SIGNAFUSE Bioactive Strip (SBS) is a bone graft substitute intended for use in bony voids or gaps of the skeletal system not intrinsic to the stability of the bony structure. These osseous defects may be surgically created or result from traumatic injury to the bone. SIGNAFUSE Bioactive Strip (SBS) is indicated to be combined with autologous bone marrow aspirate and packed into osseous defects of the extremities, pelvis, posterolateral spine, and intervertebral disc space. When used in the posterolateral spine, SIGNAFUSE Bioactive Strip (SBS) is to be used as an autograft extender. When used in intervertebral disc space. SIGNAFUSE Bioactive Strip (SBS) is to be used as an autograft extender with an intervertebral body fusion device cleared by FDA for use with a bone void filler. The device resorbs and is replaced by host bone during the healing process. SIGNAFUSE Putty is a bone void filler device intended for use in bony voids or gaps that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. SIGNAFUSE Putty is indicated to be packed gently into bony voids or gaps of the skeletal system (i.e., extremities, pelvis, posterolateral spine, and intervertebral disc space fusion procedures). SIGNAFUSE Putty can also be used with autograft as a bone graft extender in the posterolateral spine. When used in intervertebral body fusion procedures, SIGNAFUSE Putty can used on its own or as a bone graff extender, and with an intervertebral body fusion device cleared by FDA for use with a bone void filler. The device provides a bone void filler that is resorbed and replaced with host bone during the healing process.
Device Story
SIGNAFUSE Bioactive Strip (SBS) and SIGNAFUSE Putty are resorbable bone graft substitutes. SBS consists of biphasic mineral granules and 45S5 bioactive glass in a type I collagen matrix; Putty consists of the same mineral/glass components in an alkylene oxide polymer (AOP) carrier. Devices are implanted by surgeons into osseous defects or intervertebral disc spaces to fill voids and facilitate bone healing. They act as scaffolds that resorb and are replaced by host bone. The devices are single-use, sterile implants. This submission expands the existing indications for the SIGNAFUSE family to include use in the intervertebral body space.
Clinical Evidence
No clinical data; performance supported by clinical rationale for bone grafting materials in the intervertebral space and previously cleared data (K193513, K132071) for sterility, shelf-life, endotoxin, biocompatibility, and bench performance.
Technological Characteristics
Resorbable calcium salt bone void filler. SBS: Biphasic mineral granules, 45S5 bioactive glass, type I collagen matrix. Putty: Biphasic mineral granules, 45S5 bioactive glass, alkylene oxide polymer (AOP) carrier. Sterilization: Irradiation (SAL 10^-6). Single-use implant.
Indications for Use
Indicated for patients requiring bone void fillers in skeletal defects (extremities, pelvis, posterolateral spine, intervertebral disc space) not intrinsic to structural stability. Used as autograft extender in posterolateral spine and intervertebral fusion; may be used with autologous bone marrow aspirate (SBS) or intervertebral body fusion devices.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}------------------------------------------------
December 5, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
Bioventus LLC % Scott Bruder Founder & CEO Bruder Consulting & Venture Group 268 Glen Place Franklin Lakes, New Jersey 07417
Re: K233490
Trade/Device Name: SIGNAFUSE Bioactive Strip (SBS); SIGNAFUSE Putty Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable Calcium Salt Bone Void Filler Device Regulatory Class: Class II Product Code: MQV Dated: October 26, 2023 Received: October 27, 2023
Dear Scott Bruder:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Digitally signed by Je Jesse Muir -S Date: 2023.12.05 12:56:54
Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
#### Device Name
SIGNAFUSE Bioactive Strip (SBS); SIGNAFUSE Putty
Indications for Use (Describe)
- SIGNAFUSE Bioactive Strip (SBS)
SIGNAFUSE Bioactive Strip (SBS) is a bone graft substitute intended for use in bony voids or gaps of the skeletal system not intrinsic to the stability of the bony structure. These osseous defects may be surgically created or result from traumatic injury to the bone. SIGNAFUSE Bioactive Strip (SBS) is indicated to be combined with autologous bone marrow aspirate and packed into osseous defects of the extremities, pelvis, posterolateral spine, and intervertebral disc space. When used in the posterolateral spine, SIGNAFUSE Bioactive Strip (SBS) is to be used as an autograft extender. When used in intervertebral disc space. SIGNAFUSE Bioactive Strip (SBS) is to be used as an autograft extender with an intervertebral body fusion device cleared by FDA for use with a bone void filler. The device resorbs and is replaced by host bone during the healing process.
## SIGNAFUSE Putty
SIGNAFUSE Putty is a bone void filler device intended for use in bony voids or gaps that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. SIGNAFUSE Putty is indicated to be packed gently into bony voids or gaps of the skeletal system (i.e., extremities, pelvis, posterolateral spine, and intervertebral disc space fusion procedures). SIGNAFUSE Putty can also be used with autograft as a bone graft extender in the posterolateral spine. When used in intervertebral body fusion procedures, SIGNAFUSE Putty can used on its own or as a bone graff extender, and with an intervertebral body fusion device cleared by FDA for use with a bone void filler. The device provides a bone void filler that is resorbed and replaced with host bone during the healing process.
Type of Use (Select one or both, as applicable)
< | Prescription Use (Part 21 CFR 801 Subpart D)
ver-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Bioventus. The logo consists of a circular, interwoven design in shades of orange and purple, followed by the word "bioventus" in a purple sans-serif font. A registered trademark symbol is located to the upper right of the word.
# 510(k) Summary
#### Submitter
Kellie Stefaniak Vice President, Global Regulatory & Quality Bioventus LLC 4721 Emperor Blvd. Suite 100 Durham, NC 27703
## Correspondent
Scott Bruder, MD, PhD Founder and CEO Bruder Consulting & Venture Group Scott(@BruderConsulting.com 201.874.9701
## Date Prepared
December 1, 2023
## Device
| Trade Name | SIGNAFUSE Bioactive Strip (SBS)<br>SIGNAFUSE Putty |
|----------------|-----------------------------------------------------------------|
| Common Name | Bone void filler |
| Regulation | 21 CFR 888.3045 Resorbable calcium salt bone void filler device |
| Classification | Class II |
| Product Code | MQV |
| Panel | Orthopedic |
## Predicates
Primary Predicate Secondary Predicate Reference Devices
NuVasive - Attrax Putty (K203714) Prosidyan - Fibergraff™ BG Putty (K222276) Bioventus - SIGNAFUSE Bioactive Bone Graft (K193513) Bioventus -Bioactive Bone Graft Putty (K132071)
## Device Description
SIGNAFUSE Bioactive Strip (SBS) is a bioactive bone graft substitute comprising biphasic mineral granules and 45S5 bioactive glass suspended in a porous type I collagen matrix, and it identical to the device cleared in K193513. SIGNAFUSE Putty is a bioactive bone graft substitute comprising biphasic mineral granules and 45S5 bioactive glass suspended in an alkylene oxide polymer (AOP) resorbable carrier and is identical to the device cleared in K132071. The SIGNAFUSE Family of devices are single use implants in contact with bone that are sterilized by irradiation with a sterility assurance level (SAL) of 106. This submission expands the SIGNAFUSE Family indication to include use in the intervertebral space.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Bioventus logo, which consists of a stylized symbol of three interlocking rings in shades of purple and orange, followed by the word "bioventus" in a bold, sans-serif font, also in purple. A registered trademark symbol is present to the right of the word. Below the logo, the words "Indications for Use Statement" are written in a bold, serif font.
## SIGNAFUSE Bioactive Strip (SBS)
SIGNAFUSE Bioactive Strip (SBS) is a bone graft substitute intended for use in bony voids or gaps of the skeletal system not intrinsic to the stability of the bony structure. These osseous defects may be surgically created or result from traumatic injury to the bone. SIGNAFUSE Bioactive Strip (SBS) is indicated to be combined with autologous bone marrow aspirate and packed into osseous defects of the extremities, pelvis, posterolateral spine, and intervertebral disc space. When used in the posterolateral spine, SIGNAFUSE Bioactive Strip (SBS) is to be used as an autograft extender. When used in intervertebral disc space, SIGNAFUSE Bioactive Strip (SBS) is to be used as an autograft extender with an intervertebral body fusion device cleared by FDA for use with a bone void filler. The device resorbs and is replaced by host bone during the healing process.
## SIGNAFUSE Putty
SIGNAFUSE Putty is a bone void filler device intended for use in bony voids or gaps that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. SIGNAFUSE Putty is indicated to be packed gently into bony voids or gaps of the skeletal system (i.e., extremities, pelvis, posterolateral spine, and intervertebral disc space fusion procedures). SIGNAFUSE Putty can also be used with autograft as a bone graft extender in the posterolateral spine. When used in intervertebral body fusion procedures, SIGNAFUSE Putty can used on its own or as a bone graft extender, and with an intervertebral body fusion device cleared by FDA for use with a bone void filler. The device provides a bone void filler that is resorbed and replaced with host bone during the healing process.
## Substantial Equivalence
The intended use is the same for subject and predicate devices. The subject device and predicates are all indicated for use as bone void fillers in the intervertebral body space.
There are differences in the device's material composition, but these do not raise different questions of safety and effectiveness. The effect of differences in material composition is addressed in clinical rationale.
## Performance
The subject devices have been previously cleared under K193513 (SIGNAFUSE Bioactive Strip (SBS)) and K132071 (SIGNAFUSE Putty), which serve as Reference Devices. These submissions are leveraged to support the device's sterility, shelf-life, endotoxin, biocompatibility, and characterizations/bench performance as recommended in FDA's Class II Special Controls Guidance Document for Resorbable Calcium Salt Bone Void Filler Devices. The devices' performance in the intervertebral body space was supported by a clinical rationale of bone grafting materials in the intervertebral space.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Bioventus. The logo consists of a circular graphic on the left, made up of three intertwined strands in shades of purple and orange. To the right of the graphic is the word "bioventus" in purple, with a registered trademark symbol after the "s".
## Summary
The subject device and predicates have the same intended use, and the same specific indications for use in the intervertebral body space. Any differences in technological characteristics between the subject device and predicate do not raise different questions of safety and effectiveness. Based on the clinical rationale, the device is substantially equivalent to the predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.