadaveric nerve in comparison to standard suture repair according to<br>the literature | The ultimate tensile strength of the proposed device<br>nerve repairs is substantially equivalent or superior<br>to the predicate device. All samples met acceptance<br>criteria. |
| Specification Compliance | Compliance with go/no-go dimensional and visual inspection criteria for<br>all components and assemblies | All samples met acceptance criteria. |
### Table 2. Non-Clinical Testing Information
#### l. Biocompatibility Testing:
Biocompatibility of the predicate device has been established in accordance with ISO 10993-1:2018 – Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process to demonstrate that the device is safe for permanent contact (>30 days) implantation as detailed in Table 3.
### Table 3a. Device Biocompatibility Information (all testing performed on sterile product of Iargest size)
| Test | Test Method Summary |
|--------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ANSI/AAMI/ISO 10993-<br>5:2009/(R)2014 – Biological<br>evaluation of medical devices –<br>Part 5: Tests for In Vitro<br>Cytotoxicity | The purpose of the in vitro study was to assess the cytotoxicity potential of extract from the proposed<br>device using the Elution test and the mouse cell line L929. The cytotoxic potential of the test article extract<br>was assessed by the change in morphology of the cell line, which was evaluated microscopically.<br>Based on the criteria of the protocol, the proposed device met the requirements of the test. |
| ANSI/AAMI/ISO 10993-23:2021<br>– Biological evaluation of<br>medical devices – Part 23:<br>Tests for Irritation | Intracutaneous reactivity test was conducted with rabbits to determine the potential for the proposed<br>device to produce irritation from intradermal injections.<br>Under the conditions of this study, the proposed device met the requirements of the test. The positive<br>response observed in the historical positive control validation study with Hexyl Cinnamic Aldehyde (HCA)<br>validates the test system used in this study. |
| ANSI/AAMI/ISO 10993-10:2021<br>– Biological evaluation of<br>medical devices – Part 10:<br>Tests for Skin Sensitization | A guinea pig maximization sensitization test was conducted with guinea pigs to determine the potential<br>for the proposed device to invoke a dermal skin sensitization reaction.<br>The study was conducted using a two-stage induction phase and a challenge phase. An emulsion of 50%<br>v/v Freund's Complete Adjuvant (FCA) in saline and sesame oil was used during the intradermal injection<br>induction phase.<br>Based on the results of this study, the proposed device met the requirements of the test. |
| ANSI/AAMI/ISO 10993-11:2017<br>– Biological evaluation of<br>medical devices – Part 11:<br>Tests for systemic toxicity | An acute systemic toxicity test in mice was conducted to determine the potential for the proposed device<br>to produce acute systemic toxicity from a single dose administered by intravenous (IV) and intraperitoneal<br>(IP) injection.<br>Under the conditions of this study, the proposed device met the requirements of the test. |
{7}------------------------------------------------
| Test | Test Method Summary |
|--------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ANSI/AAMI/ISO 10993-<br>7:2008(R)2012 — Biological<br>Evaluation of Medical Devices -<br>Part 7: Ethylene Oxide<br>Sterilization Residuals | Quantification of sterilant gas residue was performed for the proposed device having been exposed to<br>100% Ethylene oxide (EO). Samples received post sterilization underwent immersion and exhaustive<br>extractions using purified water and evaluated using gas chromatography. Extractions taken every 24<br>hours post sterilization were pooled and reported for total mg EO and ECH (Ethylene chlorohydrin).<br>Reported values were compared against the ISO standard for acceptable limits at or below the<br>recommended average daily dose for residuals in permanent implants.<br>Under the conditions of this study, the proposed device met the requirements of the test. |
| ANSI/AAMI ST72:2019, USP<br><161>, USP <85>, EP 2.6.14,<br>and JP 4.01 | A Bacterial Endotoxins Test (BET), or Limulus Amebocyte Lysate (LAL) test, was performed to detect and<br>quantify bacterial endotoxin, a component of the cell wall of Gram-negative bacteria. Standard controls<br>and a positive product control (PPC) demonstrate a compliant assay. Acceptable detected endotoxins must<br>not exceed the maximum allowable limit for permanent implants as per the FDA Guidance Document<br>Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices<br>Labeled as Sterile.<br>Under the conditions of this study, the proposed device met the requirements of the test. |
# Table 3b. Decellularized Porcine Small Intestine Submucosa Biocompatibility Information (all testing performed on non-sterile, decellularized porcine small intestine raw material - SIS)
| Test | Test Method Summary |
|--------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ANSI/AAMI/ISO 10993-<br>5:2009/(R)2014 – Biological<br>evaluation of medical devices -<br>Part 5: Tests for In Vitro<br>Cytotoxicity | The purpose of the in vitro study was to assess the cytotoxicity potential of extract from the SIS material<br>using the Elution test and the mouse cell line L929. The cytotoxic potential of the test article extract was<br>assessed by the change in morphology of the cell line, which was evaluated microscopically.<br>Based on the criteria of the protocol, the SIS material met the requirements of the test. |
| | The purpose of the in vitro study was to assess the cytotoxicity potential of extract from the SIS material<br>using the XTT dye method and the mouse cell line L929. The cell viability and cytotoxic potential of the<br>test article extracts were determined by quantification of a formazan dye formed in active mitochondria<br>of living cells. |
| | Based on the criteria of the protocol, the SIS material met the requirements of the test. |
| ANSI/AAMI/ISO 10993-23:2021 -<br>Biological evaluation of medical<br>devices – Part 23: Tests for<br>Irritation | Intracutaneous reactivity test was conducted with rabbits to determine the potential for the SIS material<br>to produce irritation from intradermal injections. |
| | Under the conditions of this study, the SIS material met the requirements of the test and is a non-irritant<br>for both polar and nonpolar extracts. |
| ANSI/AAMI/ISO 10993-10:2021<br>– Biological evaluation of<br>medical devices – Part 10: Tests<br>for Skin Sensitization | A guinea pig maximization sensitization test was conducted with guinea pigs to determine the potential<br>for the SIS material to invoke a dermal skin sensitization reaction. |
| | Based on the results of this study, the SIS material met the requirements of the test and is not considered<br>to be a contact skin sensitizer. |
| ANSI/AAMI/ISO 10993-11:2017<br>– Biological evaluation of<br>medical devices – Part 11: Tests<br>for systemic toxicity | An acute systemic toxicity test in mice was conducted to determine the potential for the SIS material to<br>produce acute systemic toxicity from a single dose administered by intravenous (IV) and intraperitoneal<br>(IP) injection. |
| | Under the conditions of this study, the SIS material met the requirements of the test. |
| ANSI/AAMI/ISO 10993-11:2017<br>– Biological evaluation of<br>medical devices – Part 11: Tests<br>for systemic toxicity | A subacute / subchronic toxicity test in mice was conducted to determine the potential for the SIS<br>material to produce systemic toxicity in male and female rats that is likely to arise from repeated exposure<br>via a dual route approach, intraperitoneal (IP) injection and intravenous (IV) administration, over a period<br>of at least 14 days. |
| | Under the conditions of this study, and based on the toxicological endpoints evaluated, there is no<br>potential toxicity of the SIS material from repeated exposure via 10 mL/kg/day via intravenously once per<br>day or 5 mL/kg/day via intraperitoneally every three days for both male and female Sprague Dawley rats. |
| ANSI/AAMI/ISO 10993-11:2017<br>– Biological evaluation of<br>medical devices – Part 11: Tests<br>for systemic toxicity | A material-mediated pyrogenic response test in rabbits was conducted to determine the potential for the<br>SIS material to produce a pyrogenic response due to intravenous exposure. |
| | Under the conditions of this study, there were no signs of gross toxicity, adverse clinical effects, or<br>abnormal behavior. None of the animals showed increases in temperature of 0.5℃ or more than their<br>respective control temperatures following administration of the SIS material extract. Therefore, and |
{8}------------------------------------------------
| Test | Test Method Summary |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | based on interpretation of the raw data according to USP-NF<151>, the SIS material met the<br>requirements of the test. |
| ANSI/AAMI/ISO 10993-3:2014 –<br>Biological evaluation of medical<br>devices – Part 3: Tests for<br>Genotoxicity, Carcinogenicity,<br>And Reproductive Toxicity | In Vitro micronucleus testing was conducted to evaluate the potential for extract of the SIS material to<br>induce micronuclei (clastogenic response) or hypodiploidy (aneugenic response) in cultured Chinese<br>Hamster Ovary (CHO) cells in the absence and presence of metabolic activation (S9).<br>Under the conditions of this study, the SIS material extract did not induce micronuclei (via chromosome<br>breaks and/or loss of whole chromosome(s)) or hypodiploidy (chromosome loss, sub-2n nuclei) in<br>cultured CHO cells when tested up to the limit of cytotoxicity (4 hours in the presence of metabolic<br>activation) and the maximum recommended extract concentration of 0.2 g/mL (4 and 23 hours in the<br>absence of metabolic activation), and met the requirements of the test. |
| | Bacterial mutagenicity testing was conducted to evaluate the potential for extract of the SIS material to<br>induce gene mutations in bacteria using the Ames assay. Point mutations which involve substitution,<br>addition, or deletion of one or a few DNA base pairs are detected in amino acid-requiring strains of<br>Salmonella typhimurium (S. typhimurium, ST) and Escherichia coli (E. coli, EC) by their ability to<br>functionally reverse mutations. These reverse mutations result in revertant colonies of bacteria with<br>restored capability to synthesize the essential amino acid.<br>Under the conditions of this study, the SIS material extract did not elicit evidence of bacterial<br>mutagenicity in the Ames assay and met the requirements of the test. |
| ANSI/AAMI/ISO 10993-6:2016 –<br>Biological evaluation of medical<br>devices – Part 6: Tests for Local<br>Effects After Implantation | Acute intramuscular implantation (4 weeks) of the SIS material was conducted in rabbits and evaluated<br>histologically.<br>Under the conditions of this study, the SIS material was well tolerated in comparison to the control article<br>and met the requirements of the test. |
| | Chronic intramuscular implantation (16 weeks) of the SIS material was conducted in rabbits and<br>evaluated histologically.<br>Under the conditions of this study, the SIS material was well tolerated in comparison to the control article<br>and met the requirements of the test. |
| Histological Comparison<br>between Native Intestinal Tissue<br>to Decellularized Tissue | Histological evaluation was conducted for the proposed SIS material in comparison to native (non-<br>decellularized) tissue via Hematoxylin & Eosin (H&E) to semi-quantitatively assess the absence of nuclei<br>and cytoplasmic material in the sample, Masson's Trichrome to assess the integrity of the collagen<br>network, and immunohistochemistry specific to galactose-α-1,3-galactose (α-gal IHC) to assess the level<br>of the α-gal antigen present in the ECM.<br>Under the conditions of the tests, the decellularization process of the SIS material was found to have<br>removed any microscopic evidence of intact cells that could be detected by light microscopy and left no<br>intact nuclei present in any decellularized section stained by H&E and Masson's Trichrome. In addition, |
#### J. Sterilization:
The method employed to ensure sterility of the proposed device is provided in Table 4. The sterilization process is identical for the subject and predicate device.
## Table 4. Sterilization Information
| Test | Test Method Summary | Results |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Sterilization<br>validation | Validation method in conformance with ISO 11135:2014, Sterilization of healthcare products with ethylene oxide, and<br>AAMI TIR28:2016, Product adoption and process equivalence for ethylene oxide sterilization. | Pass |
#### K. Animal Studies:
In the animal study conducted, a statistically valid number of rabbits underwent implantation of the proposed NerveTape device or the predicate device on an intact tibial nerve for 4 weeks and
{9}------------------------------------------------
12 weeks, respectively. The proposed device met all acceptance criteria and was substantially equivalent or superior to the predicate device.
#### L. Clinical Studies:
No human studies were necessary to prove the safety and efficacy of the device.
#### M. Conclusion:
No new questions of safety or effectiveness were identified during device testing; therefore, the Nerve Tape device is considered substantially equivalent to the predicate device in terms of safety and effectiveness.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.