K233488 · Vioptix, Inc. · MUD · Feb 28, 2024 · Cardiovascular
Device Facts
Record ID
K233488
Device Name
Intra.Ox™ 2.0 Handheld Tissue Oximeter
Applicant
Vioptix, Inc.
Product Code
MUD · Cardiovascular
Decision Date
Feb 28, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K233488 · Feb 28, 2024
Intra.Ox™ 2.0 Handheld Tissue Oximeter
Vioptix, Inc.
Published scientific literature (Gonzalez-Jacobo et al.)
The literature was used to demonstrate that the device provides clinically relevant information regarding bowel ischemia across multiple surgical cases, providing information comparable to the existing standard of care.
Literature review; Bowel ischemia; Clinical relevance
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Gonzalez-Jacobo et al.; Published scientific literature
Clinically relevant information regarding bowel ischemia
Indications for Use
The Intra.Ox™ 2.0 Handheld Tissue Oximeter is intended to non-invasively estimate the percent oxygen saturation (StO2) in a volume of tissue, including bowel tissue. The Intra.Ox™ 2.0 Handheld Tissue Oximeter is indicated for use in monitoring patients during circulatory or perfusion examinations. The Intra.Ox™ 2.0 Handheld Tissue Oximeter is intended to be used by physicians, surgeons, nurses, or other skilled users in a medical environment. The Intra.Ox™ 2.0 Handheld Tissue Oximeter should only be used on adult patients.
Device Story
Handheld, cordless, battery-powered tissue oximeter; uses spatially resolved optical measurements to estimate StO2. Components: reusable main unit (light sources, detectors, processing electronics), single-use sterile sheath, single-use battery pack. Operated by physicians, surgeons, or nurses in medical environments. Device makes direct physical contact with patient tissue; processes reflected light to calculate StO2; displays estimate on unit. Provides real-time perfusion data to assist clinical decision-making regarding tissue viability and ischemia; benefits patient by enabling non-invasive monitoring of oxygenation status.
Clinical Evidence
Evidence includes animal studies demonstrating response to transient ischemia via arterial occlusion and human clinical data from published literature (Gonzalez-Jacobo et al.). Literature confirms device provides clinically relevant information regarding bowel ischemia across surgical cases, comparable to standard of care. Previous clinical validation (fingertip study) from K191676 also cited.
Technological Characteristics
Sterile, cordless, battery-powered handheld device. Sensing principle: spatially resolved optical measurements (reflected light). Components: reusable main unit, single-use sterile sheath, single-use battery. Biocompatible materials used for patient-contacting sheath. No specific software architecture or connectivity details provided.
Indications for Use
Indicated for adult patients requiring monitoring of tissue oxygen saturation (StO2) during circulatory or perfusion examinations, including bowel tissue.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 28, 2024
ViOptix, Inc. % Valerie Defiesta-Ng Regulatory Consultant for ViOptix, Inc. / Vice President, Regulatory Affairs, Veranex, Inc. Veranex, Inc. 224 Airport Parkway Suite 250 San Jose, California 95110
Re: K233488
Trade/Device Name: Intra.OxTM 2.0 Handheld Tissue Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: MUD Dated: January 26, 2024 Received: January 29, 2024
Dear Valerie Defiesta-Ng:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Tanisha L. Tanisha L. Hithe -S 2024.02.28 Hithe -S 18:23:01 -05'00'
Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K233488
Device Name Intra.Ox™ 2.0 Handheld Tissue Oximeter
Indications for Use (Describe)
The Intra. Ox™ 2.0 Handheld Tissue Oximeter is intended to non-invasively estimate the percent oxygen saturation (StO2) in a volume of tissue, including bowel tissue.
The Intra. Ox™ 2.0 Handheld Tissue Oximeter is indicated for use in monitoring patients during circulatory or perfusion examinations.
The Intra. Ox™ 2.0 Handheld Tissue Oximeter is intended to be used by physicians, surgeons, nurses, or other skilled users in a medical environment.
The Intra.Ox™ 2.0 Handheld Tissue Oximeter should only be used on adult patients.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| 510(k) #: | K233488 |
|-----------|-------------------------|
| | 510(k) Summary |
| | Prepared on: 2024-02-28 |
Contact Details
21 CFR 807.92(a)(1)
| Applicant Name | ViOptix, Inc. |
|-----------------------------|--------------------------------------------------|
| Applicant Address | 39655 Eureka Drive Newark CA 94560 United States |
| Applicant Contact Telephone | (510) 226-5860 |
| Applicant Contact | Mr. Scott Coleridge |
| Applicant Contact Email | secoleridge@vioptix.com |
Device Name
21 CFR 807.92(a)(2)
| Device Trade Name | Intra.Ox™ 2.0 Handheld Tissue Oximeter |
|---------------------|----------------------------------------|
| Common Name | Oximeter |
| Classification Name | Oximeter, Tissue Saturation |
| Regulation Number | 870.2700 |
| Product Code | MUD |
Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
|-------------|----------------------------------------------------------|--------------|
| K221010 | Intra.Ox™ 2.0 Handheld Tissue Oximeter | MUD |
Device Description Summary
21 CFR 807.92(a)(4)The ViOptix Intra.Ox" 2.0 Handheld Tissue Oximeter (device) is a sterile, cordless, battery-powered device that non-invasively estimates the percent oxygen saturation (StO2) in a volume of tissue. The device includes three components and an accessory: · Main Unit: a re-usable module consists of light sources, detectors, and processing electronics to convert measurements of reflected light into an estimate of StO2;
• Sheath: a single-use, sterile Sheath placed around the Main Unit during device use (provide in the Disposable Kit); and · Battery Pack: a single-use battery (provided in the Disposable Kit) that is paired with the Main Unit to provide power.
The device uses spatially resolved optical measurements. The device performs measurements on the patient by direct physical contact to the patient's tissue and displays the StO2 estimate on the Intra.Ox 2.0 Handheld Tissue Oximeter is a single-use disposable constructed from biocompatible materials that and other liquids such as disinfectants and marking materials. The device shares the same technological characteristics as the predicate device (K221010), including principle of operation, StO2 measured parameters, accuracy and range, energy delivered and power source.
### Intended Use/Indications for Use
The Intra.Ox™ 2.0 Handheld Tissue Oximeter is intended to non-invasively estimate the percent oxygen saturation (StO2) in a volume of tissue, including bowel tissue.
The Intra.Ox™ 2.0 Handheld Tissue Oximeter is indicated for use in monitoring circulatory or perfusion examinations. The Intra.Ox™ 2.0 Handheld Tissue Oximeter is intended to be used by physicians, surgeons, nurses, or other skilled users in a medical environment.
The Intra.Ox™ 2.0 Handheld Tissue Oximeter should only be used on adult patients.
# Indications for Use Comparison
21 CFR 807.92(a)(5)
21 CFR 807.92(a)(5)
{5}------------------------------------------------
The proposed Intra.Ox™ 2.0 Handheld Tissue Oximeter and the cleared Intra.Ox™ 2.0 Handheld Tissue Oximeter (K221010) have the same intended use and share the same technological characteristics. ViOptix, Inc. is proposing modifications to the predicate Intra.Ox" 2.0 Handheld Tissue Oximeter (K221010) Indications for Use statement. The differences outlined in this Traditional 510k) premarket notification do not affect the intended use or alter the fundamental scientific technology of the device. The modification to the Indications for Use statement (including bowel tissue) is particular tissue which already falls under the general tissue indication of the modification to the Precautions statement language (stool-filled tissue) provides an added condition to avoid improper device placement.
# Technological Comparison
### 21 CFR 807.92(a)(6)
The proposed Intra.Ox™ 2.0 Handheld Tissue Oximeter and the cleared Intra.Ox™ 2.0 Handheld Tissue Oximeter (K221010) share the same technological characteristics.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
The animal study demonstrates that the Intra.Ox™ 2.0 Handheld Tissue Oximeter responds appropriately to the presence of transient ischemia induced by arterial occlusion and satisfies simila to the prior clinical validation fingertip study, which was previously presented in K191676. This was the initial 510(k) that was submitted for the Intra.Ox™ 2.0 Handheld Tissue Oximeter.
The human study, Gonzalez-Jacobo et al., demonstrates that the Intra.Ox™ 2.0 Handheld Tissue Oximeter provides clinically relevant information regarding bowel ischemia across multiple surgical providing comparable information to the existing standard of care.
An animal study was conducted and human clinical data from published scientific literature was included to demonstrate that the Intra.Ox™ 2.0 Handheld Tissue Oximeter demonstrates clinically relevant differentiation in the cases of oxygenated tissue. Therefore, the proposed and cleared devices are substantially equivalent.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.