K221010 · Vioptix, Inc. · MUD · May 3, 2022 · Cardiovascular
Device Facts
Record ID
K221010
Device Name
Intra.Ox 2.0 Handheld Tissue Oximeter
Applicant
Vioptix, Inc.
Product Code
MUD · Cardiovascular
Decision Date
May 3, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2700
Device Class
Class 2
Indications for Use
The Intra.Ox™ 2.0 Handheld Tissue Oximeter is intended to non-invasively estimate the percent oxygen saturation (StO2) in a volume of tissue. The Intra.Ox™ 2.0 Handheld Tissue Oximeter is indicated for use in monitoring patients during circulatory or perfusion examinations. The Intra.Ox™ 2.0 Handheld Tissue Oximeter is intended to be used by physicians, surgeons, nurses, or other skilled users in a medical environment. The Intra.Ox™ 2.0 Handheld Tissue Oximeter should only be used on adult patients.
Device Story
Handheld, battery-powered, sterile tissue oximeter; estimates StO2 non-invasively. Components: reusable main unit (light sources, detectors, processing electronics), single-use sterile sheath, single-use battery pack. Operation: device placed in direct contact with tissue; uses spatially resolved spectrophotometric oximetry at five wavelengths (730, 760, 810, 845, 895 nm). Output: StO2 percentage displayed on built-in screen; provides both 'live' and 'average' readings. Used by physicians, surgeons, or nurses in medical environments. Assists clinicians in assessing tissue perfusion; aids in clinical decision-making regarding circulatory status.
Clinical Evidence
No clinical data. Bench testing only. Heterogeneous blood phantom study using swine blood and Intralipid solution compared device performance against a gold-standard blood co-oximeter. Results demonstrated substantial equivalence in limits of agreement, slope, and intercept compared to the predicate device.
Technological Characteristics
Handheld, battery-powered (4 Lithium AA, 6V). Spectrophotometric oximetry using 5 LED wavelengths (730, 760, 810, 845, 895 nm). Biocompatible materials. Dimensions: 7.3" x 2.7" x 3.1"; weight < 1 lb. Includes single-use sterile sheath and battery pack. Software includes averaging functionality for StO2 readings.
Indications for Use
Indicated for adult patients requiring monitoring of tissue oxygen saturation (StO2) during circulatory or perfusion examinations.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 3, 2022
ViOptix, Inc. % Valerie Defiesta-Ng Vice president, Reuglatory Affairs Veranex, Inc. (formerly Experien Group, LLC) 224 Airport Parkway, Suite 250 San Jose, California 95110
Re: K221010
Trade/Device Name: Intra.Ox 2.0 Handheld Tissue Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: MUD Dated: April 4, 2022 Received: April 5, 2022
Dear Valerie Defiesta-Ng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya, D.Eng. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221010
Device Name Intra.Ox™ 2.0 Handheld Tissue Oximeter
Indications for Use (Describe)
The Intra. Ox™ 2.0 Handheld Tissue Oximeter is intended to non-invasively estimate the percent oxygen saturation (StO2) in a volume of tissue.
The Intra. Ox™ 2.0 Handheld Tissue Oximeter is indicated for use in monitoring patients during circulatory or perfusion examinations.
The Intra. Ox™ 2.0 Handheld Tissue Oximeter is intended to be used by physicians, surgeons, nurses, or other skilled users in a medical environment.
The Intra. Ox™ 2.0 Handheld Tissue Oximeter should only be used on adult patients.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Notification K221010
#### GENERAL INFORMATION [807.92(A)(1)]
#### Applicant:
ViOptix, Inc. 39655 Eureka Drive Newark, CA 94560-4806 USA Phone: 510-226-5860 Fax: 510-226-5864
#### Contact Person:
Valerie Defiesta-Ng Vice President, Regulatory Affairs, Veranex, Inc. 224 Airport Parkway, Suite 250 San Jose, CA 95110 USA Phone: 408-400-0856 Fax: 408-400-0865 Email: Valerie.Defiesta-Ng@veranexsolutions.com
Date Prepared: April 22, 2022
#### DEVICE INFORMATION [807.92(A)(2)]
Classification: 21 CFR§870.2700
Product Code: MUD
Trade Name: Intra.Ox™ 2.0 Handheld Tissue Oximeter
Generic/Common Name: Oximeter, Tissue Saturation
PREDICATE DEVICE(S) [807.92(A)(3)] Intra.Ox™ 2.0 Handheld Tissue Oximeter (K191676)
{4}------------------------------------------------
### DEVICE DESCRIPTION [807,92(A)(4)]
The ViOptix Intra.Ox 2.0 Handheld Tissue Oximeter is a sterile, cordless, battery-powered device that non-invasively estimates the percent oxygen saturation (StO2) in a volume of tissue. The device includes three components and an accessory:
- Main Unit: a re-usable module consists of light sources, detectors, and processing ● electronics to convert measurements of reflected light into an estimate of StO>:
- Sheath: a single-use, sterile Sheath placed around the Main Unit during device use . (provide in the Disposable Kit); and
- Battery Pack: a single-use battery (provided in the Disposable Kit) that is paired with the . Main Unit to provide power.
The device uses spatially resolved optical measurements at five wavelengths. The device performs measurements on the patient by direct physical contact to the patient's tissue and displays the StO2 estimate on the built-in screen. The Intra.Ox 2.0 Handheld Tissue Oximeter is a single-use disposable constructed from biocompatible materials that can tolerate bodily fluids and other liquids such as disinfectants and marking materials.
The device shares the same indication for use and the same technological characteristics as the predicate device (K191676), including principle of operation, StO2 measured parameters, accuracy and range, energy delivered and power source.
# INDICATIONS FOR USE [807.92(A)(5)]
"The Intra.Ox™ 2.0 Handheld Tissue Oximeter is intended to non-invasively estimate the percent oxygen saturation (StO2) in a volume of tissue.
The Intra.Ox" 2.0 Handheld Tissue Oximeter is indicated for use in monitoring patients during circulatory or perfusion examinations.
The Intra.Ox"" 2.0 Handheld Tissue Oximeter is intended to be used by physicians, surgeons, nurses, or other skilled users in a medical environment.
The Intra.Ox™ 2.0 Handheld Tissue Oximeter should only be used on adult patients."
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES [807.92(a) (6)]
The technological characteristics of the modified Intra.Ox 2.0 Handheld Tissue Oximeter are substantially equivalent to the predicate device, Intra.Ox 2.0 Handheld Tissue Oximeter (K191676). Table 1 lists the technological characteristics of the predicate and modified devices and provides rationale to support a determination of substantial equivalence. Any differences between the devices do not raise any different questions of safety or efficacy.
{5}------------------------------------------------
| Table 1: Summary of Technological Characteristics | | | |
|---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | Predicate Intra.Ox 2.0<br>Handheld Tissue Oximeter | Modified Intra.Ox 2.0<br>Handheld Tissue Oximeter | Substantial<br>Equivalence<br>Rationale |
| 510(k) Number | K191676 | TBD | -- |
| Indications for<br>Use | The Intra.OxTM 2.0 Handheld<br>Tissue Oximeter is intended to<br>non-invasively estimate the<br>percent oxygen saturation (StO2)<br>in a volume of tissue.<br><br>The Intra.OxTM 2.0 Handheld<br>Tissue Oximeter is indicated for<br>use in monitoring patients during<br>circulatory or perfusion<br>examinations.<br><br>The Intra.OxTM 2.0 Handheld<br>Tissue Oximeter is intended to be<br>used by physicians, surgeons,<br>nurses, or other skilled users in a<br>medical environment.<br><br>The Intra.OxTM 2.0 Handheld<br>Tissue Oximeter should only be<br>used on adult patients. | The Intra.OxTM 2.0 Handheld<br>Tissue Oximeter is intended to<br>non-invasively estimate the<br>percent oxygen saturation (StO2)<br>in a volume of tissue.<br><br>The Intra.OxTM 2.0 Handheld<br>Tissue Oximeter is indicated for<br>use in monitoring patients during<br>circulatory or perfusion<br>examinations.<br><br>The Intra.OxTM 2.0 Handheld<br>Tissue Oximeter is intended to be<br>used by physicians, surgeons,<br>nurses, or other skilled users in a<br>medical environment.<br><br>The Intra.OxTM 2.0 Handheld<br>Tissue Oximeter should only be<br>used on adult patients. | N/A (same) |
| Measured<br>Parameters | Tissue oxygen saturation<br>(% StO2) | Tissue oxygen saturation<br>(% StO2) | N/A (same) |
| Principle of<br>Operation | Spectrophotometric oximetry | Spectrophotometric oximetry | N/A (same) |
| Energy<br>Delivered | Near-infrared light<br>Source: LED chips<br>Wavelengths: 730, 760, 810, 845,<br>and 895 nm | Near-infrared light<br>Source: LED chips<br>Wavelengths: 730, 760, 810, 845,<br>and 895 nm | N/A (same) |
| Single Patient<br>Use? | Yes, the Sheath is single patient<br>use.<br>The Main Unit is re-usable after<br>proper cleaning. | Yes, the Sheath is single patient<br>use.<br>The Main Unit is re-usable after<br>proper cleaning. | N/A (same) |
| Power Source | Battery Powered<br>Battery Type: 4 Lithium AA<br>Battery Voltage: 6 V total | Battery Powered<br>Battery Type: 4 Lithium AA<br>Battery Voltage: 6 V total | N/A (same) |
| StO2<br>Measurement<br>Range | StO2 displayed range shall be an<br>integer between 1 to 99%. | StO2 displayed range shall be an<br>integer between 1 to 99%. | N/A (same) |
| Physical Design | Ergonomic hand-held design for<br>range of hand sizes, includes<br>thumb rest, ambidextrous use.<br>Device contains a disposable kit<br>that contains a single-use battery<br>pack and sterile single-use<br>disposable sheath. | Ergonomic hand-held design for<br>range of hand sizes, includes<br>thumb rest, ambidextrous use.<br>Device contains a disposable kit<br>that contains a single-use battery<br>pack and sterile single-use<br>disposable sheath. | N/A (same) |
| Feature | Predicate Intra.Ox 2.0<br>Handheld Tissue Oximeter | Modified Intra.Ox 2.0<br>Handheld Tissue Oximeter | Substantial<br>Equivalence<br>Rationale |
| Dimensions and<br>Weight | 7.3" x 2.7" x 3.1"<br>L x W x H<br>Less than 1 lb | 7.3" x 2.7" x 3.1"<br>L x W x H<br>Less than 1 lb | N/A (same) |
| Components | Main Unit, Sheath,<br>Battery Pack and Quality Control<br>Target | Main Unit, Sheath, and<br>Battery Pack | Similar components.<br>The principle of<br>operation remains the<br>same for both devices.<br>The removal of the<br>Quality Control Target<br>does not raise different<br>questions of safety and<br>effectiveness as<br>demonstrated by<br>performance testing. |
Table 1: Summary of Technological Characteristics
{6}------------------------------------------------
#### SUBSTANTIAL EQUIVALENCE
The Intra.Ox 2.0 Handheld Tissue Oximeter is substantially equivalent to the predicate device with regard to intended use, Indications for Use, principle of operation and fundamental scientific technology. Any differences in the technological characteristics between the devices do not raise any different questions of safety or effectiveness. Thus, the modified Intra.Ox 2.0 Handheld Tissue Oximeter is substantially equivalent to the predicate device.
#### SUMMARY OF DESIGN MODIFICATION
The following design modifications have been implemented and do not affect the intended use, the Indications for Use, or alter the fundamental scientific technology of the device. ViOptix is proposing modifications to the cleared Intra. Ox 2.0 Handheld Tissue Oximeter (K191676) to optimize the design and improve manufacturability.
- The robustness of the latch-detection functions for sheath closure was improved. ●
- Adhesive was removed from the inside of the sheath window, a sterile pressure-relief vent was added to the lid, and the optical target was removed for the kit.
- An averaging functionality was added with screen modification to display both "live" and ● "average" readings.
- The removal of two non-functional components from a device Printed Circuit Board . Assembly (PCBA).
# PERFORMANCE DATA [807.92(B)]
All necessary bench testing was conducted on the modified Intra.Ox 2.0 Handheld Tissue Oximeter to support a determination of substantial equivalence to the predicate device. Tests performed included:
- Mechanical Testing ●
- Sterility Testing ●
- Latch Testing ●
- Shelf-Life and Transport Testing ●
- Software Verification and Validation ●
{7}------------------------------------------------
- Heterogeneous Blood Phantom Study
The collective performance testing demonstrates that the Intra.Ox 2.0 Handheld Tissue Oximeter does not raise any different questions of safety or effectiveness when compared to the predicate device. The results of performance testing demonstrate that the Intra.Ox 2.0 Handheld Tissue Oximeter performs as intended.
# [807.92(b)(1)]Non-clinical Testing Summary
In a heterogenous phantom study, measurements were made using the Intra.Ox 2.0 Handheld Tissue Oximeter in a heterogeneous phantom prepared with swine whole blood using an Intralipid solution to mimic tissue scattering and compared to a "gold standard" blood cooximeter. The results show that the Intra.Ox 2.0 Handheld Tissue Oximeter is substantially equivalent to the predicate device in limits of agreement to the gold standard, as well as 95% confidence intervals in slope and intercept. These data support the conclusion that the Intra.Ox 2.0 Handheld Tissue Oximeter provides as good or better agreement to the gold standard over the clinically relevant range than the predicate Intra.Ox Handheld Tissue Oximeter.
# [807.92(b)(2)]Clinical Testing Summary
Clinical testing was not required to demonstrate substantial equivalence.
# CONCLUSION [807.92 (b) (3)]
Nonclinical performance testing has been performed on the Intra.Ox 2.0 Handheld Tissue Oximeter to evaluate the overall performance of the device. The results confirm that the Intra. Ox 2.0 Handheld Tissue Oximeter meets its specifications and meets requirements for its intended use. The Intra.Ox 2.0 Handheld Tissue Oximeter is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.