K231907 · Medline Industries, LP · MEG · Oct 18, 2023 · General Hospital
Device Facts
Record ID
K231907
Device Name
Medline Safety Insulin and TB Syringes
Applicant
Medline Industries, LP
Product Code
MEG · General Hospital
Decision Date
Oct 18, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Medline Safety Insulin Syringe is intended for the delivery of U-100 insulin. The needle stick prevention feature of the device, once activated, guards against accidental needle-sticks. The Medline TB (Tuberculin) Syringe is intended for the delivery of tuberculin. The needle stick prevention feature of the device, once activated, guards against accidental needle-sticks.
Device Story
Sterile, single-use manual syringes for delivery of U-100 insulin or tuberculin. Components: barrel, plunger, permanently attached single lumen needle, needle cap, and manually operated safety shield. Operation: user pulls/pushes plunger to draw/dispense medication; after use, user manually extends safety shield over needle and locks it via twisting motion. Used in clinical or home settings by healthcare providers or patients. Safety shield prevents accidental needle-sticks. Benefits: accurate medication delivery and reduced risk of sharps injuries.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including functional performance (ISO 8537, ISO 7886-1, ISO 23908), chemical safety (USP <788>, USP <85>), and biocompatibility (ISO 10993 series). Simulated clinical usability testing performed per FDA guidance.
Technological Characteristics
Materials: Polypropylene and polyethylene. Energy source: Manual mechanical. Sterilization: Ethylene Oxide (ISO 11135:2014). Connectivity: None. Form factor: Syringe with integrated safety shield. Standards: ISO 8537 (insulin syringes), ISO 7886-1 (manual syringes), ISO 7864 (needles), ISO 9626 (tubing), ISO 23908 (sharps protection).
Indications for Use
Indicated for delivery of U-100 insulin or tuberculin in patients requiring such therapy. Includes needle-stick prevention feature for user safety.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
Kendall Monoject® Magellan Insulin and Tuberculin Safety Syringe (K061492)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 18, 2023
Medline Industries, LP Kelsey Closen Regulatory Affairs Specialist Three Lakes Drive Northfield. Illinois 60093
Re: K231907
Trade/Device Name: Medline Safety Insulin and TB Syringes Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF, MEG Dated: July 21, 2023 Received: July 21, 2023
Dear Kelsey Closen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
erely, Juliane C. Lessard -S
Juliane C. Lessard, Ph.D. Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K231907
Device Name Medline Safety Insulin and TB Syringes
Indications for Use (Describe)
The Medline Safety Insulin Syringe is intended for the delivery of U-100 insulin. The needle stick prevention feature of the device, once activated, guards against accidental needle-sticks.
The Medline TB (Tuberculin) Syringe is intended for the delivery of tuberculin. The needle stick prevention feature of the device, once activated, guards against accidental needle-sticks.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|----------------------------------------------------------|
| <div></div> | <div></div> Prescription Use (Part 21 CFR 801 Subpart D) |
| <div></div> | <div></div> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is the logo for Medline. The logo is set on a blue background. The word "MEDLINE" is in white, bold, sans-serif font. Above the word is a white starburst-like shape with four points.
ledline Industries. L Three Lakes Drive Northfield, IL 60093
# K231907- 510(k) SUMMARY
#### Submitter / 510(k) Sponsor
Medline Industries, LP Three Lakes Drive Northfield, IL 60093
#### Contact Person
Contact Person: Kelsey Closen, Regulatory Affairs Specialist Phone: 847-949-2283 Email: KClosen@medline.com
Summary Preparation Date October 19th 2023
Type of 510(k) Submission Traditional
#### Device Name / Classification
Trade Name: Medline Safety Insulin and TB Syringes Common Name: Tuberculin and Insulin Safety Syringes Classification Name: Syringe, Antistick and Syringe, Piston Product Code: MEG and FMF Classification Panel: General Hospital Regulatory Class: II Regulation Number: 21 CFR 880.5860
#### Predicate Device
Predicate Device: K061492, Kendall Monoject® Magellan Insulin and Tuberculin Safety Syringe
#### Device Description
The Medline Safety Insulin Syringe is a sterile, single-use device, intended for prescription-use only, and is labeled for delivery of U-100 Insulin. The proposed device consists of the following components: (1) a syringe barrel with a permanently attached single lumen needle; (2) a plunger; (3) a needle cap; and (4) a manually operated safety feature (i.e. shield) at the needle end of the syringe. The proposed device functions by mechanical action to deliver insulin. The syringe consists of a plunger that fits within the syringe barrel. The plunger can be linearly pulled and pushed along the inside of the barrel, allowing the syringe to take in and dispense insulin through the needle. The segmented safety shield is designed to be extended over the needle and locked. Activation is performed by a manual operation by extending
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Medline logo. The logo is set against a blue background. A white, stylized star shape is in the center, with the word "MEDLINE" in white, bold letters positioned horizontally across the star. The star shape has four points, with the top and bottom points being more elongated than the side points.
**Medline Industries, LP**
Three Lakes Drive
Northfield, IL 60093
the shield to cover the needle. Once activated, the safety shield is securely and permanently locked with a twisting motion. The Medline Safety Insulin Syringe barrel contains major and minor graduated ink markings and will be available in a 1 ml (100 units), and 0.5 ml (50 units) syringe volumes with three different needle gauge sizes.
The Medline TB Syringe is a sterile, single-use device, intended for prescription-use only, and is labeled for Tuberculin use. The proposed device consists of the following components: (1) a syringe barrel with a permanently attached single lumen needle; (2) a plunger; (3) a needle cap; and (4) a manually operated safety feature (i.e. shield) at the needle end of the syringe. The proposed device functions by mechanical action to deliver tuberculin. The syringe consists of a plunger that fits within the syringe barrel. The plunger can be linearly pulled and pushed along the inside of the barrel, allowing the syringe to take in and dispense tuberculin through the needle. The segmented safety shield is designed to be extended over the needle and locked. Activation is performed by a manual operation by extending the shield to cover the needle. Once activated, the safety shield is securely and permanently locked with a twisting motion. The syringe barrel contains graduated markings, with major graduation markings every 0.05mL and minor graduation markings every 0.01mL. The Medline TB Syringe will be available in a 1 mL syringe volume with three different needle sizes.
#### Indications for Use
The Medline Safety Insulin Syringe is intended for the delivery of U-100 insulin. The needle stick prevention feature of the device, once activated, guards against accidental needle-sticks.
The Medline TB (Tuberculin) Syringe is intended for the delivery of tuberculin. The needle stick prevention feature of the device, once activated, guards against accidental needle-sticks.
| TABLE 1: COMPARISON OF PROPOSED AND PREDICATE DEVICES | | | |
|-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Characteristic | Proposed Device | Predicate Device | Comparison<br>Analysis |
| | Medline Safety Insulin and TB<br>Syringes<br>K231907 | Kendall Monoject Magellan and<br>Tuberculin Insulin Safety Syringe<br>K061492 | |
| Product Code | MEG and FMF | FMF | Similar-<br>See comment #1 |
| Regulation Number | 21CFR 880.5860 | 21CFR 880.5860 | Same |
| Intended Use | Insulin Syringe:<br>The device is intended for the delivery<br>of U-100 insulin. The needle stick<br>prevention feature of the device, once<br>activated, guards against accidental<br>needle-sticks. | Insulin Syringe:<br>The device is intended for the delivery<br>of U-100 insulin. The needle stick<br>prevention feature of the device, once<br>activated, guards against accidental<br>needle-sticks. | Same |
| | | | |
| | TB Syringe:<br>The device is intended for the delivery<br>of Tuberculin. The needle stick<br>prevention feature of the device, once<br>activated, guards against accidental<br>needle-sticks. | TB Syringe:<br>The device is intended for the delivery<br>of Tuberculin. The needle stick<br>prevention feature of the device, once<br>activated, guards against accidental<br>needle-sticks. | |
| Design Features | -Manually operated safety feature to<br>prevent accidental needle sticks.<br>-Graduation markings | -Manually operated safety feature to<br>prevent accidental needle sticks.<br>-Graduation markings | Same |
| Mechanism of<br>Action | Insulin Syringe:<br>Mechanical delivery of U-100 insulin.<br>Needle-stick prevention feature is<br>mannually activated by the user by a<br>finger-tip or thumb operation.<br><br>TB Syringe:<br>Mechanical delivery of tuberculin.<br>Needle-stick prevention feature is<br>manually activated by the user by a<br>finger-tip or thumb operation | Insulin Syringe:<br>Mechanical delivery of U-100 insulin.<br>Needle-stick prevention feature is<br>mannually activated by the user by a<br>finger-tip or thumb operation.<br><br>TB Syringe:<br>Mechanical delivery of tuberculin.<br>Needle-stick prevention feature is<br>manually activated by the user by a<br>finger-tip or thumb operation | Same |
| Components | Sterile Syringe<br>Single lumen needle<br>Safety sheild/feature | Sterile Syringe<br>Single lumen needle<br>Safety sheild/feature | Same |
| Prescription vs.<br>OTC | Prescription Use Only | Prescription Use Only | Same |
| Sterile vs. Non-<br>Sterile | Sterile | Sterile | Same |
| Sterilization Method | EO | EO | Same |
| Shelf life | 5 years | 5 years | Same |
| Single Use vs.<br>Reusable | Single Use | Single Use | Same |
| Non-pyrogenic | Yes | Yes | Same |
| Size Configurations | Insulin Syringes:<br>1ml Syringe w/ Needle, 29G x 0.5in<br>1ml Syringe w/ Needle, 30G x 5/16in<br>1ml Syringe w/ Needle, 31G x 5/16in<br>1ml Syringe w/ Needle, 31G x 6mm<br>0.5ml Syringe w/ Needle, 29G x 0.5in<br>0.5ml Syringe w/ Needle, 30G x 5/16in<br>0.5ml Syringe w/ Needle, 31G x 5/16in<br>0.5ml Syringe w/ Needle, 31G x 6mm<br><br>TB Syringes: | Insulin Syringes:<br>1ml Syringe w/ Needle, 29G x 0.5in<br>1ml Syringe w/ Needle, 30G x 5/16in<br>0.5ml Syringe w/ Needle, 29G x 0.5in<br>0.5ml Syringe w/ Needle, 30G x 5/16in<br>0.3ml Syringe w/ Needle, 29G x 0.5in<br>0.3 ml Syringe w/ Needle, 30G x 5/16in<br><br>TB Syringes:<br>1 mL syringe w/ needle 25Gx5/8"<br>(0.508mm x 1.6cm) | Different –<br>See comment # 2<br>Reference<br>agency cleared<br>Iml syringes<br>with needle in<br>31G x6mm and<br>0.5ml syringe<br>with needle,<br>31G x6mm<br>(K220061) |
| | 1 mL Tuberculin Safety Syringe with<br>Needle, 25Gx5/8" (0.508 x 1.6 cm)<br>1 mL Tuberculin Safety Syringe with<br>Needle, 27Gx1/2" (0.356 mm x 1.3 cm)<br>1 mL Tuberculin Safety Syringe with<br>Needle, 28 G x 1/2" (0.356 mm x 1.3<br>cm | 1 mL syringe w/ needle, 28 Gx1/2"<br>(0.356mm x 1.3cm) | |
| Lubricant | Silicone | Silicone | Same |
| Needle Cap Color | Insulin: Orange<br>TB: Red, Grey and Brown | Insulin: Orange<br>TB: Orange, Red, Brown and Grey | Similar - see<br>comment #3 |
| Device Materials<br>(barrel and shield) | Polypropylene and Polyethylene | Polypropylene and Polyethylene | Same |
# Summary of Technological Characteristics
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image features the Medline logo, which is set against a blue background. The logo consists of the word "MEDLINE" in bold, white sans-serif font, positioned to the left of a stylized white starburst symbol. The starburst is oriented vertically, with its points extending upwards and downwards. A small portion of a gray rectangle is visible in the upper right corner of the image.
Medline Industries, LP Three Lakes Drive Northfield, IL 60093
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for Medline. The logo consists of the word "MEDLINE" in a sans-serif font, with a white cross-like symbol behind the text. The background is a solid dark blue color. The logo is simple and clean, with a professional look.
ledline Industries. LF Three Lakes Drive Northfield. IL 60093
Discussions of differences in technological characteristics
- . Comment #1 Medline also has MEG as a product code since the proposed devices make an anti-stick claim. Both of these product codes are defined by the same FDA requlation number (21 CFR 880.5860), and this difference does not impact Medline's ability to claim substantial equivalence.
- Comment # 2- Reference device cleared TB syringe 1mL with needle, 27Gx ½", Insulin syringe 1mL with needle 31G x 5/16″ and 0.5ml, 31Gx5/16" (K153537). Reference device cleared 1mL syringes with needle in 31Gx 6mm (K220061). The addition of 27Gx1/2", Insulin syringe 1mL with needle 31G x 5/16 including the 1ml with needle in 31Gx6mm does not raise different questions of safety and effectiveness than the predicate device.
- . Comment #3-The TB syringe cap colors are different among the proposed device and predicate device. Biocompatibility on the caps was conducted and supports that this difference does not impact the safety and effectiveness of the device to claim substantially equivalent to the predicate.
#### Shelf Life and Sterilization
The proposed Medline Safety Insulin and TB Syringe is terminally sterilized by Ethylene Oxide (EO). The sterilization validation for the proposed device has been conducted in accordance with ISO 11135:2014, Sterilization of Health Care Products – Ethylene Oxide – Requirements for Development, Validation, and Routine Control of a Sterilization Process for Medical Devices, to ensure a Sterility Assurance Level (SAL) of 1 x 10%. The proposed device has also been evaluated for EO and Ethylene Chlorohydrin (ECH) residuals in accordance with ISO 10993-7:2008(R)2012, Biological Evaluation of Medical Devices – Part 7: Ethylene Oxide Sterilization Residuals.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image features the logo for Medline, a healthcare company. The logo consists of the word "MEDLINE" in a bold, sans-serif font, with a stylized white cross symbol superimposed over the text. The background is a solid dark blue color, providing a strong contrast to the white text and cross.
ledline Industries. Three Lakes Drive Northfield II 60093
Additionally, in accordance with ASTM F1980-16, Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices, aging studies have been conducted to verify a five-year shelf life of the subject device and ensure that its functionality and sterility are successfully maintained throughout the duration of this shelf life. For additional information on the sterilization and shelf life of the proposed device, please refer to Section 14 of this submission. Test articles also underwent transportation-testing method, ISTA-3A:2008 and underwent a drop, vibration with top load, vibration without top load and a second drop test. After distribution stimulation, the test articles were tested to ASTM F1929:2015. Resistance to Dye Penetration.
#### Summary of Non-Clinical Performance Testing
Non-clinical verification of the Medline Safety Insulin and TB Syringes has been conducted to evaluate its safety, performance, and functionality. The results of these tests have demonstrated the overall safety of the proposed device and its effectiveness in accordance with relevant test methods, and ultimately support a substantial equivalence determination. Particularly, the following was conducted to adequately demonstrate the effectiveness of the proposed device in accordance with relevant test methods cited below:
#### Chemical Safety Testing
- Particulate Matter Testing in accordance with USP <788> Particulate Matter in Injections ●
- Phthalate Testing .
- Limulus Amebocyte Lysate (LAL) Bacterial Endotoxin Testing in accordance with USP <85> ● Bacterial Endotoxin Testing
#### Functional Performance Testing
- . Visual Appearance Testing in accordance with ASTM F1886:2016 Standard Test Method For Determining Integrity Of Seals For Flexible Packaging By Visual Inspection
- Package Integrity Testing in accordance with ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
- Seal Strength Testing in accordance with ASTM F88-21 Test Method for Seal Strength of Flexible ● Barrier Materials
- ISO 8537:2016 Sterile Single-Use Syringes, With or Without Needles, for Insulin ●
- ISO 23908:2011 Sharps Injury Protection Reguirements and Test Methods Sharps Protection ● Features for Single Use Hypodermic Needles, Introducers for Catheters and Needles Uses for Blood Sampling
- ISO 7864:2016 Sterile Hypodermic Needles for Single Use Requirements and Test Methods ●
- ISO 9626:2016 Stainless Steel Needle Tubing for the Manufacturing of Medical Devices -● Requirements and Test Methods
- . ISO 7886-1:2017 Sterile Hypodermic Syringes for Single Use – Part 1: Syringes for Manual Use
Usability Testing
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the Medline logo, which features the word "MEDLINE" in white, sans-serif font, stacked on top of a white, stylized starburst or cross shape. The background is a solid, dark blue color. A small portion of a gray bar is visible on the right side of the image, partially obscuring the blue background.
ledline Industries
In accordance with the FDA guidance document Medical Devices with Sharps Injury Prevention Features, a simulated clinical use (i.e. usability) testing was conducted using the Medline Safety Insulin and TB Syringes.
#### Biocompatibility Testing
The biocompatibility evaluation for the Medline Safety Insulin and TB Syringes was conducted in accordance with ANSI/AAMI/ISO 10993-1:2018 Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process, as recognized by FDA. The proposed device is classified as an externally communicating with prolonged, indirect blood contact. Final biocompatibility test protocols and reports are available for the Agency's review in Appendix D.
- Cytotoxicity in accordance with ISO 10993-5:2009 Biological Evaluation of Medical Devices-Part 5: Tests for in vitro Cytotoxicity
- . Sensitization in accordance with ISO 10993-10:2010 Biological Evaluation of Medical Devices-Part 10: Test for Irritation and Skin Sensitization
- Irritation in accordance with ISO 10993-10:2010 Biological Evaluation of Medical Devices-Part 10: Tests for Irritation and Skin Sensitization
- . Acute Systemic Toxicity in accordance with ISO 10993-11:2017 Biological Evaluation of Medical Devices-Part 11: Tests for Systemic Toxicity
- . Hemolysis Assay in accordance with ASTM F756-17 Standard Practice for Assessment of Hemolytic Properties of Materials and ISO 10993-Part 4:2017 Biological Evaluation of Medical Devices-Part 4: Selection of Tests for Interactions with Blood
- Material-Mediated Pyrogenicity in accordance with USP <151> Rabbit Pyrogen Test as ● recommended in ISO 10993-11:2017 Biological Evaluation of Medical Devices-Part 11: Tests for Systemic Toxicity
- Subacute/ Subchronic Toxicity in accordance with ISO 10993-11:2017 Biological Evaluation of Medical Devices-Part 11: Tests for Systemic Toxicity
- Complement Activation in accordance with ISO 10993-4: 2017 Biological Evaluation of Medical . Devices-Part 4: Selection of Tests for Interactions with Blood.
- Partial Thromboplastin Time in accordance with ISO 10993-4 Biological Evaluation of Medical . Devices-Part 4: Selection of Tests for Interactions with Blood.
- Platelet and Leukocyte Counts in accordance with ISO 10993-4:2017 Standard Biological ● Evaluation of Medical Devices, Part 4 – Selection of Tests for Interaction with Blood.
#### Summary of Clinical Testing
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the Medline logo. The logo is set against a blue square background. The word "MEDLINE" is in bold, white letters. A white star-like symbol is behind the word "MEDLINE".
Medline Industries, LP Three Lakes Drive Northfield, IL 60093
Not applicable.
### Conclusion
The difference between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The Medline Safety Insulin and TB Syringes are substantially equivalent to the Kendall Monoject Magellan Insulin and Tuberculin Safety Syringe with respect to the indications for use, target populations, treatment method, and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.