The VereSee Optical Veres Needle System is intended for percutaneous insertion into the peritoneal cavity for the purpose of insufflation with carbon dioxide to establish pneumoperitoneum prior to the placement of trocars during laparoscopic surgery.
Device Story
The VereSee Optical Veres Needle System facilitates closed abdominal entry for laparoscopy. The device consists of an optical Veres needle and a Camera Control Unit (CCU). The needle features three concentric stainless steel cannulas: an outer insufflation cannula with a luer fitting, a central access cannula with a clear tip, and an inner camera cannula containing a CMOS sensor and LED light fibers. During insertion, the CMOS camera provides real-time visualization of the abdominal wall penetration, which is displayed on an HDMI monitor via the CCU. A spring-loaded obturator provides haptic feedback to the surgeon during entry. The device is used by surgeons in a clinical setting to safely establish pneumoperitoneum. By providing direct visualization during the puncture, the device assists the surgeon in confirming proper needle placement, potentially reducing the risk of injury to internal structures during initial abdominal access.
Clinical Evidence
Bench testing only. Performance data included sterilization validation (ISO 11135), EO residuals (ISO 10993-7), transit simulation (ASTM D4169), shelf-life (ASTM F1980), and biocompatibility (ISO 10993-5, -10, -11, -23). Electrical safety and EMC testing were performed per IEC 60601-1 and IEC 60601-1-2. Functional bench testing evaluated optical performance (ISO 8600-3 to -5), mechanical strength, and comparative flow/leakage/penetration force against predicate devices, demonstrating comparable performance.
Technological Characteristics
Stainless steel construction; spring-loaded obturator mechanism; CMOS camera with LED illumination; HDMI output via CCU. Sterilized by ethylene oxide. Dimensions: 127 mm length, 2.4 mm outer diameter. Integrated two-component seal system. Complies with IEC 60601-1 (electrical safety) and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for percutaneous insertion into the peritoneal cavity to establish pneumoperitoneum via carbon dioxide insufflation prior to trocar placement in laparoscopic surgery.
Regulatory Classification
Identification
A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 2, 2024
Freyja Healthcare, LLC % Christine Brauer, PhD Regulatory Affairs Consultant Brauer Device Consultants. LLC 7 Trail House Court Rockville, Maryland 20850
Re: K232464
> Trade/Device Name: VereSee Optical Veres Needle System Regulation Number: 21 CFR 884.1730 Regulation Name: Laparoscopic Insufflator Regulatory Class: II Product Code: HIF Dated: April 2, 2024 Received: April 3, 2024
Dear Christine Brauer, PhD:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{3}------------------------------------------------
the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jason Roberts -S
Jason R. Roberts, PhD Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232464
Device Name
VereSee Optical Veres Needle System
Indications for Use (Describe)
The VereSee Optical Veres Needle System is intended for percutaneous insertion into the perioneal cavity for the purpose of insufflation with carbon dioxide to establish pneumoperitoneum prior to the placement of trocars copic surgery.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
## 510(k) SUMMARY VereSee Optical Veres Needle K232464
Applicant Information
| Applicant and<br>510(k) Owner: | Freyja Healthcare, LLC<br>488 Pleasant Street<br>North Andover, Massachusetts 01845 USA |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Ronald Adams<br>Chief Technology Officer<br>Freyja Healthcare, LLC<br>Telephone: 775-800-7300<br>Email: ron@freyjahealthcare.com |
Date of Preparation: 30 April 2024
510(k) Number: K232464
Device and Classification Information
| Trade Name: | VereSee Optical Veres Needle |
|--------------------|---------------------------------|
| Common Name: | Veres Needle |
| Regulation Name: | Laparoscopic Insufflator |
| Regulation Number: | 21 CFR 884.1730 |
| Product Code: | HIF (Insufflator, Laparoscopic) |
| Class: | II |
VereSee Optical Veres Needle System is comprised two components:
- 1) Optical Veres Needle, and
- 2) Camera Control Unit.
## Optical Veres Needle
The Optical Veres Needle consists of a series of three concentric, stainless steel cannulas with a handle and an umbilical cable to connect it to the VereSee Camera Control Unit. The device is provided sterile for single-patient use. Each of the three concentric, stainless steel cannulas is described below:
{6}------------------------------------------------
- 1. Outer Cannula (or Insufflation Cannula) provides a luer fitting for connection of insufflation tubing with an integral flow control stopcock and a set of seals to prevent leakage of insufflation gas.
- 2. Central Cannula (or Access Cannula) consists of a stainless steel hypo tube with a clear, point tip for penetration and visualization during abdomen penetration.
- 3. Inner Cannula (or Camera Cannula) is composed of a stainless steel hypo tube with a CMOS camera surrounded by light fibers at its tip. The light fibers carry the light from an integral LED which is mounted to a heat sink to distribute heat from the LED.
## VereSee Camera Control Unit
The VereSee Camera Control Unit connects the CMOS camera in the Optical Veres Needle to HDMI compatible monitors to provide an image for laparoscopic procedures. The VereSee Camera Control Unit (CCU) includes an LED driver to power the LED. The CCU connects to the Optical Veres Needle via an umbilical cable. The VereSee Camera Control Unit converts signals from the CMOS camera in the Optical Veres Needle to a format compatible with HDMI display input requirements.
Indication for Use
The indication for use for the VereSee Optical Veres Needle System is shown below.
The VereSee Optical Veres Needle System is intended for percutaneous insertion into the peritoneal cavity for the purpose of insufflation with carbon dioxide to establish pneumoperitoneum prior to the placement of trocars during laparoscopic surgerv.
## Predicate Device Identification
For purposes of demonstrating substantial equivalence, the following predicate device was selected:
- LaproLight Veress Needle (by Buffalo Filter, LLC) cleared via K171139 on May 18, ● 2017.
## Comparison of the Indication for Use and Intended Use
The LaproLight Veress Needle ("LaproLight") and VereSee Optical Veres Needle System share the same intended use - each device is used to enable closed abdominal entry techniques for laparoscopy. Specifically, each device has a spring-loaded tip that is inserted into the patient's peritoneal cavity to allow for insufflation with carbon dioxide gas, creating a pneumoperitoneum for laparoscopic surgery (see Table 1). Each device has the same indication for use statement except for the tradename, demonstrating the devices have the same intended use.
#### Table 1. Summary Comparison of the Intended Use and Indication for Use for the Subject Device and Predicate Device
{7}------------------------------------------------
| Characteristic | VereSee Optical Veres Needle System<br>(K232464) | LaproLight Veress Needle<br>(K171139) |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation Number<br>Product Code | 21 CFR 884.1730<br>HIF | 21 CFR 884.1730<br>HIF |
| | Laparoscopic Insufflator | Laparoscopic Insufflator |
| Class | II | II |
| Indication for Use | The VereSee Optical Veres Needle<br>System is intended for percutaneous<br>insertion into the peritoneal cavity for the<br>purpose of insufflation with carbon<br>dioxide to establish pneumoperitoneum<br>prior to the placement of trocars during<br>laparoscopic surgery. | The LaproLight is intended for<br>percutaneous insertion into the<br>peritoneal cavity for the purpose of<br>insufflation with carbon dioxide to<br>establish pneumoperitoneum prior to the<br>placement of trocars during laparoscopic<br>surgery. |
| Intended Use | Facilitate entry of laparoscope using a<br>closed (veres needle) technique | Facilitate entry of laparoscope using a<br>closed (veres needle) technique |
| Conditions of Use | Rx Only; Single Use Only | Rx Only; Single Use Only |
| Purpose | To establish pneumoperitoneum prior to<br>placement of trocars during a<br>laparoscopic procedure | To establish pneumoperitoneum prior to<br>placement of trocars during a<br>laparoscopic procedure |
| Functions | To puncture peritoneum wall<br>To allow for the flow of liquid or gas<br>through the needle<br>To insufflate the peritoneal cavity with<br>carbon dioxide<br>To provide visualization during veres<br>needle entry | To puncture peritoneum wall<br>To allow for the flow of liquid or gas<br>through the needle<br>To insufflate the peritoneal cavity with<br>carbon dioxide |
As shown in the table above, the subject device includes the function of visualization using a camera during veres needle entry. This difference does not raise different questions of safety and effectiveness.
## Comparison of the Technological Characteristics
The VereSee Optical Veres Needle System and the LaproLight share many similar technological characteristics. Both devices utilize a spring loaded obturator mechanism to provide the user with haptic feedback when entering the body. Both include a luer fitting connection for either a syringe to perform a saline flow test and/or insufflation tubing for gas delivery. For both devices, the stop cock is downstream of luer connection to control the flow of liquid or gas through the cannula with variable positions between ON and OFF. Both veres needles include a visualization element although for different purposes (see Table 2). The VereSee Optical Veres Needle System includes an optical obturator tip designed to enable visualization during body cavity entry whereas the LaproLight includes an LED, which turns on when the healthcare provider has punctured into the peritoneal cavity.
Comparison of the Technological Similarities and Differences Between Table 2. the Subject Device and the Predicate Device
| Technological Characteristic | VereSee Optical Veres Needle System | LaproLight Veress Needle<br>(K171139) |
|------------------------------|-------------------------------------------------------------------|-------------------------------------------------------------------|
| Needle Design | Spring loaded stylet mechanism which<br>provides haptic feedback. | Spring loaded stylet mechanism which<br>provides haptic feedback. |
{8}------------------------------------------------
| Technological Characteristic | VereSee Optical Veres Needle System | LaproLight Veress Needle<br>(K171139) |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Outer Cannula | Beveled needle point for cutting through abdominal wall tissue | Beveled needle point for cutting through abdominal wall tissue |
| Inner Cannula | Spring loaded inner stylet | Spring loaded inner stylet |
| Needle Length | 127 mm | 120 mm<br>150 mm |
| Outer Diameter (tip) | 2.4 mm | 2 mm |
| Material | Stainless steel | Stainless steel |
| Sterility | Single use sterile | Single use sterile |
| Seal System | Integrated two component seal system to<br>maintain pneumoperitoneum. A dome<br>seal to seal when camera or visualization<br>camera are not in use and a self-<br>adjusting back-up seal when they are in<br>use. | Standard veres needle seal system |
| Stopcocks | An integral stopcock assembly with a<br>standard luer lock fitting which provides<br>for attachment of insufflation tubing or<br>syringe for saline flow test. | An integral stopcock assembly with a<br>standard luer lock fitting which provides for<br>attachment of insufflation tubing or syringe<br>for saline flow test. |
| Visualization | CMOS camera chip to provide<br>visualization during entry | LED to indicate peritoneal cavity entry |
As shown in the table above, there are differences between the subject and predicate devices. The subject device is available in one needle length rather than two for the predicate device. The outer diameter (tip) of the subject device is different (larger) than the predicate device. The subject device has a different seal system than the predicate device. The subject device has a different mode of visualization than the predicate device. None of these technological differences raise different questions of safety and effectiveness.
## Performance Data
The following is a summary of the performance data to establish the substantial equivalence of the VereSee Optical Veres Needle System to the predicate device.
## Sterilization:
The VereSee Optical Veres Needle is sterilized by ethylene oxide (EO). Sterilization was performed as described in ISO 11135:2014, Sterilization of healthcare products -Ethylene oxide - Requirements for the development, validation, and routine control of a sterilization process for medical devices. The sterilization assurance level (SAL) is 106.
Testing of EO residuals was conducted per ISO 10993-7 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals. VereSee Optical Needle devices met the requirements of the standard.
## Transit Simulation and Shelf-Life:
{9}------------------------------------------------
Transit simulation and shelf-life testing was conducted for the VereSee Optical Needle to establish the suitability of the packaging to protect the device during simulated transportation, product shelf-life and ability of the packaging to maintain a sterile barrier.
Transit simulation testing was conducted per ASTM D4169-22. The final packaging was tested following simulated transportation, including the following: Seal Strength Test (ASTM F88), Visual Inspection (ASTM F1886), and Bubble Leak testing (ASTM F2096).
Accelerated aging was conducted per ASTM F1980-21 followed by performance testing and package integrity testing. A shelf-life of 6 months was established based upon accelerated aging testing.
## Biocompatibility:
The VereSee Optical Veres Needle has been tested according to ISO 10993-1 recommendations for externally communicating devices with bone/tissue/dentin contact of limited contact duration (< 24 hours). Biocompatibility studies were performed to show the device materials are safe, biocompatible, and suitable for their intended use. Both ISO 10993 and FDA Guidance "Use of International Standard ISO 10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process" were utilized to guide the biocompatibility testing. The following biocompatibility studies were successfully completed.
Summary of the Biocompatibility Tests and Results Table 3.
| Test Performed | Test Method | Test Results |
|---------------------------------------|-------------------|--------------|
| Cytotoxicity | ISO 10993-5:2009 | Pass |
| Intracutaneous Reactivity | ISO 10993-23:2021 | Pass |
| Guinea Pig Maximization Sensitization | ISO 10993-10:2021 | Pass |
| Pyrogenicity | ISO 10993-11:2017 | Pass |
| Systemic Toxicity | ISO 10993-11:2017 | Pass |
## Electrical Safety and EMC Testing:
The VereSee Optical Veres Needle System has undergone testing to demonstrate that the device meets the requirements for medical device safety, including electrical safety, according to the following international standards: IEC 60601-1, 3rd Edition, Medical electrical equipment -General requirements for basic safety and essential performance.
The VereSee Optical Veres Needle System Camera Control Unit has undergone testing to demonstrate the device meets the following international standard: IEC 60601-1-2, 4.1 Edition, Medical electrical equipment - Part 1-2, General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
## Bench and Functional Performance Testing:
Bench testing was performed to demonstrate that the VereSee Optical Veres Needle System functions as intended, including the following:
{10}------------------------------------------------
- > Optical performance, including ISO 8600-3 through ISO 8600-5
- > Physical characteristics (such as weight and maximum handle temperature)
- > Mechanical testing (flow and leakage testing, spring obturator testing and needle penetration tip force testing) to evaluate key performance requirements
- > Mechanical testing (destructive) to evaluate the physical strength of the bonds between components
The mechanical testing included comparative testing to legally marketed veres needles for flow and leakage testing, needle spring obturator testing and needle penetration tip force testing. The Veres Optical Veres Needle System had comparable performance to the legally marketed veres needles.
## Conclusion
The VereSee Optical Veres Needle Optical System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.