Revolution External Plating System

K233025 · New Standard Device Dba Metalogix · KTT · May 30, 2024 · Orthopedic

Device Facts

Record IDK233025
Device NameRevolution External Plating System
ApplicantNew Standard Device Dba Metalogix
Product CodeKTT · Orthopedic
Decision DateMay 30, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The Revolution External Plating System is indicated for treatment of a variety of broken or deformed bones: - Stabilizes open and/or unstable fracture of complex proximal and/or distal tibial fractures - · Fusions of the joints and bone (hand, foot, long-bone) - · Correction of bone or soft tissue deformities - · Correction of segmental or non-segmental bone, soft tissue defects or bone loss - · Neutralization of fractures stabilized with limited internal fixation - · Adult and Pediatric subgroups except newborns

Device Story

External open ring fixation system; provides stability for long bone fractures, limb lengthening, and bone deformity correction at a distance from operative focus. Used for intracapsular, intertrochanteric, supracondylar, or condylar long bone fractures; joint fusions; and limb lengthening involving osteotomy. Operated by orthopedic surgeons in clinical/surgical settings. System components include welded tibial block constructs and Z-post components. Provides mechanical stabilization of bone segments to facilitate healing and deformity correction.

Clinical Evidence

Bench testing only. Performance testing included ASTM F1541 testing and torque to failure testing to support design modifications and demonstrate equivalence to predicate and reference devices.

Technological Characteristics

External open ring fixation system. Materials include metallic components. Testing conducted per ASTM F1541. Mechanical stabilization principle. No software or electronic components.

Indications for Use

Indicated for adult and pediatric patients (excluding newborns) requiring stabilization of open/unstable complex proximal/distal tibial fractures, joint/bone fusions (hand, foot, long-bone), correction of bone/soft tissue deformities, correction of segmental/non-segmental bone/soft tissue defects or bone loss, and neutralization of fractures stabilized with limited internal fixation.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION". May 30, 2024 New Standard Device DBA Metalogix % Danielle Besal Principal Consultant MRC Global 9085 E Mineral Circle Suite 110 Centennial, Colorado 80112 Re: K233025 Trade/Device Name: Revolution External Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: KTT Dated: September 22, 2023 Received: September 22, 2023 Dear Danielle Besal: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2}------------------------------------------------ Sincerely, Image /page/2/Picture/3 description: The image shows the name "Lixin Liu-S" in a simple, sans-serif font. The text is black and appears to be against a white background. The name is likely a person's name, with "Lixin Liu" being the first and last name, and "-S" potentially being an initial or suffix. Lixin Liu, PhD Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use Submission Number (if known) K233025 Device Name Revolution External Plating System Indications for Use (Describe) The Revolution External Plating System is indicated for treatment of a variety of broken or deformed bones: - Stabilizes open and/or unstable fracture of complex proximal and/or distal tibial fractures - · Fusions of the joints and bone (hand, foot, long-bone) - · Correction of bone or soft tissue deformities - · Correction of segmental or non-segmental bone, soft tissue defects or bone loss - · Neutralization of fractures stabilized with limited internal fixation - · Adult and Pediatric subgroups except newborns Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ #### 510(k) Summary ## Revolution External Plating System K233025 May 30, 2024 | Company: | New Standard Device dba Metalogix<br>4766 Research Dr<br>San Antonio, TX 78240 | |--------------------|--------------------------------------------------------------------------------------------------------| | Primary Contact: | Danielle Besal<br>Principal Consultant MRC Global<br>901.827.8670<br>Danielle.Besal@askmrcglobal.com | | Company Contact: | Ben Coburn<br>CEO Metalogix<br>833.659.2019 | | Trade Name: | Revolution External Plating System | | Common Name: | External fixation system | | Classification: | Class II | | Regulation Number: | 21 CFR 888.3030 Single/multiple component metallic bone fixation<br>appliances and accessories | | Product Code: | KTT | | Panel: | Orthopedic | | Predicate Device: | K181630 Revolution External Plating System | | Reference Device: | K152171 Orthofix TL-HEX True Lok Hexapod System | #### Device Description: The Revolution External Plating System is an external open ring fixation system to provide stability for long bone fractures, limb lengthening, and correction of bone deformities all at a distance from the operative focus. When used with other components, this device stabilizes open and/or unstable fractures of long bones including intracapsular, intertrochanteric, supracondylar, or condylar. It is also used for joint fusions and limb lengthening of deformity corrections which involve cutting the bone. #### Indications for Use: The Revolution External Plating System is indicated for treatment of a variety of broken or deformed bones: - Stabilizes open and/or unstable fracture of complex proximal and/or distal tibial fractures - Fusions of the joints and bone (hand, foot, long-bone) - Correction of bone or soft tissue deformities - Correction of segmental or non-segmental bone, soft tissue defects or bone loss - Neutralization of fractures stabilized with limited internal fixation - Adult and pediatric subgroups except newborns #### Substantial Equivalence Discussion: The subject Revolution External Plating System is a modification to the predicate Revolution External Plating System (K181630) and includes design and material changes to individual system components and introduction of new welded tibial block constructs, and Z-post components. The subject system is similar in indications for use, materials, and design to the predicate system (K181630). Testing {5}------------------------------------------------ has demonstrated that performance of the subject devices is equivalent to that of the predicate and the reference device, and thus, the differences in geometry versus the predicate do not raise different questions of safety and effectiveness. ### Performance Testing: Performance testing (ASTM F1541, and torque to failure testing) was completed to support the modifications to the system and demonstrate its substantial equivalence to the predicate device. ### Conclusion: The subject Revolution External Plating System is substantially equivalent to the predicate Revolution External Plating System (K181630).
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...