K233109 · Civco Medical Solutions · ITX · Dec 22, 2023 · Radiology
Device Facts
Record ID
K233109
Device Name
TP Pivot Pro™ Needle Guide
Applicant
Civco Medical Solutions
Product Code
ITX · Radiology
Decision Date
Dec 22, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1570
Device Class
Class 2
Indications for Use
Device is intended to be used for instrument placement to access anatomical structures under transrectal ultrasound guidance. - Prostate - Biopsy and minimally invasive puncture - Surgical (Prostate) - Biopsy and minimally invasive puncture
Device Story
TP Pivot Pro™ Needle Guide is a mechanical tool for freehanded transperineal prostate biopsies; facilitates sampling of anterior prostate zones. Used with ultrasound probe; clinician visualizes target areas, plans access site, and guides needle for specimen collection. Device clamps to probe; provides stable path for needle insertion. Used in clinical settings by physicians. Benefits include precise access to difficult-to-reach anatomical structures.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including biocompatibility (ISO 10993-5, -10, -11), mechanical design verification (clamping force, axial loading, dislodge force, torque testing), and simulated usability testing.
Technological Characteristics
Materials: ABS thermoplastic, 304 stainless steel introducer needle. Mechanical clamping mechanism for ultrasound probe attachment. Sterilization: Ethylene Oxide. Biocompatibility: ISO 10993-1 compliant for limited contact duration. No electronic or software components.
Indications for Use
Indicated for instrument placement to access anatomical structures under transrectal ultrasound guidance for prostate biopsy and minimally invasive puncture in surgical or clinical settings.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
{0}------------------------------------------------
July 1, 2024
CIVCO Medical Solutions James Leong Regulatory Affairs Manager
102 First Street South Kalona, Iowa 52247
Re: K233109
Trade/Device Name: TP Pivot Pro™ Needle Guide Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic Ultrasonic Transducer Regulatory Class: Class II Product Code: ITX
Dear James Leong:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change for your device cleared on December 22, 2023. Specifically, FDA is updating this SE Letter because FDA inadvertently indicated that the SE determination also included review and clearance of a predetermined change control plan (PCCP). However, your 510(k) submission did not include a PCCP, so FDA is providing this administrative correction. Please see the attached revised clearance letter.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Yanna Kang, OHT8: Office of Radiological Health, (301)796-6704, Yanna.Kang@fda.hhs.gov.
Sincerely,
# Yanna S. Kang -S
Yanna Kang, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
July 1, 2024
CIVCO Medical Solutions James Leong Regulatory Affairs Manager 102 First Street South Kalona, Iowa 52247
Re: K233109
Trade/Device Name: TP Pivot Pro™ Needle Guide Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic Ultrasonic Transducer Regulatory Class: Class II Product Code: ITX Dated: September 25, 2023 Received: September 27, 2023
Dear James Leong:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change for your device cleared on December 22, 2023.
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{2}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product: and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{3}------------------------------------------------
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Yanna S. Kang -S
Yanna Kang, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use
Submission Number (if known)
K233109
Device Name
TP Pivot Pro™ Needle Guide
Indications for Use (Describe)
Device is intended to be used for instrument placement to access anatomical structures under transrectal ultrasound quidance.
- · Prostate Biopsy and minimally invasive puncture.
- · Surgical (Prostate)- Biopsy and minimally invasive puncture.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
### 510(k) Summary
This 510(k) Summary is prepared in accordance with the requirements of 21 CFR 807.92.
The assigned 510(k) numberis: _K233109 .
- 1. Submitter's Identifications:
| Establishment: | CIVCO Medical Instruments Co., Inc. |
|-----------------------------------------|--------------------------------------------|
| Address: | 102 First Street South<br>Kalona, IA 52247 |
| Registration Number: | 1937223 |
| Operations Manufacturer Owner/Operator: | CIVCO Medical Instruments Co., Inc. |
| Owner/Operator Number: | 1937223 |
| Contact Person: | Jim Leong |
| Phone: | 319-248-6502 |
| e-mail: | James.Leong@civco.com |
- 2. Date 510(k) Summary Prepared: September 25, 2023
| 3. Name of the Subject Device and Classification Information: | |
|---------------------------------------------------------------|----------------------------------|
| Trade/Device Name | TP Pivot Pro™ Needle Guide |
| Regulation Number | 21 CFR 892.1570 |
| Classification Name | Diagnostic ultrasonic transducer |
| Regulatory Class | Class II |
| Product Code | ITX |
- 4. Information for the Predicate Device:
| Tradename/Device Name | PrecisionPoint Biopsy Needle Guide |
|-----------------------|------------------------------------|
| Manufacturer | Corbin Clinical Resources, LLC |
| 510(k) Number | K160414 |
| Regulation Number | 892.1570 |
| Classification Name | Diagnostic ultrasonic transducer |
| Regulatory Class | Class II |
| Product Code | ITX |
{6}------------------------------------------------
#### 5. Device Description:
The TP Pivot Pro™ needle guide is a tool for performing freehanded transperineal prostate biopsies, taking advantage of the transperineal path to sample regions of the prostate including the difficult to reach anterior zone. Used in conjunction with an ultrasound probe, users visualize suspect target areas of the prostate, plan and position an access site, and obtain specimens from a precise point in the prostate.
#### 6. Intended Use / Indications for Use:
Device is intended to be used for instrument placement to access anatomical structures under transrectal ultrasound guidance.
- . Prostate - Biopsy and minimally invasive puncture
- . Surgical (Prostate) - Biopsy and minimally invasive puncture
#### 7. Comparison to Legally Marketed Device
| Item | Subject Device<br>TP Pivot Pro™ Needle Guide<br>K233109 | Predicate Device<br>PrecisionPoint Needle<br>Guide K160414 |
|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Material | ABS Thermoplastic | Polymeric material |
| | Introducer needle – 304<br>stainless steel | Introducer needle –<br>304 stainless steel |
| Biocompatibility | Meets ISO 10993-1<br>biocompatibility<br>requirements for limited<br>contact duration:<br>• surface devices of<br>breached or<br>compromised surface<br>• externally<br>communicating<br>tissue/bone/dentin | Meets ISO 10993-1<br>biocompatibility<br>requirements for limited<br>contact duration:<br>• surface devices of<br>breached or<br>compromised surface |
| Sterilization | Ethylene Oxide | Ethylene Oxide |
| Shelf-life | 1 year | 3 years |
| Accessory | Optional NeoGuard transducer<br>cover (K013721) | Users informed to use<br>transducer cover |
{7}------------------------------------------------
- 8. Comparison of Indications to the Legally Marketed Device:
The TP Pivot Pro™ needle guide has the same basic intended use as the predicate, which is to place needle into and anatomical structure using an ultrasound transducer. CIVCO has further added the following indications for use for the TP Pivot Pro™ needle guide which is not specifically called out in the predicate.
- · Prostate Biopsy and minimally invasive puncture
- · Surgical (Prostate) Biopsy and minimally invasive puncture
#### 9. Summary of Non-Clinical Tests Performed:
- . Biocompatibility:
The TP Pivot Pro™ needle guide and accessories are manufactured with materials which meet ISO 10993-1 biocompatibility requirements for limited contact duration for surface devices of breached or compromised surface. These same materials are already utilized on CIVCO cleared medical devices for contact in the same regions.
- Cytotoxicity testing in accordance with ISO 10993-5 O
- Sensitization testing in accordance with ISO 10993-10 O
- Irritation testing in accordance with ISO 10993-10 O
- Acute Systemic Toxicity testing in accordance with ISO 10993-11 O
- Material Mediated Pyrogen testing in accordance with ISO 10993-11 O
- Design specification: ●
The TP Pivot Pro™ needle guide and accessories have certain design functionality which was tested internally to verify they meet the established requirements. These criteria are listed as follows:
- Guide Tangential Clocking force testing This test confirms that the guide O adequately attaches to the round probe such that it, once clamped, would not rotate around the probe in a typical use.
- O Guide Axial Loading force testing - This test confirms that the guide adequately attaches to the round probe such that it, once clamped, would not slide forward in a typical use.
- Guide Dislodge Force Test This test confirms that the guide adequately O attaches to the round probe such that it, once clamped, would not dislodge when a force is applied to the worst-case location that is furthest from the areas where the guide attaches to the probe.
- Needle Holder Force (horizontal) Test This test confirms that the force to o dislodge the needle holder from a clamped horizontal position shall be at the worst case position which is furthest from where the needle holder attaches to the guide.
- Needle Holder Force (angled) Test This test confirms that the force to o dislodge needle holder from a clamped, angled position of approximately 20° shall and applied at the worst case position which is furthest from where the holder attaches to the guide.
{8}------------------------------------------------
- Assemble/Removal Force with Neoguard Cover Test This tests the force o required to assemble the guide onto or remove the guide from the simulated probe with a Neoguard cover.
- Guide Clamp Knob Torque Test This test confirms that the guide is not O damaged when the user applies torque to the guide clamp knob to attach probe.
- Extract/Insert Needle Holder in Tower Force The force to extract/insert о the needle holder from/into the guide tower.
- Introducer needle initial insertion force The initial insertion force of the O introducer needle into the needle holder.
- Introducer needle In and out of the needle holder force test The force to O move the introducer needle into or out of the needle holder.
- Simulated Usability Testing ●
- Simulated use evaluations were performed by customers to ensure the design of the needle guide conforms to the user needs and intended use.
- Guidance Standards
- The introducer needle was evaluated against BS EN ISO 9626:1995 and believe it complies with the standard.
#### 10. Clinical Test Performed:
Clinical tests were not required to demonstrate substantial equivalence.
#### 11. Conclusions:
The TP Pivot Pro™ needle guide has the same intended use as compared to the legally marketed device. The design differences do not additional risk to the use of the Therefore, the TP Pivot Pro™ needle guide is substantially equivalent to the device. legally marketed PrecisionPoint needle guide marketed by Corbin Clinical Resources, LLC.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.