LaparoLight Veress Needle
K171139 · Buffalo Filter, LLC · HIF · May 18, 2017 · Obstetrics/Gynecology
Device Facts
| Record ID | K171139 |
| Device Name | LaparoLight Veress Needle |
| Applicant | Buffalo Filter, LLC |
| Product Code | HIF · Obstetrics/Gynecology |
| Decision Date | May 18, 2017 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 884.1730 |
| Device Class | Class 2 |
| Attributes | 3rd-Party Reviewed |
Intended Use
The LaparoLight™ Veress Needle is intended for percutaneous insertion into the peritoneal cavity for the purpose of insufflation with carbon dioxide to establish pneumoperitoneum prior to the placement of trocars during laparoscopic procedures.
Device Story
LaparoLight™ Veress Needle is a single-use, sterile device for establishing pneumoperitoneum during laparoscopic surgery. It features a spring-loaded blunt stylet that retracts upon pressure to allow a sharp cannula to penetrate tissue. The device includes an integrated LED indicator activated by a slip sheet; the LED remains ON when the blunt tip is extended (indicating the needle is in a cavity) and turns OFF when the tip is retracted (indicating pressure against tissue). Physicians use the device in clinical/hospital settings. It includes a luer lock for saline flow testing and a stopcock for controlling CO2 gas delivery. The LED provides visual feedback to the surgeon regarding the needle tip position relative to the peritoneum, assisting in safe entry and reducing the risk of organ injury.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included biocompatibility (ISO 10993-1: cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, rabbit pyrogen), electrical safety (AAMI/ANSI ES 60601-1), EMC (IEC 60601-1-2), design validation (simulated tissue puncture, indicator operation, flow testing), force equivalency, tip pull strength, and spring obturator operation.
Technological Characteristics
Materials: Not specified. Sensing: Mechanical spring-loaded stylet with LED indicator. Energy: Battery-powered LED. Dimensions: 120mm and 150mm lengths. Connectivity: None. Sterilization: Sterile, single-use. Standards: AAMI/ANSI ES 60601-1:2005/(R)2012, IEC 60601-1-2 Ed 4.0 2014-02.
Indications for Use
Indicated for patients undergoing laparoscopic procedures requiring percutaneous insertion into the peritoneal cavity for CO2 insufflation to establish pneumoperitoneum.
Regulatory Classification
Identification
A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.
Predicate Devices
- Unimax Veress Needle (K111441)
Related Devices
- K150068 — Unimicro Veress Needle · Unimicro Medical Systems (Shenzhen) Co., Ltd. · Jul 31, 2015
- K172120 — Veress Needle · Wickimed (Huizhou) Medical Equipment Manufacturing Co.,Ltd · Sep 14, 2017
- K993625 — INSUFFLATION NEEDLE- 120MM, MODEL 900-200 · Genicon, LC · Jun 20, 2000
- K193339 — GTK Veress Needles · Guangzhou T.K Medical Instrument Co., Ltd. · Jan 30, 2020
- K111441 — UNIMAX VERESS NEEDLE · Unimax Medical Systems, Inc. · Aug 19, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 18, 2017
Buffalo Filter, LLC % Mark A. Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street, NW BUFFALO, MN 55313
Re: K171139
> Trade/Device Name: LaproLight™ Veress Needle Regulation Number: 21 CFR 884.1730 Regulation Name: Laparoscopic insufflator Regulatory Class: Unclassified Product Code: HIF Dated: April 17, 2017 Received: April 18, 2017
Dear Mark A. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Benjamin/R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K171139
Device Name LaparoLight™ Veress Needle
### Indications for Use (Describe)
The LaparoLight™ Veress Needle is intended for percutaneous insertion into the peritoneal cavity for the purpose of insufflation with carbon dioxide to establish pneumoperitoneum prior to the placement of trocars during laparoscopic procedures.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K171139 page 1 of 4
Image /page/3/Picture/1 description: The image shows the logo for Buffalo Filter Filtration Group. The logo features a spiral of green and blue dots on the left side. To the right of the spiral is the company name, "BUFFALO FILTER," in blue, with "Filtration Group" in gray underneath.
# 510(k) Summary For LAPAROLIGHT™ VERESS NEEDLE
### Submitter:
Buffalo Filter, LLC 5900 Genesee Street Lancaster, NY 14086 Phone: (716) 835-7000 Fax No: (716) 835-3414 FDA Registration number: 1319744
| Contact: | Kyrylo Shvetsov |
|----------|---------------------------|
| | Program Manager, NPD |
| | Senior Engineer |
| | Telephone: (716) 835-7000 |
| | Fax No: (716) 835-3414 |
Prepared Date: March 20, 2017
#### 1. Device Name
| Trade Name: | LaparoLight™ Veress Needle, models LLVN12001<br>and LLVN15001 |
|----------------------|--------------------------------------------------------------------------|
| Common/Usual Name: | Veress Needle |
| Classification Name: | Insufflator, Laparoscopic (21 CFR 884.1730,<br>Product Code: HIF and FHO |
#### 2. Predicate Device
Buffalo Filter claims substantial equivalence to:
Unimax Medical Systems Inc, Unimax Veress Needle (K111441)
This predicate device has not been subjected to a design-related recall.
{4}------------------------------------------------
#### 3. Description of Device
The LaparoLight™ Veress Needle is a Veress needle that includes an LED indicator. This product is used to establish pneumoperitoneum prior to start of laparoscopic procedure. LaparoLight™ Veress Needle is available in two lengths: 120mm and 150mm to use in different size patients and applications. LaparoLight" Veress Needle is provided sterile and intended for single-use only in a healthcare facility/hospital.
LaparoLight™ Veress Needle includes a spring loaded blunt stylet mechanism. When pressure is applied on blunt tip, it retracts and allows the sharp edge of cannula to cut through the tissue. Product also includes a slip sheet which when pulled activated the LED indicator. LED will turn ON and stay ON as long as the blunt tip of the product is extended forward past sharp edge of cannula. Thus during the puncture of peritoneum wall the LED will turn OFF while there is a pressure against blunt tip and turn back ON when cannula has entered a cavity allowing the blunt tip to spring forward.
In addition LaparoLight™ Veress Needle includes a luer lock for syringe connection to perform saline flow test and insufflation tubing for gas delivery. Stop cock is present downstream of luer lock to control the flow of liquid or gas through the needle with possible positions ON and OFF.
#### 4. Intended Use
The LaparoLight M Veress Needle is intended for percutaneous insertion into the peritoneal cavity for the purpose of insufflation with carbon dioxide to establish pneumoperitoneum prior to the placement of trocars during laparoscopic procedures.
#### ട്. Description of Substantial Equivalence
The subject and predicate device have the same intended use.
At a high level, the subject and predicate devices are based on following same technological elements:
- Spring loaded blunt stylet - used to protect internal organs from sharp of cannula.
- Blunt tip exposure indicator an indicator that shows to the user if blunt tip is ● extended or retracted
- . Luer Lock - connection point for syringe to perform saline test and insufflation tubing for CO2 gas delivery
- Stop Cock - to control gas passage through the needle
{5}------------------------------------------------
The following technological differences exist between the subject and predicate device:
- Type of Blunt tip exposure indicator LaparoLight™ Veress Needle contains ● an LED instead of a red marker painted on the needle hub that shows if blunt tip is extended
- Slip Sheet for product activation prior to use .
The differences in technology do not raise different questions of safety and effectiveness.
#### Performance Data 6.
The following performance data were provided in support of the substantial equivalence determination.
# Biocompatibility Testing
The biocompatibility evaluation for the LaparoLight™ Veress Needle was conducted in accordance with International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by the FDA. The battery of testing included the following tests:
- · In Vitro Cytotoxicity
- · Maximization Sensitization
- Intracuteneous Reactivity
- · Systemic Toxicity Testing
- · Rabbit Pyrogen Testing
The LaparoLight™ Veress Needle is considered a device that contacts the Tissue / bone / dentin for a limited contact duration (less than 24hrs). The test results demonstrate that the subject device is biocompatible.
### Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC testing were conducted on the LaparoLight™ Veress Needle. The system complies with AAMI/ANSI ES 60601-1:2005/(R)2012 And A 1:2012 standard for electrical safety and IEC 60601-1-2 Edition 4.0 2014-02 standard for EMC.
{6}------------------------------------------------
# Performance Testing - Bench
The following bench tests were performed for this submission, and the subject device passed all tests.
- LaparoLight™ Veress Needle Design Validation ●
Design Validation was performed to verify that the LaparoLight™ Veress Needle product can operate as intended. Validation test included puncturing of simulated tissue, operation of blunt tip position indicator, saline and gas delivery.
- LaparoLightTM Veress Needle Force Equivalency
Force testing was conducted to confirm that the force of the spring loaded blunt stylet and puncture force are similar to the predicate device
- LaparoLight™ Veress Needle Tip Pull Test ●
Testing to confirm equivalence to predicate device performance with regard to strength of bond between components
- LaparoLight™ Veress Needle Switch Operation Test ●
Testing to confirm equivalence to predicate device performance by testing control of gas flow with stop cock/flow controller assembly
- LaparoLight™ Veress Needle Spring Obturator Operation Test ●
Testing to confirm equivalence to predicate device performance – testing covered "click" of the needle and blunt tip extension once the needle enters open internal volume
- LaparoLight™ Veress Needle Puncture Force Test ●
Testing to confirm equivalence to predicate device performance – testing covered puncture force required to puncture tissue simulator
#### 7. Conclusion
The performance testing demonstrates that the LaparoLight™ Veress Needle is substantially equivalent to the predicate device.