Nerve Tape is indicated for the repair of peripheral nerve discontinuities where can be achieved by flexion of the extremity.
Device Story
Nerve Tape is a bioabsorbable nerve cuff used for peripheral nerve repair. It consists of a porcine small intestinal submucosa (SIS) collagen matrix integrated with nickel-titanium (NiTiNOL) microhooks. The device is supplied sterile in a dried state and rehydrated before use. During surgery, the rectangular wrap is applied around the nerve ends; the microhooks provide mechanical fixation and apposition, replacing the need for traditional sutures. The SIS scaffold supports nerve remodeling and protection during healing. It is intended for use by surgeons in clinical settings for nerve injuries where gap closure is possible through extremity flexion. The device benefits patients by providing a sutureless fixation method for nerve coaptation, potentially simplifying the surgical procedure while maintaining repair strength equivalent to standard suturing.
Clinical Evidence
No human clinical studies were performed. Safety and efficacy were established via non-clinical bench testing and an animal study. Bench testing included monotonic tensile strength (compared to suture), dimension compliance, corrosion resistance (ASTM F2129), and transformation temperature (ASTM F2004). Biocompatibility was assessed per ISO 10993-1. An animal study using a rabbit tibial nerve transection model compared Nerve Tape to AxoGuard® Nerve Connector and nylon suture at 4 and 16 weeks, evaluating macroscopic tissue response, muscle weight/girth, and histology. Results demonstrated the device is as safe and effective as the predicates.
Technological Characteristics
Bioabsorbable porcine small intestinal submucosa (SIS) matrix (collagen types I, III, IV, VI); NiTiNOL microhooks for fixation. Dimensions: 2-7mm diameter, 1.4-2.2cm length, 100-750 μm thickness. Sterilization: Ethylene Oxide (ISO 11135:2014). NiTiNOL components tested per ASTM F2129 (corrosion) and ASTM F2004 (transformation temperature).
Indications for Use
Indicated for repair of peripheral nerve discontinuities in patients where gap closure is achievable via extremity flexion.
Regulatory Classification
Identification
A nerve cuff is a tubular silicone rubber sheath used to encase a nerve for aid in repairing the nerve (e.g., to prevent ingrowth of scar tissue) and for capping the end of the nerve to prevent the formation of neuroma (tumors).
Predicate Devices
Cook Biotech Inc. AxoGuard® Nerve Connector (K162741)
Cook Biotech Inc. Nerve Cuff (AxoGuard® Nerve Protector) (K132660)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 15, 2022
BioCircuit Technologies, Inc. % Jack Griffis Scientific and Regulatory Advisor 1819 Peachtree Road NE, Suite 205 Atlanta, Georgia 30309
Re: K210665
Trade/Device Name: Nerve Tape Regulation Number: 21 CFR 882.5275 Regulation Name: Nerve Cuff Regulatory Class: Class II Product Code: JXI Dated: April 6, 2022 Received: April 7, 2022
Dear Jack Griffis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210665
Device Name Nerve Tape
Indications for Use (Describe)
Nerve Tape is indicated for the repair of peripheral nerve discontinuities where can be achieved by flexion of the extremity.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
### 510(k) Number: K210665
July 15th, 2022 Date Submitted:
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
- A. Submitter: BioCircuit Technologies, Inc. 1819 Peachtree Rd NE, Suite 205 Atlanta, Georgia 30309 470-698-2117
- B. Company Contact: Jack Griffis Scientific Advisor (404) 583-6889 (direct) jackgriffisiii@gmail.com
- C. Device Information: Trade Name: Nerve Tape Common Name: Nerve Cuff
- D. Classification: Nerve Cuff 21 CFR §882.5275 (Product Code: JXI) Class II
- E. Primary Predicate Device: Cook Biotech Inc. AxoGuard® Nerve Connector (K162741)
### F. Additional Predicate Device: Cook Biotech Inc. Nerve Cuff (Marketed as AxoGuard® Nerve Protector) (K132660)
### G. Physical Description:
The Nerve Tape device is composed of a bioabsorbable, extracellular collagen matrix derived from small intestinal submucosa (SIS). Microhooks made of a nickel-titanium alloy (NiTiNOL) are integrated into the nerve contacting side of the SIS for mechanical fixation and apposition of nerve ends. The device is packaged in a dried state and supplied sterile, and is rehydrated prior to use.
- H. Indications for Use: Nerve Tape™ is indicated for the repair of peripheral nerve discontinuities where gap closure can be achieved by flexion of the extremity.
- l. Comparison to Predicate Device:
The subject Nerve Tape device is substantially equivalent in intended use to the predicate Cook Biotech Inc. AxoGuard® Nerve Connector and Cook Biotech Inc. Nerve Cuff devices in surgical repair of peripheral nerves. The subject and predicate devices are surgically fabricated with similar or identical bioresorbable porcine SIS material (certified to ISO 22442) which serves as scaffold that protects protection and supports remodeling while the injured nerve heals. The Nerve Tape differs from the predicates in the method of nerve cuff fixation. The Nerve Tape contains embedded microhooks,
{4}------------------------------------------------
whereas the predicate devices require suturing to maintain coaptation strength. A comparison of subject and predicate devices is provided in Table 1.
| Parameter | Nerve Tape<br>(Proposed) | AxoGuard® Nerve Connector<br>(Predicate, K162741) | Nerve Cuff<br>(Predicate, K132660) |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | BioCircuit Technologies, Inc. | Cook Biotech Inc. | Cook Biotech Inc. |
| 510(k) Number | K210665 | K162741 | K132660 |
| Product Code | JXI | JXI | JXI |
| Classification<br>Number | 21 CFR 882.5275 | 21 CFR 882.5275 | 21 CFR 882.5275 |
| Common name | Cuff, Nerve | Cuff, Nerve | Cuff, Nerve |
| Intended use | Intended for peripheral nerve injuries<br>where a gap closure is achieved by<br>flexion of the extremity. | Intended for peripheral nerve<br>injuries where a gap closure is<br>achieved by flexion of the<br>extremity. | Intended for peripheral nerve<br>injuries where there is no gap or<br>where a gap closure is achieved<br>by flexion of the extremity. |
| Materials of<br>Fabrication | Porcine small intestinal submucosa;<br>primarily collagen types I, III, IV, & VI<br>(manufactured by Cook Biotech Inc.),<br>NITINOL | Porcine small intestinal<br>submucosa; primarily collagen<br>types I, III, IV, & VI<br>(manufactured by Cook Biotech<br>Inc.) | Porcine small intestinal<br>submucosa; primarily collagen<br>types I, III, IV, & VI<br>(manufactured by Cook Biotech<br>Inc.) |
| Shape | Rectangular wrap (rolls into a hollow<br>tube) | Hollow tube | Hollow tube with a slit |
| Supplied Sterile? | Yes | Yes | Yes |
| Sterilization<br>method | Ethylene Oxide (cycle by Cook Biotech<br>Inc.) | Ethylene Oxide (cycle by Cook<br>Biotech Inc.) | Ethylene Oxide (cycle by Cook<br>Biotech Inc.) |
| Intended for<br>single use? | Yes | Yes | Yes |
| Packaging<br>Configuration | Clamshell tray in Tyvek-poly pouch<br>with an outer box (packaged by Cook<br>Biotech Inc.) | Clamshell tray in Tyvek pouch<br>with an outer box<br>(packaged by Cook Biotech Inc.) | Clamshell tray in Tyvek pouch<br>with an outer box<br>(packaged by Cook Biotech Inc.) |
| Shelf Life | 18 months | 18 months | 18 months |
| Intended use | Intended for peripheral nerve injuries<br>where a gap closure is achieved by<br>flexion of the extremity. | Intended for peripheral nerve<br>injuries where a gap closure is<br>achieved by flexion of the<br>extremity. | Intended for peripheral nerve<br>injuries where there is no gap or<br>where a gap closure is achieved<br>by flexion of the extremity. |
| Dimensions<br>(Wrapped) | 2-7mm diameter x 1.4-2.2cm length | 1.5-10mm diameter x 1-5 cm<br>length | 1.5-10mm diameter x 1-5 cm<br>length |
| Thickness<br>(Wrapped) | 100-750 μm | 100-1000 μm | 100-1000 μm |
| Method of<br>fixation | NITINOL microhooks | Suture | Suture |
## Table 1. Table of Substantial Equivalence
The technological characteristics of the subject Nerve Tape device are substantially equivalent to the predicate devices as demonstrated through bench, biocompatibility and animal testing.
{5}------------------------------------------------
### J. Summary of Non-Clinical Tests:
Product characterization using known standards and/or relevant acceptance criteria were performed on the Nerve Tape device. A summary of this testing is provided in Table 2.
| Test | Test Method Summary | Results |
|----------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| User handling validation/<br>Ease of Use | Surgeon performance and handling of Nerve Tape, nylon<br>suture and AxoGuard® Nerve Connector were assessed for<br>ease of use in application to the transected rabbit tibial<br>nerves | The ease of use characteristics of<br>the Nerve Tape device are<br>substantially equivalent to the<br>predicate device. All samples met<br>acceptance criteria. |
| Monotonic tensile<br>strength | Repair strength as assessed via device retention strength<br>on repaired cadaveric nerve in comparison to standard<br>suture repair according to the literature | The tensile strength of the Nerve<br>Tape device is substantially<br>equivalent to standard suture for<br>nerve repair. All samples met<br>acceptance criteria. |
| Dimension Compliance | Compliance with dimensional criteria for all components<br>and assemblies | Pass |
| Corrosion Resistance (for<br>NiTiNOL components) | ASTM F2129: Cyclic potentiodynamic polarization<br>measurements for corrosion susceptibility | Pass |
| Transformation<br>temperature (DSC, for<br>NiTiNOL components) | Per the US FDA Guidance Document, "Technical<br>Considerations for Non-Clinical Assessment of Medical<br>Devices Containing Nitinol", and ASTM F2004,<br>Transformation Temperature of Nickel-Titanium Alloys by<br>Thermal Analysis | Pass |
| Magnetic resonance<br>imaging safety and<br>compatibility | Per the US FDA Guidance Document, "Testing and Labeling<br>Medical Devices for Safety in the Magnetic Resonance (MR)<br>Environment" | Pass |
Table 2. Non-Clinical Bench Testing Information
### K. Biocompatibility Testing:
Biocompatibility of the Nerve Tape device has been established in accordance with ISO 10993-1:2018 – Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process to demonstrate that the device is safe for permanent contact (>30 days) implantation. The biocompatibility endpoints that were assessed include cytotoxicity, sensitization, acute systemic toxicity, subchronic/chronic systemic toxicity, pyrogenicity, implantation and genotoxicity.
### L. Sterilization:
Ethylene oxide sterilization validation and ethylene oxide residuals testing of the Nerve Tape device are provided in Table 4.
| Test | Test Method Summary | Results |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Sterilization validation | Validation method in conformance with ISO 11135:2014, Sterilization of<br>healthcare products with ethylene oxide, and AAMI TIR28:2016, Product adoption<br>and process equivalence for ethylene oxide sterilization. | Pass |
| EO Residuals | ISO 10993-7:2008, Ethylene oxide sterilization residuals | Pass |
| Table 4. Sterilization Information | |
|------------------------------------|--|
|------------------------------------|--|
{6}------------------------------------------------
#### M. Animal Studies:
Testing using a rabbit tibial nerve transection model were evaluated for usability and tissue responses were assessed following implantation and treatment with either the Nerve Tape device, the AxoGuard® Nerve Connector or nylon suture, and evaluated at the 4-week or 16-week evaluation time points. Outcome measures included macroscopic assessment of the nerve implant sites, muscle weight and girth, and histological assessment of the implant sites and surrounding tissues. The studies demonstrated that the Nerve Tape device is as safe and effective as the AxoGuard® Nerve Connector and nylon suture in repairing peripheral nerve discontinuities where gap closure can be achieved by flexion of the extremity.
### N. Clinical Studies:
No human studies were necessary to prove the safety and efficacy of the device.
### Conclusion:
The Nerve Tape device has the same intended use as the predicate devices. Nonclinical testing demonstrated that the Nerve Tape is substantially equivalent to the predicate device(s).
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.