K220061 · Berpu Medical Technology Co., Ltd. · MEG · Aug 23, 2022 · General Hospital
Device Facts
Record ID
K220061
Device Name
Safety Insulin Syringes
Applicant
Berpu Medical Technology Co., Ltd.
Product Code
MEG · General Hospital
Decision Date
Aug 23, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Safety Insulin Syringe is a sterile, single use and non-reusable syringe intended for injection of U-100 insulin into the body, while reducing the risk of sharps injuries and the potential for insulin syringe reuse.
Device Story
Sterile, single-use, non-reusable piston syringe with permanently attached needle; available in 0.3ml, 0.5ml, 1ml volumes with 26G-34G needles. Device includes manual safety protector; after insulin injection, user manually slides protector over needle until locked to prevent accidental sticks. Used in clinical or home settings by patients or healthcare providers. Output is insulin delivery; safety mechanism provides physical barrier post-injection. Benefits include accurate insulin dosing and reduced risk of sharps injuries and syringe reuse.
Clinical Evidence
No clinical study included. Bench testing only. Performance validated per ISO 9626, ISO 7864, ISO 8537, and ISO 23908. Biocompatibility testing (ISO 10993) included cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, and hemocompatibility. Simulated clinical use study performed on 500 samples per FDA guidance for sharps injury prevention features; results met all pre-established criteria.
Technological Characteristics
Materials: Polypropylene (cap, barrel, plunger, safety protector, hub, end cap), Stainless Steel 304 (needle), Polyisoprene Rubber (piston), Polydimethylsiloxane (lubricant), UV Curing Adhesive. Sterilization: Ethylene Oxide (ISO 11135). Standards: ISO 9626, ISO 7864, ISO 8537, ISO 23908, ISO 10993, USP <788>, USP <85>, ASTM F1980-16, ASTM D4169-16, ASTM F1886/F1886M-16, ASTM F88/F88M-15, ASTM F1929-15.
Indications for Use
Indicated for adult and pediatric patients requiring U-100 insulin injections. Intended for single-use, non-reusable administration to reduce sharps injury risk.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 23, 2022
Berpu Medical Technology Co., Ltd. David Yu Management Representative No. 14 Xingji Road, Yongxiing Street Wenzhou. Zhejiang 325000 China
#### Re: K220061
Trade/Device Name: Safety Insulin Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF, MEG Dated: July 26, 2022 Received: July 26, 2022
Dear David Yu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K220061
Device Name Safety Insulin Syringe
Indications for Use (Describe)
The Safety Insulin Syringe is a sterile, single use and non-reusable syringe intended for injection of U-100 insulin into the body, while reducing the risk of sharps injuries and the potential for insulin syringe reuse.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K220061
# 510kSummary
#### I. Submitter
Berpu Medical Technology Co., Ltd. NO.14 Xingji Road, Yongxing Street, Longwan District, Wenzhou, Zhejiang, China 325000
Establishment Registration Number: 3004496829
Contact person: David Yu Position: Management representative Tel.: +86-18868131179 Fax: 0577-86630389 E-mail: 274269118@qq.com
Preparation date:
#### Proposed Device II.
| Trade Name of Device: | Safety Insulin Syringe |
|-----------------------|------------------------|
| Common name: | Piston Syringe |
| Regulation Number: | 21 CFR 880.5860 |
| Regulatory Class: | Class II |
| Product code: | FMF, MEG |
| Review Panel | General Hospital |
#### III. Predicate Devices
K202570
Insulin Syringe with Safety Retractable
#### IV .Device description
The Safety Insulin Syringe is a sterile, single use and non-reusable syringe with a permanently attached needle, which is intended for injection of U-100 insulin into the body, while reducing the risk of sharps injuries and the potential for insulin syringe reuse. The proposed device is available in 0.3ml, 0.5ml, 1ml volumes with a 26G-34G gauge needle. The safety feature will be manually activated to cover the needle immediately after use to minimize risk of accidental needle sticks.
The proposed device is sterilized by Ethylene Oxide to achieve a SAL of 10 ° and supplied in
1
{4}------------------------------------------------
immediate package which could maintain the sterility of the device during the shelf life of 5 years.
The proposed syringes are available in different combination of syringe volumes and/or needle sizes (refer to Table 1).
| Syringe volume | Needle sizes | | |
|----------------|------------------------|----------------|------|
| | Needle length (mm) | Needle gauge | Wall |
| 0.3ml | 4, 5, 6, 8, 9, 10, 12, | 30G, 31G, 32G, | RW |
| 0.5ml | 13 | 33G, 34G | TW |
| 1ml | | | ETW |
| 0.3ml | 4, 5, 6, 8, 9, 10, 12, | 26G, 27G, 28G, | RW |
| 0.5ml | 13 | 29G | TW |
| 1ml | | | |
Table 1 specification of proposed device
#### V. Indication for use
| Characteristic | Proposed Device<br>Safety Insulin Syringe<br>K220061 | Predicate Device<br>Insulin Syringe with Safety Retractable<br>K202570 |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use | The Safety Insulin Syringe is a sterile, single use and non-reusable syringe intended for injection of U-100 insulin into the body, while reducing the risk of sharps injuries and the potential for insulin syringe reuse. | The Insulin Syringe with Safety Retractable is a sterile, single use and non-reusable, manual retractable safety insulin syringe intended for injection of U-100 insulin into the body, while reducing the risk of sharps injuries and the potential for insulin syringe reuse |
Discussions of differences in Indications for Use statement
There are only editorial differences to the indications for use statement between the predicate and the subject device which do not change the indications.
# VI. Comparison of technological characteristics with the predicate devices
The comparison and discussion between the subject device and the predicate devices are listed in below table 2:
| Item | Proposed device<br>(K220061) | Predicate device<br>(K202570) | Discussion |
|------|------------------------------|-------------------------------|------------|
| | | | |
| Table 2. Technological Characteristics | |
|----------------------------------------|--|
|----------------------------------------|--|
{5}------------------------------------------------
| Prescription/ove<br>r-the-counter<br>use | Prescription Use | Prescription Use | Same |
|------------------------------------------|-----------------------------------------------|------------------------------------------------|-----------------------------------------------------------------|
| Configuration<br>and material | 1) (1) needle cap<br>(Polypropylene) | 1) Needle Cap<br>(Polypropylene) | Same |
| | 2) (2) needle tube (Stainless<br>Steel 304) | 2) Needle Tube<br>(SUS304) | Same |
| | 3) (3) Rubber piston<br>(Polyisoprene Rubber) | 3) Plunger Stopper<br>(Polyisoprene<br>Rubber) | It's the same<br>component,<br>just with a<br>different<br>name |
| | 4) barrel (Polypropylene) | 4) Barrel<br>(Polypropylene) | Same |
| | 5) plunger (Polypropylene) | 5) Plunger<br>(Polypropylene) | Same |
| | 6) Safety protector<br>(Polypropylene) | 6) Safety-shield<br>(Polypropylene) | It's the same<br>component,<br>just with a<br>different<br>name |
| | 7) Fixed hub<br>(Polypropylene) | / | Different<br>See comment<br>#1 |
| | 8) end cap (Polypropylene) | / | |
| | 9) Lubricant<br>(Polydimethylsiloxane) | unknown | |
| | 10) Adhesive (UV Curing<br>Adhesive) | unknown | |
| Size | 0.3ml, 0.5ml, 1ml | 0.5ml, 1ml | Different- |
| Needle Gauge | 26G to 34G | 27G to 31G | See comment<br>#2 |
| Needle Length<br>(mm) | 4, 5, 6, 8, 9, 10, 12, 13 | 1/2", 5/16", 1/4" | |
| Wall type | RW, TW, ETW | / | |
{6}------------------------------------------------
| Needle<br>performance | ISO 9626<br>ISO 7864 | ISO 9626<br>ISO 7864 | Same |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| Syringe<br>performance | ISO 8537 | ISO 8537 | Same |
| Safety feature<br>performance | The force to activate the<br>safety mechanism is less than<br>5N<br><br>The force that the safety<br>mechanism is destroyed is<br>greater than 20N | In-Safe mode force: not<br>be more than 5N<br><br>Resist force:<br>60s with 20N weights,<br>and the protective device<br>shall not be opened | Difference<br>See comment<br># 3 |
| Intended user<br>population | Adult and Pediatric | Adult and Pediatric | Same |
# Discussions of differences in technological characteristics
#### Comment# 1
The configurations and raw materials of Safety Insulin Syringe are different from predicate device. These differences do not raise new questions of safety or effectiveness. The biocompatibility for the subject device has been evaluated and the results comply with the requirements of ISO 10993.
# Comment# 2
The needle gauge and length for proposed devices is different from the predicate devices. The differences do not raise new questions of safety or effectiveness.
# Comment# 3
The Safety feature performance specification for predicate device is a little difference. The safety mechanism of proposed device is implemented by the upper and lower buckle of the vertical axis (After the injection is completed, manually pull the safety protector toward the needle tip until it cannot slide anymore, the protection mechanism is triggered, the safety protector can't return, and the needle tube is completely protected by the safety protector). while the safety mechanism of predicate device is implemented by the convex points of the vertical axis + horizontal axis. However, the safety feature performance test for proposed device has been evaluated and the test result conforms to requirements of ISO 23908:2011 standards. Therefore, the differences on configuration and materials does not affect substantially equivalence.
{7}------------------------------------------------
#### VII .Non-Clinical Testing
The sterile single lumen hypodermic needles described in this summary were tested and demonstrated to be in conformance with the following FDA recognized standards:
- ISO 9626 Second edition: Stainless steel needle tubing for the manufacture of medical . devices- Requirements and methods
- ISO 7864:2016 Sterile hypodermic needles for single use-Requirements and test ● methods
- ISO 8537: 2016 Sterile single-use syringes, with or without needle, for insulin. ●
- ISO 23908- First edition: Sharps injury protection- Requirements and test methods-● Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling
#### Biocompatibility Testing:
In accordance with ISO 10993-1, the needle is classified as: Externally Communicating Device, Blood Path Indirect, Prolonged Contact (>24hours to 30days). The following testing was conducted:
- . Cytotoxicity
- Sensitization ●
- Irritation or intracutaneous reactivity ●
- Acute systemic toxicity .
- Pyrogenicity .
- . Hemocompatibility
- Subacute Systemic Toxicity .
Particulate Testing, USP<788>
#### Sterility, Shipping and Shelf Life
The sterilization method has been validated per ISO 11135, which has thereby determined the routine control and monitoring parameters. The shelf life of the Safety Insulin Syringe is 5 years, determined based on stability studies which includes accelerated aging.
Sterilization and shelf life testing listed were performed on the proposed device.
| Item | Standard |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| EO residue | ISO 10993-7:2008 |
| ECH residue | ISO 10993-7:2008 |
| Bacteria Endotoxin Limit | USP <85> |
| Shelf Life Evaluation | Physical, Mechanical, Chemical, Package Tests were<br>performed on aging samples to verify the claimed shelf<br>life of the device. Shelf-Life of 5 years is validated<br>using ASTM F1980-16 |
{8}------------------------------------------------
- Simulated transportation Test was conducted in accordance with ASTM D4169-16 used ● final packaged sterile device after accelerated aging, which were stored for 72 hours under the environment of temperature and humidity (-35 °C±2°C, 24 hours, 40°C±2°C, 90%RH±5%RH, 24 hours, 60°C±2°C, 24 hours).
- Sterile Barrier Packaging Testing performed on the proposed device: ●
- Visual Inspection in accordance with ASTM F1886 / F1886M-16 1
- Seal Strength in accordance with ASTM F88/F88M-15 .
- Dye Penetration in accordance with ASTM F1929-15 1
#### Simulated Clinical Use
A simulated clinical use study was performed on 500 device samples for the safety insulin syringe according to FDA Guidance, Guidance for Industry and FDA Staff: Medical Device with Sharps Injury Prevention Feature, issued on August 9, 2005 and ISO 23908 to evaluate the safety mechanism of the proposed device. The results demonstrated that the proposed device met the pre-established criteria.
# VIII. Clinical Testing
No clinical study is included in this submission.
# IX. Conclusion
The proposed device has the same indication for use and has similar design features and technological characteristics as the predicate device. Performance testing data demonstrates that the proposed device is substantially equivalent. Accordingly, the proposed device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.