The biopsy-needle is intended for soft-tissue biopsy (such as breast, kidney, liver, prostate).
Device Story
The Automatic Biopsy Needle (BAM) and Semi-automatic Biopsy Needle (BIM) are sterile, spring-loaded, disposable percutaneous soft tissue biopsy systems. The device consists of a handle, spring, cannula, stylet, trigger, and spring guide. It is used by a qualified physician under MR image guidance to obtain core biopsy samples. The BAM utilizes an automatic gun with a guillotine system where the mandrin moves automatically; the BIM uses a similar guillotine system but with manual mandrin movement. The physician inserts the needle into the target tissue; the spring-loaded mechanism triggers the cannula to cover the stylet notch, capturing a tissue sample. The device provides intact, cylindrical core samples for diagnostic evaluation. The system is designed for MRI compatibility, allowing for precise needle placement during image-guided procedures.
Clinical Evidence
No clinical data. Bench testing only. Performance evaluated via needle penetration, firing mechanism (shot) testing, sample quantity/length measurement, and sample quality/integrity observation under video microscopy. MRI safety evaluation conducted. All acceptance criteria met.
Indicated for soft-tissue biopsy (breast, kidney, liver, prostate) in patients requiring tissue sampling.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 18, 2024
ITP Innovative Tomography Products GmbH Dominik Främke Project Manager Universitätsstraße 136 Bochum, 44799 Germany
Re: K233294
Trade/Device Name: Automatic Biopsy Needle BAM, Semi-Automatic Needle BIM Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: Class II Product Code: KNW Dated: May 21, 2024 Received: May 21, 2024
Dear Dominik Främke:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Jessica Carr -S
Jessica Carr, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K233294
Device Name Automatic Biopsy Needle - BAM Semi-automatic Biopsy Needle - BIM
Indications for Use (Describe)
The biopsy-needle is intended for soft-tissue biopsy (such as breast, kidney, liver, prostate).
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Innotom. The logo consists of the letters "ITP" in bold, black font, with a small orange square above the "I". Below the letters, the website address "innotom.com" is written in a smaller, sans-serif font. The logo is simple and modern in design.
## 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of 21CFR Section 807.92.
The assigned 510(k) Number: K233294
Date of Preparation: 06/17/2024
| Submitter | Innovative Tomography Products GmbH<br>Universitaetsstr. 136<br>44799 Bochum<br>Germany<br>Phone: +49 234 54621940<br>Fax: +49 234 54622444 | Device & Predicate Device(s): | K233294<br>Automatic (BAM)<br>Semi-Automatic (BIM) | K222462 (Predicate)<br>Semi-Automatic (BIM) | | |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| E-Mail: | qm@innotom.com | General Device Characteristics | | | | |
| Contact: | Dominik Främke ( project manager) | Material | Needle: Inconel 625, Nitinol<br>Handle: ABS | Needle: Inconel 625, Nitinol<br>Handle: ABS | | |
| Device Trade<br>Name: | Automatic Biopsy Needle - BAM<br>Semi-automatic Biopsy Needle - BIM | Size<br>Diameter (G)/Length (mm)/Notch (mm) | BAM/BIM Diameter: 18, 16<br>BAM/BIM Length: 100, 150, 175,<br>200<br>Notch: 18 (BAM), 10/18 (BIM) | Diameter: 18, 16, 14<br>Length: 100, 150, 200<br>Notch: 10/20 (BIM) | | |
| Common<br>Name: | Gastroenterology-urology biopsy instrument | | | Design | BAM: Cutting is performed with an<br>automatic gun with guillotine<br>system; mandrin is moved<br>automatically with the help of a<br> | Cutting is performed with an<br>automatic gun with guillotine<br>system; mandrin is moved<br>manually. |
| Classification: | Instrument, Biopsy | | Single Use | Yes | Yes | |
| Regulation: | 876.1075 | | Cannula Covers the notch of stylet | Yes | Yes | |
| Product Code: | KNW | Biopsied Tissue Sample Length<br>(Cylindrical, Intact) | >10 mm (BAM, 18 mm notch)<br>>10 mm (BIM, 18 mm notch)<br>> 7 mm (BIM, 10 mm notch)<br>Tissue Types: Muscle, Liver, Kidney | In Muscle:<br>~6.72 mm for 10 mm Notch (BIM)<br>~11.77 for 20 mm Notch (BIM)<br>Overall (Muscle, Liver, Apple)<br>~8.82 mm for 10 mm Notch (BIM)<br>~17.22 for 20 mm Notch (BIM) | | |
| Primary<br>Predicate: | K222462 | Monitoring | MRI | MRI | | |
| Device<br>Description<br>Summary: | The Semi-automatic Biopsy-Needle - BIM and Automatic Biopsy Needle - BAM are<br>sterile, spring loaded, disposable percutaneous soft tissue biopsy systems. It consists of<br>the following major components: handle, spring, cannula, stylet, trigger and spring<br>guide. It is used to obtain multiple core biopsy samples from soft tissue such as breast,<br>kidney, liver and prostate. The needle has to be inserted by a qualified physician under<br>MR image guidance. | Sterlization | EO | EO | | |
| Indications for<br>Use: | The biopsy-needle is intended for soft-tissue biopsy (such as breast, kidney, liver,<br>prostate). | Packaging | Sterile pouches made of<br>sterilization-resistant 70 g/m²<br>heavy-duty paper and a plastic film | Tyvek bag | | |
| Indications for<br>Use<br>Comparison | The indications for use statement of the subject device is identical to the indications for<br>use statement of the primary predicate device. | | | | | |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for ITP, which stands for Innotom. The logo is black, except for a small orange square in the upper left corner. The website address, innotom.com, is printed in black below the logo.
Technological Comparison
The Semi-automatic Biopsy Needle - BIM and Automatic Biopsy Needle - BAM are identical or similar in technology, design and material to the predicate device. Both devices are sterile and for single-use.
The material of the needles are identical to the predicate/reference device. Based on the same intended use and the similarities in technology, design and materials the proposed devices are substantially equivalent to their predicate/reference devices. Biocompatibility, MR-testing, sterility and packaging demonstrate the safety and effectiveness of the proposed device.
#### Non-Clinical Tests
The following tests were performed to evaluate the performance of the subject device and support substantial equivalence with the predicate device.
- 1) Penetration of the needle
- 2) Shot of the needle
- 3) Quantity of the sample taken
- 4) Quality/integrity of the sample taken
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Innotom. The logo consists of the letters "iTP" in black, with a small orange square above the "i". Below the letters is the text "innotom.com" in a smaller font. The logo is simple and modern, with a focus on the company's name.
5) MRI Safety Evaluation
For the present report different Semiautomatic Biopsy Needles - BIM and Automatic Biopsy Needles - BAM has been used. 18G and 16G biopsy needles were used. Two lengths were used: 150 mm and 200 mm. The testing was conducted after the validated EO sterilization of all devices. Therefore the tests were conducted on final finished devices.
Description of the tests and Results
1) For the stylet perforation capacity, a biopsy needle is inserted in three different tissues (i.e., muscle, liver, kidney) according to the IFU. It is then visually tested to ensure that the stylet tip could easily and effortlessly penetrate the tissue without causing tears or lacerations to create a good entry path for the cannula. The needle is inserted in different tissues to assess the penetration capacity of the stylet tip. For all tested biopsy devices the insertion was appropriate and conducted effortlessly with no difficulties at all. The acceptance criteria were met. This demonstrates that the device works as intended.
2) To test the shot of the needle the release mechanism of the biopsy device was triggered in air as well as in the animal tissue. The test showed that the device spring has sufficient power for the cannula to cover the tip of the stylet that contains the notch completely. The tests were conducted for the 10 mm and 18 mm notch.
3) Different tissue samples were taken with biopsy devices for the 10 mm and 18 mm notch. After that the length of the collected samples was measured with a ruler. The length was written down and the results were analyzed. The results showed that the device met the acceptance criteria (e.g., ≥ 5 mm for the 10 mm notch and ≥ 9 mm for the 18 mm notch) and the dimensions of the samples were comparable to the dimensions of the sample acquired using the predicate.
4) The taken tissue samples in test 3 are observed under video microscope. The samples were carefully observed to see if they are cylindrical, intact and abundant. This was done for both 10 mm and 18 mm notch samples. All samples were cylindrical, intact and abundant. The acceptance criteria were met. The samples collected using the subject device were comparable to the samples acquired using the predicate.
5) MRI safety evaluation was performed to develop conditions of safe use.
- Clinical Tests Not needed.
- Conclusion Based on the comparison and analysis above, the proposed subject device is determined to be Substantially Equivalent (SE) to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.