METICULY Patient-specific titanium maxillofacial mesh implant is intended for bone fixation and reconstruction, restoration of bone defects and intended to provide continuity in regions where the bone is missing and/or to augment the bone by means of an onlay device in the maxillofacial skeleton and midface.
Device Story
Patient-specific titanium mesh implant; designed for maxillofacial/midface trauma and reconstruction. Input: patient CT scan data. Transformation: additive manufacturing (Laser Powder Bed Fusion) of titanium alloy (Ti-6Al-4V ELI) based on patient-specific digital design. Output: custom-fit titanium mesh implant. Usage: clinical setting; operated by surgeons. Surgeon approves design prior to fabrication. Output usage: fixation to bone using commercially available titanium screws; provides structural continuity or augmentation. Benefit: patient-specific anatomical fit for non-bending maxillofacial applications.
Clinical Evidence
Bench testing only. Validation included: material composition/powder recycling analysis; density consistency across build plates; dynamic fatigue strength testing; and verification that printed output matches digital design/CT inputs. Biocompatibility, sterilization (moist heat, SAL 10^-6), packaging, and transportation integrity validated per ISO and ASTM standards.
Technological Characteristics
Material: Titanium Ti-6Al-4V ELI (Grade 23) per ASTM F3001-14. Manufacturing: Laser Powder Bed Fusion (additive). Form factor: Patient-specific mesh, 0.4-0.9 mm thickness. Fixation: Non-locking, uses external titanium screws. Sterilization: Non-sterile (steam/moist heat validation provided). Connectivity: None (standalone).
Indications for Use
Indicated for adults (≥22 years) requiring bone fixation, reconstruction, or restoration of bone defects in the maxillofacial skeleton and midface. Not intended for orbital roof defects, frontal bone defects (e.g., supra-orbital ridge), or scenarios requiring bone support/stability under functional loading.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
Predicate Devices
TruMatch CMF Titanium 3D Printed Implant (K173039)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.
April 30, 2024
Meticuly Co., Ltd. % Paweena U-Thainual CEO MDR Solutions Co., Ltd. 1435 Kanjanapisek Road Bang Khae Nuea, Bang Khae Bangkok, 10160 THAILAND
Re: K232889
Trade/Device Name: METICULY Patient-specific titanium maxillofacial mesh implant Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: Class II Product Code: JEY Dated: August 21, 2023 Received: April 1, 2024
Dear Paweena U-Thainual:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Sherrill Lathrop Blitzer
for Andrew Steen Assistant Director DHT1B: Division of Dental and
{2}------------------------------------------------
ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K232889
Device Name
METICULY Patient-specific titanium maxillofacial mesh implant
Indications for Use (Describe)
METICULY Patient-specific titanium maxillofacial mesh implant is intended for bone fixation and reconstruction, restoration of bone defects and intended to provide continuity in regions where the bone is missing and/or to augment the bone by means of an onlay device in the maxillofacial skeleton and midface.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"> <span> <div style="display:inline-block; vertical-align:middle;"> <div style="display:flex; align-items:center;"> <span> <div style="display:inline-block; vertical-align:middle;"> <span> <div style="display:inline-block; vertical-align:middle;"> <span> <div style="display:inline-block; vertical-align:middle;"> <span> <div style="display:inline-block; vertical-align:middle;"> <span> <div style="display:inline-block; vertical-align:middle;"> <span> <div style="display:inline-block; vertical-align:middle;"> <span> <div style="display:inline-block; vertical-align:middle;"> <span> <div style="display:inline-block; vertical-align:middle;"> <span> <div style="display:inline-block; vertical-align:middle;"> <span> <div style="display:inline-block; vertical-align:middle;"> <span> <div style="display:inline-block; vertical-align:middle;"> <span> <div style="display:inline-block; vertical-align:middle;"> <span> <div style="display:inline-block; vertical-align:middle;"> <span> <div style="display:inline-block; vertical-align:middle;"> <span> <div style="display:inline-block; vertical-align:middle;"> <span> <div style="display:inline-block; vertical-align:middle;"> <span> <div style="display:inline-block; vertical-align:middle;"> <span> <div style="display:inline-block; vertical-align:middle;"> <span> <div style="display:inline-block; vertical-align:middle;"> <span> <div style="display:inline-block; vertical-align:middle;"> <span style="font-size:16px;">☑</span> </div> </span> </div> </span> </div> </span> </div> </span> </div> </span> </div> </span> </div> </span> </div> </span> </div> </span> </div> </span> </div> </span> </div> </span> </div> </span> </div> </span> </div> </span> </div> </span> </div> </span> </div> </span> </div> </span> </div> </span> </div> </div> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div style="display:flex; align-items:center;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{4}------------------------------------------------
# 510(k) Summary – K232889
The following information is provided in accordance with 21 CFR 807.92 for the Premarket 510(k) Summary of the METICULY Patient-specific titanium maxillofacial mesh implant:
# 1. Submitter Information
| Company/Applicant: | Mrs. Peeranoot Lohwongwatana<br>Managing Director<br>Meticuly Co., Ltd.<br>924 B bldg., Rm. B116-118, B210-212<br>Soi Chula 7, Wang Mai, Pathum Wan Bangkok, Thailand 10330<br>Telephone: +6688-834-7777<br>Email: peeranoot@meticuly.com |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Paweena U-Thainual, PhD<br>MDR Solutions Co., Ltd.<br>1435, Kanjanapisek Rd., Bang Khae Nuea<br>Bang Khae, Bangkok 10160 Thailand<br>Email: paweena@mdrsolutions.co.th |
| Date Summary Prepared: | April 30, 2024 |
| 2. Device Name | |
| Trade Name: | METICULY Patient-specific titanium maxillofacial mesh implant |
| Common Name: | Plate, Bone |
| Classification Name: | Bone Plate |
| Review Panel: | Dental |
| Regulation: | 872.4760 |
| Class: | Class II |
| Product Code: | IFY |
{5}------------------------------------------------
# 3. Predicate Device
METICULY Patient-specific titanium maxillofacial mesh implant is substantially equivalent to the following legally marketed predicate devices.
Table 1.1 Primary Predicate device
| Applicant | Device Name | 510(k) Number |
|----------------|------------------------------------------|---------------|
| Materialise NV | TruMatch CMF Titanium 3D Printed Implant | K173039 |
Table 1.2 Reference devices*
| Applicant | Device Name | 510(k) Number |
|--------------------------|-------------------------------------------------|---------------|
| KLS-Martin L.P. | KLS Martin Individual Patient Solutions | K191028 |
| Meticuly Co., Ltd. | METICULY Patient-specific Titanium Mesh Implant | K210099 |
| Jeil Medical Corporation | LeForte System Bone Plate & Screw | K112457 |
*These devices are referred to support the similar mesh design and the use of similar material and manufacturing for the subject device.
## 4. Description
The METICUL Y Patient-specific titanium maxillofacial mesh implant is a device designed individually for each patient and intended for use in selective trauma of the maxillofacial skeleton, maxillofacial surgery, and reconstructive procedures. It is specifically designed with a focus on applications of non-bending related scenarios. The implant is made of titanium alloys produced via additive manufacturing (Laser Powder Bed Fusion) and is intended to be used with titanium screws. All additive manufacturing and other post-processing steps are only to take place under Meticuly manufacturing control. The device is not intended to substitute for bone reconstruction in clinical situations where bone is needed for support and stability of the maxillofacial skeleton under functional loading conditions. It is intended for adults only (at least 22 years of age). The surgeon approves the design of the mesh implant prior to fabrication of the implant device by the sponsor. The proposed FaciMesh models are intended to treat the maxillofacial region, while the OrbiMesh models treat the Orbital region. The device is not intended for reconstruction of the orbital roof defects or for any frontal bone defects, such as the supra-orbital ridge. Additionally, the FDA-cleared commercially available titanium screws that can be used with the subject devices include the TITANIUM MINIPLATE SYSTEM (K951690) and MICRO TITANIUM PLATE SYSTEM (K951688).
#### 5. Indications for Use
METICULY Patient-specific titanium maxillofacial mesh implant is intended for bone fixation and reconstruction. restoration of bone defects and intended to provide continuity in regions where the bone is missing and/or to augment the bone by means of an onlay device in the maxillofacial skeleton and midface.
{6}------------------------------------------------
## 6. Comparison of Technological Characteristics with the Predicate Device
The subject device is substantially equivalent to the following legally marketed predicate devices. METICUL Y Patient-specific titanium maxillofacial mesh implant and the predicate devices have the similar characteristics, for example, indication for use, intended use, sterilization method, manufacturing method, device design, and device performance. The differences of these characteristics have been addressed with the provided performance test data in this submission and do not raise different questions of safety and effectiveness.
| Table 2: Technical Characteristics in Comparison to the Predicate and Reference Devices | | | | | |
|-----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>comparison | Subject Device:<br>METICULY<br>Patient-specific<br>titanium<br>maxillofacial mesh<br>implant | Predicate Device:<br>TruMatch CMF<br>Titanium 3D Printed<br>Implant | Reference Device:<br>LeForte System<br>Bone Plate &<br>Screw | Reference Device:<br>METICULY<br>Patient-specific<br>titanium mesh<br>implant | Reference Device:<br>KLS Martin<br>Individual Patient<br>Solutions |
| 510(K) number | K232889 | K173039 | K112457 | K210099 | K191028 |
| Product<br>Code(s) | JEY | JEY | JEY | GXN | JEY |
| Classification | Class II | Class II | Class II | Class II | Class II |
| Indications for<br>use | METICULY<br>Patient-specific<br>titanium<br>maxillofacial mesh<br>implant is intended<br>for bone fixation and<br>reconstruction,<br>restoration of bone<br>defects and intended<br>to provide continuity<br>in regions where the<br>bone is missing<br>and/or to augment<br>the bone by means<br>of an onlay device in<br>the maxillofacial<br>skeleton and<br>midface. | The TruMatch CMF<br>Titanium 3D Printed<br>Implant is a patient<br>specific implant and<br>is intended for bone<br>fixation and<br>reconstruction,<br>restoration of bone<br>defects and intended<br>to provide continuity<br>in regions where the<br>bone is missing<br>and/or<br>to augment the bone<br>by means of an onlay<br>device in the<br>maxillofacial<br>skeleton, midface<br>and chin. | This device is<br>intended for use in<br>selective trauma of<br>the mid-face,<br>reconstruction<br>procedures and<br>selective<br>orthognathic<br>surgery of the<br>maxilla and chin. | Meticuly Patient-<br>specific Titanium<br>Mesh Implant is a<br>device that is<br>designed<br>individually for each<br>patient. This device<br>is intended for use<br>in selective trauma<br>of the cranial and<br>craniofacial skeleton<br>(e.g., frontal bone,<br>temporal bone,<br>occipital bone,<br>parietal bone,<br>sphenoid bone,<br>supraorbital process,<br>vomer), cranial and<br>craniofacial surgery,<br>and reconstructive<br>procedures. | KLS Martin<br>Individual Patient<br>Solutions implant<br>devices are intended<br>for use in the<br>stabilization,<br>fixation, and<br>reconstruction of the<br>maxillofacial /<br>midface and<br>mandibular skeletal<br>regions. |
| Material(s) | Titanium Ti-6Al-4V<br>ELI (Grade23) | Commercially pure<br>titanium | pure Titanium<br>(ASTM F67 Grades<br>1,2, and 3) | Titanium Ti6Al4V<br>ELI (Grade23) | Titanium (ASTM<br>F67) or Ti-6Al-4V<br>(ASTM F136) |
| Technical<br>Specifications | Custom-made to<br>each patient using<br>CT data | Custom-made to<br>each patient using<br>CT data | Commercially sized | Custom-made to<br>each patient using<br>CT data | Custom-made to<br>each patient using<br>CT data |
| Table 2: Technical Characteristics in Comparison to the Predicate and Reference Devices | | | | | |
| Device<br>comparison | Subject Device:<br>METICULY<br>Patient-specific<br>titanium<br>maxillofacial mesh<br>implant | Predicate Device:<br>TruMatch CMF<br>Titanium 3D Printed<br>Implant | Reference Device:<br>LeForte System<br>Bone Plate &<br>Screw | Reference Device:<br>METICULY<br>Patient-specific<br>titanium mesh<br>implant | Reference Device:<br>KLS Martin<br>Individual Patient<br>Solutions |
| Manufacturing<br>Method | 3D printed using<br>laser powder bed<br>fusion additive<br>manufacturing | Additive<br>manufacturing,<br>Selective Laser<br>Melting | Machined and<br>surface treated by<br>anodization | 3D printed using<br>laser powder bed<br>fusion additive<br>manufacturing | Epoxy/ Acrylic<br>Resins:<br>Stereolithography<br>(SLA)<br>CP Titanium:<br>Traditional(Subtract<br>ive)<br>Ti-6A1-4V:<br>Traditional and 3D<br>(Additive; Selective<br>Laser Melting) |
| Fixation<br>Method | Commercially<br>available titanium<br>screws systems | Commercially<br>available titanium<br>screws system<br>(Synthes) | Own plate and<br>screw system | Commercially<br>available titanium<br>screws systems | Own plate and<br>screw system |
| Sterilization | Non-sterile | Non-sterile | Non-sterile | Non-sterile | Non-sterile |
| Plate<br>Dimension | OrbiMesh<br>Length: 10-100 mm<br>Width: 10-90 mm<br>Height: 0.4-125 mm<br>FaciMesh<br>Length: 10-120 mm<br>Width: 10-100 mm<br>Height: 0.4-125 mm | Length: 10-294mm<br>Width: Unknown<br>Height: Unknown | Length: 5.2-223.5<br>mm<br>Width: Unknown<br>Height: 4.2 - 46.8<br>mm | Length: 20 - 105<br>mm<br>Width: 20 - 105 mm<br>Height: 0.4 - 125<br>mm | Length: 18-350 mm<br>Width:<br>Min: $\geq$ 4.5 mm<br>(around<br>screw holes)<br>Min: $\geq$ 3 mm<br>(not around<br>screw hole)<br>Max: Dependent on<br>screw-hole<br>Height: Unknown |
| Overall<br>Thickness | OrbiMesh<br>0.4 – 0.9 mm<br>FaciMesh<br>0.4 - 0.9 mm | 0.4—10 mm | 0.2-2.5 mm | 0.4-0.9 mm | Orbital:<br>0.3-1.0 mm<br>Maxillofacial /<br>Midface:<br>0.6-10.0 mm |
| Internal angle<br>at edge<br>boundary | 45-180 deg | 0-12 deg/mm length | Unknown | 45-180 deg | In Plane:<br>30-180 deg<br>Out of Plane:<br>15-180 deg |
| Plate Holes | Non-locking | Unknown | Non-locking | Non-locking | Locking threads<br>optional |
| Screw<br>Diameter | 1.5-2.3 mm | Unknown | 0.7-1.6 mm | 1.4-1.8 mm | 1.5-2.3 mm |
| | | | | | |
| Device<br>comparison | Subject Device:<br>METICULY<br>Patient-specific<br>titanium<br>maxillofacial mesh<br>implant | Predicate Device:<br>TruMatch CMF<br>Titanium 3D Printed<br>Implant | Reference Device:<br>LeForte System<br>Bone Plate &<br>Screw | Reference Device:<br>METICULY<br>Patient-specific<br>titanium mesh<br>implant | Reference Device:<br>KLS Martin<br>Individual Patient<br>Solutions |
| Screw Length | 4.0-7.0 mm | Unknown | 4.0-18.0 mm | 3.0-4.0 mm | 3.5-22 mm |
{7}------------------------------------------------
{8}------------------------------------------------
#### 7. Performance Tests
The submission provided validation information to support the use of the leftover powder material recycling/reuse between print runs based on representative testing that evaluated maintained composition, particle size, flowability, density, and visual characteristics within the same acceptance criteria ranges as virgin powder. Testing has been provided to show that the density of printed samples representative of the subject devices is consistently within the same acceptance criteria throughout a comprehensive array of build plate locations. Dynamic fatigue strength testing of the subject device to account for the worst-case consideration of print orientation. Validation testing was provided to demonstrate the printed patient-matched device form output consistently matches the digital design and CT scan data inputs under the representative manufacturing conditions across a repeatable process. According to the results, the METICULY Patient-specific titanium maxillofacial mesh implant is considered substantially equivalent to the predicate device.
Materials and manufacturing method quality of METICULY Patient-specific titanium maxillofacial mesh implant were assessed through physical properties and mechanical properties. The device testing was designed to validate the manufacturing process and to ensure that the subject device complicable voluntary consensus standards for biocompatibility, packaging, transportation. The device was validated using the Overkill method for steam sterilization with moist heat at 135°C for 10 minutes to achieve a SAL of 10-6.
Verification and validation testing confirms that the product specifications have been met, demonstrating that the device will perform as intended. There were no unexpected results which indicate the suitable material used and manufacturing process compared to the standards for medical devices.
| Table 3: Testing and compliance standards summary table | |
|---------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Test | Standard (FDA recognition number) |
| Materials and manufacturing method | ASTM F3001-14 (8-439)<br>Standard Specification for Additive Manufacturing<br>Titanium-6 Aluminum-4 Vanadium ELI (Extra Low<br>Interstitial) with Powder Bed Fusion |
| Biological evaluation and Biocompatibility | ISO 10993-1 Fifth edition 2018-08 (2-258)<br>Biological evaluation of medical devices - Part 1:<br>Evaluation and testing within a risk management process<br>ISO 10993-3 Third edition 2014-10-1 (2-228) |
| Table 3: Testing and compliance standards summary table | |
| Test | Standard (FDA recognition number) |
| | Biological evaluation of medical devices - Part 3:Tests<br>for genotoxicity carcinogenicity and reproductive<br>toxicity |
| | ISO 10993-5 Third edition 2009-06-01 (2-245)<br>Biological evaluation of medical devices - Part 5:Tests<br>for in vitro cytotoxicity |
| | ISO 10993-6 Third edition 2016-12-01 (2-247)<br>Biological evaluation of medical devices -- Part 6:<br>Tests for local effects after implantation |
| | ISO 10993-10 Third Edition 2010-08-01 (2-174)<br>Biological evaluation of medical devices - Part 10:Tests<br>for irritation and skin sensitization |
| | ISO 10993-11 Third edition 2017-09 (2-255)<br>Biological evaluation of medical devices - Part 11:<br>Tests for systemic toxicity |
| | ISO 10993-23 First Edition 2021-01 (2-291)<br>Biological evaluation of medical devices - Part 10:Tests<br>for irritation |
| | USP-NF M98900_01_01 (2-295)<br><151> Pyrogen Test (USP Rabbit Test) |
| Sterilization process control and validation | ANSI AAMI ST72:2019 (14-541)<br>Bacterial endotoxins - Test methods routine<br>monitoring and alternatives to batch testing |
| | USP-NF M98830_02_01 <85> Bacterial<br>Endotoxins Test (14-570) |
| | USP-NF M98910_01_01 <161> Medical Devices-<br>Bacterial Endotoxin and Pyrogen Tests (14-564) |
| | USP-NF M98810_01_01 <71> Sterility Tests (14-569) |
| | ISO 17665-1 First edition 2006-08-15 (14-333)<br>Sterilization of health care products - Moist heat - Part1:<br>Requirements for the development validation and routine<br>control of a sterilization process for medical devices |
| Table 3: Testing and compliance standards summary table | |
| Test | Standard (FDA recognition number) |
| | ISO 11737-1 Third edition 2018-01 (14-514)<br>Sterilization of health care products - Microbiological<br>methods - Part 1: Determination of a population of<br>microorganisms on product |
| | ISO 11737-2 Third edition 2019-12 (14-540)<br>Sterilization of medical devices - Microbiological<br>methods - Part 2: Tests of sterility performed in the<br>definition validation and maintenance of a sterilization<br>Process |
| | ISO 19227 First edition 2018-03 (11-348)<br>Implants for surgery - Cleanliness of orthopedic implants<br>- General requirements |
| Packaging and transportation control and<br>validation | ASTM F88/F88M-15 (14-482)<br>Standard Test Method for Seal Strength of Flexible<br>Barrier Materials |
| | ASTM D7386-16 (5-113)<br>Standard Practice for Performance Testing of Packages<br>for Single Delivery Systems |
| | F1886/F1886M-16 (14-501)<br>Standard Test Method for Determining Integrity of Seals<br>for Flexible Packaging by Visual Inspection |
| | F1929-15 (14-484)<br>Standard Test Method for Detecting Seal Leaks in<br>Porous Medical Packaging by Dye Penetration |
{9}------------------------------------------------
{10}------------------------------------------------
#### 8. Conclusion
Based upon testing and comparison to the predicate device, the METICULY Patient-specific titanium mesh implant has the same intended use and similar technological characteristics. Thus, the subject device is concluded to be substantially equivalent to the legally commercialized predicate devices for the purposes of this 510(k) submission.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.