K232780 · Cc-Instruments, Inc. · HTN · Dec 18, 2023 · Orthopedic
Device Facts
Record ID
K232780
Device Name
CC-Clip® Implant System
Applicant
Cc-Instruments, Inc.
Product Code
HTN · Orthopedic
Decision Date
Dec 18, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CC-Clip® device is intended as an adjunct in coracoclavicular ligament disruption repair. Specifically, the CC-Clip® device is intended to provide fixation during the healing process following a syndesmotic trauma, such as fixation of acromioclavicular separations due to coracoclavicular ligament disruptions.
Device Story
CC-Clip® is an implantable system for arthroscopic coracoclavicular (CC) and acromioclavicular (AC) ligament reconstruction. The system comprises a titanium implant for the clavicle, a titanium counterpart for the coracoid, and non-resorbable suture (e.g., Arthrex #5 FiberWire) connecting them. The device is used by surgeons in an OR setting to provide fixation during the healing process following AC joint separation, which causes superior and posterior clavicle translation. The system includes reusable surgical instruments (Straight and Curved Lasso guides) and requires user-supplied general instrumentation (lasso wire, guide pins, drill bits). The implants are single-use and sterile; instruments are non-sterile and require steam sterilization. The surgeon uses the device to secure the clavicle to the coracoid, restoring joint stability and facilitating ligament healing.
Clinical Evidence
No clinical data was conducted. Substantial equivalence is supported by non-clinical bench testing, including static (ultimate load) and dynamic mechanical testing, which met all performance requirements and acceptance criteria.
Technological Characteristics
Implants: Titanium alloy (ASTM F136). Instruments: Stainless steel. System includes two titanium buttons and non-resorbable suture. Implants are single-use, gamma-sterilized. Instruments are reusable, steam-sterilized. Dimensions: Clavicle clips (16mm, 20mm); Coracoid clips (10mm, 12mm round/rectangular).
Indications for Use
Indicated for patients requiring fixation during healing of syndesmotic trauma, specifically acromioclavicular separations due to coracoclavicular ligament disruptions.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The Department of Health & Human Services logo is a stylized representation of a human figure.
June 28, 2024
CC-Instruments Inc % Mason Makulinski Project Engineer Jalex Medical 27865 Clemens Rd Suite 3 Westlake, Ohio 44145
Re: K232780
Trade/Device Name: CC-Clip® Implant System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HTN
Dear Mason Makulinski:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change for your device cleared on December 18, 2023. Specifically, FDA is updating this SE Letter because FDA inadvertently indicated that the SE determination also included review and clearance of a predetermined change control plan (PCCP). However, your 510(k) submission did not include a PCCP, so FDA is providing this administrative correction. Please see the attached revised clearance letter.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Limin Sun, OHT6: Office of Orthopedic Devices, 301-796-7056, Iimin.sun@fda.hhs.gov.
Sincerelv.
# Christopher Ferreira -S
for
Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters FDA in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 28, 2024
CC-Instruments Inc % Mason Makulinski Project Engineer Jalex Medical 27865 Clemens Rd Suite 3 Westlake, Ohio 44145
Re: K232780
Trade/Device Name: CC-Clip® Implant System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HTN Dated: September 11, 2023 Received: September 11, 2023
Dear Mason Makulinski:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change for your device cleared on December 18, 2023.
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{2}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{3}------------------------------------------------
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Christopher Ferreira -S
for
Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use
510(k) Number (if known)
#### K232780
Device Name CC-Clip® Implant System
#### Indications for Use (Describe)
The CC-Clip® device is intended as an adjunct in coracoclavicular ligament disruption repair.
Specifically, the CC-Clip® device is intended to provide fixation during the healing process following a syndesmotic trauma, such as fixation of acromioclavicular separations due to coracoclavicular ligament disruptions.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a metal clip to the left of the text "CC-CLIP". The clip is silver and has a u-shape. The text is in a sans-serif font and is also silver. The background is white.
Page 1 of 4 510(k) Submission - CC-Clip® Implant System
K232780
## 510(k) Summary
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Safety and Effectiveness for the use of the CC-Clip® Implant System.
| Submitted By: | CC-Instruments Inc.<br>850 New Burton Road, Suite 201<br>Dover, Delaware 19904 |
|-------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| Date: | 18DEC2023 |
| Contact Person:<br>Contact Telephone:<br>Contact Fax: | Mason Makulinski, Project Engineer<br>(440) 242-9127<br>(440) 933-7839 |
| Device Trade Name: | CC-Clip® Implant System |
| Common Name:<br>Device Classification Name: | Button/Suture<br>21 CFR 888.3030, Single/multiple bone fixation appliances and<br>accessories |
| Device Classification:<br>Reviewing Panel:<br>Product Code: | Class II<br>Orthopedic<br>HTN |
| Primary Predicate Device: | Arthrex TightRope® Acromioclavicular (AC) device (K052776) |
#### Device Description:
CC-Clip® is an implantable device for an arthroscopic coracoclavicular (CC) ligament reconstruction and acromioclavicular (AC) ligament reconstruction technique. The original ligaments fail due to an AC joint separation and cause superior and posterior translation of the clavicle.
CC-Clip® device consists of a titanium implant on the clavicle, a titanium counterpart under the coracoid, and non-resorbable suture between the implants. A USP size 5 suture is not provided with CC-Clip®, but the Arthrex #5 FiberWire is recommended with the CC-Clip® device.
Instrumentation included with the CC-Clip® system includes two instruments: the CC-Clip® Straight Lasso guide and CC-Clip® Curved Lasso guide. Additional general-use, readily available instrumentation required for the CC-Clip® procedure includes lasso wire, guide pins, and drill bits which are to be supplied by the end user per the surgical technique. Both CC-Clip® guides are re-usable surgical instruments provided non-sterile to the user. The CC-Clip® implants are supplied sterile to the user and are single use.
#### Indications for Use:
The CC-Clip® device is intended as an adjunct in coracoclavicular ligament disruption repair.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a metal clip on the left and the text "CC-CLIP" on the right. The clip is a small, U-shaped piece of metal with a flat top. The text is in a sans-serif font and is gray in color. The text is slightly larger than the clip.
K232780 Page 2 of 4
Specifically, the CC-Clip® device is intended to provide fixation during the healing process following a syndesmotic trauma, such as fixation of acromioclavicular separations due to coracoclavicular ligament disruptions.
#### Summary of Technological Characteristics:
The CC-Clip® Implant System is similar in its intended use to the predicate device, Arthrex TightRope® Acromioclavicular (AC) device System. The proposed devices are composed of two buttons that are fabricated from titanium alloy per ASTM F136, and the design features for the implant system are similar to the predicate device, including dimensions, shape, and sizes.
#### Performance Data - Nonclinical:
Substantial equivalence is supported by the results of mechanical testing including engineering analysis of static (ultimate load) and dynamic testing.
| Table 6.1. Mechanical Testing Program for CC-Clip Implant System | |
|------------------------------------------------------------------|------------|
| Verification Method | Results |
| Static Testing | Acceptable |
| Dynamic Testing | Acceptable |
### oblo 6 1, Mochanicol Tosting Program for CC Clip® Implont System
Testing performed on the CC-Clip® Implant System demonstrated that all performance-based requirements and acceptance criteria were met to support safety and effectiveness and is considered substantially equivalent to the predicate device.
#### Performance Data - Clinical:
Clinical testing is not required and therefore not conducted as part of this submission.
#### Substantial Equivalence:
The CC-Clip® Implant System has the same intended use and fundamental scientific technology as the Arthrex TightRope® Acromioclavicular (AC) device (K052776) primary predicate device. Both devices compare similarly in:
- Design features and instrumentation
- Intended use and indications for use
- Material composition
- Dimensions
- Function ●
- . Sterilization methods
Refer to Table 8.2 below for additional information. The potential impact on substantial equivalence of each technological difference was addressed through risk analysis and verification and validation testing. Any differences between the CC-Clip® Implant System and the predicate device do not raise any questions of safety and effectiveness.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a metal clip on the left and the text "CC-CLIP" on the right. The clip is a small, U-shaped piece of metal with a flat top. The text is in a sans-serif font and is gray in color. The text is slightly larger than the clip.
## 510(k) Submission – CC-Clip® Implant System
K232780 Page 3 of 4
The table below summarizes the features of the subject device and its equivalence to the predicate device.
| Table 8.2 Subject and Predicate Device Comparison | | |
|---------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Subject Device:<br>CC-Clip® Implant System | Predicate Device:<br>Arthrex TightRope®<br>Acromioclavicular (AC) device |
| Regulation Name | Single/multiple component metallic<br>bone fixation appliances and<br>accessories | Single/multiple component metallic<br>bone fixation appliances and<br>accessories |
| Regulation<br>Number | 21 CFR 888.3030 | 21 CFR 888.3030 |
| Product Code | HTN | HTN |
| Device<br>Description | CC-Clip® is an implantable device for<br>an arthroscopic coracoclavicular (CC)<br>ligament reconstruction and<br>acromioclavicular (AC) ligament<br>reconstruction technique. The original<br>ligaments fail due to an AC joint<br>separation and cause superior and<br>posterior translation of the clavicle.<br><br>CC-Clip® device consists of a titanium<br>implant on the clavicle, a titanium<br>counterpart under the coracoid, and<br>non-resorbable suture between the<br>implants. Suture is not provided with<br>CC-Clip®, but the Arthrex #5<br>FiberWire is recommended with the<br>CC-Clip® device.<br><br>Instrumentation included with the CC-<br>Clip® system includes two<br>instruments: the CC-Clip® Straight<br>Lasso guide and CC-Clip® Curved<br>Lasso guide. Additional general-use,<br>readily available instrumentation<br>required for the CC-Clip® procedure<br>includes lasso wire, guide pins, and<br>drill bits which are to be supplied by<br>the end user per the surgical<br>technique. Both CC-Clip® guides are<br>re-usable surgical instruments<br>provided non-sterile to the user. The<br>CC-Clip® implants are supplied sterile<br>to the user and are single use. | The TightRope™ Acromioclavicular<br>(AC) Device is designed as two<br>differently sized metal buttons, both<br>stainless steel or both titanium, and<br>FiberWire™ suture. The buttons are<br>pre-threaded with FiberWire Suture,<br>looped twice through the buttonholes.<br>Pull-Through FiberWire sutures are<br>also looped through each button.<br><br>The TightRope™ Acromioclavicular<br>(AC) Device is intended as an adjunct<br>in fracture repair involving<br>metaphyseal and periarticular small<br>bone fragments where screws are not<br>indicated, and as an adjunct in external<br>and intramedullary fixation systems<br>involving plates and rods, with fracture<br>braces and casting. Specifically, the<br>Arthrex TightRope Acromioclavicular<br>(AC) Device is intended to provide<br>fixation during the healing process<br>following a syndesmotic trauma, such<br>as fixation of acromioclavicular<br>separations due to coracoclavicular<br>ligament disruptions. |
| Indications for<br>Use | The CC-Clip® device is intended as<br>an adjunct in coracoclavicular<br>ligament disruption repair. | The TightRope™ Acromioclavicular<br>(AC) Device is intended as an adjunct<br>in fracture repair involving<br>metaphyseal and periarticular small bone |
| | Specifically, the CC-Clip® device is intended to provide fixation during the healing process following a syndesmotic trauma, such as fixation of acromioclavicular separations due to coracoclavicular ligament disruptions. | fragments where screws are not indicated and as an adjunct in external and intramedullary fixation systems involving plates and rods, with fracture braces and casting.<br>Specifically, the Arthrex TightRope Acromioclavicular (AC) Device is intended to provide fixation during syndesmotic trauma, such as fixation of acriomioclavicualr separations due to coracoclavicular ligament disruptions. |
| Device Model | Image: CC-Clip Device Model | Image: Arthrex TightRope Acromioclavicular (AC) Device |
| Material | Implants: Titanium Alloy<br>Instruments: Stainless Steel | Implants: Titanium Alloy and stainless steel |
| Sterilization | Implants: Gamma Radiation<br>Instruments: End User cleaning and steam sterilization | Unknown |
| Clavicle Clip Sizes | 16mm and 20mm Length | Unknown |
| Coracoid Clip Sizes | 10mm and 12mm Round<br>10mm and 12mm Rectangular | Unknown |
| Suture Size Indication | #5 Suture | #5 Suture |
## Table 8,2 Subject and Predicate Device Comparison
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the words "CC-CLIP" in a sans-serif font. To the left of the text is a diagram of a metal clip. The text and diagram are in gray. The image is simple and clear.
## 510(k) Submission – CC-Clip® Implant System
#### Conclusion:
The CC-Clip® Implant System is as safe and effective as its predicate device, as it has the same intended use, indication for use, material composition, design features, manufacturing method, sterilization method, performance requirements and principles of operation as the primary predicate device (K052776). The design differences do not alter the intended use of the device and do not affect its safety and effectiveness when used as labeled. Non-clinical testing demonstrates that the CC-Clip® Implant System is substantially equivalent to the primary predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.