K232171 · Tecnoideal America · KDI · Feb 7, 2024 · Gastroenterology, Urology
Device Facts
Record ID
K232171
Device Name
Purema® H Hemoconcentrator (EtO Sterilized)
Applicant
Tecnoideal America
Product Code
KDI · Gastroenterology, Urology
Decision Date
Feb 7, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Purema® H Hemoconcentrator is designed to remove excess fluid from the blood in order to maintain proper hematocrit and protein concentration during cardiopulmonary bypass and to enable reinfusion of blood remaining in the circuit after bypass. It is intended to be used in a hospital setting, in connection with a suitable circuit for extracorporeal blood circulation and a pump that regulates its flow. There are no other accessories. The Purema® H Hemoconcentrator Models DP07HC, DP09HC, and DP12HC are intended to be used for adult patients.
Device Story
Purema® H Hemoconcentrator is a single-use extracorporeal device for fluid removal during cardiopulmonary bypass. It consists of a hollow-fiber polyethersulfone membrane housed in a polycarbonate cartridge. Blood is pumped through the membrane; a hydrostatic transmembrane pressure gradient drives water and small molecules across the membrane via convective filtration, similar to glomerular filtration. The device is used in hospital settings by clinical staff. It helps maintain hematocrit and protein levels and allows for the reinfusion of blood remaining in the bypass circuit after the procedure. The device is available in three models (DP07HC, DP09HC, DP12HC) varying by membrane surface area.
Clinical Evidence
Bench testing only. Performance evaluated via biocompatibility (ISO 10993-1, ISO 10993-5/11, USP <151>), mechanical integrity (blood compartment/structural integrity), and performance characteristics (solute clearance, albumin sieving coefficient, ultrafiltration coefficient, pressure drop). Hemocompatibility assessed via plasma-free hemoglobin, leukocyte/platelet counts, and clotting assays (ASTM F2382-18, ASTM F2888-19). All tests met predetermined criteria.
Indicated for adult patients undergoing cardiopulmonary bypass requiring removal of excess fluid to maintain hematocrit and protein concentration and to enable reinfusion of residual circuit blood.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 7, 2024
Tecnoideal America % Ahmad Bayat Sr. Director Amarex Clinical Research, LLC 20201 Century Blvd, Fourth Floor Germantown, Maryland 20874
Re: K232171
Trade/Device Name: Purema® H Hemoconcentrator (EtO Sterilized) Regulation Number: 21 CFR 876.5860 Regulation Name: High Permeability Hemodialysis System Regulatory Class: Class II Product Code: KDI Dated: January 9, 2024 Received: January 9, 2024
Dear Ahmad Bayat:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Maura Rooney -S
Maura Rooney Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices
{3}------------------------------------------------
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232171
Device Name Purema® H Hemoconcentrator
## Indications for Use (Describe)
The Purema® H Hemoconcentrator is designed to remove excess fluid from the blood in order to maintain proper hematocrit and protein concentration during cardiopulmonary bypass and to enable reinfusion of blood remaining in the circuit after bypass.
It is intended to be used in a hospital setting, in connection with a suitable circuit for extracorporeal blood circulation and a pump that regulates its flow. There are no other accessories.
The Purema® H Hemoconcentrator Models DP07HC, DP09HC, and DP12HC are intended to be used for adult patients.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
#### 6 510(K) SUMMARY 6.1 OWNER INFORMATION
This Premarket Notification is submitted by:
# Tecnoideal America
7600 Standish Place Derwood, Maryland 20855
Phone: +1 (240) 403 4063 Fax: +1 (240) 801-8875 Email: office@tecnoidealamerica.com Website: https://tecnoidealamerica.com
Tecnoideal Contact: Paul Rodbhajon Email: prodbhajon@tecnoidealamerica.com
# MEDICA S.p.A is the parent company of Tecnoideal America.
## Correspondent Contact Information:
## Ahmad Bayat, M.D.
Tecnoideal America Authorized Representative Sr. Director, Regulatory Affairs Amarex Clinical Research, LLC 20201 Century Blvd, Fourth Floor Germantown, MD 20874
Phone: (301) 956-2523 Fax: (301) 528-2300 Email: ahmadb(@amarexcro.com
Date Prepared:
July 21, 2023
{6}------------------------------------------------
#### 6.2 DEVICE NAME
Device Trade/Proprietary Name: Purema® H Hemoconcentrator Common Name: Hemoconcentrator Classification: Dialyzer, High Permeability with or without Sealed Dialysate System Regulatory Class: Class 2 Product Code: KDI Regulation Number: 21 CFR Part 876.5860
#### DEVICE DESCRIPTION 6.3
Purema® H hemoconcentrators are single use hemoconcentrators containing filters composed of a hollow fiber made of polyethersulfone (PUREMA® H). Purema® H hemoconcentrators are hemoconcentrators of various dimensions that can be used in treatments which require the removal of liquids that are in excess. Blood is pumped through a membrane that has high permeability, and the pressure gradient, through the membrane (TMP) determines the passage of water and molecules with a mechanism that is similar to glomerular filtration (convective mechanism). The fraction of filtrated liquid depends on the osmotic pressure, hydrostatic transmembrane pressure, the surface, membrane permeability and patient hematocrit. The Purema® H hemoconcentrators are designed to be used in a healthcare facility.
Purema® H hemoconcentrators can be sterilized using Ethlyene Oxide (EtO). The EtO sterilized Purema® H hemoconcentrators consist of hollow fiber made of polyethersulfone (PUREMA® H), cartridge, rings, connectors, potting (fiber closure seals), and O-rings.
The EtO sterilized Purema® H Hemoconcentrator is available in three models: DP07HC, and DP12HC. The main differences in these models are their total membrane surface areas. Details the technical features of each model.
| | DP07HC | DP09HC | DP12HC |
|-------------------------------|------------------------------|--------|--------|
| Fiber material | Polyethersulfone (Purema® H) | | |
| Surface area (m²) | 0.7 | 0.9 | 1.2 |
| Blood Ports | Twist Lock | | |
| Filtrate Ports | Female Luer Lock | | |
| Max TMP (mmHg) | 500 | | |
| Max Blood Flow (ml/min) | 300 | 300 | 400 |
| Min. Blood Flow Rate (ml/min) | 100 | 100 | 100 |
| Priming Volume (ml) | 49 | 57 | 75 |
| Sterilization | EtO | | |
| Fiber internal diameter (µm) | 200 | | |
| Fiber external diameter (µm) | 260 | | |
| Fiber Wall thickness (µm) | 30 | | |
| Fiber length (mm) | 140 | | 225 |
| Number of fibers | 9.300 | 11.000 | 9.300 |
Table 1. Technical Features of Purema® H Hemoconcentrator Models DP07HC, and DP12HC
{7}------------------------------------------------
| | DP07HC | DP09HC | DP12HC |
|---------------------------------------|--------------------|--------|--------|
| Filter housing internal diameter (mm) | 38 | | |
| Potting compound | Polyurethane | | |
| Housing material | Polycarbonate (PC) | | |
| Cap material | Polycarbonate (PC) | | |
| Total length (mm) | 180 | | 265 |
| Ulltrafiltration rate (ml/h/mmHg) | 32 | 41 | 55 |
#### 6.4 INTENDED USE
The Purema® H Hemoconcentrator is designed to remove excess fluid from the blood to maintain proper hematocrit and protein concentration during cardiopulmonary bypass and to enable reinfusion of blood remaining in the circuit after bypass.
It is intended to be used in a hospital setting, in connection with a suitable circuit for extracorporeal blood circulation and a pump that regulates its flow. There are no other accessories.
The Purema® H Hemoconcentrator Models DP07HC, and DP12HC are intended to be used for adult patients.
#### 6.5 INDICATIONS OF USE
Same as intended use.
{8}------------------------------------------------
#### 6.6 PREDICATE DEVICE
| Name of Device: | CAPIOX® Hemoconcentrator |
|-------------------|---------------------------------------------------------------------|
| Common Name: | Hemoconcentrator |
| Classification: | Dialyzer, High Permeability with or without Sealed Dialysate System |
| Regulatory Class: | Class 2 |
| Product Code: | KDI |
The proposed predicate device for the Purema® H Hemoconcentrator is the CAPIOX® Hemoconcentrator [510(k) number K973516] from Terumo Medical Corp., a Class II medical device cleared by the FDA and marketed in the US under regulation 21CFR Part 876.5860 (Product Code: KDI). The CAPIOX® Hemoconcentrator was used to established substantial equivalence.
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE 6.7 PREDICATE DEVICE
The EtO sterilized Purema® H hemoconcentrators and the predicate device, the Terumo Capiox Hemoconcentrator have the same intended use and indication for use, as shown in Table 2. Both the Purema® H hemoconcentrators Terumo Capiox Hemoconcentrator have similar specifications as seen in Table 2 below. Both devices are based on the same technological elements: they provide ultra filtration rates which permit the sufficient removal of excess fluid by a slight hydrostatic pressure differential across the membrane without loss of essential plasma proteins.
| | Purema® H Hemoconcentrator<br>(EtO Sterilized) | | | Terumo Capiox Hemoconcentrator<br>[K973516] | |
|-------------------------------|------------------------------------------------|--------|--------|---------------------------------------------------------------------|-------|
| | DP07HC | DP09HC | DP12HC | HC05 | HC11 |
| Fiber material | Polyethersulfone | | | Polysulfone | |
| Surface area (m²) | 0.7 | 0.9 | 1.1 | 0.5 | 1.1 |
| Intended Use | Adult | | | Pediatric | Adult |
| Blood Ports | Twist Lock | | | 1/4" (6,4 mm) slip | |
| Filtrate Ports | Female Luer Lock | | | 1/2" (12,7 mm) (1/4" 6,4 mm adapter) | |
| Max TMP (mmHg) | 500 | | | 500 | |
| Max Blood Flow (ml/min) | 300 | 300 | 400 | 500 | |
| Min. Blood Flow Rate (ml/min) | 100 | 100 | 100 | 100 | 100 |
| Priming Volume (ml) | 49 | 57 | 75 | 34 | 67 |
| Sterilization | EtO | | | EtO | |
| Intended Use | The Purema® H Hemoconcentrator is designed to | | | The CAPIOX Hemoconcentrator is designed to remove excess fluid from | |
# Table 2. Technological Characteristics, Intended Use, and Indications for Use Compared to Predicate Device
{9}------------------------------------------------
| | Purema® H Hemoconcentrator<br>(EtO Sterilized) | | | Terumo Capiox Hemoconcentrator<br>[K973516] | |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| | DP07HC | DP09HC | DP12HC | HC05 | HC11 |
| | remove excess fluid from the<br>blood to maintain proper<br>hematocrit and protein<br>concentration during<br>cardiopulmonary bypass and to<br>enable reinfusion of blood<br>remaining in the circuit after<br>bypass. | remove excess fluid from the<br>blood to maintain proper<br>hematocrit and protein<br>concentration during<br>cardiopulmonary bypass and to<br>enable reinfusion of blood<br>remaining in the circuit after<br>bypass. | remove excess fluid from the<br>blood to maintain proper<br>hematocrit and protein<br>concentration during<br>cardiopulmonary bypass and to<br>enable reinfusion of blood<br>remaining in the circuit after<br>bypass. | the blood in order to maintain proper<br>hematocrit and protein concentration<br>during cardiopulmonary bypass and to<br>enable reinfusion of blood remaining<br>in the circuit after bypass. | |
| | It is intended to be used in a<br>hospital setting, in connection<br>with a suitable circuit for<br>extracorporeal blood circulation<br>and a pump that regulates its flow.<br>There are no other accessories. | | | It is intended to be used during and<br>after surgical procedures requiring<br>cardiopulmonary bypass (up to 6<br>hours) when the removal of excess<br>fluid from blood is required. It should<br>not be used as a dialyzer, hemofilter or<br>other device. | |
| Indications for Use | Same as intended use. | | | Same as intended use. | |
The following biological safety, mechanical characteristics, performance characteristics, and hemocompatibility studies were performed, the results of which demonstrate that the EtO sterilized Purema® H Hemoconcentrators are substantially equivalent to the predicate device, the Terumo Capiox Hemoconcentrator:
Biological Safety:
- In vitro sterility ●
- Non pyrogenicity
Mechanical Characteristics:
- Blood Compartment Integrity ●
- Structural integrity
- Hemoconcentrator blood and filtrate ports
Performance Characteristics:
- Solute clearance
- Sieving coefficient (SC) for albumin
- Ultrafiltration coefficient
- Pressure drop of the blood compartment
Hemocompatibility:
- Plasma free Hemoglobin
- White Blood Cells and Platelets counts
- o Blood Clotting at minimum flow rate
Confidential
{10}------------------------------------------------
#### 6.8 MARKETING HISTORY
Purema® H Hemoconcentrators are CE marked in Europe.
#### 6.9 PERFORMANCE DATA
The biocompatibility evaluation for the EtO sterilized Purema H Hemoconcentrator was conducted in accordance with ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
A summary of the biocompatibility tests condcuated is shown in
| Test | Standard |
|--------------------------------------------------------------------------|----------------------------------------------------------------|
| Cytotoxicity | ISO 10993-5 (2009) and ISO 10993-11<br>(2017) |
| Acute systemic toxicity tests | ISO 10993-5 (2009) and ISO 10993-11<br>(2017)) |
| Intracutaneous reactivity irritation | ISO 10993-12 |
| Guinea pig maximization sensitization<br>(delayed hypersensitivity test) | ISO 10993-12 |
| Material-mediated pyrogen | USP <151> |
| Reverse Mutation Assay using Bacteria | ISO 10993-1; ISO 10993-3; ISO 10993-12;<br>and ISO/TR 10993-33 |
| Mammalian Cell Gene Mutation Assay | ISO 10993-1; 10993-3; and 10993-12 |
| Sc5b-9 complement activation assay | FDA Biocompatibility Guidance and ISO<br>10993-4:2017 |
| Non-activated Partial Thromboplastin Time (PTT)<br>assay | ASTM F2382-18 |
| Platelet and Leukocyte Count assay | ASTM F2888-19 |
All testing met predetermined testing criteria.
#### 6.10 CONCLUSIONS
The comparison of performances, technological characteristics, and intended use/indications for use between the EtO Sterilized Purema® H Hemoconcentrator (models DP07HC, and DP12HC) and the comparator, the CAPIOX® Hemoconcentrator (models HC05 and HC11), indicate that the Purema® H Hemoconcentrator is substantially equivalent to the predicate device.
{11}------------------------------------------------
#### 7 FINANCIAL CERTIFICATION OR DISCLOSURE STATEMENT
This section is not applicable.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.