The ROSA® Shoulder System, for use with the ROSA® RECON platform, is indicated as a stereotaxic instrumentation system for Total Shoulder Arthroplasty (TSA) surgery. It is used to assist the surgeon in providing software-defined spatial boundaries for orientation and reference to identifiable anatomical structures for the accurate placement of the shoulder implant components. The robotic arm placement is performed relative to anatomical landmarks and bony anatomy as recorded using the system intraoperatively, and based on a three-dimensional representation of the bone structures determined pre-operatively using compatible CT based imaging technology. It includes a robotic arm, an optical sensor navigation system and accessories, software system, surgical instruments and accessories.
Device Story
ROSA Shoulder System is a stereotaxic instrumentation system for Total Shoulder Arthroplasty (TSA). It utilizes the ROSA RECON platform, incorporating a robotic arm, optical sensor navigation, and software. Pre-operative CT-based 3D bone models are uploaded via the Zimmer Biomet Portal. Intraoperatively, the system registers patient anatomy; the surgeon uses the robotic arm to guide humeral resection and glenoid reaming. The system provides software-defined spatial boundaries for implant component placement relative to anatomical landmarks. Used in clinical settings by surgeons, the device assists in reproducing pre-operative plans, aiming for accurate implant positioning and improved surgical outcomes.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing, engineering analysis, usability engineering, and a validation lab study using cadaveric specimens to simulate full surgical use. Biocompatibility testing followed ISO 10993. Electrical safety and EMC testing confirmed compliance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-6.
Technological Characteristics
Stereotaxic instrumentation system comprising a robotic arm, optical navigation, and software. Connectivity via Zimmer Biomet Portal (MDDS). Complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-6. Software follows IEC 62304 standards. System uses optical tracking for real-time instrument guidance relative to CT-derived anatomical landmarks.
Indications for Use
Indicated for skeletally mature patients undergoing Total Shoulder Arthroplasty (TSA) using compatible humerus and glenoid implants. Contraindicated for patients with significant bone loss, severe range of motion limitations, active shoulder joint infections, or those requiring revision surgery. Also contraindicated if strong infrared sources/reflectors are present or if pre-existing radio-opaque elements interfere with imaging/registration.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
Predicate Devices
ExactechGPS Total Shoulder Application, Equinoxe Planning Software (K213546)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 21, 2024
Orthosoft Inc (d/b/a Zimmer CAS) Sankalp Dere Regulatory Affairs Project Manager 75 Queen Street Suite 3300 Montreal, QC H3C 2N6 Canada
Re: K233199
Trade/Device Name: ROSA® Shoulder System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO, LLZ Dated: January 18, 2024 Received: January 22, 2024
Dear Sankalp Dere:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tejen D. Soni -S
For Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair
{3}------------------------------------------------
and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known) K233199
Device Name ROSA® Shoulder System
#### Indications for Use (Describe)
The ROSA® Shoulder System, for use with the ROSA RECON platform, is indicated as a stereotaxic instrumentation system for Total Shoulder Arthroplasty (TSA) surgery. It is used to assist the surgeon in providing software-defined spatial boundaries for orientation and reference to identifiable anatomical structures for the accurate of the shoulder implant components.
The robotic arm placement is performed relative to anatomical landmarks and bony anatomy as recorded using the system intraoperatively, and based on a three-dimensional representation of the bone structures determined pre-operatively using compatible CT based imaging technology.
It includes a robotic arm, an optical sensor navigation system and accessories, software system, surgical instruments and accessories.
The ROSA Shoulder System is designed for use on a skeletally mature patient population. The target population has the same characteristics as the population targeted by the implants compatible with the ROSA Shoulder System.
The ROSA Shoulder System is to be used with the following shoulder replacement systems in accordance with their indications and contraindications:
· Humerus implants: Comprehensive® Total Shoulder System, Comprehensive® Reverse Shoulder System, Identity® Shoulder System and Identity® Reverse Shoulder System.
· Glenoid implants: Alliance™ Glenoid and Comprehensive® Reverse Shoulder System.
Type of Use (Select one or both, as applicable)
| <div style="display:flex; align-items:center;"><span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> |
|-------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) Summary
In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the ROSA Shoulder System 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on September 13, 2019.
| Sponsor: | Orthosoft Inc (d/b/a. Zimmer CAS)<br>75 Queen Street, Suite 3300<br>Montreal, QC, Canada H3C 2N6<br>Establishment Registration Number: 9617840 | | | | | | | | | | | | | | | |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| Contact Person: | Sankalp Dere<br>Regulatory Affairs Project Manager<br>SankalpSanjay.Dere@zimmerbiomet.com | | | | | | | | | | | | | | | |
| Date: | 27 SEP 2023 | | | | | | | | | | | | | | | |
| Subject Device: | Trade Name: ROSA® Shoulder System<br>Common Name: ROSA Shoulder System, ROSA Shoulder, ROSA TSA System<br>Classification Name:<br>• OLO - Orthopedic Stereotaxic Instrument (21 CFR 882.4560)<br>• LLZ - Medical image management and processing system (21 CFR 892.2050) | | | | | | | | | | | | | | | |
| Predicate Device: | Device Manufacturer 510(k) Number Predicate or Reference ExactechGPS Total<br>Shoulder Application,<br>Equinoxe Planning<br>Software Blue Ortho K213546 Primary<br>Predicate ROSA Knee System Zimmer CAS K230243 Secondary<br>Predicate | | | | | | | | | | | | | | | |
| Purpose and Device Description: | The ROSA® Shoulder system (RSS) for use with ROSA® RECON Platform (cleared via K230243) is used to assist surgeons in performing Total Shoulder Arthroplasty (TSA) for both anatomic and reverse techniques. It features humeral resection and glenoid reaming capabilities to reproduce the preoperative plan intraoperatively with | | | | | | | | | | | | | | | |
{6}------------------------------------------------
| | The RSS uses a Non-Device Medical Device Data System (MDDS)<br>called the Zimmer Biomet Portal, which manages the creation and<br>tracking of surgical cases. The cases with the pre-operative planning<br>based on surgeon preferences reside on the portal until they are<br>uploaded to the ROSA RECON Platform before surgeries.<br><br>The intra-operative workflow and surgical concepts implemented in<br>the system remain close to the conventional TSA workflow. As such,<br>at the time of the surgery, the system mainly assists the surgeon in (1)<br>determining reference alignment axes in relation to anatomical<br>landmarks, (2) planning the orthopedic implants location intra-<br>operatively based on these reference alignment axes and orthopedic<br>implant geometry, and (3) precisely position the humeral cut guide and<br>glenoid reamer relative to the planned orthopedic implant location by<br>using a robotic arm. |
|---------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use: | The ROSA® Shoulder System, for use with the ROSA® RECON<br>platform, is indicated as a stereotaxic instrumentation system for Total<br>Shoulder Arthroplasty (TSA) surgery. It is used to assist the surgeon<br>in providing software-defined spatial boundaries for orientation and<br>reference to identifiable anatomical structures for the accurate<br>placement of the shoulder implant components.<br><br>The robotic arm placement is performed relative to anatomical<br>landmarks and bony anatomy as recorded using the system<br>intraoperatively, and based on a three-dimensional representation of<br>the bone structures determined pre-operatively using compatible CT<br>based imaging technology.<br><br>It includes a robotic arm, an optical sensor navigation system and<br>accessories, software system, surgical instruments and accessories. |
| | The ROSA® Shoulder System is designed for use on a skeletally<br>mature patient population. The target population has the same<br>characteristics as the population targeted by the implants compatible<br>with the ROSA® Shoulder System.<br><br>The ROSA® Shoulder System is to be used with the following<br>shoulder replacement systems in accordance with their indications and<br>contraindications:<br>• Humerus implants: Comprehensive® Total Shoulder System,<br>Comprehensive® Reverse Shoulder System, Identity® Shoulder<br>System and Identity® Reverse Shoulder System.<br>• Glenoid implants: Alliance™ Glenoid and Comprehensive® Reverse Shoulder System. |
| | |
| Contraindications: | The ROSA® Shoulder System may not be suitable for use in case of:<br>• Shoulder pathologies with significant bone loss<br>• Shoulder pathology severely limiting range of motion<br>• Active infections of the shoulder joint area<br>• Shoulder replacement revision surgery<br>• Presence of strong infrared sources or infrared reflectors in the<br>vicinity of the trackers<br>• Contraindications for the implants as given by the implant<br>manufacturer<br>• Implants that are not compatible with the system<br>• Presence of pre-existing radio-opaque elements that could<br>interfere with pre-operative image acquisition or intra-operative<br>registration |
| Summary of<br>Technological<br>Characteristics: | The rationale for substantial equivalence is based on consideration of<br>the following characteristics:<br>• Indications for Use/Intended Use: The proposed device has the<br>same intended use and similar indications for use as the<br>predicates. The devices are intended to assist surgeons by<br>providing software-defined spatial boundaries for orientation, and<br>reference information to relative anatomical structures for<br>accurate implant component placement.<br>• The proposed and the predicates allow users for patient bone<br>registration within the software and use it to guide instruments<br>intra-operatively in real time using similar navigation system.<br>• The proposed and ROSA predicate device share the same<br>previously cleared ROSA RECON Platform with its major core<br>hardware and software components, tracking system, various<br>instrumentation including reusable and disposable. The proposed<br>and ROSA predicate use the robotic arm on the ROSA RECON<br>Platform to assist the guidance of instruments.<br>• The proposed and primary predicate assist surgeon using similar<br>intra-operative instruments guiding to perform the glenoid<br>reaming based on the surgeon's plan. |
| Summary of<br>Performance Data<br>(Nonclinical and/or<br>Clinical): | The following performance data was provided in support of the<br>substantial equivalence determination:<br>Biocompatibility Testing<br>The biocompatibility evaluation for ROSA Shoulder System was<br>conducted in accordance with ISO 10993. The evaluation shows that<br>the ROSA Shoulder devices meets biocompatibility requirements.<br>Electromagnetic Compatibility (EMC) and Electrical Safety |
| Substantial Equivalence Conclusion: | Both the proposed device and predicate devices have the same intended use and similar indications for use and are used to guide instrumentation to assist with orthopedic surgery. The technological characteristics between the proposed device and the primary predicate are similar with bone registration and intra-operatively instrument guidance with live tracking. The proposed device does not perform any pre-operative surgical planning as is the case with primary predicate. The proposed ROSA Shoulder System and the secondary ROSA predicate utilize the same ROSA RECON Platform with its core hardware and software components. In addition, the proposed device and ROSA predicate use the robotic arm to assist the guidance of instruments. In summary, any differences between the devices do not raise new questions of safety and effectiveness and verification and validation |
| | |
| | |
| | Electrical Safety and EMC testing was conducted on the ROSA RECON Platform. The device complies with recognized standards: IEC 60601-1, IEC 60601-1-2 & IEC 60601-1-6 standards.<br>Device Performance Testing<br>Verification and Validation Testing for ROSA Shoulder System was conducted with the following aspects:<br>• Performance Tests- to ensure the performance of the implemented features and verify related design inputs<br>• Engineering Analysis- to ensure the performance of the implemented features and verify related design inputs<br>• Usability Engineering- addressed user interactions with ROSA Shoulder System<br>• Validation Lab- performed to validate that using ROSA Shoulder System is safe and effective and that the performances of the system are acceptable under full simulated use on cadaveric specimens<br>Software Verification and Validation Testing<br>Software tests were conducted to satisfy requirements of the FDA Guidance for the Content Premarket Submissions for Device Software Functions and IEC 62304 (Medical Device Software - Life Cycle Processes). The software was considered an "Enhanced" Documentation Level, since a failure of the software could result in serious injury or death to the patient. The testing demonstrates that the ROSA Shoulder System does not raise any new issues of safety and effectiveness as compared to the predicate devices. |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
| activities demonstrate that the proposed device is at least as safe and |
|-------------------------------------------------------------------------|
| effective as the legally marketed predicate devices. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.