K231903 · Orthocon, Inc. · MQV · Aug 25, 2023 · Orthopedic
Device Facts
Record ID
K231903
Device Name
Montage-QS Settable, Resorbable Bone Putty
Applicant
Orthocon, Inc.
Product Code
MQV · Orthopedic
Decision Date
Aug 25, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Orthocon Montage-QS Settable, Resorbable Bone Putty is indicated to fill bony voids or gaps in the skeletal system (i.e., extremities and pelvis). These defects may be surgically created, or osseous defects created as the result of traumatic injury to the bone. Montage-QS is indicated only for filling bony voids or gaps that are not intrinsic to the integrity of the bony structure. When hardened in situ, Montage-QS may be used to augment provisional hardware (e.g., k-wires, plates and screws) and to help support bone fragments during the surgical procedure. The hardened putty acts only as a temporary support medium and is not intended to provide structural support during the healing process. Montage-QS can be drilled and tapped, and hardware can be placed through it at any time during process.
Device Story
Sterile, resorbable, two-component putty-like bone void filler; composed of granular calcium phosphate, calcium stearate, vitamin E acetate, triglyceride, polyalcohols, and lactide-diester/polyester-based polymers. Components manually mixed at surgery to form cohesive, settable material; adheres to bone surface; hardens in situ within 5 minutes without significant exothermic reaction. Used by surgeons in clinical settings to fill bone voids and provide temporary support for provisional hardware (k-wires, plates, screws) during procedures. Can be drilled and tapped after hardening. Material slowly resorbed and replaced by bone during remodeling. Benefits include temporary stabilization of bone fragments and hardware during surgical fixation.
Clinical Evidence
No human clinical data. Bench testing verified device drillability and temporary hardware support. Animal study (rabbit critical-sized femoral defect, n=not specified) compared Montage-QS to HydroSet and empty control. At 12 weeks, Montage-QS showed 17.2% new bone formation vs. 12.4% for HydroSet and 10% for control; ~75% implant material remained in both treatment groups.
Technological Characteristics
Two-part putty-like material; components: granular calcium phosphate (hydroxyapatite/beta-tricalcium phosphate), calcium stearate, vitamin E acetate, triglyceride, polyalcohols, lactide/caprolactone polymers. Hardens in situ via manual mixing. Sterile (gamma irradiation, SAL 10^-9). Biocompatibility per ISO 10993. No exothermic reaction.
Indications for Use
Indicated for filling surgically created or traumatic bony voids/gaps in extremities and pelvis not intrinsic to bony structure integrity. Used to augment provisional hardware (k-wires, plates, screws) and support bone fragments during surgery. Not for permanent structural support.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}------------------------------------------------
August 25, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Orthocon, Inc. Howard Schrayer Consultant 8 Lookout Hilton Head Island, South Carolina 29928
Re: K231903
Trade/Device Name: Montage-QS Settable, Resorbable Bone Putty Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV Dated: June 28, 2023 Received: June 28, 2023
Dear Howard Schrayer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Sara S. Thompson -S
For
Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K231903
Device Name Montage-QS Settable, Resorbable Bone Putty
#### Indications for Use (Describe)
Orthocon Montage-QS Settable, Resorbable Bone Putty is indicated to fill bony voids or gaps in the skeletal system (i.e., extremities and pelvis). These defects may be surgically created, or osseous defects created as the result of traumatic injury to the bone. Montage-QS is indicated only for filling bony voids or gaps that are not intrinsic to the integrity of the bony structure.
When hardened in situ, Montage-QS may be used to augment provisional hardware (e.g., k-wires, plates and screws) and to help support bone fragments during the surgical procedure. The hardened putty acts only as a temporary support medium and is not intended to provide structural support during the healing process.
Montage-QS can be drilled and tapped, and hardware can be placed through it at any time during process.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other sploct of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY (Per 21 CFR 807.92)
#### General Company Information
| Name: | Orthocon, Inc. |
|------------------------------------------------------------|----------------------------------------------------------------------------------|
| Contact: | Howard Schrayer<br>Regulatory Affairs Consultant |
| Address: | 700 Fairfield Avenue, Suite 1<br>Stamford, CT 06902 |
| Telephone: | (609) 273 - 7350 |
| Date Prepared | August 24, 2023 |
| General Device Information | |
| Product Name: | Montage-QS Settable, Resorbable Bone Putty |
| Common Name: | Resorbable calcium salt bone void filler device |
| Classification: | Product code: MQV |
| Regulation: | 21 CFR 888.3045 |
| Predicate Devices<br>Primary Predicate:<br>Orthovita, Inc. | HydroSet XT™<br>[510(k) Number K161447] |
| Reference Devices: | |
| Orthocon, Inc. | Montage-QS Settable, Resorbable Hemostatic Bone Putty<br>[510(k) Number K191140] |
| Orthocon, Inc. | Montage Settable, Resorbable Bone Putty<br>[510(k) Number K222063] |
#### Description
Montage-QS Settable, Resorbable Bone Putty is a sterile, biocompatible, resorbable material for use in filling bony voids or gaps in skeletal bones of the extremities. The Montage-QS device comprises two separate components of putty-like consistency containing granular calcium phosphate, calcium stearate, vitamin E acetate, a triglyceride, polyalcohols and a mixture of a lactide-diester and polyester-based polymers. When mixed together, the components of the Montage-QS device form a
{4}------------------------------------------------
cohesive putty-like material that adheres to the bone surface and remains in place following application. The resulting hardened, resorbable material is primarily calcium phosphate and is slowly resorbed and replaced with bone during the remodeling process. Montage-QS can be drilled and tapped, and hardware can be placed through it at any time during the setting process. The performance of Montage-QS was compared to HydroSet in a rabbit critical sized femoral defect model. At the 12-week timepoint, animal study data demonstrated new bone formation averages of 17.2% in the Montage-QS group, 12.4% in the HydroSet proup, and 10% in the empty defect negative control group. Animal study data suggest that approximately 75% of implant material remained in both the Montage-QS group and the HydroSet group at 12 weeks following implantation.
## Indications for Use
Orthocon Montage-QS Settable, Resorbable Bone Putty is indicated to fill bony voids or gaps in the skeletal system (i.e., extremities and pelvis). These defects may be surqically created, or osseous defects created as the result of traumatic injury to the bone. Montage-QS is indicated only for filling bony voids or gaps that are not intrinsic to the integrity of the bony structure.
When hardened in situ. Montage-QS may be used to augment provisional hardware (e.g., k-wires, plates and screws) and to help support bone fragments during the surgical procedure. The hardened putty acts only as a temporary support medium and is not intended to provide structural support during the healing process.
MONTAGE-QS can be drilled and tapped, and hardware can be placed through it at any time during the setting process.
## Purpose of Submission
Orthocon is proposing to modify the labeling (Instructions for Use) to provide for use of Montage-QS as a bone void filler that can be used to augment provisional hardware (e.g., k-wires, plates and screws) and to help support bone fragments during the surqical procedure.
## Substantial Equivalence
This submission supports the position that Orthocon Montage-QS Settable, Resorbable Bone Putty is substantially equivalent to a number previously cleared devices, including the HydroSet referenced predicate and is exactly the same composition as the reference predicate Orthocon Montage-QS Settable, Resorbable Bone Putty [cleared under 510(k) K191140].
{5}------------------------------------------------
## SUBSTANTIAL EQUIVALENCE INFORMATION
| Orthocon, Inc.<br>Montage-QS Settable,<br>Resorbable Bone Putty | Orthocon, Inc.<br>Montage-QS Settable, Resorbable<br>Hemostatic Bone Putty | Orthovita, Inc.<br>HydroSet XTTM<br>Bone Void Filler |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) - TBD | 510(k) – K191140 | 510(k) - K161447 |
| | Comparisons | |
| Device is indicated for use as a bone graft<br>substitute to fill voids in damaged bone that<br>are not intrinsic to the stability of the bony<br>structure. | Device is indicated for use in the control of<br>bleeding from cut or damaged bone surfaces<br>by acting as a tamponade | Device is indicated for use as a bone graft<br>substitute to fill voids in damaged bone that<br>are not intrinsic to the stability of the bony<br>structure. |
| Hardened device can be drilled and tapped<br>to provide temporary support for the<br>placement of provisional hardware during the<br>surgical procedure. | Not currently cleared for use in temporary<br>support of provisional hardware, but exactly<br>the same formulation as the subject of this<br>510(k). | Hardened device can provide temporary<br>support for the placement of provisional<br>hardware during the surgical procedure. |
| At the time of application, device is in the<br>form of a putty-like material | At the time of application, device is in the<br>form of a putty-like material | At the time of application, device is in the<br>form of a paste-like material |
| Device is designed to be manually applied<br>and spread onto voids in bone tissue | Device is designed to be manually applied<br>and spread onto bleeding bone tissue | Device is designed to be manually applied<br>or injected with a syringe and spread onto<br>voids in bone tissue |
| Montage-QS Settable, Resorbable Bone<br>Putty is formulated as a two-part putty/putty<br>device that forms a "settable" (hardening)<br>material when manually mixed at the time of<br>surgery | Montage-QS Settable, Resorbable Bone<br>Putty is formulated as a two-part putty/putty<br>device that forms a "settable" (hardening)<br>hemostatic putty when manually mixed at the<br>time of surgery | HydroSet XT settable, resorbable bone void<br>filler device is formulated as a two-part<br>powder/liquid device that forms a "settable"<br>(hardening) material when manually mixed<br>at the time of surgery |
| Sterile mixture of two separate components<br>of putty-like consistency comprised of | Sterile mixture of two separate components<br>of putty-like consistency comprised of | Sterile mixture of two separate components,<br>a powder comprised of dicalcium phosphate |
| granular calcium phosphate, (hydroxyapatite<br>and β-tricalcium phosphate), calcium<br>stearate, vitamin E acetate, a triglyceride,<br>polyalcohols and a mixture of a lactide-<br>diester and polyester-based (lactide and<br>caprolactone) absorbable polymers.<br>Montage-QS is to be mixed immediately<br>prior to use. Resulting settable material from<br>the two putties is primarily comprised (~70%<br>by weight) of calcium phosphate similar to<br>the mineral phase of native bone tissue | granular calcium phosphate, (hydroxyapatite<br>and β-tricalcium phosphate), calcium<br>stearate, vitamin E acetate, a triglyceride,<br>polyalcohols and a mixture of a lactide-<br>diester and polyester-based (lactide and<br>caprolactone) absorbable polymers.<br>Montage-QS is to be mixed immediately prior<br>to use. Resulting settable material from the<br>two putties is primarily comprised (~70% by<br>weight) of calcium phosphate similar to the<br>mineral phase of native bone tissue | dihydrate, tetracalcium phosphate and tri-<br>sodium citrate; and a liquid comprised of<br>sodium phosphate, polyvinylpyrrolidone and<br>water. HydroSet XT is to be manually<br>mixed immediately prior to use. Resulting<br>settable material from the two components<br>is primarily comprised of calcium phosphate<br>similar to the mineral phase of native bone<br>tissue. |
| Implanted device is resorbable in greater<br>than 30 days primarily due to presence of<br>calcium phosphate. | Implanted device is resorbable in greater<br>than 30 days primarily due to presence of<br>calcium phosphate. | Implanted device is resorbable in greater<br>than 30 days primarily due to presence of<br>calcium phosphate. |
| The non-calcium salt and non-polymeric<br>components degrade via dissolution; the<br>polymer degrades via hydrolysis and calcium<br>salts degrade via chemical dissolution and/or<br>cellular removal. | The non-calcium salt and non-polymeric<br>components degrade via dissolution; the<br>polymer degrades via hydrolysis and calcium<br>salts degrade via chemical dissolution and/or<br>cellular removal. | The non-calcium salt components degrade<br>via dissolution and calcium salts degrade<br>via chemical dissolution and/or cellular<br>removal. |
| Single-patient-use device is provided sterile<br>by gamma irradiation. | Single-patient-use device is provided sterile<br>by gamma irradiation. | Single-patient-use device is provided sterile<br>by gamma irradiation and ethylene oxide. |
| The bone putty is available in individual<br>and/or multi-pack patient use sizes. | The bone putty is available in individual<br>and/or multi-pack patient use sizes. | The device is available in individual; and/or<br>multi-pack patient use sizes. |
| Each putty is placed into a separate inner foil<br>"blister" which are contained within a single<br>outer foil pouch. The outer foil pouch<br>contains a desiccant. The inner blister and<br>outer pouch is heat sealed and sterilized. | Each putty is placed into a separate inner foil<br>"blister" which are contained within a single<br>outer foil pouch. The outer foil pouch<br>contains a desiccant. The inner blister and<br>outer pouch is heat sealed and sterilized. | Each kit contains one liquid-filled glass<br>syringe and one plastic bowl of powder<br>packaged within a double pre-formed tray<br>with a Tyvek lid. |
| Mixing for homogeneity takes 30 sec. | Mixing for homogeneity takes 30 sec. | Mixing for homogeneity takes 45 sec. |
| Material is hardened within 5 minutes of application | Material is hardened within 5 minutes of application | Material is hardened within 10 minutes of application |
| Device hardens with no appreciable exothermic reaction. | Device hardens with no appreciable exothermic reaction. | Device hardens with no appreciable exothermic reaction |
{6}------------------------------------------------
{7}------------------------------------------------
The differences described in the table above do not impact substantial equivalence.
{8}------------------------------------------------
## Testing Completed
## Efficacy Evaluation
The performance of Montage-QS as a bone void filler was compared to HydroSet in a rabbit critical size femoral defect model. Micro-CT and histopathology / histomorphometry assessments were performed on defects treated with each material to quantify device resorption and new bone formation. At the 12-week timepoint, animal study data demonstrated new bone formation averages of 17.2% in the Montage-QS group, 12.4% in the HydroSet predicate group, and 10% in the empty defect negative control group. Animal study data demonstrated that approximately 75% of implant material remained in both the Montage-QS group and the predicate group at 12 weeks following implantation.
## Performance Data
Testing was conducted to verify that the device may be drilled when hardened without fragmenting or being displaced. In addition, an in vitro study was conducted to demonstrate that once placed as indicated, the device provides temporary support to a complex repair until permanent hardware fixation is accomplished.
## Biocompatibility Testing
Testing was conducted to evaluate the device's biocompatibility in accordance with the recommendations of ISO 10993. The following biocompatibility studies were conducted on the final, finished, gamma-irradiation sterilized device and in accordance with the GLP requirements: cytotoxicity, irritation, sensitization, systemic toxicity, local tissue toxicity, hemolysis, endotoxicity and pyrogenicity.
#### Sterility
The gamma sterilization process has been validated to provide a SAL of 10-9. Each lot of finished devices is tested for bacterial endotoxin for lot release.
## Conclusions
The information provided establishes that the Orthocon Montage-QS Settable, Resorbable Bone Putty performs substantially equivalent to the predicate device for the same intended use. All results demonstrate that any differences in technology do not impact substantial equivalence.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.