PINION™ PDO Knotless Sutures are indicated for use in general soft tissue approximation where use of an absorbable suture is appropriate. PINION™ PGA-PCL Knotless Sutures are indicated for use in general soft tissue approximation where use of an absorbable suture is appropriate.
Device Story
PINION™ PDO and PINION™ PGA-PCL Knotless Sutures are sterile, synthetic, absorbable monofilament sutures featuring unidirectional or bidirectional barbs. Barbs eliminate the need for manual surgical knots by anchoring in tissue. Devices are supplied with attached stainless steel needles. Used by surgeons for general soft tissue approximation. PDO variant is poly(p-dioxanone); PGA-PCL variant is poly(glycolide-co-caprolactone). Both are MR safe. Barbing process reduces tensile strength compared to non-barbed equivalents, necessitating size adjustments. Devices provide wound support during healing and are absorbed over time.
Clinical Evidence
Bench testing only. Performance evaluated per USP <861> (diameter), <881> (tensile strength), and <871> (needle attachment). Biocompatibility testing per ISO 10993-1 included cytotoxicity, sensitization, systemic toxicity, genotoxicity, hemolysis, implantation, pyrogenicity, and chronic toxicity/carcinogenicity risk assessment.
Technological Characteristics
Materials: Poly(p-dioxanone) (PDO) or poly(glycolide-co-caprolactone) (PGA-PCL). Monofilament, uncoated. Sterilization: Ethylene Oxide. Form factor: Barbed suture with stainless steel needles. MR Safe. Complies with USP requirements for synthetic absorbable sutures.
Indications for Use
Indicated for general soft tissue approximation in patients requiring absorbable sutures.
Regulatory Classification
Identification
An absorbable polydioxanone surgical suture is an absorbable, flexible, sterile, monofilament thread prepared from polyester polymer poly (p-dioxanone) and is intended for use in soft tissue approximation, including pediatric cardiovascular tissue where growth is expected to occur, and ophthalmic surgery. It may be coated or uncoated, undyed or dyed, and with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for the device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
Predicate Devices
STRATAFIX™ Symmetric PDS™ Plus Knotless Tissue Control Device (K182873)
STRATAFIX™ Spiral MONOCRYL™ Knotless Tissue Control Device (K151200)
STRATAFIX™ Spiral MONOCRYL™ Plus Unidirectional Knotless Tissue Control Device (K221744)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size, also in blue.
April 25, 2024
M/s. Meril Endo Surgery Private Limited Asma Shaikh Assistant Manager- Regulatory Affairs Third Floor, E1-E3, Meril Park Survey No. 135/2/B & 174/2. Muktanand Marg. Chala Vapi, Gujarat 396191 India
Re: K232246
Trade/Device Name: 1) PINION™ PDO Knotless Suture - Absorbable Poly (P-Dioxanone) Knotless Suture: 2) PINION™ PGA-PCL Knotless Suture - Absorbable Polyglycolic Acid - Poly (Glycolide-Co-Caprolactone) Knotless Suture Regulation Number: 21 CFR 878.4840 Regulation Name: Absorbable Polydioxanone Surgical Suture Regulatory Class: Class II Product Code: NEW, GAM Dated: July 28, 2023 Received: March 25, 2024
Dear Asma Shaikh:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{2}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for
{3}------------------------------------------------
more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Tek N. Tek N. Lamichhane -S Lamichhane -S Date: 2024.04.25 15:43:22 -04'00' Tek N. Lamichhane, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use
Submission Number (if known)
K232246
Device Name
1) PINION™ PDO Knotless Suture - Absorbable Poly (P-Dioxanone) Knotless Suture; 2) PINION™ PGA-PCL Knotless Suture - Absorbable Poly (Glycolide-Co-Caprolactone) Knotless Suture
Indications for Use (Describe)
1. PINION™ PDO Knotless Sutures are indicated for use in general soft tissue approximation where use of an absorbable suture is appropriate.
2. PINION™ PGA-PCL Knotless Sutures are indicated for use in general soft tissue approximation where use of an absorbable suture is appropriate.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
#### I. SUBMITTER
M/s. Meril Endo Surgery Private Limited. Third Floor, E1 – E3, Meril Park, Survey No. 135/2/B & 174/2, Muktanand Marg, Chala, Vapi – 396191 Gujarat, India. Tel. No: +91-260-3052100, Fax: +91-260-3052125 Web site: www.merillife.com
#### Applicant Information
Mr. Umesh Sharma
General Manager – Quality Assurance / Regulatory Affairs
E-mail: umesh.sharma@merillife.com
#### Primary Correspondent Information
Asma Shaikh
Assistant Manager – Regulatory Affairs
E-mail: asma.shaikh@merillife.com
Date Prepared: April 24th, 2024
### II. SUBJECT DEVICE
| Trade / Proprietary Name | PINION™ PDO Knotless Suture |
|--------------------------|--------------------------------------------|
| Common Name | Suture, Surgical, Absorbable Polydioxanone |
| Classification | Absorbable Polydioxanone Surgical Suture |
| Regulatory Class | II |
| Product Code | NEW |
| Regulation Number | 21 CFR 878.4840 |
| Review Panel | General & Plastic Surgery Devices Panel |
{6}------------------------------------------------
| Trade / Proprietary Name | PINION™ PGA-PCL Knotless Suture |
|--------------------------|------------------------------------------------------|
| Common Name | Suture, Surgical, Absorbable Polyglycolic Acid |
| Classification | Absorbable poly(glycolide/l-lactide) surgical suture |
| Regulatory Class | II |
| Product Code | GAM |
| Regulation Number | 21 CFR 878.4493 |
| Review Panel | General & Plastic Surgery Devices Panel |
### III. PREDICATE DEVICE
| Subject device<br>(K232246) | Trade Name | Predicate device<br>Manufacturer | 510 (K) No. |
|------------------------------------|--------------------------------------------------------------------------------------------|----------------------------------|-------------|
| PINION™ PDO<br>Knotless Suture | STRATAFIX™ Symmetric<br>PDS™ Plus Knotless Tissue<br>Control Device | Ethicon Inc. | K182873 |
| PINION™ PGA-PCL<br>Knotless Suture | STRATAFIX™ Spiral<br>MONOCRYL™ Knotless<br>Tissue Control<br>Device(Primary Predicate) | Ethicon Inc. | K151200 |
| | STRATAFIX™ Spiral<br>MONOCRYL™ Plus<br>Unidirectional Knotless<br>Tissue<br>Control Device | | K221744 |
### IV. Device Description
#### PINION™ PDO KNOTLESS SUTURE
The PINION™ PDO Knotless Suture consists of barbed suture material armed with a surgical needle on each end. Barbs allow for tissue approximation without the need to tie surgical knots.
PINION™ PDO Knotless Suture is a sterile, synthetic, absorbable monofilament suture comprised of Poly (p- Polydioxanone). The empirical molecular formula of which is (G.HგO3)x. PINION™ PDO Knotless Sutures are available in undyed or dyed with D & C violet No.2 (21 CFR 74. 3602) at a level not exceeding 0.3 % by weight of the suture in sizes USP (EP) 3-0 (2), 2-0 (3) & 0 (3.5). It is supplied with the attached stainless steel surgical needles. Polydioxanone has been found to be nonantigenic, nonpyrogenic and to elicit only a slight tissue reaction during absorption.
The PINION™ PDO Knotless Suture is designed with small unidirectional and bidirectional barbs along the length of the device. It consists of an absorbable thread with unidirectional anchors, equipped with a
{7}------------------------------------------------
surgical needle at one end and a fixation tab at the other while the bidirectional anchors equipped with a surgical needle at both end. The anchors and fixation tab design allows for tissue approximation without the need to tie surgical knots.
The USP designations for diameter are applicable to the PINION™ PDO Knotless Suture material prior to barbing. After the creation of barbs, the PINION™ PDO Knotless Suture is identified as one size smaller than the non-barbed suture. This modification reduces the tensile strength of the suture similar to effect of knot tying in non-barbed suture. Therefore, the straight pull tensile strength of the PINION™ PDO Knotless Suture is comparable to the USP knot pull tensile strength for a non barbed suture of the equivalent size. The non-barbed size and equivalent size of the PINION™ PDO Knotless Suture is further clarified in Table 1.
The maximum oversize for non barbed suture material from the USP Size is further detailed in Table 2.
The PINION™ PDO Knotless Suture meets requirements established by the United States Pharmacopoeia (USP) for Synthetic absorbable sutures for needle attachments only.
| PINION™ PDO<br>Knotless Suture | Pre-Barbing Suture Size | Equivalent Non-barbed<br>Suture Size<br>(USP)/Tensile Strength<br>(kgf) |
|--------------------------------|-------------------------|-------------------------------------------------------------------------|
| 3-0 | 2-0 | 3-0/1.77 |
| 2-0 | 0 | 2-0/2.68 |
| 0 | 1 | 0/3.90 |
#### Tensile Strength/Size Equivalency Table 1
Maximum Oversize for non-barbed suture material in diameter (mm) from USP Table 2
| PINION™ PDO<br>Knotless Suture | USP Size Designations<br>(mm) | PINION™ PDO<br>Knotless Suture<br>Diameter Specification<br>(mm) |
|--------------------------------|-------------------------------|------------------------------------------------------------------|
| 3-0 | 0.200-0.249 | 0.300-0.349 |
| 2-0 | 0.300-0.349 | 0.350-0.399 |
| 0 | 0.350-0.399 | 0.400-0.499 |
The PINION™ PDO Knotless Suture is MR Safe.
#### PINION™ PGA-PCL KNOTLESS SUTURE
The PINION™ PGA-PCL Knotless Suture consists of barbed suture material armed with a surgical needle on each end. Barbs allow for tissue approximation without the need to tie surgical knots.
PINION™ PGA-PCL Knotless Suture is a sterile, synthetic, absorbable monofilament suture comprised of poly (glycolide-co-caprolactone).
{8}------------------------------------------------
PINION™ PGA-PCL Knotless Suture is provided undyed or dyed with D & C violet No.2 (21 CFR 74. 3602) at a level not exceeding 0.25 % by weight of the suture in sizes USP (EP) 4-0 (1.5), 3-0 (2) & 2-0 (3). It is supplied with the attached stainless steel surgical needles. The polymer has been found to be nonantigenic, nonpyrogenic and eliciting only a slight tissue reaction during absorption.
The PINION™ PGA-PCL Knotless Suture is designed with small unidirectional and bidirectional barbs along the length of the device. It consists of an absorbable thread with unidirectional anchors, equipped with a surgical needle at one end and a fixation tab at the other while the bidirectional anchors equipped with a surgical needle at both end. The anchors and fixation tab design allows for tissue approximation without the need to tie surgical knots.
The USP designations for diameter are applicable to the PINION™ PGA-PCL Knotless Suture material prior to barbing. After the creation of barbs, the PINION™ PGA-PCL Knotless Suture is identified as one size smaller than the non-barbed suture. This modification reduces the tensile strength of the suture similar to effect of knot tying in non-barbed suture. Therefore, the straight pull tensile strength of the PINION™ PGA-PCL Knotless Suture is comparable to the USP knot pull tensile strength for a non barbed suture of the equivalent size. The non-barbed size and equivalent size of the PINION™ PGA-PCL Knotless Suture is further clarified in Table 1.
The maximum oversize for non barbed suture material from the USP Size is further detailed in Table 2.
The PINION™ PGA-PCL Knotless Suture meets requirements established by the United States Pharmacopoeia (USP) for Synthetic absorbable sutures for needle attachments only.
|--|
| PINION™ PGA-<br>PCL Knotless<br>Suture | Pre-Barbing Suture Size | Equivalent Non-barbed<br>Suture Size<br>(USP)/Tensile Strength<br>(kgf) |
|----------------------------------------|-------------------------|-------------------------------------------------------------------------|
| 4-0 | 3-0 | 4-0/0.95 |
| 3-0 | 2-0 | 3-0/1.77 |
| 2-0 | 0 | 2-0/2.68 |
Maximum Oversize for non-barbed suture material in diameter (mm) from USP Table 2
| PINION™ PGA-<br>PCL Knotless<br>Suture | USP Size Designations<br>(mm) | PINION™ PGA-PCL<br>Knotless Suture<br>Diameter Range (mm) |
|----------------------------------------|-------------------------------|-----------------------------------------------------------|
| 4-0 | 0.150-0.199 | 0.200-0.249 |
| 3-0 | 0.200-0.249 | 0.300-0.349 |
| 2-0 | 0.300-0.349 | 0.350-0.399 |
The PINION™ PGA-PCL Knotless Suture is MR Safe.
{9}------------------------------------------------
### V. Indications for Use
PINION™ PDO Knotless Sutures are indicated for use in general soft tissue approximation where use of an absorbable suture is appropriate.
PINION™ PGA-PCL Knotless Sutures are indicated for use in general soft tissue approximation where use of an absorbable suture is appropriate.
### VI. Comparison of Technological Characteristics with the predicate device
The PINION™ PDO & PINION™ PGA-PCL Knotless Sutures are substantially equivalent to marketed predicate device with respect to intended use and physical characteristics. The PINION™ PGA-PCL Knotless Sutures have same material, mechanism of action, performance and packaging as predicate devices.
The following characteristics were compared between the subject device and the predicate device in order to demonstrate the substantial equivalence.
{10}------------------------------------------------
The below table provides the comparison of key features of the subject device and predicate devices.
#### Table 1: PINION™ PDO Knotless Suture
| Sr.<br>No. | Characteristics | Subject Device | Predicate Device | Safety / Effectiveness | |
|------------|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. | Device Name | PINION™ PDO Knotless Suture | STRATAFIX™ Symmetric PDS™ Plus<br>Knotless Tissue Control Device | - | |
| 2. | Manufacturer | Meril EndoSurgery Pvt. Ltd. | Ethicon Inc. | - | |
| 3. | 510(K) Number | K232246 | K182873 | - | |
| 4. | Common Name | Absorbable Polydioxanone surgical<br>suture | Absorbable Polydioxanone surgical<br>suture | No change | |
| 5. | Class | II | II | No change | |
| 6. | Product Code | NEW | NEW | No change | |
| 7. | Regulation Number | CFR 878.4840 | CFR 878.4840 | No change | |
| 8. | Intended use | PINION™ PDO Knotless Suture is<br>indicated for use in general soft tissue<br>approximation where use of an<br>absorbable suture use is appropriate. | STRATAFIX™ Symmetric PDS™ Plus<br>Knotless Tissue Control Devices are<br>indicated for general soft tissue<br>approximation where use of an<br>absorbable suture is appropriate. | Identical | |
| 9. | For single use only | Single use only | Single use only | No change | |
| 10 | Mode of Actions | Two important characteristics<br>describe the in vivo performance of<br>absorbable sutures: first, tensile<br>strength retention and second, the<br>absorption rate (loss of mass).<br>PINION™ PDO Knotless Suture has<br>been formulated to minimize the<br>variability of these characteristics and<br>to provide wound support through<br>the critical wound healing period and | Two important characteristics describe<br>the in vivo performance of absorbable<br>sutures: first, tensile strength retention<br>and second, the absorption rate (loss<br>of mass). STRATAFIX™ Symmetric<br>PDS™ Plus Knotless Tissue Control<br>Devices has been formulated to<br>minimize the variability of these<br>characteristics and to provide wound<br>support through the critical wound | Identical | |
| | K232246 | | | | |
| Sr.<br>No. | Characteristics | Subject Device | Predicate Device | Safety / Effectiveness | |
| | | an extended healing period.<br>The results of in vitro study using<br>PINION™ PDO Knotless Suture<br>indicate that approximately 70-80% of<br>the original strength remains after<br>two weeks and four weeks of study.<br>At six weeks approximately 40-50% of<br>the original strength is retained.<br>The absorption of PINION™ PDO<br>Knotless Suture is minimal until about<br>120 days and complete absorption<br>usually takes place within 180 days<br>(six months). | healing period and an extended<br>healing period.<br>The results of implantation studies in<br>animals using STRATAFIX™ Symmetric<br>PDS™ Plus Knotless Tissue Control<br>Devices indicates that for sizes 3-0 and<br>larger, approximately 80 % of the<br>original strength remains after two<br>weeks and four weeks of implantation.<br>At six weeks post implantation,<br>approximately 40% to 70% of the<br>original strength is retained. For Size 4-<br>0, approximately 67 % of the original<br>tensile strength remains after two<br>weeks, approximately 50% at 4 weeks,<br>and approximately 37 % at 6 weeks.<br>Data obtained from implantation<br>studies show that absorption of<br>STRATAFIX™ Symmetric PDS™ Plus<br>Knotless Tissue Control Devices is<br>minimal until about 120 days and is<br>essentially complete within 180 days<br>(six months). | | |
| 11 | Material composition | poly (p-dioxanone) | poly (p-dioxanone) | No change | |
| 12 | Body Contact | Tissue/Bone/Blood | Tissue/Bone/Blood | No change | |
| 13 | Shelf Life | 5 Year | 5 Year | No change | |
| 14 | Size availability | 3-0 to 0 | 3-0 to 1 | Identical; Suture size is<br>covered by the<br>Predicate Suture Range | |
| 15 | Absorbable/Non Absorbable | Absorbable | Absorbable | No change | |
| 16 | Braided/ | Monofilament | Monofilament | No change | |
| Sr.<br>No. | Characteristics | Subject Device | Predicate Device | Safety / Effectiveness | |
| | Monofilament | | | | |
| 17 | Dyed / Undyed | Undyed or Dyed Suture Strands | Undyed or Dyed Suture Strands | No change | |
| 18 | Colorant | D&C Violet No. 2 | D&C Violet No. 2 | No change | |
| 19 | Coated / Uncoated | Uncoated | Uncoated | No change | |
| 20 | Types of Barbs | Uni-directional, Bi-directional | Uni-directional, Bi-directional | No change | |
| 21 | No. of barbs per linear length<br>of Suture | 13-17 barbs per cm | 10 barbs per cm | No. of Barbs is more in<br>subject device than<br>predicate device which<br>provides more safety to<br>device. | |
| 22 | Barb Angle | 3-0 | | Different; While there are<br>difference in cut angle<br>between the subject device<br>and the predicate device, these<br>do not raise the new questions<br>of safety and effectiveness<br>as the average barb holding<br>strength for the subject is<br>equivalent to that of the<br>predicate device. | |
| | | 2-0 | $40-50 (∠°)$ | | $23 -28 (∠°)$ |
| | | 0 | $35-45 (∠°)$ |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.