The Spark™ Clear Aligner System is indicated for the alignment of teeth during orthodontic treatment of malocclusion.
Device Story
Spark™ Clear Aligner System consists of custom-manufactured, thin, clear, plastic removable appliances; used for orthodontic tooth movement. Clinician-prescribed; treatment planning, design, and manufacturing supported by proprietary software. Clinician monitors treatment; patient wears aligners in a specific sequential order. Features include attachments (via templates or prefilled templates), hooks/cutouts for elastics, bite ramps, pontics, and posterior bite turbos. Aligners apply gentle pressure to move teeth incrementally. Used in clinical settings; patient removes aligners for eating/cleaning. Output is physical tooth movement over time; benefits include aesthetic, removable orthodontic correction.
Clinical Evidence
No clinical data conducted. Substantial equivalence established via non-clinical bench testing and adherence to international consensus standards, including ISO 10993 (biocompatibility), ISO 14971 (risk management), and ASTM F1980 (accelerated aging).
Technological Characteristics
Constructed from TruGEN materials. Features include attachments, hooks, cutouts, bite ramps, pontics, and posterior bite turbos. Proprietary software used for treatment planning and design. Biocompatibility tested per ISO 10993-1, -5, -10, -11, ISO 7405, and ANSI ADA No. 41. Accelerated aging tested per ASTM F1980-21.
Indications for Use
Indicated for the alignment of teeth during orthodontic treatment of malocclusion in patients requiring orthodontic correction.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 11, 2024
Ormco Corporation Samantha Britt Principal Specialist, Regulatory Affairs 200 S. Kraemer Blvd. Brea, California 92821
Re: K240501
Trade/Device Name: Spark™ Clear Aligner System Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: September 16, 2024 Received: September 16, 2024
Dear Samantha Britt:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{3}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael E. Adjodha -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use
Submission Number (if known)
K240501
Device Name
Spark™ Clear Aligner System
Indications for Use (Describe)
The Spark™ Clear Aligner System is indicated for the alignment of teeth during orthodontic treatment of malocclusion.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
| 510(k) #: | K240501 | | 510(k) Summary | | Prepared on: | 2024-10-09 |
|-----------|---------|--|----------------|--|--------------|------------|
|-----------|---------|--|----------------|--|--------------|------------|
| Contact Details | | | | 21 CFR 807.92(a)(1) |
|-----------------------------|--------------------------------------------------|--|--|---------------------|
| Applicant Name | Ormco Corporation | | | |
| Applicant Address | 200 S. Kraemer Blvd. Brea CA 92821 United States | | | |
| Applicant Contact Telephone | 714.817.7252 | | | |
| Applicant Contact | Ms. Samantha Britt | | | |
| Applicant Contact Email | samantha.britt@envistaco.com | | | |
| Device Name | 21 CFR 807.92(a)(2) |
|---------------------|-----------------------------|
| Device Trade Name | Spark™ Clear Aligner System |
| Common Name | Orthodontic plastic bracket |
| Classification Name | Aligner, Sequential |
| Regulation Number | 872.5470 |
| Product Code(s) | NXC |
| Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) |
|------------------------------------|---------------------|
|------------------------------------|---------------------|
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
|-------------|----------------------------------------------------------|--------------|
| K203737 | Spark™ Clear Aligner System | NXC |
| K203688 | Clear Aligner | NXC |
| Device Description Summary | 21 CFR 807.92(a)(4) |
|----------------------------|---------------------|
|----------------------------|---------------------|
The Spark™ Clear Aligner System consists of a series of doctor-prescribed, custom manufactured, thin, clear, plastic removable
orthodontic appliances (aligners) that gently move the participant's teeth in small increments from their original state to a more optimal,
treated state. Treatment planning, aligner design, and aligner manufacture are supported by a proprietary software system. The
clinician receives the aligners and provides them in sequential "stages" to the patient, confirming fit and monitoring treatment from the
placement of the first aligner to the removal of the final aligner. The aligners are held in place by pressure and can be removed by the
patient at any time. The aligners are individually identified and dispensed to patients and are to be worn in a specific prescribed
sequence. Several treatment options may be integrated into the Spark Clear Aligners consisting of:
• Attachments: Attachments are small protuberances of tooth-colored dental bonding material and may be prescribed by the dental
clinician to act as anchor points on the teeth to focus the forces of the aligner to help with teeth movement.
o Templates: Attachments are placed onto the teeth via Templates. The dental
practitioner may choose a standard dental composite, to fill into the Template themselves, and
adhesive to bond the attachments onto the dentition. The option for composite material to
be prefilled into the Template during manufacturing is also available via Prefilled
Attachment (PFA) Templates.
• Hooks & Cutouts: Hooks and Cutouts may be designed into the aligner to accommodate elastics for aiding in applying additional
aligner forces. Integrated Hooks are an optional alternative to the traditional Hooks.
• Bite ramps: Bite ramps are step features built into the lingual surfaces of the upper and lower aligner arch that are used by the clinician
to prop open the bite temporarily so that certain teeth have a clearer path for movement without interference.
• Pontics: Pontics are cavity spaces (that may or may not be "tooth-shaped") built into the aligner per clinician's request to fill voids of
missing teeth or other gaps that the clinician may wish to retain during treatment.
{6}------------------------------------------------
· Posterior Bite Turbos: Posterior Bite Turbos (PBT) are designed to prevent complete closure of jaws.
# Intended Use/Indications for Use
The Spark™ Clear Aligner System is indicated for the alignment of teeth during orthodontic treatment of malocclusion.
### Indications for Use Comparison
The Subject Device and Predicate Device have the same indications for use.
# Technological Comparison
This section provides the substantial equivalence rationale for Spark™ Clear Aligner System and the predicates with regards to Indications for Use, Technology, and Performance Testing. As described in this section, the differences between the Subject Device and the Predicate Device do not raise different questions of substantial equivalence.
Details of the similarities between Subject and Predicate Devices:
The similarities between Subject Device are the Indications for Use/ Intended Use, technological characteristics, and mode of action. Biocompatibility testing and material testing processes are similar to the Predicate Device. The Subject Device's Integrated Hook and analogous feature on the Reference Device share a rounded design and same protrusion height. Therefore, the two features are similar shape and profile, with respect to the risk of causing irritation in the oral mucosa from the surface protrusions.
Details of the differences between Subject and Predicate Devices:
There are no major differences between the Subject Device and predicates. However, there are two minor differences between Subject Device and Predicate Device. The Predicate Device includes Templates which are used by the dental clinician to place attachments onto the teeth at specific locations. The composite material is filled into the field, whereas the Subject Device includes an optional Prefilled Attachment (PFA) Template. Additionally, the Subject Device includes Integrative to the Predicate Device's traditional Hooks, that accommodate elastics for aiding in applying additional aligner forces.
Conclusion:
Based on a comparison of intended use, indications for use, technological characteristics, materials, principle of operation, features, and performance data, the Spark™ Clear Aligner System is deemed to the Predicate Device as it satisfies all criteria of substantial equivalence and does not raise new concerns regarding substantial equivalence: (1) Indications for Use, (2) Technological Characteristics, and (3) Performance Data. Spark™ Clear Aligner System does not introduce a fundamentally new scientific technology and the nonclinical tests demonstrate that the device is substantially equivalent.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Non-Clinical performance bench testing to international standards and FDA recognized consensus standards for Aligner, Sequential Device has been conducted to determine conformance in regards to:
· ASTM F 1980-21 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices · ISO 14971:2019-12 Third Edition: Medical devices - Application of risk management to medical devices
Biocompatibility testing were complete accessory, such as Cytotoxicity, Sensitization, and on the Aligner materials (TruGEN materials), such as Cytotoxicity, Sensitization, and Acute Systemic & Pyrogenicity, in accordance with the following standards:
· ISO 10993-1 2018-08 Fifth Edition: Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
## 21 CFR 807.92(a)(5)
### 21 CFR 807.92(a)(6)
{7}------------------------------------------------
· ISO 10993-5 2009-06-01 Third Edition: Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity
· ISO 10993-10 Third Edition 2010-08-01:Biological evaluation of medical devices -Part 10: Tests for irritation and skin sensitization
· ISO 10993-11 2017-09 Third Edition: Biological evaluation of medical devices-Part 11: Tests for systemic toxicity
- ISO 7405 Third Edition 2018-10, Corrected version 2018-12: Dentistry - Evaluation of biocompatibility of medical devices used in dentistry
· ANSI ADA Standard No. 41-2020: Evaluation of Biocompatibility of Medical Devices Used in Dentistry
### Performance Testing Clinical:
Clinical data is not needed to characterize performance and establish substantial equivalence. The non-clinical test data characterize all performance aspects of the device based on well-established scientific and engineering principles. Clinical testing has not been conducted on this product.
### Conclusion:
Based on a comparison of intended use, indications for use, material composition, technological characteristics, principle of operation, features and performance data, the Spark™ Clear Aligner System is deemed to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.