3M™ Clarity™Aligners with Quick Attachments

K233257 · 3M Company, Unitek Orthodontic Products · NXC · Sep 29, 2023 · Dental

Device Facts

Record IDK233257
Device Name3M™ Clarity™Aligners with Quick Attachments
Applicant3M Company, Unitek Orthodontic Products
Product CodeNXC · Dental
Decision DateSep 29, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.5470
Device ClassClass 2
AttributesTherapeutic, 3rd-Party Reviewed

Intended Use

3M™ Clarity™ Aligners with Quick Attachments are indicated for the alignment of teeth during orthodontic treatment of malocclusion.

Device Story

3M™ Clarity™ Aligners with Quick Attachments are clear, removable plastic sequential aligners used for orthodontic treatment of malocclusions. The system provides an alternative delivery mechanism for orthodontic attachments, which are used to assist aligner retention and optimize force systems for tooth movement. Attachments are either formed using an empty template filled with adhesive and cured on the teeth or applied as pre-formed, 3D-printed custom attachments. The device is used in a clinical setting under the direction of an orthodontist or dental professional. The aligners apply controlled forces to teeth to achieve alignment. By providing more accurately shaped attachments, the system aims to improve the efficiency and effectiveness of the orthodontic treatment process compared to traditional methods.

Clinical Evidence

Bench testing only. Performance testing included abrasion, bond strength, flexural strength, and stain resistance to validate the Quick Attachment delivery mechanism. No clinical data was required or provided.

Technological Characteristics

Clear plastic sequential aligners. Features pre-formed, pre-cured, or template-formed 3D-printed attachments. Materials are consistent with previously cleared 3M Clarity Aligners. System utilizes orthodontic adhesive for attachment bonding. Software is used for treatment planning and aligner design, consistent with the predicate device.

Indications for Use

Indicated for the alignment of teeth during orthodontic treatment of malocclusion in patients requiring orthodontic correction.

Regulatory Classification

Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font. September 29, 2023 3M Company, Unitek Orthodontic Products % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114 Re: K233257 Trade/Device Name: 3M™ Clarity™Aligners with Ouick Attachments Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic plastic bracket Regulatory Class: Class II Product Code: NXC Dated: September 28, 2023 Received: September 28, 2023 Dear Prithul Bom: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerelv. # Michael E. Adjodha -S Michael E. Adjodha, MChE, RAC, COIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic. Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. Submission Number (if known) K233257 KXXXXX Device Name 3M™ Clarity TM Aligners with Quick Attachments Indications for Use (Describe) 3M™ Clarity™ Aligners with Quick Attachments are indicated for the alignment of teeth during orthodontic treatment of malocclusion. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ | 510(k) #: KXXXX | K233257 | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------| | 510(k) Summary | Prepared on: 2023-09-27 | | | 21 CFR 807.92(a)(1) | | Contact Details | | | Applicant Name | 3M Company, Unitek Orthodontic Products | | Applicant Address | 2510 Conway Avenue St.Paul MN 55144 United States | | Applicant Contact Telephone | 612-710-9847 | | Applicant Contact | Mrs. Hillary Peetsch | | Applicant Contact Email | hpeetsch@mmm.com | | Device Name | | | | 21 CFR 807.92(a)(2) | | Device Trade Name | 3M™ Clarity™ Aligners with Quick Attachments | | Common Name | Orthodontic plastic bracket | | Classification Name | Aligner, Sequential | | Regulation Number | 872.5470 | | Product Code | NXC | | Legally Marketed Predicate Devices | | | | 21 CFR 807.92(a)(3) | | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | | K231464 | 3M™ Clarity™ Aligners-Force, 3M™ Clarity™ Aligners-Flex | | | Product Code | | | NXC | | Device Description Summary | | | | 21 CFR 807.92(a)(4) | | 3M™ Clarity™ Aligners with Quick Attachments, is a clear plastic sequential aligner that offer a solution for patients who want an aesthetic orthodontic treatment by utilizing a set of removable aligners to correct tooth malocclusions without the use of conventional wire and bracket orthodontic technology. 3M™ Clarity™ Aligners with Quick Attachments utilize the option of pre-formed and pre-cured attachments offering a more accurately shaped attachment. Attachments help create forces on the tooth which can assist aligner retention or optimized aligner force system for tooth movement. 3M Clarity Aligners with Quick Attachments provide the option to form the attachments via an empty attachment template that is filled with adhesive and cured on the teeth or by pre-formed 3D printed custom attachments. The aligners, with or without attachments, are for orthodontic treatment of malocclusions. | | | Intended Use/Indications for Use | | | | 21 CFR 807.92(a)(5) | | 3M™ Clarity™ Aligners with Quick Attachments are indicated for the alignment of teeth during orthodontic treatment of malocclusion. | | | Indications for Use Comparison | | | | 21 CFR 807.92(a)(5) | | Both predicate and subject device have the same indications for use. | | | Technological Comparison | | | | 21 CFR 807.92(a)(6) | | 3M™ Clarity™ Aligners with Quick Attachments are substantially equivalent to the predicate device. 3M™ Clarity™ Aligners with Quick Attachments are comparable to previously cleared device, 3M™ Clarity™ Aligners (3M™ Clarity™ Aligners-Force, 3M™ Clarity™ Aligners- | | {4}------------------------------------------------ Flex) (K231464) as they have the same indications for use, same manufacturing processes, same design, same device feature, software, and are considered substantially equivalent (21 CFR 807.100). #### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) The proposed Quick Attachment alternative attachment delivery mechanism completed functional and performance testing; such tests include but are not limited to: abrasion, bond strength, flexural strength, and stain resistance. Not applicable The performance testing showed the alternate Quick Attachment delivery mechanism does not impact the safety and effectiveness profile of the 3M™ Clarity™ Aligners-Force, 3M™ Clarity™Aligners-Flex and is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...