K232536 · Embolx, Inc. · DQO · Feb 23, 2024 · Cardiovascular
Device Facts
Record ID
K232536
Device Name
Soldier Microcatheter
Applicant
Embolx, Inc.
Product Code
DQO · Cardiovascular
Decision Date
Feb 23, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Soldier Microcatheter is indicated for use in the blood vessels of the peripheral vasculature. It is intended to assist in the delivery of diagnostic agents and therapeutic agents into the target treatment area. The Soldier Microcatheter is intended for Prescription Use Only. The Soldier is intended for the infusion of contrast media into all peripheral vessels up to and including the cervical vessels, all vessels in the lower and upper extremities. The Soldier is also intended for drug infusion in intra-arterial therapy and the infusion of embolic materials for hemostasis. The Soldier is contraindicated for use in the cerebral and coronary vessels. It is not intended for use in embolectomy or angioplasty procedures.
Device Story
Soldier Microcatheter is a single-lumen, disposable device designed to access small, tortuous peripheral vasculature. It features a metal coil-reinforced multi-layer polymer tube with a hydrophilic coating for lubricity and a PTFE-lined inner lumen to facilitate guidewire movement. The distal tip is radiopaque for fluoroscopic visualization. Operated by physicians in clinical settings, the device is used to deliver diagnostic agents (contrast media) and therapeutic agents (drugs, embolic materials) into target treatment areas. The device is provided sterile (EtO) and is compatible with guidewires up to 0.018". It does not include accessories. By providing a trackable, lubricious pathway, the catheter enables precise navigation to peripheral sites, facilitating targeted therapy and diagnostic imaging.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included biocompatibility (ISO 10993-1), visual inspection, dimensional verification, hub integrity, burst/leakage pressure, flow, tensile strength, torsion, bend/buckle, radiopacity, particulate testing, and simulated use in peripheral vasculature. Sterilization validated per ISO 10993-7. Shelf-life validated to 12 months per ASTM F1980.
Technological Characteristics
Single-lumen catheter; metal coil-reinforced multi-layer polymer tubing; hydrophilic coating; PTFE-lined inner lumen; radiopaque distal tip. Sizes: 2.0 Fr and 2.5 Fr. Lengths: 130 cm and 155 cm. Sterilization: Ethylene Oxide (EtO). Shelf-life: 12 months. Standards: ISO 10993-1, ISO 10993-7, ASTM F1980.
Indications for Use
Indicated for use in peripheral vasculature blood vessels to assist in delivery of diagnostic and therapeutic agents. Contraindicated for use in cerebral and coronary vessels. Not intended for embolectomy or angioplasty.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
Sniper Infusion Catheter with Balloon Occlusion (K180904)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in large, bold, blue letters, and the word "ADMINISTRATION" in smaller, blue letters below.
February 23, 2024
Embolx, Inc. Louise Musante Regulatory Compliance Consultant 530 Lakeside Dr. Suite 200 Sunnyvale, California 94085
Re: K232536
Trade/Device Name: Soldier Microcatheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: DQO Dated: January 18, 2024 Received: January 25, 2024
Dear Louise Musante:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Lydia S.
Glaw -S
Digitally signed by Lydia S. Glaw -S Date: 2024.02.23 20:43:43 -05'00'
Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices
{3}------------------------------------------------
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232536
Device Name Soldier Microcatheter
Indications for Use (Describe)
The Soldier Microcatheter is indicated for use in the blood vessels of the peripheral vasculature. It is intended to assist in the delivery of diagnostic agents and therapeutic agents into the target treatment area.
The Soldier Microcatheter is intended for Prescription Use Only.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for EMBOLX. The logo consists of a spiral of black and red dots on the left, followed by the word "EMBOLX" in black, sans-serif font. The dots are arranged in a circular pattern, with the black dots on one side and the red dots on the other.
K232536
# Section 5 - 510(k) Summary
# 510(k) Summary Embolx, Inc.'s Soldier
February 23, 2024 DATE PREPARED:
# 1. COMPANY NAME/CONTACT Embolx, Inc. 530 Lakeside Dr. #200 Sunnyvale, CA 94085
- 2. CONTACT: Louise Musante Regulatory Compliance Consultant Email: louisemusante@gmail.com Cell Phone: (650) 242-5563
# 3. DEVICE INFORMATION
Device Trade Name: Soldier Microcatheter Common Name: Microcatheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Product Code: DQO Device Class: Class II 510(k) Review Panel: Cardiovascular
# 4. PREDICATE DEVICE
Trade name: Progreat
510(k) submitter/holder: Terumo Medical Corporation
510(k) Numbers: K033583, K170223
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for EMBOLX. The logo consists of a spiral of circles on the left and the word "EMBOLX" in bold, sans-serif font on the right. The circles in the spiral are of varying sizes and are colored in shades of black and red.
# 5. REFERENCE DEVICE USED IN NON-CLINICAL PERFORMANCE TESTING
Sniper Infusion Catheter with Balloon Occlusion (K180904)
# 6. DEVICE DESCRIPTION
The Soldier is a single lumen catheter designed to access small, tortuous vasculature. It is available in a variety of outer and inner diameters. Each configuration has a hydrophilic coating to provide lubricity for navigation of vessels. The inner lumen is lined with lubricious PTFE to facilitate movement of guidewires and other devices. The distal tip of the catheter is radiopaque to aid in visualization under fluoroscopy.
# 7. INDICATION FOR USE STATEMENT
The Soldier Microcatheter is indicated for use in the blood vessels of the peripheral vasculature. It is intended to assist in the delivery of diagnostic agents and therapeutic agents into the target treatment area.
The Soldier Microcatheter is intended for Prescription Use Only.
# 8. INTENDED USE STATEMENTS:
The Soldier is intended for the infusion of contrast media into all peripheral vessels up to and including the cervical vessels, all vessels in the lower and upper extremities. The Soldier is also intended for drug infusion in intra-arterial therapy and the infusion of embolic materials for hemostasis.
The Soldier is contraindicated for use in the cerebral and coronary vessels. It is not intended for use in embolectomy or angioplasty procedures.
# 9. SUBSTANTIAL EQUIVALENCE COMPARISON
The Soldier catheter is substantially equivalent to the claimed predicate device; the Progreat catheter (K033583), based on comparisons of the intended use and technological characteristics.
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for EMBOLX. The logo consists of a circular design made up of smaller circles, with the circles transitioning from black to red. To the right of the circular design is the word "EMBOLX" in a bold, sans-serif font.
# Table 1 – Comparison Table of Subject Device to Predicate Device
| Attributes | Soldier<br>(Subject Device) | Progreat Catheter<br>(Predicate Device)<br>(K033583) | Differences | Differences<br>raise any<br>additional<br>safety issues? |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| General Information | | | | |
| Regulation # | 21 CFR 870.1200 | 21 CFR 870.1200 | Same | N/A |
| Regulation<br>Name | Diagnostic Intravascular<br>Catheter | Diagnostic Intravascular Catheter | Same | N/A |
| Regulatory<br>Class | Class II | Class II | Same | N/A |
| Product Code | DQO | DQO | Same | N/A |
| Indications<br>for Use | The Soldier Microcatheter is<br>indicated for use in the blood<br>vessels of the peripheral<br>vasculature. It is intended to<br>assist in the delivery of<br>diagnostic agents and<br>therapeutic agents into the<br>target treatment area.<br>The Soldier Microcatheter is<br>intended for Prescription Use<br>Only. | The Progreat is intended for the<br>infusion of contrast media into all<br>peripheral vessels up to and<br>including the cervical vessels, all<br>vessels in the lower and upper<br>extremities and all coronary<br>vessels. The Progreat is also<br>intended for drug infusion in intra-<br>arterial therapy and the infusion<br>of embolic materials for<br>hemostasis. The Progreat should<br>not be used in cerebral vessels. | Soldier<br>should not<br>be used in<br>coronary<br>vessels. It is<br>also not<br>intended for<br>use in<br>embolectomy<br>or<br>angioplasty<br>procedures. | No |
# Table 2 – Comparison Table of Technological Characteristics: Subject Device to Predicate Device
| | Attributes | Soldier<br>(Subject Device) | Progreat Catheter<br>(Predicate Device)<br>(K033583) | Differences | Differences<br>raise any<br>additional<br>safety<br>issues? |
|-------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|-------------------------------------------------------------|
| | Device Specifications | | | | |
| Catheter | Catheter Size<br>Fr (mm) | 2.0 Fr (0.67 mm)<br>2.5 Fr (0.84 mm) | 2.0 Fr (0.67 mm)<br>2.4 Fr (0.80 mm)<br>2.7 Fr (0.90 mm)<br>2.8 Fr (0.93 mm) | Dimensions | None |
| | Catheter OD<br>Distal End<br>Fr (mm) | 2.0 Fr (0.67 mm)<br>2.5 Fr (0.84 mm) | 2.0 Fr (0.67 mm)<br>2.4 Fr (0.80 mm)<br>2.7 Fr (0.90 mm)<br>2.8 Fr (0.93 mm) | Dimensions | None |
| Attributes | Soldier<br>(Subject Device) | Progreat Catheter<br>(Predicate Device)<br>(K033583) | Differences | Differences<br>raise any<br>additional<br>safety<br>issues? | |
| Catheter OD<br>Proximal End<br>Fr (mm) | 2.0 Fr (0.67 mm)<br>2.5 Fr (0.84 mm) | 2.7 Fr (0.90 mm)<br>2.9 Fr (0.97 mm)<br>2.9 Fr (0.97 mm)<br>3.0 Fr (1.00 mm) | Dimensions | None | |
| Maximum OD<br>Fr (mm) | 2.5 Fr (0.84 mm)<br>2.9 Fr (0.97 mm) | 2.7 Fr (0.90 mm)<br>2.9 Fr (0.97 mm)<br>2.9 Fr (0.97 mm)<br>3.0 Fr (1.00 mm) | Dimensions | None | |
| Catheter<br>Effective<br>Lengths | 130 cm<br>155 cm | 100 cm<br>110 cm<br>130 cm<br>150 cm | Dimensions | None | |
| Tip Design | Atraumatic | Atraumatic | None | N/A | |
| Hub Design | Female Luer, Tapered<br>Funnel Access | Female Luer, Tapered<br>Funnel Access | None | N/A | |
| Hydrophilic<br>Coating Length<br>(Catheter<br>length/Coating<br>length, from the<br>distal tip) | 130 cm/ 70 cm<br>155 cm/ 70 cm | 100 cm/ 50 cm<br>110 cm/ 50 cm<br>130 cm/ 70 cm<br>150 cm/ 90 cm | Dimensions | None | |
| Coating | Hydrophilic Coating | Hydrophilic coating | None | N/A | |
| Guidewire | Wire Diameter | Up to 0.018" | Up to 0.021" | Dimensions | None |
| Accessories | | None | Packaged with:<br>Guidewire<br>Guidewire Inserter<br>Catheter Mandrel (stylet)<br>2.5mL Syringe with Lock<br>Wire Stopper<br>Catheter Stopper XS<br>Hemostatic Valve | Soldier<br>packaged<br>without<br>Accessories | None |
| General Information | | | | | |
| Design/Construction | Single lumen catheter<br>consisting of metal coil<br>reinforced multi-layer<br>polymer tubing with<br>hydrophilic coating. | Single lumen catheter<br>consisting of metal coil<br>reinforced multi-layer<br>polymer tubing with<br>hydrophilic coating. | None | N/A | |
| Pouch | Polyester/Polyethylene/<br>Tyvek* | Unknown | N/A | N/A | |
| Attributes | Soldier<br>(Subject Device) | Progreat Catheter<br>(Predicate Device)<br>(K033583) | Differences | Differences<br>raise any<br>additional<br>safety<br>issues? | |
| Hoop | Polyethylene | Unknown | N/A | N/A | |
| Display Carton | SBS Paperboard | Unknown | N/A | N/A | |
| Configuration | Single Use/Disposable | Single Use/Disposable | None | N/A | |
| Sterilization | EtO | EtO | None | N/A | |
| Shelf-life | 12 months | 24 months | Soldier<br>Shelf Life<br>Validated to<br>12 Months | No | |
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the word "EMBOLX" in bold, black letters. To the left of the word is a circular design made up of small circles. The circles are arranged in a spiral pattern, with the circles on the left side of the spiral being a dark red color and the circles on the right side of the spiral being black.
# Section 5 – 510(k) Summary
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for EMBOLX. The logo consists of a spiral of circles on the left and the word "EMBOLX" on the right. The circles are arranged in a spiral pattern, with the circles on the outside of the spiral being larger than the circles on the inside. The word "EMBOLX" is written in a bold, sans-serif font.
# Section 5 - 510(k) Summary
# 10. SUMMARY OF TECHNOLOGICAL DIFFERENCES
#### Indications for Use:
The subject device: Soldier Microcatheter, has been validated for use in the blood vessels of the peripheral vasculature, with the exception of the coronary vessels. It is also not intended for use in embolectomy or angioplasty procedures. The limitations of use for the Soldier Microcatheter are clearly identified as contraindications in the product label. The predicate device: Progreat catheter, has been cleared for use in the peripheral vasculature, including the coronary vessels. It is also cleared for the infusion of embolic materials for hemostasis.
The differences described do not raise any additional safety or efficacy concerns for the subject device as the indications for use of the subject device, falls within the intended use of the cleared predicate device, therefore the two devices have the same intended use.
#### Dimensions:
The subject device: Soldier Microcatheter, offers fewer catheter configurations than the cleared predicate device: Progreat catheter. However, the catheter sizes, with the exception of the 155 cm length, fall within the largest and smallest catheter dimensions cleared to market for the predicate device. Performance data for the longer catheter length, in addition to the other sizes, support the substantial equivalency claim and therefore do not raise any additional safety or efficacy concerns for the subject device.
#### Accessories:
The subject device: Soldier Microcatheter, is packaged without accessories, unlike the predicate device: Progreat catheter, which is packaged with a guidewire, a guidewire
{10}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for EMBOLX. The logo consists of a circular design made up of smaller circles in varying shades of pink and black on the left side. To the right of the circular design is the word "EMBOLX" in bold, sans-serif font.
inserter, a catheter mandrel (stylet), a 2.5mL syringe with lock, a wire stopper, a catheter stopper XS, and a hemostatic valve. Performance testing for the Soldier Microcatheter was performed using a third-party guidewire without incidence. Device compatibility specifications are outlined in the product labeling. Therefore, the absence of accessories packaged with the subject device, does not raise any additional safety or efficacy concerns for the subject device.
### Shelf-life:
The subject device: Soldier Microcatheter, has been validated to a shelf-life of 12months (1 year) though accelerated aging, which will be included on the product label. The predicate device: Progreat catheter, has been validated to a shelf-life of 24-months (2 years). The difference in shelf-life does not raise any additional safety or efficacy concerns for the subject device.
# 11.SUMMARY OF NON-CLINICAL PERFORMANCE TESTING
As required under Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990. A summary of any information regarding substantial equivalence of the device follows.
Included in this section are summary descriptions of the testing which substantiates the performance of the subject device: Śoldier Microcatheter as well as its substantial equivalence to the predicate device: Progreat;
- Biocompatibility Testing ●
- Design Verification (Performance Bench-Top Testing) ●
- Sterilization Validation
- Shelf-life
- a) Biocompatibility Testing
Biocompatibility testing was conducted in accordance with the FDA recognized consensus standard; ISO 10993-1, Biological Evaluation of Medical Devices, Part 1 Evaluation and Testing within a risk management process (recognition #2-258), to ensure that patient contact materials are biocompatible for their intended use. All studies were conducted pursuant to 21 CFR, Part 58, Good Laboratory Practices. Testing included the following:
- > Cytotoxicity: Cytotoxicity using the ISO L929 MEM Elution Test
- Sensitization: ISO Kligman Maximum Sensitization Test A
- > Irritation: ISO Intracutaneous Injection Test
- A Systemic Toxicity (Acute): ISO Systemic Injection Test, ISO Rabbit Pyrogen Test
- > Hemocompatibility: ASTM Partial Thromboplastin Time (PTT) Test; ISO C3a and SC5b-9 Complement Activation Test; ISO Dog Thrombogenicity Study in
{11}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for EMBOLX. The logo consists of a spiral of circles on the left and the word "EMBOLX" in bold, sans-serif font on the right. The circles in the spiral are colored in shades of red and black.
Dogs Test
- Chemical/Material Review A
Results from each test met the defined acceptance criteria and support compliance to ISO 10993-1, Biological Evaluation of Medical Devices, Part 1 Evaluation and Testing within a risk management process (recognition #2-258).
- b) Design Verification (Performance Bench-top Testing)
Performance testing was conducted to ensure the safety and effectiveness of the Soldier Microcatheter and to demonstrate substantial equivalence to the predicate device: Progreat catheter. The following bench tests were:
- A Visual Inspection
- Dimensional Verification A
- Hub Integrity A
- A Catheter Burst & Leakage Pressure
- Catheter Flow A
- > Tensile Strength
- > Torsion
- Bend & Buckle ("flexibility and kink test") A
- > Materials Test Report
- > Radiopacity ("Radio-detectability")
- Particulate >
- > Simulated Use in Peripheral Vasculature
- Shipping and Packaging >
- Pouch Seal Strength A
- Pouch Seal Integrity
- Corrosion Resistance A
- c) Sterilization
Sterilization of the subject device: Soldier catheter, is performed using ethylene oxide, an established sterilization method, as identified in the FDA guidance; Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile, January 21, 2016. The predicate device: Progreat catheter, is also sterilized using ethylene oxide.
All testing data met the defined acceptance criteria for sterility assurance level (SAL), EtO residuals, bacterial endotoxins and material mediated pyrogenicity, in accordance with the FDA recognized consensus standard; ISO 10993-7:2008, Biological evaluation of medical devices – Part 7: Ethylene oxide sterilization residuals (recognition #2-275), in addition to other international sterility standards.
{12}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for EMBOLX. The logo consists of a circular design made up of smaller circles that transition from black to red. To the right of the circular design is the word "EMBOLX" in a bold, sans-serif font.
- d) Shelf-Life
A shelf-life of 12 months (1 year) has been validated though accelerated aging of the subject device: Soldier catheter. The predicate device: Progreat catheter, has been validated to 24 months (2 years). Testing was conducted on final device assemblies, packaged, and sterilized with ethylene oxide. Testing was performed in accordance with the FDA recognized consensus standard; ASTM F1980:2016: Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices (recognition # 14-497). All tests met the pre-defined performance criteria.
#### 12. CONCLUSION
Embolx, Inc. believes that the Soldier Microcatheter is substantially equivalent in intended use and technological characteristics to the Progreat catheter (k033583). The Soldier catheter therefore meets the Federal Food, Drug and Cosmetic Act criteria for 510(k) clearance of this device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.