The biopsy-needle is intended for soft-tissue biopsy (such as breast, kidney, liver, prostate).
Device Story
Semi-automatic biopsy needle; sterile, single-use, spring-loaded system. Components include handle, stylet, trigger, and coring guide. Used by clinicians for percutaneous soft-tissue biopsy (breast, kidney, liver, prostate). Device operates by manual insertion followed by a spring-loaded trigger mechanism that advances the cannula over the stylet notch to excise tissue. Output is a cylindrical tissue sample for histological analysis. MR-compatible design allows for use under MRI guidance. Benefits include improved sample quality and integrity compared to predicate, with clear visibility under MRI without significant artifacts or heating.
Clinical Evidence
Bench testing only. Evaluated penetration capacity, spring-loaded shot mechanism, and sample quality/integrity in animal tissue and apples. Sample sizes were measured and compared against acceptance criteria (5mm and 10mm). MR-compatibility assessed at 1.5T and 3T, confirming visibility without artifacts or heating. Biocompatibility testing included cytotoxicity, sensitization, acute systemic toxicity, and pyrogenicity. Results showed the device met all acceptance criteria and performed effectively.
Technological Characteristics
Sterile, single-use, spring-loaded biopsy needle. Materials differ from predicate to enable MR-compatibility. Dimensions include 14G, 16G, and 18G gauges with 10mm and 20mm notch sizes. Operates via mechanical spring-loaded trigger. Sterilized using validated ethylene oxide (EO) process.
Indications for Use
Indicated for soft-tissue biopsy (e.g., breast, kidney, liver, prostate) to obtain diagnostic tissue samples for histological examination during percutaneous procedures.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
ITP Innovative Tomography Products GmbH Dominik Fraemke Project Manager Universitaetsstr. 136 Bochum. 44799 Germany
January 20, 2023
#### Re: K222462
Trade/Device Name: Semi-automatic Biopsy-Needle (BIM 18/20); Semi-automatic Biopsy-Needle (BIM 18.15); Semi-automatic Biopsy-Needle (BIM 18/10);Semi-automatic Biopsy-Needle (BIM 16/20);Semi-automatic Biopsy-Needle (BIM 16/15);Semiautomatic Biopsy-Needle (BIM 16/10);Semi-automatic Biopsy-Needle (BIM 14/20);Semi-automatic Biopsy-Needle (BIM 14/15);Semi-automatic Biopsy-Needle (BIM 14/10) Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology-Biopsy Instrument Regulatory Class: Class II Product Code: KNW Dated: October 24, 2022 Received: October 24, 2022
Dear Dominik Fraemke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Colin K. Chen -S
for
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Submission Number (if known)
| Device Name | Semi-automatic Biopsy-Needle (BIM 18/20);<br>Semi-automatic Biopsy-Needle (BIM 18/15);<br>Semi-automatic Biopsy-Needle (BIM 18/10);<br>Semi-automatic Biopsy-Needle (BIM 16/20);<br>Semi-automatic Biopsy-Needle (BIM 16/15);<br>Semi-automatic Biopsy-Needle (BIM 16/10);<br>Semi-automatic Biopsy-Needle (BIM 14/20);<br>Semi-automatic Biopsy-Needle (BIM 14/15);<br>Semi-automatic Biopsy-Needle (BIM 14/10) | | |
|-----------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| Indications for Use (Describe) | The biopsy-needle is intended for soft-tissue biopsy (such as breast, kidney, liver, prostate). | | |
| Type of Use (Select one or both, as applicable) | <table><tr><td><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</td><td><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</td></tr></table> | <input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
| <input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 510(k)#: K222462 | | 510(k) Summary | Prepared on: 2022-12-22 |
|------------------------------------|--|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| Contact Details | | | 21 CFR 807.92(a)(1) |
| Applicant Name | | ITP Innovative Tomography Products GmbH | |
| Applicant Address | | Universitaetsstr. 136 Bochum 44799 Germany | |
| Applicant Contact Telephone | | +4923454621940 | |
| Applicant Contact | | Dr. Heinz-Wemer Henke | |
| Applicant Contact Email | | henke@innotom.com | |
| Correspondent Name | | ITP Innovative Tomography Products GmbH | |
| Correspondent Address | | Universitaetsstr. 136 Bochum 44799 Germany | |
| Correspondent Contact Telephone | | +4923454621940 | |
| Correspondent Contact | | Mr. Dominik Fraemke | |
| Correspondent Contact Email | | fraemke@innotom.com | |
| Device Name | | | 21 CFR 807.92(a)(2) |
| Device Trade Name | | Semi-automatic Biopsy-Needle (BIM 18/20);<br>Semi-automatic Biopsy-Needle (BIM 18/15);<br>Semi-automatic Biopsy-Needle (BIM 18/10);<br>Semi-automatic Biopsy-Needle (BIM 16/20);<br>Semi-automatic Biopsy-Needle (BIM 16/15);<br>Semi-automatic Biopsy-Needle (BIM 16/10);<br>Semi-automatic Biopsy-Needle (BIM 14/20);<br>Semi-automatic Biopsy-Needle (BIM 14/15);<br>Semi-automatic Biopsy-Needle (BIM 14/10) | |
| Common Name | | Gastroenterology-urology biopsy instrument | |
| Classification Name | | Instrument, Biopsy | |
| Regulation Number | | 876.1075 | |
| Product Code | | KNW | |
| Legally Marketed Predicate Devices | | | 21 CFR 807.92(a)(3) |
| Predicate # | | Predicate Trade Name (Primary Predicate is listed first) | |
| K130616 | | Semi-Automatic Biopsy Needle | |
| | | | KNW |
| Device Description Summary | | | 21 CFR 807.92(a)(4) |
The semiautomatic Biopsy needle– BM is a sterile, spring loaded, disule biopsy system. It consists of the
following maior components: handle stylet. tincer and coring nuide.
{4}------------------------------------------------
# Intended Use/Indications for Use
The biopsy-needle is intended for soft-tis sue biopsy (such as breast, kidney, liver, prostate).
# Indications for Use Comparison
The proposed devices are working equivalently to the predicate device. They are all intended to obtain diagnostic samples of tissue for histological examination during a percutaneaous biopsy procedure. They differ only in optical design aspects. In addition the proposed device is MR-compatible.
## Technological Comparison
The BM is identical or similar in technology, design and material to the predicate devices are sterile and for single-use. The only difference between the two devicesare the materials of the needles as well as optical design aspects. Based on the same intended use and the similarities in technology, design and materials the proposed devices are substantially equivale/ reference devices. Biocompatibility, MR-testing, sterility and packaging demonstrate the safety and effectiveness of the proposed device
# Non-Clinical and/or Clinical Tests Summary & Conclusions
The purpose of the tests is to evaluate:
1) the penetration of the needle
2) The shot of the needle
3) The quantity of the sample taken
4) The quality/integrity of the sample taken
5) MR tests for artificat
6) Heating and induction test
7) Biocompatibility
For the present report different Semiautomatic Biopsy Need. The 5 samples have been manufactured in 5 different lots. 14G, 16G and 18G biopsy needles were used. The lengths ranged from 100 mm. The testing was conducted after the validated EO sterilization of all devices. Therefore the tests were conducted on final devices
Description of the tests
1) For the stylet perforation capacity, a biopsy needle in three different tissues (animal tissue and apples) according to the IFU. It isthen visually tested to ensure that the style tip could easily and effortlessly penetrate the tissue without causing tears or lacerations to create a good enty path for the cannula. The need in different tissues to assess the penetration capacity of the stylet tip. 2) To test the shot of the needle the release mechanism of the biopsy device were triggered in the air as well as in the animal tissue. If the device works well, the cannula would cover the tip of the stylet that contains then otch completely. With this you can test if the spinq has enough power to push the cannula over the notch to cuttissue. This was conducted for the 10 mm and 20 mm notch.
3) Different tissue samples were taken with biopsy devices for the 10mm and 20 mm notch. After that the collected samples was measured with aruler. The length was written down and the results were analyzed.
4) The taken tissue samples in test 3 are observed under video microscope.
The samples were carefully observed to seeif they are cylindrical, intact and abundant. This was also done for 10 mm and 20 mm notch samples.
5&6) Furthermore MR-compatibility tests were conducted.
For the tests we used similar needles that use the biopsy devices and tested them under a 37 and 1.5T MRI-System. The needle is clearly visible under MR-quidance without any major attifacts. Furthermore no heating occured in any unit 7) Biocompatibility tests for Cytotoxicity, Sensitization, Acute Systemice Toxicity and Pyrogenicy were conducted.
1.Needle penetration capacity
For all tested biopsy devices the insertion was optimal and conducted effortles sty with no difficulties at all. The acceptance criteria were met. This demonstrates that the devices works for penetration and is safe.
2.Shot of the needle.
For all tested hions v device the not the not the case for hotch notch sizes (10 mm and 20 mm ) From this
### Page 2 of 3
### 21 CFR 807.92(a
21 CFR 807 92(a)(5)
#### 21 CFR 807.92(a)(6)
{5}------------------------------------------------
============================================================================================================================================================================= it can be concluded that the spring is strong enough. The acceptance criteria were met and the device works and is safe.
#### 3.Quantity of the collected samples
Each sample for the liver and apple equals the notch size. For muscle tissue taken with the 10 mm notch the average length is 6.72 mm and for 20 mm notch 11.77 mm. Every sample was larger than the acceptance criteria of 5 mm and 10 mm. Therefore the criteria is met. The average sample size over all tissues is 8.82mm for the 10 mm notch. The device is therefore is therefore effective to take proper samples.
4.Quality/integrity of the collected samples
All samples are cylindrical, intact and abundant. The acceptance criteria were met. Because of this the device is effective and safe.
Based on the performed tests the proposed device is as safe and as effective as the predicate device. Furthe more the average sample size of the collected tissue is larger for the proposed device. Proposed device: 10 mm notch: 8.82 mm samples, 20 mm notch: 17.22 mm samples
Predicate device: 10 mm notch 7 cm samples, 20 mm notch: 13 mm samples
5&6.
Based on the MR-tests the device is MR-compatible. The new any major artificats and is cleary visible especially compared to the needle with stainless steel. In addition no heating or cured during these tests.
7. Biocompatibility were proofen with the conducted tests.
Conclusion:
The Semi-automatic Biopsy-Needle functions as intended. The results drawn from the non-dinical tests as well as the literature presented in this 510(k) submission demonstrate the proposed device is as safe, effectiveand performs as well as the legally marketed predicate device. The proposed device has a similiar intended use and design characteristics. The M-compatbility were proven with MR tests and the different needlemateliality. Therefore the proposed device is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.