K232346 · GE Healthcare · LNH · Oct 27, 2023 · Radiology
Device Facts
Record ID
K232346
Device Name
Digital Expert Access with Remote Scanning
Applicant
GE Healthcare
Product Code
LNH · Radiology
Decision Date
Oct 27, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Digital Expert Access with Remote Scanning is intended as a remote collaboration tool to view and review MR images, to remotely control MR Imaging Devices and to initiate MRI scans remotely.
Device Story
Digital Expert Access with Remote Scanning is a remote collaboration tool enabling real-time interaction between an onsite technologist and a remote expert. The system facilitates remote guidance, procedure assessment, and scanning parameter management. It allows the remote expert to view scanner screens, provide live annotations, and initiate MRI scans on compatible GE HealthCare MRI systems. The device requires an onsite technologist to be present throughout the procedure for patient support. It does not acquire images or perform post-processing; all imaging functions remain with the host MRI system. The system is used in clinical environments where GE HealthCare MRI systems are installed. It benefits patients by providing access to expert support for technologists, potentially improving scan quality and workflow efficiency. The onsite user maintains control by granting access and validating the connection via session password.
Clinical Evidence
No clinical data. Bench testing only. Verification and validation testing were conducted by integrating the device with a subset of GE HealthCare MRI systems to confirm the remote expert's ability to perform remote scanning and ensure proper implementation of design requirements.
Technological Characteristics
Remote collaboration software platform; PC/tablet-based interface; supports touch, keyboard, and touchpad input. Connectivity via hospital network. Operates as an add-on to GE HealthCare MRI systems. Designed under 21 CFR 820 and ISO 13485 standards. No direct patient interface.
Indications for Use
Indicated for use by trained radiologic technologists and experts to facilitate remote collaboration, guidance, and scan initiation on GE HealthCare MRI systems. Requires an onsite technologist to be present for patient support during scanning. Not for diagnostic use or patient safety-related management.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
MAGNETOM Systems with syngo Expert-i option (K052423)
Submission Summary (Full Text)
{1}------------------------------------------------
October 27, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
GE Healthcare % Allena Holzworth Regulatory Affairs Leader 500 W. Monroe Street Chicago, Illinois 60661
Re: K232346
Trade/Device Name: Digital Expert Access with Remote Scanning Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH, LLZ Dated: August 4, 2023 Received: August 4, 2023
Dear Allena Holzworth:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{2}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Samul for
Daniel Krainak, Ph.D. Assistant Director Magnetic Resonance and Nuclear Medicine Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K232346
Device Name Digital Expert Access with Remote Scanning
#### Indications for Use (Describe)
Digital Expert Access with Remote Scan assistance solution which allows remote control of an MR Imaging Device including the ability to initiate a scan remotely. This access provides real time communication mechanisms between the remote and onsite users to facilitate the acquisition occurring on the device. Access must be granted by the system. Images reviewed remotely are not for diagnostic use.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a blue circle. The circle has a wavy, water-like design around the letters, giving it a dynamic and fluid appearance.
# 510(k) Summary
This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.92:
| Date: | August 4, 2023 |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE HealthCare<br>500 W. Monroe Street<br>Chicago, IL 60661 |
| Primary Contact: | Allena Holzworth<br>Regulatory Affairs Leader<br>GE HealthCare<br>Phone: +1 (414) 208-5162<br>Email: Allena.Holzworth@ge.com |
| Secondary Contact: | Christopher Paulik<br>Senior Manager Regulatory Affairs<br>GE HealthCare<br>Phone: +1 (262) 548-2010<br>Email: Christopher.A.Paulik@ge.com |
| Subject Device Name: | Digital Expert Access with Remote Scanning |
| Device Classification: | Class II |
| Regulation Number: | 21 CFR 892.1000 Magnetic Resonance Diagnostic Device<br>21 CFR 892.2050 Medical image management and processing system |
| Product Code: | LNH, LLZ |
### Predicate Device Information
| Predicate Device 1 | |
|--------------------|------------------------------------------------------|
| Device Name: | SIGNA Hero |
| Manufacturer: | GE Medical Systems, LLC |
| 510(k) Number: | K213668 |
| Regulation Number: | 21 CFR 892.1000 Magnetic Resonance Diagnostic Device |
| Product Code: | LNH |
| Predicate Device 2 | |
|--------------------|-------------------------------|
| Device Name: | Customer Remote Console (CRC) |
| Manufacturer: | GE Medical Systems, LLC |
| 510(k) Number: | K150193 |
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized font, enclosed within a circular border. There are small, curved lines emanating from the circle, giving the impression of movement or energy. The logo is colored in a light blue hue.
| Regulation Number:<br>Product Code: | 21 CFR 892.2050 Medical image management and processing system<br>LLZ |
|-------------------------------------|-----------------------------------------------------------------------|
| Predicate Device 3 | |
| Device Name: | MAGNETOM Systems with syngo Expert-i option |
| Manufacturer: | Siemens Medical Solutions, INC. |
| 510(k) Number: | K052423 |
| Regulation Number: | 21 CFR 892.1000 Magnetic Resonance Diagnostic Device |
| Product Code: | LNH |
### Device Description
Digital Expert Access is a variation of the Customer Remote Console cleared under K150193. It is a remote scan assistance solution designed to address the skill variability in technologists and their need for ondemand support by allowing them to interact directly with a remote expert connected to the hospital network. By enabling the collaboration between an Onsite Technologist and Remote Expert, Digital Expert Access helps the onsite technologist to seek guidance and real time support on scanning related queries including but not limited to training, procedure assessment, and scanning parameter management.
Digital Expert Access with Remote Scanning introduces a feature that enables the Remote Expert to initiate a scan and make changes in real time during the scanning session. This remote scan feature is only available when Digital Expert Access is connected to a compatible GE HealthCare MRI system.
Digital Expert Access with Remote Scanning enables the following capabilities for the Onsite Technologist and the Remote Expert:
- 1. Collaborative session between an Onsite Technologist and Remote Expert
- 2. Real-time scanner screen share and live annotation
- ന് Remote console access and control
- 4. Remote Scan Initiation
Digital Expert Access with Remote Scanning is not intended for diagnostic use or patient safety-related management. This solution is not intended to be used by individuals who are not properly trained in the operation of GE HealthCare Medical Imaging systems. Digital Expert Access with Remote Scanning does not directly interface with any patients and requires the Onsite Technologist to be continuously present throughout the scanning procedure. Digital Expert Access with Remote Scanning does not acquire any MRI images, nor does it do any post image processing. All image acquisition and image processing is conducted by the GE HealthCare MRI system.
### Intended Use
Digital Expert Access with Remote Scanning is intended as a remote collaboration tool to view and review MR images, to remotely control MR Imaging Devices and to initiate MRI scans remotely.
#### Indications for Use
{6}------------------------------------------------
# GE HealthCare 510(k) Premarket Notification Submission
Image /page/5/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in white inside. The letters are stylized and connected, and there are small white swirls around the edge of the circle. The logo is simple and recognizable, and it is often used to represent the company.
Digital Expert Access with Remote Scanning is a remote scan assistance solution which allows remote control of an MR Imaging Device including the ability to initiate a scan remotely. This access provides real time communication mechanisms between the remote and onsite users to facilitate the acquisition occurring on the device. Access must be granted by the onsite user operating the system. Images reviewed remotely are not for diagnostic use.
### Technology
The proposed device, Digital Expert Access with Remote Scanning, employs similar fundamental scientific technology as its predicate devices.
### Comparisons
Digital Expert Access with Remote scanning is substantially equivalent to the predicate devices, SIGNA Hero (K213668), Customer Remote Console (CRC) (K150193) and MAGNETOM Systems with syngo Experti option (K052423). The proposed device is intended to be a remote scan assistance solution, allowing the onsite technologist to connect to a remote expert to view/review images, provide real time guidance and initiate an MRI scan remotely. The table below summarizes the substantive feature/technological similarities and differences between the predicate devices and the proposed device:
| Items | Predicate 1:<br>SIGNA Hero<br>(K213668) | Predicate 2:<br>Customer Remote<br>Console (CRC)<br>K150193 | Predicate 3:<br>MAGNETOM Systems<br>with syngo Expert-i<br>option (K052433) | Proposed:<br>Digital Expert Access<br>with Remote Scanning |
|------------------------------------------------------|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| User Interface<br>Functions | PC-based | PC-based | PC-based | Tablet-based/PC-based |
| Operator<br>Interface | Keyboard, mouse | Keyboard, mouse | Keyboard, mouse | Touch LCD, Keyboard,<br>touchpad |
| Remote Viewer<br>interactive<br>control | N/A | User profile<br>(preferences) is<br>associated with an<br>Administrator, User<br>with View role, or<br>User with Guidance<br>role. | The Expert-I status<br>dialog is used to<br>determine if<br>the remote user<br>should have Full<br>Access or View Only<br>access to the syngo<br>Workplace | The onsite user initiates<br>a request through DEA<br>which initiates an<br>acceptance screen on<br>the user interface. Once<br>the request is<br>acknowledged on the<br>scanner, the onsite user<br>must validate the<br>connection by entering<br>the session password<br>for the remote session. |
| Imaging System<br>Modality for<br>Remote<br>Scanning | N/A | N/A | MR | MR |
| Intended Users | Radiologist, Expert<br>Technologist | Radiologist, Expert<br>Technologist | Radiologist, Expert<br>Technologist | Radiologist, Expert<br>Technologist |
Table 5-1: Comparison Table for Digital Expert Access with Remote Scanning and its Predicate Devices
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a blue circle. There are three curved lines, resembling water droplets or stylized commas, surrounding the circle. The logo is simple, recognizable, and has been used by GE for many years.
# GE HealthCare 510(k) Premarket Notification Submission
| Scan Initiation | Scan initiated by<br>onsite technologist | Scan initiated by<br>onsite technologist | Remote user can<br>initiate scan on MR<br>scanners remotely | Remote user can initiate<br>scan on MR scanners<br>remotely |
|--------------------------------------------------|-----------------------------------------------------------------------------------|------------------------------------------------------|-----------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| Patient/Scanner<br>Support during<br>Acquisition | Onsite technologist<br>oversees patient<br>support during<br>scanning acquisition | N/A – CRC is not<br>intended for scan<br>acquisition | Onsite technologist<br>oversees patient<br>support during<br>scanning acquisition | Onsite technologist<br>oversees patient<br>support during scanning<br>acquisition |
## Clinical and Non-Clinical Tests
## Summary of Non-Clinical Tests
Digital Expert Access with Remote Scanning has successfully completed the design control testing per GE HealthCare's quality system. It was designed under the Quality System Regulations of 21CFR 820 and ISO 13485. No new questions of safety and effectiveness and no unexpected test results were observed.
The following quality assurance measures have been applied to the development of the system:
- . Requirement Definition
- Risk Analysis and Control
- Technical Design Reviews
- Formal Design Reviews
- Software Development Lifecycle
- Safety Testing (Verification) ●
- Performance Testing (Verification, Validation)
- Software Release
Digital Expert Access with Remote Scanning verification was conducted to demonstrate proper implementation of its design requirements. The testing and results did not raise any new issues of safety and effectiveness.
In addition, verification and validation testing was conducted on the bench by integrating Digital Expert Access with Remote Scanning on a subset of GE HealthCare MRI systems to ensure a Remote Expert was able to perform remote scanning on GE HealthCare MRI systems.
Design verification and validation testing confirmed that the safety and effectiveness of the device has not been affected. The test plans and results have been executed with acceptable results.
## Summary of Clinical Testing
The subject of this premarket submission, Digital Expert Access with Remote Scanning did not require clinical studies to support substantial equivalence to enable the remote scanning feature on GE HealthCare MRI systems.
## Substantial Equivalence Conclusion
The introduction of Digital Expert Access with Remote Scanning does not result in any new potential safety risks and uses the same technology as the predicate devices, including the initiation of an MRI scan remotely. After analyzing design verification and validation testing on the bench it is the conclusion of GE
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo of General Electric (GE). The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circle. The circle has decorative swirls around the letters, giving it a classic and recognizable appearance.
# GE HealthCare
# 510(k) Premarket Notification Submission
HealthCare that Digital Expert Access with Remote Scanning to be as safe, as effective, and performance is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.